eClinics

K260075 · Ceiba Healthcare US, Inc. · MSX · Oct 7, 2026 · Cardiovascular

Device Facts

Record IDK260075
Device NameeClinics
ApplicantCeiba Healthcare US, Inc.
Product CodeMSX · Cardiovascular
Decision DateOct 7, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

eClinics is intended to allow trained healthcare professionals to capture and display waveforms (such as ECG, SpO2, ART), discrete parameters, patient demographics, alarms, and other information from connected medical devices and electronic health record systems on centralized, multi-patient views. This software is intended as a diagnostic aid to provide access for professional staff to view patient clinical data when the professional cannot observe the source system directly. eClinics does not store Protected Healthcare Information on the device accessing the data. Data available on the web viewer is in “near real time” and is displayed to show a full range of clinical data available, even when the data sources are obtained from different devices.

Device Story

eClinics is a software-only, web-based platform for centralized, multi-patient monitoring; it aggregates physiologic data (ECG, SpO2, ART, etc.), EHR data, and alarms from connected bedside monitors, ventilators, and infusion pumps. Data is ingested via eConnect middleware (hardware/gateway interfaces) and HL7/HTTPS EHR connections. The system displays near real-time waveforms and parameters on desktop browsers for clinicians. It functions as a secondary, adjunct display; it does not control or alter source device settings or therapy. It provides advisory clinical decision support (CDS) outputs (e.g., SOFA/qSOFA scores) and supports telehealth audio/video. Used in hospitals and clinics by healthcare professionals to assess patient status when direct observation is unavailable. It benefits patients by extending monitoring reach and facilitating remote clinical decision-making.

Clinical Evidence

No clinical performance data was necessary to claim substantial equivalence. Evidence consists of bench testing, including software verification and validation (per IEC 62304), electronic interface/integration testing, alarm system testing (per IEC 60601-1-8), cybersecurity testing (per FDA 2025 guidance), and summative human factors validation (per IEC 62366-1).

Technological Characteristics

Software-only, web-based SaMD; server-side services (Windows Server 2016+, PostgreSQL 13+). Connectivity via HL7, APIs, and eConnect middleware (CON-1/CON-2 interfaces). Architecture includes front-end web app, backend services, Gatekeeper (authentication), and Telehealth container. Role-based access control (RBAC) and audit trails. Operates as a distributed information system for alarm conditions (IEC 60601-1-8).

Indications for Use

Indicated for pediatric, adult, and geriatric patients to allow trained healthcare professionals to visualize physiological data, waveforms, and patient parameters from networked medical devices and EHRs as a diagnostic aid when the source system cannot be observed directly. Used for remote consultation, viewing historic/near real-time data, and managing secondary alarm transmission. Contraindications: None.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{1} ![LOGO](U.S. Food & Drug Administration) October 7, 2026 Ceiba Healthcare US, Inc. % Mary Vater Director of Regulatory Affairs Innolitics, LLC 1101 W. 34th St. #550 Austin, Texas 78705 Re: K260075 Trade/Device Name: eClinics Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX Dated: September 11, 2026 Received: September 11, 2026 Dear Mary Vater: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K260075 - Mary Vater Page 2 Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {3} K260075 - Mary Vater Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JENNIFER W. SHIH -S Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} # Indications for Use Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K260075 Please provide the device trade name(s). eClinics Please provide your Indications for Use below. eClinics is intended to allow trained healthcare professionals to capture and display waveforms (such as ECG, SpO2, ART), discrete parameters, patient demographics, alarms, and other information from connected medical devices and electronic health record systems on centralized, multi-patient views. This software is intended as a diagnostic aid to provide access for professional staff to view patient clinical data when the professional cannot observe the source system