← Product Code [JOW](/productcode/JOW) · K254217

# Aerloom (K254217)

_Evolym, LLC · JOW · Aug 31, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K254217

## Device Facts

- **Applicant:** Evolym, LLC
- **Product Code:** [JOW](/productcode/JOW.md)
- **Decision Date:** Aug 31, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: • Chronic edema • Lymphedema • Venous insufficiency • Wound healing

## Device Story

Aerloom is a portable, battery-powered pneumatic compression system for clinical or home use. It consists of a controller and multi-chambered garment worn over a liner. The controller inflates/deflates garment chambers in a distal-to-proximal sequence to apply graduated compression (20–80 mmHg), facilitating fluid movement from impaired lymphatic regions. The device features Bluetooth connectivity for an optional mobile application interface. It is operated by patients or clinicians to manage edema and promote wound healing. The system provides sequential calibrated gradient pressure, mirroring the therapeutic action of the predicate device.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including pressure verification, biocompatibility (ISO 10993), electrical safety (ANSI/AAMI ES 60601-1), and electromagnetic compatibility (IEC 60601-1-2).

## Technological Characteristics

Pneumatic compression system; controller with rechargeable Li-ion battery; 7-chambered garment (nylon/polyester/polyurethane); pressure range 20-80 mmHg. Connectivity via Bluetooth Low Energy (BLE). Compliant with ANSI/AAMI ES 60601-1, IEC 60601-1-11, and IEC 60601-1-2. Software/firmware controlled.

## Regulatory Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

## Predicate Devices

- entré, Model PD08-U ([K143185](/device/K143185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**U.S. FOOD & DRUG**
ADMINISTRATION

August 31, 2026

Evolym, LLC
Alex Chang
Regulatory Consultant
1209 Orange St.
Wilmington, Delaware 19801

Re: K254217

Trade/Device Name: Aerloom
Regulation Number: 21 CFR 870.5800
Regulation Name: Compressible Limb Sleeve
Regulatory Class: Class II
Product Code: JOW
Dated: December 24, 2025
Received: December 29, 2025

Dear Alex Chang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

K254217 - Alex Chang

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K254217 - Alex Chang

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Meaghan Erlewein -S

For Nicole Gillette

Assistant Director

DHT2B: Division of Circulatory Support,

Structural, and Vascular Devices

OHT2: Office of Cardiovascular Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254217 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Aerloom  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: • Chronic edema • Lymphedema • Venous insufficiency • Wound healing  |   |   |
|  Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |
|  Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ?  |

{4}

Evolym LLC

Aerloom 510(k) Submission

K254217

# 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

I. SUBMITTER: Evolym LLC
1209 Orange St. Wilmington, DE 19801 USA
Establishment Registration: 33-4391469

CONTACT: Alex Chang
Regulatory Consultant
P: 408-839-5826
E-mail: info@evolym.com

DATE PREPARED: December 23 2025

## II. DEVICE:

TRADE NAME: AERLOOM
CLASSIFICATION NAME: Compressible limb sleeve
DEVICE CLASSIFICATION: CLASS II
PRODUCT CODE: JOW

## III. PREDICATE DEVICE:

Manufacturer: Tactile Systems Technology Inc.
Trade Name: entré, Model PD08-U
510(k): K143185

510(k) Summary

Page 1

{5}

Evolym LLC

Aerloom 510(k) Submission

# IV. DEVICE DESCRIPTION:

The Aerloom system consists of two primary components: a controller and a multi-chambered compression garment. The system delivers sequential pneumatic compression therapy to the affected limb to support the treatment of chronic edema, lymphedema, venous insufficiency, and wound healing.

The garment is worn over lightweight clothing or a liner to avoid direct skin contact and is wrapped securely around the treatment area. It contains independently controlled air chambers that inflate and deflate in a programmed sequence, applying graduated compression in the range of 20–80 mmHg.

The controller, which houses the compression logic and pump system, is compact, portable, and powered by a rechargeable lithium-ion battery, allowing for limited mobility and untethered use during treatment. Aerloom is designed to be easy to use and suitable for use in both clinical and home settings under medical supervision.

# V: INDICATIONS FOR USE:

The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions:

- Chronic edema
- Lymphedema
- Venous insufficiency
- Wound healing

510(k) Summary

Page 2

{6}

Evolym LLC

Aerloom 510(k) Submission

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The Aerloom device and the predicate entré device are pneumatic compression devices that use the same fundamental technology, mode of action, and operating principles.

