PulseAPI SmartWatch is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects (when prescribed or used under the care of a physician). The device supports analyzing data recorded in compatible formats from ECG devices from lead I or lead II recorded using wet-electrodes, or compatible dry-electrode devices. The device is not intended for use in patients who have pacemakers or ICDs or other implanted electronic devices. The device is not intended for use in life supporting, or sustaining systems, or ECG monitors, or as part of an alarm system.
Device Story
PulseAPI SmartWatch is a software analysis library integrated into third-party device software; processes ECG waveform data and metadata from lead I/II wet-electrode or compatible dry-electrode devices (e.g., Apple Watch, Samsung Galaxy, Fitbit, Withings, KardiaMobile). Operates as a programmable diagnostic computer; performs noise filtering, heart rate measurement, rhythm analysis (sinus, AFib, bradycardia, tachycardia), QRS detection, and beat classification (PVC/non-PVC). Used under physician care; provides automated analysis to support clinical diagnosis. Benefits include automated rhythm assessment and beat classification to assist clinicians in evaluating ambulatory ECG data.
Clinical Evidence
Bench testing only. Performance evaluated using standard reference datasets (ANSI/AAMI EC57) and additional clinical datasets containing wet-electrode and model-specific dry-electrode ECGs. Testing verified algorithm performance for noise filtering, heart rate measurement, rhythm analysis, and beat classification across diverse populations and device types. Results demonstrated substantial equivalence to the predicate.
Technological Characteristics
Software analysis library; programmable diagnostic computer (Class II). Inputs: Lead I/II ECG waveforms/metadata. Processing: Proprietary automated ECG algorithms for noise filtering, rhythm analysis, and beat classification. Connectivity: API-based integration into third-party device software. Compatible with various wet and dry-electrode hardware.
Indications for Use
Indicated for adult patients requiring assessment of ambulatory ECG rhythms. Contraindicated for patients with pacemakers, ICDs, or other implanted electronic devices. Not for use in life-supporting/sustaining systems, ECG monitors, or alarm systems.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Omron Model BP7900 Blood Pressure Monitor + EKG (K182579)
Submission Summary (Full Text)
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
October 27, 2026
PulseAI, Ltd.
Alan Kennedy
CEO
58 Howard Street
Belfast, United Kingdom BT1 6PL
Re: K254226
Trade/Device Name: PulseAPI SmartWatch
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK
Dated: October 26, 2026
Received: October 26, 2026
Dear Alan Kennedy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K254226 - Alan Kennedy
Page 2
When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Office of Industry and Consumer Education (OICE) to ask a question about a specific regulatory topic. See the OICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K254226 - Alan Kennedy
Page 3
assistance/contact-us-office-industry-and-consumer-education-oice) for more information or contact OICE by email (AskOICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254226 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| PulseAPI SmartWatch | | |
| Please provide your Indications for Use below. | | ? |
| PulseAPI SmartWatch is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects (when prescribed or used under the care of a physician). The device supports analyzing data recorded in compatible formats from ECG devices from lead I or lead II recorded using wet-electrodes, or compatible dry-electrode devices. The device is not intended for use in patients who have pacemakers or ICDs or other implanted electronic devices. The device is not intended for use in life supporting, or sustaining systems, or ECG monitors, or as part of an alarm system. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
510(k) Notification K254226
# GENERAL INFORMATION [807.92(a)(1)]
Applicant:
PulseAI Ltd.
58 Howard Street
Belfast
United Kingdom
BT1 6PL
Phone: +447714640144
Applicant Contact Person:
Rebecca Funston
Senior Consultant, RAQA
58 Howard Street
Belfast
United Kingdom
BT1 6PL
Phone: +447795546264
Email: rebecca@pulseai.io
Date Prepared: October 9th, 2026
# DEVICE INFORMATION [807.92(a)(2)]
Trade Name:
PulseAPI SmartWatch
Generic/Common Name:
Programmable Diagnostic Computer
Classification:
21 CFR§870.1425, Computer, Diagnostic, Programmable, Class II
Product Code:
DQK
Subsequent Product Code(s):
N/A
{6}
510(k) SUMMARY
K254226
# **PREDICATE DEVICE(S) [807.92(a)(3)]**
K201985 – AliveCor KardiaAI V2
# **DEVICE DESCRIPTION [807.92(a)(4)]**
PulseAPI SmartWatch is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects (when prescribed or used under the care of a physician). The PulseAPI SmartWatch software receives ECG waveform data and associated metadata from lead I or lead II recorded using wet-electrodes, or compatible dry-electrode devices. The software is intended to be integrated into other device software.