directly. eClinics does not store Protected Healthcare Information on the device accessing the data. Data available on the web viewer is in “near real time” and is displayed to show a full range of clinical data available, even when the data sources are obtained from different devices. eClinics gathers data from the medical devices that are connected to the patient. eClinics is not directly connected to the individual, but to a networked medical device receiving data, and connects that data to individuals by medical record number or defined patient ID. eClinics is intended to be used by trained healthcare professionals for the following purposes: - To visualize physiological data using a web viewer to assess clinical status of a patient when the source system cannot be observed directly by the clinician. - To view other near real-time and historic waveforms. - To view other near real-time patient data from a monitored system. - To provide a request for remote consultation regarding a patient’s waveform or other data. - To allow trained healthcare professionals to manage secondary alarm transmission. eClinics is appropriate for use in pediatric, adult and geriatric populations, as the networked medical devices eClinics collects data from are already cleared for these populations. eClinics software can display the following physiologic parameters and waveforms captured by other cleared medical devices, or hospital information systems: - ECG Waveform - Heart Rate - Respiratory Rate - Oxygen Saturation - Central Venous Pressure - Pulmonary Artery Pressure - Temperature - Body Weight - Systolic Blood Pressure Invasive - Mean Arterial Pressure Invasive - Diastolic Blood Pressure Invasive - Systolic Blood Pressure Cuff - Mean Arterial Pressure Cuff - Diastolic Blood Pressure Cuff - Vasoactive Infusions - Ventilators {5} - Allergies - IV Fluid Delivery Rate - Pulse Rate - Monitored Events - Laboratory Data, including o Blood Gas o Chemistry o Hematology o Coagulation - Medications - IV Volume Infused - VTBI - ETCO2 Warning: eClinics is an adjunct to and not a replacement for direct viewing of the monitoring medical device(s). Precautions: eClinics requires an internet or cellular connection to provide data; this may be the hospital's internal network (LAN), or an internet/cellular connection where the hospital network is not used or not available. eClinics is not intended for use where cellular telephones or other wireless devices are prohibited unless connectivity is instead provided via the hospital's wired/internal network in such areas. Please select the types of uses (select one or both, as applicable). ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ? {6} CEIBA HEALTH K260075 # eClinics 510(k) Summary # 1. CONTACT INFORMATION | Company Name | Ceiba Healthcare US Inc. | | --- | --- | | Address | 100 Park Avenue Suite 1635, New York, NY 10017 | | Phone Number | +1-646-476-0384 | | Company Representative | Afsar Alp, CEO/President | | Email | afsar.alp@ceiba-healthcare.com | | Primary Correspondent | Innolitics LLC | | Correspondent Contact | Mary Vater, Director of Regulatory Affairs | | Primary Correspondent Email | fda@innolitics.com | | Additional Contributors | J. David Giese, Partner at Innolitics Ameera Roubi, Regulatory Consultant at Innolitics Ilya Spivakov, Senior Software Engineer at Innolitics | | Date Summary Prepared | 9 September 2026 | # 2. DEVICE INFORMATION | Trade Name | eClinics | | --- | --- | | Common Name | Physiological Monitors Network and Communication System | | Product Code | MSX | | Regulation Number | 21 CFR 870.2300 | | Class | Class II | | Panel | Cardiovascular | # 3. PREDICATE DEVICE INFORMATION | Predicate Device Name | AirStrip ONE Web Client with ACM (AirStrip Technologies, Inc.) | | --- | --- | Page 1 of 11 {7} CEIBA HEALTH K260075 # eClinics 510(k) Summary | Predicate Device K Number | K211949 | | --- | --- | | Product Code | MSX | | Regulation Number | 21 CFR 870.2300 | | Class | Class II | | Panel | Cardiovascular | ## 4. DEVICE DESCRIPTION eClinics is a software-only, web-based device that receives, organizes and displays physiologic and other patient information from connected cleared medical devices and from hospital electronic health record (EHR) systems on centralized single-patient and multi-patient views. It is deployed as a set of containerized server-side services, accessed by trained healthcare professionals through a desktop web browser, in either a cloud-hosted or an on-premises configuration. eClinics does not