← Product Code [MQC](/productcode/MQC) · K260031

# THERMEO® I MSI (K260031)

_Pro3Dure Medical GmbH · MQC · Sep 29, 2026 · DE · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K260031

## Device Facts

- **Applicant:** Pro3Dure Medical GmbH
- **Product Code:** [MQC](/productcode/MQC.md)
- **Decision Date:** Sep 29, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE

## Indications for Use

The THERMEO® I MSI blank is a disc intended for the fabrication of dental appliances such as mouthguards, nightguards, bruxism and TMJ splint appliances, and bite splints using CAD/CAM milling machines.

## Device Story

THERMEO® I MSI is a pre-manufactured disc blank used for the fabrication of dental appliances; specifically mouthguards, nightguards, bruxism/TMJ splints, and bite splints. The device is processed by dental professionals using CAD/CAM milling machines. It is composed of a methyl methacrylate-, BPA-, and silicone-free material. The device serves as the raw material substrate for custom-milled dental prosthetics. Clinical benefit is derived from the patient-specific fit and therapeutic function of the resulting dental appliance.

## Clinical Evidence

Bench testing only. Performance testing conducted per ISO 20795-2 (flexural modulus, flexural strength, tensile strength, elongation at break). Biocompatibility testing conducted per ISO 10993-1. Shelf-life testing conducted per ASTM F1980-16.

## Technological Characteristics

Disc blank for CAD/CAM milling. Material is methyl methacrylate-, BPA-, and silicone-free. Performance testing per ISO 20795-2. Biocompatibility per ISO 10993. Shelf life of 2 years validated per ASTM F1980-16.

## Predicate Devices

- Thermeo System ([K212017](/device/K212017.md))

## Reference Devices

- GR Resin System MSI ([K231775](/device/K231775.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG
ADMINISTRATION

September 29, 2026

Pro3dure Medical GmbH
% Patricia Kontoudis
Regulatory Specialist
Kontoudis Consulting, LLC
804 Umbra St.
Baltimore, Maryland 21224

Re: K260031

Trade/Device Name: Thermeo® I MSI

Regulatory Class: Unclassified

Product Code: MQC

Dated: August 14, 2026

Received: August 14, 2026

Dear Patricia Kontoudis:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K260031 - Patricia Kontoudis

Page 2

Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}

K260031 - Patricia Kontoudis

Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA

Assistant Director

DHT1B: Division of Dental and
ENT Devices

OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices

Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260031 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  THERMEO® I MSI  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The THERMEO® I MSI blank is a disc intended for the fabrication of dental appliances such as mouthguards, nightguards, bruxism and TMJ splint appliances, and bite splints using CAD/CAM milling machines.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

### 3. 510(k) Summary - K260031

Device Trade Name: THERMEO® I MSI
Manufacturer: pro3dure Medical GmbH
Am Burgberg 13
58642 Iserlohn Germany

Contact: Mr. Frank Gischer
Phone: +49 (0)2374 920050-14
Fax: +49 (0)2374 920050-50
Frank.gischer@pro3dure.com

Prepared by: Ms. Patricia Kontoudis
Regulatory Affairs Specialist
Kontoudis Regulatory Consulting, LLC 804 Umbra Street
Baltimore, MD 21224
Phone: (443)722-0126
Email: tricia@kontoudisconsulting.com

Date Prepared January 5, 2026

Classification Unclassified

Product Code MQC

Predicate Device Thermeo System (K212017)

Reference Device GR Resin System MSI (K231775)

Indications for Use: The THERMEO® I MSI blank is a disc intended for the fabrication of dental appliances such as mouthguards, nightguards, bruxism and TMJ splint appliances, and bite splints using CAD/CAM milling machines.

Device Description: The THERMEO® I MSI blanks are pre-made by pro3dure using the two-part mixable device. The THERMEO® I MSI blank is intended to be used in conjunction with CAD/CAM milling technology systems.

Performance Testing: Performance testing for the THERMEO® I MSI was performed in accordance with ISO 20795-2 including Flexural modulus, Flexural Strength, Tensile Strength, and Elongation at Break.

Biocompatibility: Biocompatibility testing was conducted in accordance with ISO 10993-1.

{5}

# **Shelf-Life:**

The shelf life of the THERMEO® I MSI is 2 years. Testing was performed in accordance with ASTM F1980-16.

# **Conclusion:**

The THERMEO® I MSI is substantially equivalent in indications, technical characteristics, function, material, performance, biocompatibility, and shelf life to the predicate devices.

# **Comparison to Predicate Device:**

|  Device | Subject Device: THERMEO® I MSI | Primary Predicate Device: Thermeo System (blank) (K212017)  |
| --- | --- | --- |
|  Manufacturer | Pro3dure Medical GmbH | Pro3dure Medical GmbH  |
|  Indications for use | The THERMEO® I MSI blank is a disc intended for the fabrication of dental appliances such as mouthguards, nightguards, bruxism and TMJ splint appliances, and bite splints using CAD/CAM milling machines. | The Thermeo blank is a disc intended for the fabrication of dental appliances such as mouthguards, nightguards, bruxism and TMJ splint appliances, and bite splints using CAD/CAM milling machines.  |
|  Chemical Description | Methyl methacrylate-, BPA- and silicone-free material | Methyl methacrylate-, BPA- and silicone-free material  |
|  Manufacturing Technology | CAD/CAM milling technology systems. | CAD/CAM milling technology systems.  |
|  Product State | Blank | Blank  |
|  Performance Testing | ISO 20795-2 | ISO 20795-2  |
|  Biocompatibility | ISO 10993 | ISO 10993  |

---

**Source:** [https://fda.innolitics.com/device/K260031](https://fda.innolitics.com/device/K260031)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
