K260051 · CorVent® Medical, Inc. · MOD · May 12, 2026 · Anesthesiology
Device Facts
Record ID
K260051
Device Name
CorVision® (91-205-Z0014)
Applicant
CorVent® Medical, Inc.
Product Code
MOD · Anesthesiology
Decision Date
May 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
CorVision® is only compatible with and intended for use exclusively with the CorVent® RESPOND® Ventilator (K241135) by qualified, trained personnel under the direction of a licensed clinician. This application wirelessly interfaces with the RESPOND® Ventilator to receive data and provides a supplemental display of ventilator parameters, settings, measurements, status, waveforms, and event log information. The application operates in read-only mode and has no ventilator settings or alarm control functions. The RESPOND® Ventilator is intended for pediatrics through adult patients weighing at least 38 kg (84 lbs). CorVision® is intended for use in hospital and institutional environments, including intra-hospital transport, long-term acute care, skilled nursing facilities, long-term care, and subacute care facilities (same as the RESPOND® Ventilator). CorVision® is intended to supplement and not replace any part of the current ventilator monitoring procedures.
Device Story
CorVision® is a mobile iOS software accessory for the CorVent® RESPOND® Ventilator; receives real-time physiological data, waveforms, and status via Bluetooth (BLE 5.x) from the ventilator; displays information on iPad (9th/10th gen, Air M3) for supplemental clinical monitoring; operates in read-only mode; no control of ventilator settings or alarm management; used in hospital/institutional environments by trained clinicians; requires physical proximity (~10m) and manual serial-number pairing; enhances clinical workflow by providing secondary visualization of ventilator parameters; does not replace primary ventilator monitoring.
Clinical Evidence
No clinical data. Evidence consists of bench testing, system validation, usability testing, and cybersecurity verification. Usability testing (summative) confirmed successful task completion by respiratory therapists with no hazardous use errors. Reliability testing included 168 hours of uninterrupted operation.
Technological Characteristics
Mobile iOS application; Bluetooth Class 1 (BLE 5.x) connectivity; AES-128 CCM link-layer encryption; read-only architecture; compatible with iPad 9th/10th gen and Air M3; software class A per IEC 62304; cybersecurity controls include iOS code signing and secure storage.
Indications for Use
Indicated for pediatrics through adult patients (≥38 kg) requiring mechanical ventilation; for use by qualified, trained personnel under clinician direction in hospital and institutional settings to provide supplemental, read-only display of ventilator data.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 12, 2026
CorVent® Medical, Inc.
Amed Ayubi
Vice President of Regulatory and Quality
4837 Amber Valley Pkwy. S Suite #2
Fargo, North Dakota 58104
Re: K260051
Trade/Device Name: CorVision® (91-205-Z0014)
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous Ventilator
Regulatory Class: Class II
Product Code: MOD
Dated: April 8, 2026
Received: April 8, 2026
Dear Amed Ayubi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K260051 - Amed Ayubi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260051 - Amed Ayubi
Page 3
Sincerely,
John S. Bender -S
for Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260051
Device Name
CorVision®
Indications for Use (Describe)
CorVision® is only compatible with and intended for use exclusively with the CorVent® RESPOND® Ventilator (K241135) by qualified, trained personnel under the direction of a licensed clinician. This application wirelessly interfaces with the RESPOND® Ventilator to receive data and provides a supplemental display of ventilator parameters, settings, measurements, status, waveforms, and event log information. The application operates in read-only mode and has no ventilator settings or alarm control functions. The RESPOND® Ventilator is intended for pediatrics through adult patients weighing at least 38 kg (84 lbs).