When used with wet-electrode ECGs from lead I and II and dry-electrode Apple Watch series 6 it provides the following automated analysis:
- Filtering ECG noise
- Reporting heart rate measurement from ECGs
- Detecting noisy ECGs
- Reporting ECG rhythm analysis for the presence of sinus rhythm, atrial fibrillation, bradycardia, and tachycardia
- ECGs detected as sinus rhythm, detecting normal sinus rhythm and sinus rhythm with premature ventricular contractions (PVC)
- Detecting QRS complexes in an ECG, For ECGs detected as sinus rhythm, classifying individual beats as a PVC or non-PVC beat and generating an average beat from an ECG
When used with ECGs from other compatible dry-electrode hardware (i.e., Samsung Galaxy Watch, Fitbit Sense, Withings ScanWatch, AliveCor KardiaMobile and KardiaMobile 6L, Lead I only) it provides the following automated analysis:
- Filtering ECG noise,
- Reporting heart rate measurement from ECGs,
- Detecting noisy ECGs,
- Reporting ECG rhythm analysis for the presence of sinus rhythm and atrial fibrillation
- Generating an average beat from an ECG
# **INDICATIONS FOR USE [807.92(a)(5)]**
PulseAPI SmartWatch is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects (when prescribed or used under the care of a physician). The device supports analyzing data recorded in compatible formats from ECG devices from lead I or lead II recorded using wet-electrodes, or compatible dry-electrode devices.
The device is not intended for use in patients who have pacemakers or ICDs or other implanted electronic devices. The device is not intended for use in life supporting, or sustaining systems, or ECG monitors, or as part of an alarm system.
Page 2 of 6
{7}
510(k) SUMMARY
K254226
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE[807.92(a)(6)]
| Feature | PulseAPI SmartWatch | KardiaAI V2 |
| --- | --- | --- |
| 510(k) | K254226 | K201985 |
| Regulation Number(s) | 21 CFR 870.1425 | 21 CFR 870.1425 |
| Regulation Name | • Programmable Diagnostic Computer | • Programmable Diagnostic Computer |
| Regulatory Class | Class II | Class II |
| Classification Product Code | DQK | DQK |
| Subsequent Product Code(s) | N/A | DPS |
| Over The Counter | No | No |
| Indications for use | PulseAPI SmartWatch is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects (when prescribed or used under the care of a physician). The device supports analyzing data recorded in compatible formats from ECG devices from lead I or lead II recorded using wet-electrodes, or compatible dry-electrode devices. The device is not intended for use in patients who have pacemakers or ICDs or other implanted electronic devices. The device is not intended for use in life supporting, or sustaining systems, or ECG monitors, or as part of an alarm system. | KardiaAI is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects (when prescribed or used under the care of a physician). The device supports analyzing data recorded in compatible formats from any ambulatory ECG devices such as event recorders, or other similar devices. The library is intended to be integrated into other device software. The library is not intended for use in life supporting, or sustaining systems, or ECG monitors, or cardiac alarm, or OTC use only devices. The KardiaAI library provides the following capabilities: Filtering ECG noise, Reporting heart rate measurement from ECGs, Detecting noisy ECGs, Reporting ECG rhythm analysis for the presence of sinus rhythm, atrial fibrillation, bradycardia, and tachycardia; for ECGs detected as sinus rhythm, detecting normal sinus rhythm, sinus rhythm with wide QRS, sinus rhythm with premature ventricular contractions (PVC), and sinus rhythm with supraventricular ectopy, Detecting QRS complexes in an ECG, For ECGs detected as sinus rhythm, classifying individual beats as a PVC or non-PVC beat, and |
Page 3 of 6
{8}
510(k) SUMMARY
K254226
| Feature | PulseAPI SmartWatch | KardiaAI V2 |
| --- | --- | --- |
| | | generating an average beat from an ECG The device is not intended for use in patients who have pacemakers, ICDs, or other implanted electronic devices |
