Vitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK)

K254261 · CELL STORE (BEIJING) BIOTECHNOLOGY CO., LTD. · MQL · Aug 28, 2026 · Obstetrics/Gynecology

Device Facts

Record IDK254261
Device NameVitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK)
ApplicantCELL STORE (BEIJING) BIOTECHNOLOGY CO., LTD.
Product CodeMQL · Obstetrics/Gynecology
Decision DateAug 28, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2

Indications for Use

The Vitrification Freeze Kit (Model. CS-IVF-FK) is indicated for use in the vitrification of oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. The Thawing Kit (Model. CS-IVF-TK) is indicated for use in the thawing of vitrified oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos.

Device Story

Vitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK) consist of six media solutions for cryopreservation and warming of oocytes and embryos. Freeze kit uses equilibration and vitrification solutions containing ethylene glycol and sucrose to replace water in cells prior to liquid nitrogen storage. Thawing kit uses four stepwise solutions (TS, DS1, DS2, DS3) to remove cryoprotectants. Used in clinical IVF laboratories by embryologists. Output is preserved/thawed oocytes or embryos for subsequent reproductive procedures. Benefits include successful cryopreservation of reproductive cells for future use.

Clinical Evidence

Bench testing only. Includes aseptic filtration validation, radiation sterilization validation, shelf-life testing (accelerated aging per ASTM F1980-21), transportation testing (ASTM D4169-22), and Mouse Embryo Assay (MEA) per FDA guidance (≥ 80% development to expanded blastocyst at 96 hours).

Technological Characteristics

Media contain water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, polymeric substance, gentamicin, ethylene glycol, and sucrose. Provided in polypropylene vials. Sterilized via aseptic filtration; vials sterilized by radiation (ISO 11137). Shelf-life 6 months at 2-8°C. No software or electronic components.

Indications for Use

Indicated for vitrification and thawing of MII oocytes, day 3 cleavage stage embryos, and blastocyst stage embryos. Prescription use only.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Submission Summary (Full Text)