CorVision® is intended for use in hospital and institutional environments, including intra-hospital transport, long-term acute care, skilled nursing facilities, long-term care, and subacute care facilities (same as the RESPOND® Ventilator). CorVision® is intended to supplement and not replace any part of the current ventilator monitoring procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
510(k) Summary
K260051
| Date Prepared | March 31, 2026 |
| --- | --- |
| Submitter | CorVent® Medical Inc. 4837 Amber Valley Pkwy S, Suite #2 Fargo, ND 58104 (443) 928-9195 |
| Submitter Contact | Amed Ayubi VP of Regulatory and Quality Management Representative & PRRC P: 937-728-3704 E: aayubi@corventmedical.com |
| Device Name | CorVision® |
| Device Regulation Number | 868.5895 |
| Device Regulation Name | Continuous Ventilator |
| Device Product Code | MOD |
| Device Product Code Name | Accessory to Continuous Ventilator (Respirator) |
| Device Regulatory Class | Class II |
| Predicate Device Name | Care Cycle Connect Application – K161411 |
| Predicate Regulation Number | 868.5895 |
| Predicate Regulation Name | Continuous Ventilator |
| Predicate Product Code | MOD |
| Predicate Product Code Name | Accessory to Continuous Ventilator (Respirator) |
| Predicate Regulatory Class | Class II |
1. Indications for Use
CorVision® is only compatible with and intended for use exclusively with the CorVent® RESPOND® Ventilator (K241135) by qualified, trained personnel under the direction of a licensed clinician. This application wirelessly interfaces with the RESPOND® Ventilator to receive data and provides a supplemental display of ventilator parameters, settings, measurements, status, waveforms, and event log information. The application operates in read-only mode and has no ventilator settings or alarm control functions. The RESPOND® Ventilator is intended for pediatrics through adult patients weighing at least 38 kg (84 lbs).
CorVision® is intended for use in hospital and institutional environments, including intra-hospital transport, long-term acute care, skilled nursing facilities, long-term care, and subacute care facilities (same as the RESPOND® Ventilator). CorVision® is intended to supplement and not replace any part of the current ventilator monitoring procedures.
2. Predicate citation and legal market confirmation
Predicate device: Respironics Care Cycle Connect Application, 510(k) K161411. The predicate 510(k) decision (K161411) was cleared on 2016-10-14 and remains legally marketed in the U.S. as of the submission date of this 510(k).
3. Device Description
CorVision® is an accessory to a continuous ventilator. CorVision® is intended for use with RESPOND® Ventilator cleared by the FDA under K241135. CorVision® is a mobile iOS software accessory to the RESPOND® Ventilator. It is designed to display ventilator waveforms, parameters, settings, status and event log information in real time in the hospital and institutional - healthcare settings.
In its current implementation, CorVision® connects wirelessly to the RESPOND® Ventilator via a Bluetooth® Class 1 (BLE 5.x) interface with a nominal range of approximately 10 meters (30.8 feet). The application is compatible with iPad 9th, 10th, and Air M3 generations.
CorVision® operates in read-only mode; it transmits no commands or control data to the ventilator. All ventilator functionality, alarm management, and parameter adjustments remain exclusively under the ventilator's primary
{5}
control interface. The RESPOND® Ventilator's performance and safety functions are not altered in any way by pairing with the CorVision®.
Data is transmitted periodically from the RESPOND® Ventilator and displayed on the CorVision® for supplemental viewing by trained clinical personnel. The application is intended for use in hospital and institutional environments where the ventilator is physically present and remains the primary source of patient monitoring and alarm management.
## 4. Technology Comparison Summary
Data flow architecture: RESPOND® Ventilator → BLE (BLE 5.x, GATT profile, custom service UUID) → CorVision® (iPad application). CorVision® only subscribes to ventilator read-characteristics; no written characteristics are exposed.
Bluetooth details: BLE version: 5.x (specified), radio class: Bluetooth Class 1, nominal range: ~10 m. Pairing method: Out-of-box device serial-number confirmation + on-device visual ventilator ID. Pairing prevents connection to non-matching serial numbers.
Encryption and key management: Link-layer encryption using AES-128 CCM (BLE Security Mode 1, LE Secure Connections where available). Session keys are negotiated by BLE specification; no persistent long-term keys are stored outside the iOS Keychain. Keys are protected by iOS secure storage and application-level access control.
Failure modes and mitigations: BLE dropouts -> connection-loss warning and automatic reconnect attempts; wrong-device pairing -> serial number confirmation and visual ID; intercepted traffic -> AES-128 encryption and application-level payload validation; tampered app -> iOS code signing enforcement and app-store distribution. Comparison to predicate (Care Cycle Connect K161411):
Both use iOS platform and BLE with link-layer encryption. Differences: BLE version (predicate unspecified) and limited hospital-only usage for CorVision®. These differences do not alter the read-only architecture nor affect safety/effectiveness.
