K260055 · Siemens Medical Solutions USA, Inc. · QIH · Sep 25, 2026 · Radiology
Device Facts
Record ID
K260055
Device Name
syngo.CT Brain Quantification
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
QIH · Radiology
Decision Date
Sep 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260055 · Sep 25, 2026
syngo.CT Brain Quantification
Siemens Medical Solutions USA, Inc.
Retrospective anonymized non-contrast head CT datasets from multiple clinical institutions
The device performance was evaluated in a retrospective study using clinical datasets to assess the accuracy of automated midline shift measurements and intracranial hyperdensity segmentation compared to neuroradiologist ground-truth.
Retrospective, stand-alone clinical performance study; Retrospective, stand-alone study
Patients aged 18-98 years with non-contrast head CT scans; Sample Size: 500 total cases (300 for Midline Shift, 200 for Brain Hyperdensity); Number of Sites: Multiple clinical institutions in the United States and Europe
Not applicable for this study
Volumetric accuracy and Dice Similarity Coefficient (DSC) for hyperdensities; absolute measurement difference for midline shift
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Intracranial hyperdensities
Supervised voxel classification with Convolutional Neural Networks
—
Mean Dice Similarity Coefficient 0.72 (95% CI 0.68, 0.75); Mean absolute volume difference 4.84 ml
Supervised voxel classification with Convolutional Neural Networks
—
Mean absolute measurement difference 0.87 mm
—
—
Retrospective, stand-alone study: 300 head CT cases.
>1 (board-certified neuroradiologists)
Indications for Use
syngo.CT Brain Quantification is a radiological post-processing application for the analysis of non-contrast head CT images. The device is intended for automatic labeling, visualization, and quantification of intracranial structures. The software is indicated for use in the analysis of: • Intracranial hyperdensities • Midline Shift The device is intended to be used in patients aged 18 years or above. No surgical signs must be present in the images.
Device Story
Radiological post-processing application; analyzes non-contrast head CT images. Inputs: DICOM images from hosting CT systems. Processing: Deep learning (CNN) segments intracranial hyperdensities and quantifies midline shift. Outputs: Annotated DICOM images with volumetric data returned to host system. Used in hospital networks by clinicians. Benefits: Automates labeling, visualization, and quantification of brain structures; assists clinical decision-making by providing objective measurements of hyperdensities and midline shift. Operates on SOMARIS-X and syngo.via platforms.
Clinical Evidence
Retrospective study of 300 head CT cases (midline shift) and 200 cases (hyperdensities). Midline shift accuracy: mean absolute difference 0.87 mm (SD 1.04). Hyperdensity segmentation: mean Dice Similarity Coefficient 0.72 (95% CI 0.68, 0.75); mean absolute volume difference 4.84 ml (SD 6.85). Compared against board-certified neuroradiologist ground-truth.
Technological Characteristics
Standalone software; DICOM compatible. Deep learning (CNN) for voxel classification. Operates on SOMARIS-X and syngo.via platforms. Complies with IEC 62304, ISO 14971, IEC 62366-1, and NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for patients aged 18+ requiring analysis of non-contrast head CT images for intracranial hyperdensities and midline shift. Contraindicated for images containing surgical signs.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 25, 2026
Siemens Medical Solutions USA, Inc.
Kenny Bello
Regulatory Affairs Professional
810 Innovation Dr.
Knoxville, Tennessee 37932
Re: K260055
Trade/Device Name: syngo.CT Brain Quantification
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: August 27, 2026
Received: August 28, 2026
Dear Kenny Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260055 - Kenny Bello
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K260055 - Kenny Bello
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260055 | ? |
| Please provide the device trade name(s). | | ? |
| syngo.CT Brain Quantification | | |
| Please provide your Indications for Use below. | | ? |
| syngo.CT Brain Quantification is a radiological post-processing application for the analysis of non-contrast head CT images. The device is intended for automatic labeling, visualization, and quantification of intracranial structures. The software is indicated for use in the analysis of: • Intracranial hyperdensities • Midline Shift The device is intended to be used in patients aged 18 years or above. No surgical signs must be present in the images. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
SIEMENS
Healthineers
K260055
# 510(k) Summary
for
syngo.CT Brain Quantification with software version SOMARIS/8 VB80
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92".