control, program or alter the function, settings, therapy parameters or alarm configuration of any connected medical device, and it does not store protected health information on the device used to access it. It is a secondary, adjunct display: the source medical device at the point of care remains the primary monitor and the primary alarm annunciator. eClinics receives data from three types of cleared medical device — bedside physiologic monitors, mechanical ventilators, and infusion and syringe pumps — through the Ceiba eConnect middleware, over one of two electronic interfaces: CON-1, from a manufacturer central-station or middleware gateway to eConnect; and CON-2, from an eConnect hardware (eConnect Box / Uno / Nano) at the point of care to eConnect. Laboratory, blood gas, demographic, medication, and admission-discharge-transfer data are received from the EHR over HL7 and HTTPS. eConnect is receive-only: it transmits toward a device only the protocol-defined messages required to establish and maintain the data connection, and implements no message that writes, modifies or controls device settings, therapy parameters or alarm configuration. The manufacturers and models declared compatible with eClinics, each with its FDA clearance number, are listed in the Compatible Device List section of the Instructions for Use. ### 4.1. Software Modules The eClinics software comprises a front-end web application providing the browser-based user interface and its clinical modules; the eClinics backend, providing the application services that receive, persist and serve patient data, including the clinical decision support calculation services; Gatekeeper, providing authentication and session management; a Telehealth container providing audio/video consultation; the eConnect middleware, providing device integration; and the datastores. The architecture is provided in the System and Software Architectural Design Chart. The clinical modules presented to the user are CMS Prime; CMS Telemetry, including CMS Device Telemetry, Wavestrip Events and CMS History; Population Management; Patient Timeline; Blood Gas; Diagnosis; Laboratory Results; Patient Score; User Assignment; Flowsheets; Order Management; Doctor Page 2 of 11 {8} CEIBA HEALTH K260075 # eClinics 510(k) Summary Notes; Nurse Forms, including Restriction Tracking; Medication Administration Record; Fluid Balance; Telehealth; eConnect; and the Admin Platform. Access to each module is governed by role-based permissions. # 4.2. Alarm Display Functions and Limitations eClinics is a distributed information system about alarm conditions within the meaning of IEC 60601-1-8. Alarm text, severity and color are displayed as received from the source device, and alarm priority is manufacturer-defined; eClinics does not generate, evaluate, prioritize, suppress or recalculate any alarm condition, and provides no alarm configuration. Where several alarms are active for a patient, the highest-priority alarm is displayed continuously and the remainder are presented in rotation, with the number active at each priority level shown alongside. Limitations, stated in the Instructions for Use: eClinics is an adjunct to, and not a substitute for, the alarms on the originating medical device or direct patient observation, is not to be relied upon for the receipt of alarm signals, and does not guarantee delivery of alarm information. Annunciation occurs within 5.0 seconds of receipt of the alarm message at the eConnect interface; the delay before the source device transmits that message is determined by that device. On loss of a device data feed, waveform rendering stops, numeric values freeze, and a "No device data" notification/indicator is presented for the affected bed. # 4.3. Clinical Decision Support Outputs eClinics computes advisory clinical decision support outputs from data already present in the system. None of the CDS outputs are alarm signals under IEC 60601-1-8, none carries auditory annunciation, and none is presented to the user as a device alarm. # 5. INTENDED USE eClinics is intended to allow trained healthcare professionals to capture and display waveforms (such as ECG, SpO2, ART), discrete parameters, patient demographics, alarms, and other information from connected medical devices and electronic health record systems on centralized, multi-patient views. eClinics does not store Protected Healthcare Information on the device accessing the data. Data available on the web viewer is in "near real time" and is displayed to show a full range