## 5. Risk Management Summary
Risk management activities were conducted in accordance with ISO 14971:2019 and AAMI TIR57, addressing system-level, software, usability, communication, and cybersecurity risks. Identified hazards included incorrect data display, BLE communication loss, wrong-device pairing, incompatible hardware/software, cybersecurity threats, and use-related risks. All hazards were mitigated through established design controls, including:
- Read-only architecture
- Serial-number confirmation and visual ventilator identification
- Encrypted BLE communication
- Connection-loss warnings
- Hardware/software compatibility enforcement
- Event logging and IFU-aligned warnings
All Risk Control Measures were verified and validated. No new hazards were introduced, and all residual risks were determined to be "Acceptable and Reduced As-Far-As-Possible (AFAP) consistent with Generally Accepted State of the Art".
Residual Risk: All identified risks were reduced as far as possible and judged acceptable based on benefit to risk analysis. Residual risks and the rationale for acceptability are documented in the Risk Management File.
## 6. Performance and Verification and Validation Testing
- FDA software guidance
- IEC 62304 Class A software
{6}
- IEC 62366-1
- Applicable human factors and cybersecurity standards
## Testing Included:
| Verification Testing | BLE pairing, encryption, data accuracy, labeling, compatibility, and design input conformance |
| --- | --- |
| System Validation | Accurate visualization across ventilation modes, BLE disconnection/recovery behavior, event synchronization |
| Usability Testing | Summative testing with Respiratory Therapists demonstrating successful completion of all critical tasks with no hazardous use errors |
| Reliability Testing | 168 hours of uninterrupted operation |
| Cybersecurity Testing | No vulnerabilities identified |
All testing demonstrated that CorVision® performs safely and effectively for its intended use.
## 7. Principle of Operation
1. RESPOND Ventilator transmits physiological and status data via BLE.
2. CorVision, as a bonded client, receives decrypts, and parses the data stream.
3. The application displays values in real-time graphical and tabular formats.
4. Users navigate between Plots, Parameters, Event Log, and System Info tabs.
5. Ventilator therapy and alarms remain unaffected by CorVision operation.
## Inputs into CorVision:
- From RESPOND Ventilator (BLE 5.x, read-only):
- Waveform data: Pressure, Flow, Volume, Leak (time-based).
- Breath-by-breath values: Tidal Volume (Vti, Vte), Respiratory Rate (RR), Minute Ventilation (MVe), etc.
- Ventilator settings: PEEP, FiO₂ %, Inspiratory Pressure, Tidal Volume, Backup settings, etc.
- Alarm settings: High/low thresholds, Apnea, Disconnect, O₂ margin etc.
- System information: Device ID, software version, serial number, operational hours, state (Run/Standby).
- From User (iPad navigation):
- Bluetooth pairing key entry during initial setup.
- Screen navigation (button presses, tab selection, waveform pause/resume, scaling, cursor movement).
## Outputs from CorVision:
- To User (on iPad display):
- Real-time waveform plots (Airway Pressure, Flow, Volume, Leak).
- Loop plots (P–V, F–V).
- Measurement and settings tiles.
- Ventilator settings.
- Alarm limits (visual only, no annunciation).
- Chronological ventilator even- log (10,000 events).
- System/device information (serial number, software build, ventilated hours).
- To RESPOND Ventilator:
- Ventilator Ping (visual identification only; no effect on therapy).
## Summary:
All I/O is one-way Ventilator → CorVision except for the optional Ping. No ventilator therapy or safety-critical controls are affected by CorVision inputs or outputs.
{7}
Accessories / Interoperability
- Compatible System: RESPOND Ventilator only.
- Optional Accessory: iPad arm and iPad securement enclosure for pairing a given iPad with a single ventilator.
Risk Controls & Cybersecurity
- Read-only design eliminates ventilator control risk.
- Bonding/authentication requires ventilator-side confirmation.
- Encryption via BLE AES-128/256 CCM.
- Hazard mitigations:
- Communication loss – ventilator remains alarm/display master. CorVision displays messages stating communication loss.
- Incorrect display – validated against ICD scaling; ventilator remains primary source.
- Wrong ventilator connection – mitigated by user confirming serial number confirmation and Ping invoked RESPOND Ventilator flashing LCD.
- iPad/app failure – ventilator unaffected; user alerted by frozen/absent data.
