K254246 · Zhejiang Longterm Medical Technology Co., Ltd. · NAD · Sep 10, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254246
Device Name
LT Antibacterial Foam with Silver Dressing
Applicant
Zhejiang Longterm Medical Technology Co., Ltd.
Product Code
NAD · General, Plastic Surgery
Decision Date
Sep 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4020
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
LT Antibacterial Foam with Silver Dressing is indicated for management acute and chronic wounds with moderate to heavy exudate, including stage II, III and IV pressure ulcers, leg ulcers and diabetic foot ulcers. The dressing can also be used for partial thickness burns (including first- and superficial second-degree burns), donor sites, post operative wounds and skin abrasions. LT Antibacterial Foam with Silver Dressing is intended for external use only.
Device Story
Advanced wound care dressing; porous polyurethane foam structure; silver-impregnated for antimicrobial activity; activated carbon for odor control; optional polyurethane film backing. Absorbs and retains wound exudate; maintains moisture balance; silver inhibits bacterial growth (4-log activity for up to 7 days). Used in clinical or home settings; applied by healthcare professionals or patients. Output is physical wound coverage and exudate management; aids healing by maintaining optimal environment and reducing bacterial load.
Clinical Evidence
Bench testing included antimicrobial performance (AATCC 100), physical performance (EN 13726-1, ASTM D3574), and biocompatibility (ISO 10993). Animal testing: GLP-compliant porcine wound healing study compared subject device to Biatain Ag Foam and dry gauze control; demonstrated comparable wound healing with no adverse tissue response.
Technological Characteristics
Materials: Polyurethane film, silver-impregnated polyurethane foam, activated carbon. Energy: None. Form factor: Multi-size foam sheets with/without film backing. Sterilization: Gamma irradiation (ISO 11137-2). Connectivity: None. Mechanical properties: Water absorbency (EN 13726-1), wet tensile strength (ASTM D3574).
Indications for Use
Indicated for management of acute and chronic wounds with moderate to heavy exudate in patients with stage II, III, and IV pressure ulcers, leg ulcers, diabetic foot ulcers, partial thickness burns (first- and superficial second-degree), donor sites, post-operative wounds, and skin abrasions. Intended for external use only.
Regulatory Classification
Identification
An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gases such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing. This classification does not include an occlusive wound dressing that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 10, 2026
Zhejiang Longterm Medical Technology Co., Ltd.
Claudia Zsang
Director of Regulatory Affairs
Shuangshan Rd. 277, Fuxi St.
Deqing, Zhejiang 313200
China
Re: K254246
Trade/Device Name: LT Antibacterial Foam with Silver Dressing
Regulatory Class: Unclassified
Product Code: FRO
Dated: August 8, 2026
Received: August 10, 2026
Dear Claudia Zsang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254246 - Claudia Zsang
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K254246 - Claudia Zsang
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MUSTAFA A.
MAZHER -S
For Yu-Chieh Chiu, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K254246
Device Name
LT Antibacterial Foam with Silver Dressing
Indications for Use (Describe)
LT Antibacterial Foam with Silver Dressing is indicated for management acute and chronic wounds with moderate to heavy exudate, including stage II, III and IV pressure ulcers, leg ulcers and diabetic foot ulcers. The dressing can also be used for partial thickness burns (including first- and superficial second-degree burns), donor sites, post operative wounds and skin abrasions.
LT Antibacterial Foam with Silver Dressing is intended for external use only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
510(k) #: k254246
# 510(k) Summary
Prepared on: 2026-08-08
## Contact Details
21 CFR 807.92(a)(1)
Applicant Name
Zhejiang Longterm Medical Technology Co., Ltd
Applicant Address
ShuangShan Road 277, Fuxi Street Deqing Zhejiang 313200 China
Applicant Contact Telephone
1-647-991-3316
Applicant Contact
Ms. Claudia Zsang
Applicant Contact Email
claudia.zsang@gmail.com
## Device Name
21 CFR 807.92(a)(2)
Device Trade Name
LT Antibacterial Foam with Silver Dressing
Common Name
Foam dressing with silver
Classification Name
Occlusive wound dressing
Regulation Number
Unclassified
Product Code(s)
FRO
## Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K100218
Biatain Ag Foam Dressing
FRO
## Device Description Summary
21 CFR 807.92(a)(4)
LT Antibacterial foam with Silver Dressing is an advanced wound care dressing and is available in two configurations: with or without a polyurethane film backing. Multiple sizes are available to accommodate various wound dimensions.
The porous structure provides rapid absorption and retention of wound exudate, helping to maintain an optimal moisture balance. The activated carbon incorporated in the foam absorbs wound odor. Upon contact with wound fluid, the silver inhibits bacterial growth within the dressing. The soft, elastic foam conforms to wound contours.