1. Identification of the Submitter
Submitter
SIEMENS MEDICAL SOLUTIONS USA, INC.
810 INNOVATION DR.
KNOXVILLE, TN 37932
Registration Number: 1034973
Importer/Distributor
SIEMENS MEDICAL SOLUTIONS USA, INC.
40 Liberty Boulevard
Malvern, PA 19355
Registration Number: 2240869
Location of Manufacturing Site
Siemens Healthineers AG
SIEMENSSTRASSE 1 -OR-
Rittigfeld 1
FORCHHEIM Bavaria, DE 91301
Registration Number: 3004977335
Note: Descriptions in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.
Submitter Contact Person:
Kenny M Bello
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
Molecular Imaging
810 Innovation Drive
Knoxville, TN 37932
Phone: (205) 856-6099
monsuru.bello@siemens-healthineers.com
Backup Contact:
Alaine Medio
Regulatory Affairs Manager
Siemens Medical Solutions USA, Inc.
Molecular Imaging
810 Innovation Drive
Knoxville, TN 37932
Phone: +1 (865)206-0337
alaine.medio@siemens-healthineers.com
K260055 - Siemens Medical Solutions USA, Inc.
1
{6}
SIEMENS Healthineers
## 2. Device Name and Classification
Product Name: syngo.CT Brain Quantification
Propriety Trade Name: syngo.CT Brain Quantification
Regulation Description: Medical Image Management and Processing System
Classification Name: Automated Radiological Image Processing Software
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Product Code: QIH
## 3. Predicate Devices
Predicate Device:
Trade Name: qER-Quant
Manufacturer: Qure.Ai Technologies
510(k) Number: K211222
Clearance Date: 07/30/2021
Regulation Description: Medical Image Management and Processing System
Classification Name: Automated Radiological Image Processing Software
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Product Code: QIH
## 4. Device Description
syngo.CT Brain Quantification is a radiological post-processing application intended for the analysis of non-contrast head datasets acquired with computed tomography (CT) imaging systems. syngo.CT Brain Quantification is intended for automatic labeling, visualization, and quantification of intracranial structures. The subject device is intended to be used in the analysis of intracranial hyperdensities and in the quantification of the midline shift. The device receives images from a hosting system and returns DICOM images back to the hosting system, which can display the results within its user interface.
K260055 - Siemens Medical Solutions USA, Inc.
2
{7}
SIEMENS
Healthineers
## 5. Indications for Use
syngo.CT Brain Quantification is a radiological post-processing application for the analysis of non-contrast head CT images. The device is intended for automatic labeling, visualization, and quantification of intracranial structures.
The software is indicated for use in the analysis of:
- Intracranial Hyperdensities
- Midline Shift
The device is intended to be used in patients aged 18 years or above. No surgical signs must be present in the images.
## 6. Indications for Use Comparison to the Predicate Device
| Subject Device | Predicate Device |
| --- | --- |
| Siemens syngo.CT Brain Quantification | Qure.Ai Technologies qER-Quant |
| syngo.CT Brain Quantification is a radiological post-processing application for the analysis of non-contrast head CT images. The device is intended for automatic labeling, visualization, and quantification of intracranial structures. The software is indicated for use in the analysis of: • Intracranial Hyperdensities • Midline Shift The device is intended to be used in patients aged 18 years or above. No surgical signs must be present in the images. | The qER-Quant device is intended for automatic labeling, visualization and quantification of segmentable brain structures from a set of Non-Contrast head CT (NCCT) images. The software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on NCCT images. qER-Quant provides volumes from NCCT images acquired at a single time point and provides a table with comparative analysis for two or more images that were acquired on the same scanner with the same image acquisition protocol for the same individual at multiple time points. The qER-Quant software is indicated for use in the analysis of the following structures: Intracranial Hyperdensities, Lateral Ventricles and Midline Shift. |
Both devices share the same fundamental intended use: post-processing of non-contrast head CT images for automatic labeling, visualization, and quantification of brain structures.
Both devices are:
K260055 - Siemens Medical Solutions USA, Inc.