{0} [LOGO] FDA U.S. FOOD & DRUG ADMINISTRATION August 28, 2026 Cell Store (Beijing) Biotechnology Co., Ltd. Kang Zhang RA Director Rm. 201 B2 Diamond Center, #66 Xixiaokou Rd. Haidian District Beijing, 100192 CHINA Re: K254261 Trade/Device Name: Vitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK) Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and supplements Regulatory Class: II Product Code: MQL Dated: July 20, 2026 Received: July 21, 2026 Dear Kang Zhang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K254261 - Kang Zhang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K254261 - Kang Zhang Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Monica D. Garcia -S Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K254261 Device Name Vitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK) Indications for Use (Describe) The Vitrification Freeze Kit (Model. CS-IVF-FK) is indicated for use in the vitrification of oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. The Thawing Kit (Model. CS-IVF-TK) is indicated for use in the thawing of vitrified oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} K254261 Page 1 of 5 # 510(k) Summary K254261 # 1. Submitter Information Applicant: CELL STORE (BEIJING) BIOTECHNOLOGY CO., LTD. Address: Room 101, 5th Floor, Building 10, Yard 6, Dongsheng Science Park North Street, Haidian District, Beijing, China # 2. Submission Correspondent Contact: Kang Zhang Role: RA Director Company: CELL STORE (BEIJING) BIOTECHNOLOGY CO., LTD. Address: Room 101, 5th Floor, Building 10, Yard 6, Dongsheng Science Park North Street, Haidian District, Beijing, China Phone: +86 13811621417 Email: zhangkang@cellstorebio.com 3. Date prepared: August 27, 2026 # 4. Device Information | Device Name: | Vitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK) | | --- | --- | | Common Name: | Assisted Reproduction Media | | Regulation Number: | 21 CFR 884.6180 | | Regulation Name: | Reproductive Media and Supplements | | Product Code: | MQL (Media, Reproductive) | | Regulatory Class: | Class II | # 5. Predicate Device Information | Device Name: | Vitrification Kit and Thawing Kit | | --- | --- | | 510(k) Number: | K171748 | | Sponsor: | Kitazato Corporation. | The predicate device has not been subject to a design-related recall. # 6. Device Description The Vitrification Freeze Kit and Thawing Kit (CS-IVF-FK; CS-IVF-TK) are composed of a set of six media to vitrify and warm MII oocytes, day 3 cleavage stage embryos, and blastocyst stage embryos. The Vitrification Freeze Kit includes two media, equilibration solution and vitrification solution, containing the cryoprotectants ethylene glycol, sucrose. During the vitrification process, the oocytes and embryos are first exposed to the equilibration solution and then to vitrification solution. Using this methodology, the permeating cryoprotectants can replace water in the oocytes and embryos in the cleavage stage and blastocyst stage prior to vitrification and storage in the liquid nitrogen. The Vitrification Freeze Kit comes prepackaged {5} K254261 Page 2 of 5 with one 4.5ml vial of equilibration solution (ES) and one 4.5ml vial of vitrification solution (FS). The Thawing Kit is composed of four media used stepwise for thawing and removing cryoprotectants from vitrified oocytes, cleavage stage embryos, and blastocyst stage embryos. The Thawing Kit is composed of TS (Thawing Solution), DS1 (Dilution Solution 1), DS2 (Dilution Solution 2), and DS3 (Dilution Solution 3). The Thawing kit comes prepackaged one 4.5ml vial thawing solution, one 4.5ml vial dilution solution 1, one 4.5ml vial dilution solution 2, and one 4.5ml vial dilution solution 3. The base medium for the vitrification freeze and thawing media includes water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, polymeric substance, and gentamicin. Cryoprotectants in the media include ethylene glycol, and sucrose. The Six solutions in the Vitrification Freeze Kit and Thawing Kit are aseptically filtered (storage vials sterilized by radiation) and provided in polypropylene vials. They have a shelf-life of 6 months when stored at 2-8°C. # 7. Indications for Use Statement The Vitrification Freeze Kit (Model. CS-IVF-FK) is indicated for use in the vitrification of oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. The Thawing Kit (Model. CS-IVF-TK) is indicated for use in the thawing of vitrified oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. # 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device The table below includes a comparison of the intended use and technological characteristics of the subject and predicate devices. | | **Vitrification Freeze Kit and Thawing Kit K254261 (Subject Device)** | **Vitrification Kit and Thawing Kit K171748 (Predicate Device)** | **Comparison** | | --- | --- | --- | --- | | Regulation Number | 21 CFR 884.6180 | 21 CFR 884.6180 | Same | | Product Code | MQL | MQL | Same | | Classification | Class II | Class II | Same | | Indications for Use | The Vitrification Freeze Kit (Model. CS-IVF-FK) is indicated for use in the vitrification of oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. The Thawing Kit (Model. CS-IVF-TK) is indicated for use in the thawing of vitrified oocytes (MII), day 3 cleavage stage embryos, and blastocyst stage embryos. | The Vitrification Kit is indicated for use in the preparation, vitrification, and storage of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos. The Thawing Kit is indicated for use in the preparation and thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage | **Different:** There are differences in the wording of the indications for use statements for the subject and predicate device; however, the intended uses of the subject and predicate device media are the same (i.e., vitrification and warming of oocytes and embryos). | {6} K254261 Page 3 of 5 | | | embryos, and blastocyst stage embryos. | | | --- | --- | --- | --- | | Conditions for Use | Prescription Use Only | Prescription Use Only | Same | | Kit components | The Vitrification Freeze Kit includes Equilibration Solution (ES) and Vitrification Solution (FS). The Thawing Kit includes Thawing Solution (TS) and Dilution Solution 1 (DS1), Dilution Solution 2 (DS2), Dilution Solution 3 (DS3). | The Vitrification Kit includes three media components, Basic Solution (BS), Equilibration Solution (ES) and Vitrification Solution (VS). The Vitrification Kit comes prepackaged with 4 Cryotop devices (Cryotop CL, Cryotop SC, or Cryotop US), and 2 Repro Plates. The Thawing Kit is composed of TS (Thawing Solution), DS (Dilution Solution) and WS (Wash Solution). The Thawing Kit comes pre-packaged with one Repro Plate, and two 35 mm dishes. | **Different:** The predicate device includes an additional basic solution component, equipment to conduct the procedure (plates), and a closed cryopreservation storage device that are different than the subject device. These differences do not raise different questions of safety and effectiveness (S&E). | | Freeze Kit Formulation | Base medium (water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, polymeric substance, gentamicin) with Poly(vinyl alcohol), Human albumin ,Cryoprotectants Ethylene Glycol and Sucrose | Medium 199, Ethylene Glycol, DMSO, Sucrose, Dextran Serum Supplement, Gentamicin | **Different:** The formulations of the subject and predicate devices are not the same. Differences in device formulations do not raise different questions of S&E. | | Thaw Kit Formulation | Base medium (water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, polymeric substance, gentamicin) with Poly(vinyl alcohol), Human albumin ,Cryoprotectant Sucrose | Medium 199; Sucrose, Dextran Serum Supplement, Gentamicin | **Different:** The formulations of the subject and predicate devices are not the same. Differences in device formulations do not raise different questions of S&E. | | pH | 7.2 – 7.6 | 7.2 – 7.6 | Same | | Osmolality (mOsm/kg) | Equilibration Solution:1500-1900 (1:1 dilution) Vitrification Solution: 2045-2445 (1:2 dilution) | Not available publicly | **Different:** The subject device and predicate devices have differences in osmolality specifications. These differences in osmolality | {7} K254261 Page 4 of 5 | | Thawing Solution: 1546-1946 Dilution Solution 1: 860-960 Dilution Solution 2: 530-610 Dilution Solution 3: 275-300 | | specifications do not raise different questions of S&E. | | --- | --- | --- | --- | | Bacterial Endotoxin | <0.25 EU/mL | ≤ 0.25 EU/mL | **Similar** | | Mouse Embryo Assay (MEA) | 1-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours. | 1-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours. | Same | | Sterilization method | Aseptic Filtration | Aseptic Filtration | Same | | Storage | 2 - 8°C | 2 - 8°C | Same | | Shelf-life | 6 months | 1 year | **Different:** Differences in shelf-life do not raise different questions of S&E. | As shown in the table above, there are differences in the indications for use statements and technological characteristics of the subject and predicate devices including differences in vitrification and thaw kit formulation, specifications, and shelf life. However, as stated in the table above, the differences in indications for use do not represent a new intended use and the differences in technological characteristics do not raise different questions of safety and effectiveness. ### 9. Summary of Non-Clinical Performance Testing The following studies have been conducted in support of the substantial equivalence to the predicate device. - Aseptic filtration and aseptic filling validation, per ISO 13408-1:2023 and ISO 13408-2:2018. - Vial radiation sterilization, per ISO 11137-1:2025 Sterilization of health care products — Radiation — Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2:2013 - Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose [Including Amendment 1 (2022)]. - Shelf-life testing was conducted to support a 6-month shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after accelerated aging in accordance with ASTM F1980-21: i. Appearance: Uniform color, clear and transparent, free of particles and precipitates ii. pH per USP <791>: 7.2 – 7.6 for all solutions {8} K254261 Page 5 of 5 iii. Osmolality per USP <785>: Equilibration Solution:1500-1900 (1:1 dilution), Vitrification Solution: 2045-2445 (1:2 dilution), Thawing Solution: 1546-1946, Dilution Solution 1: 860-960, Dilution Solution 2: 530-610, Dilution Solution 3: 275-300 iv. Sterility per USP <71>: No microbial growth v. Bacterial endotoxin per USP <85>: < 0.25 EU/mL vi. MEA per the 2021 FDA guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices: 1-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours. - Transportation testing per ASTM D4169-22 and cap/seal leak testing using a method equivalent to USP <1207.2> on transportation-conditioned devices. ### 10. Conclusion The results of the performance testing described above demonstrate that the subject devices are as safe and effective as the predicate device and support a determination of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...