|  Feature | Subject Device (Aerloom) | Predicate Device (entré)  |
| --- | --- | --- |
|  Indications for Use | The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - Chronic edema - Lymphedema - Venous insufficiency - Wound healing | The entré System is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - Chronic edema - Lymphedema - Venous insufficiency - Wound healing  |
|  Product Type | Pneumatic Compression Device | Pneumatic Compression Device  |
|  Components | Controller Garments Hoses Power Adaptor | Controller Garments Hoses Power adapter  |
|  Mode of Action | Helps direct and move excess fluid from impaired lymphatic regions to healthier areas, where it can be naturally absorbed and processed by the body. | Helps direct and move excess fluid from impaired lymphatic regions to healthier areas, where it can be naturally absorbed and processed by the body.  |
|  Principles of Operation | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal, using a pressure gradient that delivers higher pressure distally. Each chamber is inflated in sequence and held at pressure, after which all chambers are deflated simultaneously. | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal, using a pressure gradient that delivers higher pressure distally. Each chamber is inflated in sequence and held at pressure, after which all chambers are deflated simultaneously.  |
|  Environment of Use | Hospital, clinical, and home settings | Hospital, clinical, and home settings  |
|  Expected Service Life | 5 Years | 5 Years  |
|  Controller unit size and weight | 5.7" x 1.5" x 1.6" 0.75 lbs | 11.0" x 6.0" x 8.0" 4 lbs  |
|  Electrical Requirements | Rechargeable Li-ion Battery Pack, with max 1.00-0.60 A input from 90-264 VAC | 100-240 VAC 50/60 Hz to AC Adapter with output voltage of 12.0V DC and 3.0A  |

510(k) Summary

Page 3

{7}

Evolym LLC

Aerloom 510(k) Submission

|   | 50/60 Hz to AC Adapter, with output voltage of 5.0V DC and 5.0A |   |
| --- | --- | --- |
|  Controller Enclosure Material | Plastic | Plastic  |
|  User Interface | Pushbuttons and a mobile application interface for use on a mobile device. | Pushbuttons  |
|  Wireless Functions | Bluetooth Low Energy (BLE) | None  |
|  Software / Hardware | Analog and digital electronic with microprocessor | Analog and digital electronic with microprocessor  |
|  Software Concern Level | Basic Documentation Level | Moderate Level of Concern  |
|  Device Total Pressure Range | 20-80 mmHg | 20-80 mmHg  |
|  Output | Sequential calibrated gradient Pressure | Sequential calibrated gradient Pressure  |
|  Treatment Time | 1 hour maximum | 1 hour maximum  |
|  Garment-Controller Connection | Hose attached to garments | Hose attached to garments  |
|  Garment Port | Plastic push to connect | Plastic push to connect  |
|  Garment Material | Nylon and polyester fabric Polyurethane coated polyester and nylon fabric | Nylon and polyester fabric Polyurethane coated polyester and nylon fabric  |
|  Garment Chambers | 7 | 8  |
|  Device Cleanability | Can be cleaned using commonly available mild household cleaners and disinfectants. | Can be cleaned using commonly available mild household cleaners and disinfectants.  |

Technological differences from the predicate device, include:

- Bluetooth connectivity for optional use with a mobile application.
- An internal battery to support portability.
- A smaller controller to support portability.

These differences do not raise new questions of safety or effectiveness for the Aerloom system compared to the predicate device.

510(k) Summary

Page 4

{8}

Evolym LLC

Aerloom 510(k) Submission

### VII. PERFORMANCE DATA:

The following performance data were provided in support of substantial equivalence determination.

#### Biocompatibility Testing

The Aerloom device can be supplied with either upper or lower garments that are applied over the provided circular-knit liners. These liners are biocompatible and have been tested in accordance with ISO 10993-1, ISO 10993-5, and ISO 10993-10. All patient-contacting materials are compliant with applicable biocompatibility safety standards.

#### Sterilization & Shelf-life Testing

The subject device is non-sterile, and components are unlikely to deteriorate with age.

#### Electrical Safety and Electromagnetic Compatibility (EMC)

The subject device was evaluated based on the following applicable performance and safety standards: o ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) [Including Amendment 2 (2021)], IEC 60601-1-11:2015, EN 60601-1-2:2015/A1:2021, BS EN 60601-1-2:2015+A1:2021, and IEC 60601-1-2:2014/AMD1:2020. Results demonstrated that the subject device was compliant to all applicable performance and safety standards.

#### Software Verification and Validation Testing

The subject device includes embedded firmware in the controller which has the ability to connect with custom software installed on a mobile device. The system software exhibits a moderate level of concern and because there is no risk of death or serious injury, this corresponds to the Basic recommended documentation level.

Software lifecycle planning and documentation as well verification and validation testing were performed in accordance with IEC 62304:2015 and as recommended by the following FDA Guidance documents for Industry and FDA Staff:

- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
- Guidance for Industry and Food and Drug Administration Staff Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- Guidance for Industry, FDA Reviewers and Compliance on Postmarket Management of Cybersecurity in Medical Devices

#### Mechanical Testing

The subject device was evaluated based on the following benchtop performance tests.

|  Performance Study | Overview | Status  |
| --- | --- | --- |
|  Packaging Transportation Simulation | ASTM D4169-23e1 | Completed / Pass  |
|  Pressure Verification Testing | Pressure capable of delivering 20-80 mmHg compression pressures. | Completed / Pass  |

#### Animal Study

Animal performance testing was not required to demonstrate safety and effectiveness of the Aerloom system.

510(k) Summary

Page 5

{9}

Evolym LLC

Aerloom 510(k) Submission

# Clinical Study

Clinical testing was not required to demonstrate the safety and effectiveness of the Aerloom system.

# VIII. CONCLUSION:

The data included in this submission demonstrates that the Aerloom system is substantially equivalent to the primary predicate device, entré (K143185).

510(k) Summary

Page 6

---

**Source:** [https://fda.innolitics.com/device/K254217](https://fda.innolitics.com/device/K254217)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