| Software Functionalities | An interface that provides tools to process and analyze ECGs through various algorithms. The automated proprietary ECG algorithms provide supportive information for ECG diagnosis. The library can be accessed by directly connecting to the PulseAPI Smartwatch's Application Programming Interface The PulseAPI SmartWatch library provides the following capabilities: Filtering ECG noise, Reporting heart rate measurement from ECGs, detecting noisy ECGs, and provide the following analysis: Wet electrodes and compatible dry-electrode devices (Apple Watch Series 6*) - Reporting ECG rhythm analysis for the presence of sinus rhythm, atrial fibrillation, bradycardia*, and tachycardia*; for ECGs detected as sinus rhythm, detecting normal sinus rhythm and sinus rhythm with premature ventricular contractions (PVC)*, Detecting QRS complexes in an ECG, For ECGs detected as sinus rhythm, classifying individual beats as a PVC* or non-PVC beat, and generating an average beat from an ECG. Compatible dry-electrode devices: (i.e., Samsung Galaxy Watch, Fitbit Sense, Withings ScanWatch, AliveCor KardiaMobile and KardiaMobile 6L, Lead I only) it provides the following automated analysis: • Filtering ECG noise, • Reporting heart rate | An interface that provides tools to process and analyze ECGs through various algorithms The automated proprietary ECG algorithms provide supportive information for ECG diagnosis. The library can be accessed by directly connecting to the KardiaAI's Application Programming Interface |
Page 4 of 6
{9}
510(k) SUMMARY
K254226
| Feature | PulseAPI SmartWatch | KardiaAI V2 |
| --- | --- | --- |
| | measurement from ECGs, • Detecting noisy ECGs, • Reporting ECG rhythm analysis for the presence of sinus rhythm and atrial fibrillation • Generating an average beat from an ECG | |
| Compatible dry-electrode ECG Devices | Apple Watch Series 6* • Samsung Galaxy Watch • Fitbit Sense • Withings ScanWatch • AliveCor KardiaMobile and KardiaMobile 6L, Lead I only | • Triangle System (K183319) KardiaMobile System (K182396) • KardiaBand System (K171816) • Omron Model BP7900 Blood Pressure Monitor + EKG (K182579) |
### SUBSTANTIAL EQUIVALENCE
The PulseAPI SmartWatch subject device has the same intended use and similar technological characteristics as the KardiaAI V2 predicate device. The differences in technological characteristics have been analyzed and addressed through performance testing. The evaluation and testing results showed that differences between the subject and predicate device do not raise different questions of safety or effectiveness.
### PERFORMANCE DATA [807.92(b)]
All necessary performance testing was conducted on the PulseAPI Smartwatch to support a determination of substantial equivalence to the predicate device.
### [807.92(b)(1)] Nonclinical Testing Summary:
Nonclinical testing, consistent with that performed for the predicate device, was conducted to evaluate algorithm performance and verify that PulseAPI SmartWatch functions as intended. Analyses were conducted using standard reference datasets, including those specified in ANSI/AAMI EC57. Additional clinical datasets were used to support performance assessment including wet-electrodes ECGs and model-specific dry-electrode ECGs. All analysis outputs supported the claims of substantial equivalence. The results confirm that PulseAPI SmartWatch performs according to its design specifications and intended use, supporting a determination of substantial equivalence to the predicate device.
### SUMMARY
The results of the validation testing demonstrate that the PulseAPI SmartWatch Software is
Page 5 of 6
{10}
510(k) SUMMARY
K254226
substantially equivalent to the predicate device.
Across multiple independent datasets comprising segments from diverse populations and device types, the algorithm consistently demonstrated substantially equivalent performance. The PulseAPI SmartWatch subject device has similar intended use and technological characteristics as the KardiaAI V2 predicate device. The evaluation and testing results showed that any differences between the subject and predicate device do not raise different questions of safety or effectiveness. The PulseAPI SmartWatch is substantially equivalent to the predicate device.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.