of clinical data available, even when the data sources are obtained from different devices. # 6. INDICATIONS FOR USE eClinics is intended to allow trained healthcare professionals to capture and display waveforms (such as ECG, SpO2, ART), discrete parameters, patient demographics, alarms, and other information from connected medical devices and electronic health record systems on centralized, multi-patient views. This software is intended as a diagnostic aid to provide access for professional staff to view patient clinical data when the professional cannot observe the source system directly. eClinics does not store Protected Healthcare Information on the device accessing the data. Data available on the web viewer is in "near real Page 3 of 11 {9} CEIBA HEALTH K260075 # eClinics 510(k) Summary time" and is displayed to show a full range of clinical data available, even when the data sources are obtained from different devices. eClinics gathers data from the medical devices that are connected to the patient. eClinics is not directly connected to the individual, but to a networked medical device receiving data, and connects that data to individuals by medical record number or defined patient ID. eClinics is intended to be used by trained healthcare professionals for the following purposes: - To visualize physiological data using a web viewer to assess clinical status of a patient when the source system cannot be observed directly by the clinician. - To view other near real-time and historic waveforms. - To view other near real-time patient data from a monitored system. - To provide a request for remote consultation regarding a patient's waveform or other data. - To allow trained healthcare professionals to manage secondary alarm transmission. eClinics is appropriate for use in pediatric, adult and geriatric populations, as the networked medical devices eClinics collects data from are already cleared for these populations. eClinics software can display the following physiologic parameters and waveforms captured by other cleared medical devices, or hospital information systems: ECG waveform; heart rate; respiratory rate; oxygen saturation; central venous pressure; pulmonary artery pressure; temperature; body weight; systolic blood pressure invasive, mean arterial pressure invasive, diastolic blood pressure invasive, systolic blood pressure cuff, mean arterial pressure cuff, diastolic blood pressure cuff; vasoactive infusions; ventilators; allergies; IV fluid delivery rate; pulse rate; monitored events; laboratory data, including blood gas, chemistry, hematology, coagulation; medications; IV volume infused; VTBI; ETCO2. Warning: eClinics is an adjunct to and not a replacement for direct viewing of the monitoring medical device(s). Precautions: eClinics requires an internet or cellular connection to provide data; this may be the hospital's internal network (LAN), or an internet/cellular connection where the hospital network is not used or not available. eClinics is not intended for use where cellular telephones or other wireless devices are prohibited unless connectivity is instead provided via the hospital's wired/internal network in such areas. ## 7. CONTRAINDICATIONS FOR USE There are no contraindications for the eClinics software. ## 8. SUBSTANTIAL EQUIVALENCE DISCUSSION ### 8.1. Predicate Device Indications for Use AirStrip ONE Web Client with ACM is intended to allow clinicians to capture and display waveforms (such as ECG, SpO2, ART), discrete parameters, patient demographics, alarms, and other information from medical devices on a desktop device. This software is intended as a diagnostic aid to provide access for professional staff to view patient clinical data when the professional cannot observe the source system directly. AirStrip ONE Web Client with ACM does not store Protected Healthcare Information on the Page 4 of 11 {10} CEIBA HEALTH K260075 # eClinics 510(k) Summary device accessing the data. Data available on the web viewer is in “near real time” and is displayed to show a full range of clinical data available, even when the data sources are obtained from different devices. AirStrip ONE Web Client with ACM gathers data from the medical devices that are connected to the patient. AirStrip ONE Web Client with ACM is not directly connected to the individual, but to a networked medical device receiving data, and connects that data to individuals by medical record number or defined patient ID. AirStrip ONE Web Client with ACM is intended to be used by clinicians