K254238 · Hans Biomed Corporation · NEW · Sep 18, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254238
Device Name
MINT MONO Sutures
Applicant
Hans Biomed Corporation
Product Code
NEW · General, Plastic Surgery
Decision Date
Sep 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MINT MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate. The suture is not intended for lifting or filling aesthetic uses.
Device Story
MINT MONO Sutures are synthetic, absorbable, sterile, monofilament surgical sutures made of polydioxanone (PDO). They are provided in various lengths and USP sizes (7-0, 6-0, 5-0) with stainless steel (SUS 304) needles or cannulas. Used by clinicians for soft tissue approximation; the device is implanted and degrades/resorbs over time. The device is not intended for aesthetic lifting or filling procedures. Substantial equivalence is established through non-clinical performance testing, including biocompatibility, tensile strength, and sterilization validation, confirming the device meets USP standards and performs similarly to existing absorbable polydioxanone sutures.
Clinical Evidence
No clinical data was necessary to determine substantial equivalence. Evidence is based on bench testing, including biocompatibility (ISO 10993 series), sterilization validation (ISO 11135-1), shelf-life (ASTM F88/F88M, ASTM F1929), and mechanical performance (USP <881> tensile strength, dimension verification, degradation profile, and cannula testing).
Technological Characteristics
Synthetic absorbable monofilament polydioxanone (PDO) suture. Stainless steel (SUS 304) needle/cannula. Sterilized via ethylene oxide. Compliant with USP <881> for tensile strength and USP standard suture sizes. Implantable, absorbable, contact duration >30 days.
Indications for Use
Indicated for soft tissue approximation in patients requiring an absorbable suture. Not indicated for lifting or filling aesthetic uses.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 18, 2026
Hans Biomed Corporation
% Sarah Fitzgerald
Senior Consultant
Emergo by UL
2500 Bee Caves Rd.
Bldg. 1, Suite 300
Austin, Texas 78746
Re: K254238
Trade/Device Name: MINT MONO Sutures
Regulation Number: 21 CFR 878.4840
Regulation Name: Absorbable Polydioxanone Surgical Suture
Regulatory Class: Class II
Product Code: NEW
Dated: August 14, 2026
Received: August 14, 2026
Dear Sarah Fitzgerald:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K254238 - Sarah Fitzgerald
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K254238 - Sarah Fitzgerald
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254238
Device Name
MINT MONO Sutures
Indications for Use (Describe)
MINT MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate.
The suture is not intended for lifting or filling aesthetic uses.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
K254238
# 510(k) Summary -K254238
## MINT® MONO Sutures
### 1. Submission Sponsor
Hans Biomed Corporation
64, Yuseong-Daero 1628 Beon-Gil
Yuseong-Gu, Daejeon 34054 Korea
Contact: Mark Bae
Title: Regulatory Affairs Deputy Manager
Telephone: 82-02-466-2266
Email: mhbae@hansbiomed.com
### 2. Submission Correspondent
Emergo by UL
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327-9997
Email: LST.US.EmergoFDASubmissions@ul.com
Contact: Sarah Marie Fitzgerald
Title: Senior Consultant
### 3. Date Prepared
September 18, 2026
### 4. Device Identification
Trade/Proprietary Name: MINT MONO Sutures
Common/Usual and Classification Name: Polydioxanone absorbable surgical suture
Regulation Number(s): 878.4840
Product Code(s): NEW
Class: Class II
Classification Panel: General & Plastic Surgery
### 5. Legally Marketed Predicate Device
Device name: DemeDIOX Absorbable Surgical Suture
510(k) number: K181582
Manufacturer: Demetech Corporation
Page 1 of 4
{6}
K254238
# 6. Device Description
MINT® Product Family Sutures, including MINT® MONO Sutures, are synthetic, absorbable, sterile, surgical sutures comprised of polydioxanone. They are available in a range of lengths and thicknesses. They are provided sterile. They degrade and resorb over time.