3
{8}
SIEMENS Healthineers
- Standalone post-processing software applications
- Using non-contrast head CT images as input
- Intended for automatic labeling, visualization and quantification of brain structures
- Indicated for use in the analysis of intracranial hyperdensities and midline shift
## 7. Comparison of Technological Characteristics with the Predicate Device
The differences and similarities between the above referenced predicate device are listed at a high-level in the following table:
| Feature | Subject Device | Predicate Device | Comparison Table |
| --- | --- | --- | --- |
| | syngo.CT Brain Quantification | Qure.AI qER-Quant | |
| Regulation | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Product Code | QIH | QIH | Same |
| Regulation Description | Medical Image Management and Processing System | Medical Image Management and Processing System | Same |
| Device type | Automated Radiological Image Processing Software | Automated Radiological Image Processing Software | Same |
| Intended User | Hospital networks and qualified clinicians | Hospital networks and qualified clinicians | Same |
| Input images | Non-contrast CT | Non-contrast CT | Same |
| Body Part | Head | Head | Same |
| Target structures analyzed on NCCT scans | Intracranial hyperdensities and midline shift | Intracranial hyperdensities, lateral ventricles and midline shift | **Similar** Lateral ventricles are not target structures for syngo.CT Brain Quantification |
| Support of data acquired at multiple time points | No | Yes | **Different** Subject device operates on data acquired at single time point and does not provide functionality for analysis of images acquired at multiple timepoints. |
| Artificial Intelligence algorithm | Segmentation by deep learning (supervised voxel classification with Convolutional Neural Networks) | Segmentation by deep learning (supervised voxel classification with Convolutional Neural Networks) | Same |
K260055 - Siemens Medical Solutions USA, Inc.
4
{9}
SIEMENS Healthineers
| Feature | Subject Device | Predicate Device | Comparison Table |
| --- | --- | --- | --- |
| | syngo.CT Brain Quantification | Qure.AI qER-Quant | |
| Segmentation by deep learning (supervised voxel classification with Convolutional Neural Networks) | Yes | Yes | Same |
| Limited to analysis of imaging data | Yes | Yes | Same |
| Scanner Manufacturer of Input Data | Multi-vendor | Multi-vendor | Same |
| Output | - Annotated DICOM Images with volumetric information of brain structures and midline shift | - Annotated DICOM Images with volumetric information of structures and midline shift - Multiple electronic reports (PDF) | **Similar** The subject device offers annotated DICOM images only without producing additional PDF reports. |
| Communication | DICOM Compatible | DICOM Compatible | Same |
| Reference Standard for Performance testing | Accuracy Manual labeled images for all structures | Accuracy Manually labeled images for all structures Reproducibility Test-retest images | **Similar** The subject device provides tests for accuracy. |
| Deployment | SOMARIS-X platform, syngo.via platform | Proprietary qure.AI Platform | **Similar** The predicate device operates on its proprietary platform while the subject device operates on the syngo compatible host systems SOMARIS-X and syngo.via since this device is a part of the Siemens application universe. |
K260055 - Siemens Medical Solutions USA, Inc.
5
{10}
SIEMENS Healthineers
## 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Software Verification and Validation
Software documentation at an Enhanced Documentation level per FDA's Guidance Document "Content of Premarket Submissions for Software Contained in Medical Devices" issued on June 14, 2023 is included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
### Non-Clinical Testing Summary
Non-clinical testing submitted in support of substantial equivalence included software verification and validation testing, algorithm performance testing, cybersecurity testing, system integration testing, and risk management activities. Testing was conducted to verify that the device performs as intended and meets predefined acceptance criteria. All non-clinical test results supported the substantial equivalence of the device.
### Clinical Testing Summary
Clinical performance of the device was evaluated in a retrospective, stand-alone study using anonymized non-contrast head CT datasets collected from multiple clinical institutions in the United States and Europe. The study assessed the performance of the Midline Shift Quantification module and the Brain Hyperdensity Segmentation module by comparison to ground-truth measurements established by experienced board-certified neuroradiologists.