for the following purposes: - To visualize physiological data using a web viewer to assess clinical status of a patient when the source system cannot be observed directly by the clinician. - To view other near real-time and historic waveforms. - To view other near real-time patient data from a monitored system. - To provide a request for remote consultation regarding a patient’s waveform or other data. - To send electronic images of the cardiac rhythm to the medical record. - To allow clinicians to manage secondary alarm transmission by suppressing non-critical alarms. AirStrip ONE Web Client with ACM is appropriate for use in pediatric (0-21 years), adult and geriatric populations, as the networked medical devices AirStrip collects data from are already cleared for these populations. AirStrip ONE Web Client with ACM software can display the following physiologic parameters and waveforms captured by other cleared medical devices: - ECG Waveform - Heart Rate - Respiratory Rate - Oxygen Saturation - Intracranial Pressure - Central Venous Pressure - Pulmonary Capillary Wedge Pressure - Pulmonary Artery Pressure - Cardiac Index - Cardiac Output - Cerebral Perfusion Pressure - Temperature - Temperature Source - Bispectral index (BIS) - Umbilical Artery - Body Weight - Systolic Blood Pressure Invasive - Mean Arterial Pressure Invasive - Diastolic Blood Pressure Invasive - Systolic Blood Pressure Cuff - Mean Arterial Pressure Cuff - Diastolic Blood Pressure Cuff - Vasoactive Infusions Page 5 of 11 {11} CEIBA HEALTH K260075 # eClinics 510(k) Summary - Ventilators - Allergies - IV Fluid Delivery Rate - IV Medication Dose - Pulse Rate - Bispectral Parameters - Monitored Events - Right Atrial Pressure - Left Atrial Pressure - Laboratory Data, including - Blood Gas - Chemistry - Hematology - Coagulation - Hemodynamic Calculated Values - Medications - IV Volume Infused - ETCO2 - Fractional Inspired Concentration O2, CO2 - Case Consumption Liquid Medications - Case Consumption Respiratory Gas Contraindications: There are no contraindications. Warning: AirStrip ONE Web with ACM is an adjunct to and not a replacement for direct viewing of the monitoring medical device(s). Precautions: AirStrip ONE Web with ACM requires an internet or cellular connection to provide data. AirStrip ONE Web with ACM is not intended for use where cellular telephones or other wireless devices are prohibited. ### 8.2. Indications for Use Equivalence Discussion The Ceiba eClinics system shares the same fundamental intended use as the predicate device: providing trained healthcare professionals with access to physiologic and patient-related data from connected cleared medical devices, to support clinical decision-making in a secondary capacity. Like AirStrip ONE Web Client with ACM, the subject device displays patient information as a diagnostic aid and serves as an adjunct rather than a replacement for bedside monitoring or alarm systems, and both emphasise remote access and a broad range of data types including waveforms, laboratory values and medications. eClinics additionally incorporates integration with EHR systems, centralized multi-patient views, and advisory clinical decision support score outputs; the predicate additionally provides management of secondary alarm transmission by suppressing non-critical alarms. These differences reflect workflow adaptations and technological enhancements. They do not alter the core intended use of secondary display of patient data, and they do not raise new questions of safety or effectiveness. The subject device is therefore substantially equivalent to the predicate device with respect to indications for use. Page 6 of 11 {12} CEIBA HEALTH K260075 # eClinics 510(k) Summary 8.3. Device Comparison Table | Attribute | Ceiba eClinics (Subject Device) | AirStrip ONE | Comparison (Subject vs. Predicate) | | --- | --- | --- | --- | | 510(k) # | K260075 | K211949 | - | | Class | II | II | same | | Primary Product Code | MSX | MSX | same | | All Product Codes | MSX | MSX | same | | Regulation | 21 CFR 870.2300 — Cardiac Monitor (Including Cardiotachometer & Rate Alarm) | 21 CFR 870.2300 — Cardiac Monitor (Including Cardiotachometer & Rate Alarm) | same | | Indications for Use | See Above | See Above | Both are clinician-facing secondary displays for near real-time patient data when the source cannot be observed directly; subject adds centralized multi-patient views, care-team communication, EHR integration; predicate adds secondary alarm transmission management. | | Environment