The presence of silver can provide 4-log antibacterial activity for up to 7 days within the dressing against gram-negative and gram-positive bacteria, such as Klebsiella pneumoniae, Pseudomonas Aeruginosa, Escherichia coli, Streptococcus epidermidis, Enterococcus faecium and Staphylococcus aureus.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
LT Antibacterial Foam with Silver Dressing is indicated for management acute and chronic wounds with moderate to heavy exudate, including stage II, III and IV pressure ulcers, leg ulcers and diabetic foot ulcers. The dressing can also be used for partial thickness burns (including first- and superficial second-degree burns), donor sites, post operative wounds and skin abrasions.
LT Antibacterial Foam with Silver Dressing is intended for external use only.
{5}
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device has the following indications for use, which are similar to those of the predicate device (Biatain Foam Silver, K100218):Indications for Use:LT Antibacterial Foam with Silver Dressing is indicated for management acute and chronic wounds with moderate to heavy exudate, including stage II, III and IV pressure ulcers, leg ulcers and diabetic foot ulcers. The dressing can also be used for partial thickness burns (including first- and superficial second-degree burns), donor sites, post operative wounds and skin abrasions. The device is intended for external use only.
# Technological Comparison
21 CFR 807.92(a)(6)
Both devices share similar indications for use, fundamental scientific technology, major materials composition, and performance characteristics.
Indications for Use: Both the subject and predicate devices are indicated for the management of moderate to heavily exuding acute and chronic wounds, including pressure ulcers (Stage II, III, and IV), leg ulcers, diabetic foot ulcers, donor sites, postoperative wounds, skin abrasions, and partial-thickness burns. Both devices are intended for external use only and provide antimicrobial activity within the dressing.
Technological Characteristics: The subject device is composed of Polyurethane film and silver-impregnated polyurethane foam with activated carbon for odor control, and the predicate consists of PU film and silver-impregnated PU foam. Both devices utilize the same antimicrobial agent (silver), and the same base material (polyurethane foam). The subject device is available in two configurations—with and without polyurethane film backing—to accommodate different wound types, identical to the predicate device's product line. The configuration with film backing provides a semi-permeable barrier to prevent exudate strike through. These design features and clinical applications are identical to those of the predicate device.
The minor difference in material composition, specifically the inclusion of activated carbon, does not impact the foam's ability to absorb wound exudate or the antibacterial activity of silver within the dressing. Activated carbon is a well-established safe material with a history of safe use in wound dressings, where it is commonly incorporated for wound odor management.
Performance Characteristics: Comparative performance testing demonstrates that the subject device performs equivalently to the predicate device across all critical performance parameters, including absorption capacity, fluid retention under pressure, silver elution rate, antimicrobial efficacy, and tensile strength.
Biocompatibility evaluation: The biocompatibility evaluation for the LT Antibacterial Foam Dressing with Silver was conducted in accordance with the FDA Guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices-part1: Evaluation and testing within a risk management process'", both devices are categorized as prolonged contact with breached/ compromised skin.
Sterility: Both devices are terminally sterilized using gamma irradiation in accordance with ISO 11137. The devices are packaged in sealed pouches that maintain sterility throughout the labeled shelf life.
The LT Antibacterial Foam with Silver Dressing is substantially equivalent to the predicate device Biatain Foam Silver (K100218) based on similar indications for use, the similar fundamental scientific technology, comparable materials composition, equivalent performance characteristics, and similar biocompatibility profiles. The minor differences in material composition and sizes do not alter the device's fundamental technological characteristics or raise different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device and is as safe and effective as the legally marketed predicate for its intended use.
# Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Antimicrobial Performance Testing - Pass
- Antimicrobial Activity (AATCC 100)
- Minimum Effective Concentration (AATCC 100)
- Silver Content (AAS)
- Silver Elution (AAS)
Physical Performance Testing - Pass
- Water Absorbency (EN 13726-1)
- Fluid Retention
- Wet Tensile Strength (ASTM D3574:2008)-
- pH Value (USP <791>, ISO 10993-12)
Animal Testing
A GLP-compliant porcine wound healing study compared the subject device to the predicate device (Biatain Ag Foam) and a dry gauze control. The subject device demonstrated wound healing comparable to the predicate device with adequate performance and no evidence of delayed healing or adverse tissue response.
{6}
# Sterilization and Packaging
- Sterilization Method (ISO 11137-2:2013, VDmax25)
- Sterility Assurance Level (ISO 11137-2:2013)
- Package Seal Strength (ASTM F88/F88M-23)
- Package Integrity (ASTM F3039-15)
# Biocompatibility Testing - Pass
The LT Antibacterial Foam with Silver Dressing was evaluated for biocompatibility per ISO 10993-1 requirements for a surface device with prolonged contact. Testing included:
- Cytotoxicity (ISO 10993-5:2009, MEM Elution)
- Sensitization (ISO 10993-10:2010, Guinea Pig Maximization Test)
- Irritation (ISO 10993-10:2010, Intracutaneous)
- Implantation (ISO 10993-6:2016, 4-week intramuscular)
- Acute Systemic Toxicity (ISO 10993-11:2017)
- Subacute/Subchronic Toxicity (ISO 10993-11:2017, 28-day rat study)
- Pyrogenicity (ISO 10993-11:2017 / USP <151>)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.