Continue next page
{8}
8. Predicate Device Substantial Equivalence Summary
| Characteristic | Subject Device: CorVent® Connect Application (CCAP) | Predicate Device: Care Cycle Connect (CCC) Application | Determination |
| --- | --- | --- | --- |
| Applicable Regulations | 21CFR820 Quality System Regulation 21CFR807 Subpart E – Premarket Notification 21CFR868 Anesthesiology Devices 21CFR868:5895 Continuous Ventilator | 21CFR820 Quality System Regulation 21CFR807 Subpart E – Premarket Notification 21CFR868 Anesthesiology Devices 21CFR868:5895 Continuous Ventilator | Equivalent |
| Review Panel | Anesthesiology | Anesthesiology | Equivalent |
| Device Classification | (b) Class II (Performance Standards). | (b) Class II (Performance Standards). | Equivalent |
| Classification Name | 21CFR868:5895 Continuous Ventilator | 21CFR868:5895 Continuous Ventilator | Equivalent |
| Device Identification | (a) Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device. | (a) Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device. | Equivalent |
| Performance Standards | General Controls & Special Controls | General Controls & Special Controls | Equivalent |
| Classification Code | MOD | MOD | Equivalent |
| Classification Code Name | Accessory to Continuous Ventilator (Respirator) | Accessory to Continuous Ventilator (Respirator) | Equivalent |
| Indications for Use | CorVision® is only compatible with and intended for use exclusively with the CorVent® RESPOND® Ventilator (K241135) by qualified, trained personnel under the direction of a licensed clinician. This application wirelessly interfaces with the RESPOND® Ventilator to receive data and provides a supplemental display of ventilator parameters, settings, measurements, status, waveforms, and event log information. The application operates in read-only mode and has no ventilator settings or alarm control functions. The RESPOND® Ventilator is intended for pediatrics through adult patients weighing at least 38 kg (84 lbs). CorVision® is intended for use in hospital and institutional environments, including intra-hospital transport, long-term acute care, skilled nursing facilities, long-term care, and subacute care facilities (same as the RESPOND® Ventilator). CorVision® is intended to supplement and not replace any part of the current ventilator monitoring procedures. | The Care Cycle Connect software application is intended for use with Trilogy Series ventilators by both caregivers and clinicians. The application pairs with the Trilogy device via a Bluetooth connection. The application provides the caregiver remote patient monitoring, and alarm surveillance. Alarm surveillance consists of both an audible tone and a visible alert if an alarm condition exists. The application provides the clinician with the ability to view, collect and store patient ventilator usage data. Care Cycle Connect (CCC) also provides educational information on ventilator use to both caregivers and clinicians. It is intended to be used in the home, and hospital/ institutional settings. The Care Cycle connect application is intended to supplement and not replace any part of the current device monitoring procedures | Equivalent 1) The CCAP is not intended to control or display alarms in real time (only through event log summary). 2) The CCAP is not intended to provide clinicians with educational information on ventilator use. 3) The CCAP is not intended to be used in the home environment. The differences in indication for use do not impact function and do not raise different questions of safety and/or effectiveness. Therefore, the intended use of the CCAP is equivalent to that of the predicate device: to provide a supplemental display of ventilator data without controlling ventilation. |
| User Population | Same as the RESPOND® Ventilator (cleared by FDA under K241135) is intended for use by qualified, trained personnel under the direction of a licensed clinician. | Clinicians (respiratory and non-respiratory) and physicians. | Equivalent, Both devices are intended for use by trained healthcare professionals (respiratory therapists, clinicians) under clinical supervision. |
| Hardware Install Platform | iPad 9®, 10® and/or Air M3 Generation (via Bluetooth direct connection) | Apple iPad; iPad mini (via Bluetooth direct connection) | Equivalent |
| Viewable Ventilator Patient Data | | | |
| Ventilator Settings | Yes | Yes | Equivalent |
| Ventilator/Patient Waveforms | Yes | Yes | Equivalent |
| Device Logs | Yes | Yes | Equivalent |
| Alarms | Yes (through event log summary) | Yes | Equivalent |
5
{9}
6
| Characteristic | Subject Device: CorVent® Connect Application (CCAP) | Predicate Device: Care Cycle Connect (CCC) Application | Determination |
| --- | --- | --- | --- |
| | | | The difference in patient data does not impact function and does not raise different questions of safety and/or effectiveness. |
| Application Features | | | |
| Calendar | No | Yes | Equivalent The difference in calendar feature does not impact function and does not raise different questions of safety and/or effectiveness. |
| Patient Information Record | No (CCAP does not support any entry of patient data. Nor is any patient information available from the RESPOND® Ventilator for display). | Yes | Equivalent The difference in patient information does not impact function and does not raise different questions of safety and/or effectiveness. |