of Use | Hospitals and clinics; remote/telehealth interactions supported. | Hospital/clinical environments with networked medical devices. | same | | Intended User | Healthcare professionals (physicians, nurses, dietitians, etc.). | Healthcare professionals. | same | | Alarms/Alerts | Receives device alarms and provides near real-time notifications; severity colors; advisory CDS indicators. | Displays alarms from networked devices and allows suppression of non-critical secondary alarms. | Both support secondary alarm/notification for clinicians; subject adds CDS-based advisory alerts; predicate focuses on routing/suppression. | | Operating System / Deployment | Server OS: Windows Server 2016+; | Web client on standard desktop browsers; server | Both are software-only solutions on existing IT | Page 7 of 11 {13} CEIBA HEALTH K260075 # eClinics 510(k) Summary | Attribute | Ceiba eClinics (Subject Device) | AirStrip ONE | Comparison (Subject vs. Predicate) | | --- | --- | --- | --- | | | Database: PostgreSQL 13+. | platform defined by AirStrip. | infrastructure; differences are implementation details. | | Communication Method | HL7 over TCP, APIs, EHR integration, web access, A/V calls, SSO with Telehealth. | Receives data from networked devices and hospital systems for web viewing and alarm management. | Both rely on networked data from cleared devices/systems; transports differ but role is the same. | | Monitoring Type | Near real-time monitoring; telemetry; population management; CDS. | Near real-time remote visualization of waveforms and parameters. | Both provide near real-time remote viewing as an adjunct to bedside monitoring; subject also supports population/CDS views. | | Interoperability | HL7 interfaces; APIs; EHR integration; device integration via interface servers; eClinics-eConnect. | Integrates with networked medical devices and hospital systems. | Both depend on interoperable connections to existing cleared systems; differences are in breadth of integrations. | | Devices it Communicates With | Bedside monitors, ventilators, infusion pumps; EHR systems; mobile A/V carts/units. | Networked medical devices and primary monitors providing the listed parameters and alarms. | Same clinical role of aggregating data from primary devices; subject also connects to additional IT systems. | | Configurations / Architecture | Admin Platform, CMS Prime, CMS Telemetry, Population Management, clinical modules, CDS, Telehealth; interface and alert servers. | Web client with ACM rules engine on the AirStrip ONE platform. | Both provide a clinician-facing viewer with secondary alarm support; architectures differ, but intended use is the same. | | Hardware Requirements | ≥8-core CPU; ≥32 GB RAM; Windows Server 2016+; PostgreSQL 13+. | Standard desktops/laptops and server hardware per AirStrip deployment. | Both use general-purpose IT hardware; no patient-connected hardware differences. | | Communications Operation | Near real-time updates (10–15 s), instant alarm/event tracking, dashboards. | Near real-time acquisition and display with secondary alarm routing. | Same clinical behavior of near real-time central monitoring and notifications. | | Administration / User Management | Role-based access, configuration, audit trails. | Institution-configured access and ACM rules. | Both allow institutional configuration and controlled access; subject also | Page 8 of 11 {14} CEIBA HEALTH K260075 # eClinics 510(k) Summary | Attribute | Ceiba eClinics (Subject Device) | AirStrip ONE | Comparison (Subject vs. Predicate) | | --- | --- | --- | --- | | | | | emphasizes detailed RBAC/audit. | | Connection to Network | Web-based over hospital network/internet; HL7 TCP to EHR/devices. | Requires network (internet or cellular) to connect to hospital systems and devices. | Same need for reliable, secure network connectivity; differences are in specific topologies. | | Logging / Audit | Audit trails of user actions. | Usage and alarm reporting via platform tools. | Both support post-event review and quality activities; subject describes explicit per-user audit logging. | | User Interface | Web dashboards (Population Management, Patient Timeline, Flowsheets, MAR, Orders, Notes). | Web viewer showing waveforms, parameters, and alarms on desktop. | Both are web-based clinician UIs; subject adds multi-patient and documentation views, predicate focuses on waveform/parameter display. | | Alarm Presentation | Severity displays exactly as received from the source device and CDS- indicators; thresholds