The Midline Shift Quantification module was evaluated using 300 head CT cases to assess the accuracy of automated midline shift measurements. The data contains patients from 18 to 93 years of age (average 61.1 +/-19.4 years) with 51% males and 42% females and an average ground-truth midline shift of 2.4 +/-2.6 mm.
The Brain Hyperdensity Segmentation module was evaluated using 200 head CT cases containing acute intracranial hyperdensities. Algorithm-generated segmentation and volume measurements were compared to expert-generated ground-truth segmentations. Performance was assessed using volumetric accuracy and Dice Similarity Coefficient (DSC). The data contains patients from 18 to 98 years of age (average 62.4 +/-19.3 years) with 47% males and 40% females and an average ground-truth hyperdensity volume of 21.4 +/- 33.4 ml.
The clinical study demonstrated that the device met all predefined acceptance criteria.
| | Absolute measurement difference | | Dice Similarity Coefficient |
| --- | --- | --- | --- |
| Structure (Number of Series) | Mean (standard deviation) | Median (10^{th}-90^{th} percentile) | Mean (95% confidence interval) |
| Intracranial hyperdensities (200) | 4.84 (6.85) ml | 2.32 (0.16-12.60) ml | 0.72 (0.68, 0.75) |
| Midline shift (300) | 0.87 (1.04) mm | 0.65 (0.16-1.78) mm | Not applicable |
K260055 - Siemens Medical Solutions USA, Inc.
6
{11}
SIEMENS Healthineers
In addition, subgroup performance analyses were performed across relevant patient demographics (i.e., age, gender, origin), technical acquisition parameters (i.e., scanner manufacturer, model, slice thickness, kernel, etc.), clinical aspects (i.e. location, volume, confounders), and image quality parameters to demonstrate that algorithm performance is consistent across the intended-use population. Additionally, truther variability analyses and linear regression analyses were conducted.
### Conclusion from non-clinical and clinical tests
The non-clinical testing submitted in support of substantial equivalence demonstrated that the device meets its design specifications and performs as intended under the specified conditions of use. Software verification and validation activities, performance testing, cybersecurity testing, and other applicable non-clinical evaluations provided evidence that the device is safe and effective for its intended use.
Clinical testing demonstrated that the subject device can perform effective labeling, visualization, and quantification of intracranial hyperdensities and midline shift.
Collectively, the non-clinical and clinical test results support the conclusion that the subject device is as safe and effective as the predicate device and does not raise different questions of safety and effectiveness. Therefore, the data support a determination of substantial equivalence.
### Risk Analysis
The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
K260055 - Siemens Medical Solutions USA, Inc.
7
{12}
**SIEMENS**
**Healthineers**
### Standards
Siemens hereby certifies that *syngo*.CT Brain Quantification meets the following FDA Recognized Consensus standards listed below:
| Standard | Version | Content | FDA Recognition Number (if applicable) |
| --- | --- | --- | --- |
| IEC 62304 | 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | Medical device software - Software life cycle processes [Including Amendment 1 (2016)] | 13-79 |
| NEMA PS 3.1 - 3.20 2023e | 2023 | Digital Imaging and Communications in Medicine (DICOM) Set | 12-352 |
| ISO 14971 | Third Edition 2019-12 | Application of Risk Management to Medical Devices | 5-125 |
| IEC 62366-1 | Edition 1.1 2020-06 CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability engineering to medical devices | 5-129 |
| ISO 15223-1 | Fourth edition 2021-07 | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements | 5-134 |
| ISO 20417:2021 | First edition 2021-04 Corrected version 2021-12 | Medical devices - Information to be supplied by the manufacturer | 5-135 |
### 9. Conclusion
*syngo*.CT Brain Quantification has the same intended use and similar indication for use as the predicate device. The result of all testing conducted was found acceptable to support the claim of substantial equivalence. The comparison of technological characteristics, clinical and non-clinical performance data, and software validation demonstrates that the subject device is as safe and effective when compared to the predicate device that is currently marketed for the same intended use. Siemens considers *syngo*.CT Brain Quantification to be as safe, as effective and with performance substantially equivalent to the commercially available predicate device.
K260055 - Siemens Medical Solutions USA, Inc.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.