| Note Taking | No (CCAP does not support any entry of data) | Yes | Equivalent The difference in note taking features does not impact function and does not raise different questions of safety and/or effectiveness. |
| Educational Reference | Yes, limited to CCAP usability (i.e., IFU, Privacy Policy, and Supplemental display). | Yes, Care Cycle Connect (CCC) Application provides its IFU, Privacy Policy, and Supplemental display information and also provides educational information on ventilator use to both caregivers and clinicians. | Equivalent The difference in education does not impact function and does not raise different questions of safety and/or effectiveness. |
| Software Application Platform | Apple iOS | Apple iOS | Equivalent Both devices use iOS platform security controls and Bluetooth LE with AES-128 encryption; cybersecurity verification. |
| Hardware Install Platform | iPad 9®, 10® and/or Air M3 Generation (via Bluetooth direct connection) | Apple iPad; iPad mini (via Bluetooth direct connection) | Equivalent The difference in hardware platforms does not impact function and does not raise different questions of safety and/or effectiveness. |
## Substantial Equivalence Discussion
The table above compares the key features of the proposed subject device with the identified predicate device. The following decision points are per FDA guidelines:
**Decision 1** – Is the predicate device legally marketed? Yes.
**Decision 2** – Do the devices have the equivalent intended use? Yes differences do not raise different or new questions of safety and/or effectiveness.
Rationale for differences in Indications and Features:
- **Home-use exclusion**: The subject device (CorVision®) is limited to hospital and institutional environments because pairing requires physical presence and manual serial-number confirmation with the RESPOND® ventilator; there is no remote cloud connectivity or persistent patient location tracking. This limitation reduces exposure to uncontrolled home networks and does not
{10}
change the intended use of providing a supplemental, read-only ventilator display. Supporting evidence: BLE range limitation testing (see V&V traceability table) and labeling warnings (IFU Section X).
- Alarm functionality: CorVision® operates in read-only mode and does not provide real-time audible alarm control. Alarms are viewable only via event-log summaries and transient waveform/parameter displays. The predicate's alarm surveillance features include audible alerts; CorVision® lack of real-time alarm control is mitigated by design (ventilator maintains primary alarm responsibility) and does not introduce new safety questions. Supporting evidence: architecture diagrams, usability results demonstrating clinician reliance on ventilator primary interface.
- Educational and patient data features: CorVision® provides only device IFU and supplemental display help (limited educational content) and does not permit patient data entry or storage. This reduces the risk of misattributed patient data and does not change the device's fundamental intended purpose as a read-only display. Conclusion: Differences are limited to implementation and environment restrictions and do not raise new questions of safety and effectiveness via FDA Decision 2 guidance.
Decision 3 – Do the devices have equivalent technological characteristics? Yes differences do not raise different or new questions of safety and/or effectiveness.
CorVision® and the predicate both present ventilator data in a read-only fashion via iOS BLE connections and use link-layer encryption. Differences in BLE version, platform generation, and environmental use (hospital-only vs. predicate home/hospital) are implementation differences that do not raise new questions of safety or effectiveness.
Based on the review and assessment of table above Subject Device vs Predicate Device Comparison Analysis, Determinations, and the answers to decisions 1, 2, and 3 are YES, the Respironics Care Cycle Connect (CCC) Application is a valid "Predicate Device", and the CorVent® Connect Application (CCAP) is considered to be "Substantially Equivalent to the Care Cycle Connect (CCC) Application ("The Predicate")" and differences do not raise different or new questions of safety and/or effectiveness.
Continue next page
{11}
9. Conclusion
The CorVision® software is a read-only accessory intended to display ventilator data and does not alter the intended use, performance, alarm behavior, or control functions of the RESPOND® Ventilator. Differences between the subject device and the predicate are limited to implementation details that do not raise new questions of safety or effectiveness.
Comprehensive risk management, verification, validation, usability, reliability, and cybersecurity testing demonstrate that CorVision® meets all applicable design and performance requirements and performs as intended in its clinical environment.
Based on equivalent intended use, comparable technological characteristics, and satisfactory performance data, CorVision® is substantially equivalent to the Care Cycle Connect App predicate device.
End of Document
8
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.