set on source devices. | ACM for suppression of non-critical secondary alarms. | Both configure secondary notification behavior; neither changes primary monitor alarm algorithms. | | Performance Testing (selected) | Testing per IEC 62304, IEC 62366-1, and cybersecurity/network guidance. | Software V&V and performance testing for the AirStrip ONE platform. | Both are software-validated SaMD central-monitor solutions; subject explicitly references current SaMD standards. | ### 8.4. Summary of Technological Characteristics Comparison Both eClinics and AirStrip ONE + ACM are clinician-facing remote viewing and central monitoring solutions that aggregate physiologic data from bedside and networked medical devices, display near-real-time multi-parameter data when the source system cannot be observed directly, and support clinical decision-making as diagnostic aids rather than replacing bedside monitors. Ceiba achieves this as a software-only, web-deployed platform that integrates bedside and EHR data (for example, through HL7 interfaces and APIs), provides configurable multi-patient views, secondary notifications, and advisory clinical decision support (for example, or SOFA, qSOFA scores and Sepsis Status indicator), and supports A/V telehealth, RBAC, and audit trails. AirStrip ONE Web Client with ACM is a web-based viewer that receives data from already-cleared networked medical devices, links those data to specific patients, and allows clinicians to visualize near-real-time and historic waveforms, parameters, and alarms, manage secondary alarm transmission, request remote consultation, and transmit images of cardiac rhythm for Page 9 of 11 {15} CEIBA HEALTH K260075 # eClinics 510(k) Summary inclusion in the medical record. These differences in deployment architecture, data sources, and ancillary features do not alter the shared clinical role—providing remote access to patient monitoring data as an adjunct to, and not a replacement for, direct observation—and with appropriate software V&V, usability, interoperability, cybersecurity, and network-performance testing, they do not introduce new risks or change the intended use, supporting a conclusion of substantial equivalence. ## 9. PERFORMANCE DATA ### 9.1. Software Verification and Validation Testing Software verification and validation testing was conducted and documentation was provided at the Enhanced Documentation Level, as recommended by the 2023 FDA guidance “Content of Premarket Submissions for Device Software Functions”. Testing also confirmed that the device maintains basic safety and essential performance under single-fault conditions. ### 9.2. Electronic Interface and Device Integration Testing Interoperability of the electronic interfaces was verified by execution of the Medical Device Integration Verification Procedure. ### 9.3. Alarm System Testing Alarm presentation was verified against the applicable requirements of IEC 60601-1-8 for a distributed information system about alarm conditions across all declared device categories (bedside physiologic monitors, ventilators, and infusion/syringe pumps), including testing at the declared maximum configuration and verification of the alarm notification latency criteria. ### 9.4. Cybersecurity Testing Cybersecurity documentation and testing were provided in accordance with the 2025 FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”. Third-party penetration testing was performed. ### 9.5. Human Factors and Usability Testing Summative human factors validation was conducted in accordance with the FDA guidance “Applying Human Factors and Usability Engineering to Medical Devices” and IEC 62366-1. ### 9.6. Clinical Performance Testing Page 10 of 11 {16} CEIBA HEALTH K260075 # eClinics 510(k) Summary No clinical performance data was necessary to claim substantial equivalence. # 10. CONCLUSION eClinics shares similar technological characteristics, intended use and functionality with the predicate device, AirStrip ONE Web Client with ACM (K211949). There are no differences between the devices that raise new questions of safety or effectiveness. Furthermore, the non-clinical performance test data and the software verification and validation demonstrate that eClinics performs comparably to the predicate device in terms of safety and effectiveness. Based on the device comparisons and the acceptable testing results, it is determined that eClinics is substantially equivalent to the predicate device. Page 11 of 11
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...