MINT® MONO Sutures are smooth sutures. MINT® MONO Sutures are part of the MINT® Product Family of Sutures.
# 7. Indication for Use Statement
MINT® MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate.
The suture is not intended for lifting or filling aesthetic uses.
# 8. Substantial Equivalence Discussion
MINT® MONO Sutures are equivalent to DemeDIOX Absorbable Surgical Sutures cleared in K181582. MINT® Sutures (K220549) and PDREX Sutures (K173779) are provided as reference devices.
The following table compares the MINT® MONO Sutures to the predicate device and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
Comparison of Characteristics
| Attribute | Subject MINT® MONO Sutures | Predicate DemeDIOX Sutures K181582 | Comparison |
| --- | --- | --- | --- |
| Product Code / Regulation | NEW / 878.4840 | NEW / 878.4840 | Same |
| Intended Use | Absorbable surgical sutures for use in soft tissue | Absorbable surgical sutures for use in soft tissue | Same |
| Indications for Use | MINT® MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate. The suture is not intended for lifting or filling aesthetic uses. | DemeDIOX Absorbable Surgical Suture is indicated for use in general soft tissue approximation including pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery, but not for use in adult cardiovascular, microsurgery and neural tissue. These sutures are useful where absorbable sutures with extended | Similar Provides fewer details compared to predicate but equivalent details compared to reference devices. |
Page 2 of 4
{7}
K254238
| Attribute | Subject MINT® MONO Sutures | Predicate DemeDIOX Sutures K181582 | Comparison |
| --- | --- | --- | --- |
| | | wound support (up to six weeks) is desirable. | |
| Suture Material | Polydioxanone | Polydioxanone | Same |
| Suture Characteristics | Synthetic absorbable monofilament | Synthetic absorbable monofilament | Same |
| Sterilization | Ethylene oxide | Ethylene oxide | Same |
| Size (USP) | Compliant with USP Standard Suture Sizes | Compliant with USP Standard Suture Sizes | Same |
| Contact Type | Implant, Absorbable | Implant, Absorbable | Same |
| Contact Duration | >30 days | >30 days | Same |
| Suture Diameter | 7-0, 6-0, and 5-0 USP sizes | 6-0 USP size | Similar Within range of reference devices. |
| Suture Tensile Strength | Compliant with USP <881> requirements | Compliant with USP <881> requirements except smallest is too short to perform knot pull test | Same |
| Accessories | Needle / Cannula | Needle / Cannula | Same |
| Accessory Material (Patient Contacting) | Stainless Steel (SUS 304) | Stainless Steel | Same |
This submission is to add MINT® MONO Sutures, which are smooth, to cleared sutures in the MINT® Product Family of Sutures. All previously cleared MINT® Sutures are barbed.
### 9. Non-Clinical Performance Data
Testing was conducted in alignment with the FDA guidance “Class II Special Controls Guidance Document: Surgical Sutures,” confirming that the design output meets the design input and specifications for the sutures.
Testing conducted included biocompatibility evaluation and testing per ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, ISO 10993-23, sterilization validation in accordance with ISO 11135-1, shelf-life testing in accordance with ASTM F88/F88M and ASTM F1929, tensile strength testing to confirm USP <881> monograph strength, dimension verification (length, diameter / gauge), degradation profile / degradation, residual tensile strength, cannula elasticity, cannula bending strength, and cannula pull-out testing.
The results of non-clinical performance testing demonstrates substantial equivalence between the subject and predicate device.
### 10. Animal Performance Data
No animal testing was conducted for this submission.
Page 3 of 4
{8}
K254238
# 11. Clinical Performance Data
No clinical data was necessary to determine the substantial equivalence of this device.
# 12. Statement of Substantial Equivalence
All testing and evidence confirms that the subject device does not raise any new questions related to the safety or effectiveness and that the device is at least as safe and at least as effective as the predicate device for the intended use and indications for use. Therefore, the device can be considered substantially equivalent to the predicate device.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.