K254206 · Ningbo Zhishan Medical Technology Co., Ltd. · ITI · Mar 27, 2026 · Physical Medicine
Device Facts
Record ID
K254206
Device Name
Power wheelchair (ZS-EW8026)
Applicant
Ningbo Zhishan Medical Technology Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Mar 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Battery-powered four-wheeled vehicle; provides mobility for disabled or elderly users. Components include frame, lithium battery, brushless DC motors, seat, back cushion, armrests, foot pedal, and anti-tip devices. Operated via joystick controller; direct drive to rear wheels. Features automatic electromagnetic braking system; brakes engage when joystick is released. Used in indoor and outdoor environments; patient-operated. Provides battery status and abnormal condition displays via controller interface. Benefits include improved mobility and transportation for individuals with limited physical capacity.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including static/dynamic stability, brake effectiveness, energy consumption, obstacle climbing, and electromagnetic compatibility (EMC) per ISO 7176 series and IEC 60601 standards.
Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 27, 2026
Ningbo Zhishan Medical Technology Co., Ltd.
% Ariel Xiang
Official Correspondent
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Bldg., 1500# Century Ave.
Shanghai,
China
Re: K254206
Trade/Device Name: Power wheelchair (ZS-EW8026)
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: December 29, 2025
Received: December 29, 2025
Dear Ariel Xiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254206 - Ariel Xiang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254206 - Ariel Xiang
Page 3
Sincerely,
Digitally signed by
MARY S. KESZLER
Date: 2026.03.27
10:49:28 -04'00'
for, Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254206 | |
| Device Name Power wheelchair (ZS-EW8026) | |
| Indications for Use (Describe) It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
# 510(k) Summary
Document Prepared Date: 2025/12/02
## Applicant Information:
Ningbo Zhishan Medical Technology Co., Ltd.
Address: Linmu Chang, Yunlong Town, Yinzhou, Ningbo, Zhejiang, China
Contact person: Stone Zhang
Tel: 0086 15968993997
Mail: stone@jindaomedical.com
## Submission Correspondent:
Primary Contact: Ariel Xiang
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Central Ave., Shanghai 200122, China
Tel: 0086-21-58817802
Email: shouqiu.xiang@sungoglobal.com
## Subject Device:
Trade name: Power wheelchair
Classification name: Powered wheelchair
Model: ZS-EW8026
Regulatory Information
Classification: Class II
Product code: ITI
Regulation Number: 890.3860
{5}
Review Panel: Physical Medicine
## Substantial Equivalence Information:
Zhejiang Cleisman Industry and Trade Co.,LTD
Device Name: Electric Wheelchair
Models: ZH-W001, ZH-W002, ZH-W003
510(K): K250158
## Device Description
The product is intended only carry one person and used as a means of transportation for the disabled, the sick and the infirm.
The power wheelchair is classified as class A and the maximum occupant mass is 150kg.
The power wheelchair is a battery powered four wheeled vehicle.
It consists one Lithium battery with an off-board battery charger, frame, controller, motors, seat, back cushion, control system, armrest, two rear wheels, two casters (front wheels), foot pedal, anti-tip devices.
The power wheelchair can easily fold and unfold for transportation or storage.
## Indication for Use
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
{6}
# 1. Product Parameter
Table 1 General Comparison
| Elements of Comparison | Subject device (K254206) | Predicate Device (K250158) | Remark |
| --- | --- | --- | --- |
| Manufacturer | Ningbo Zhishan Medical Technology Co., Ltd | Zhejiang Cleisman Industry and Trade Co.,LTD | -- |
| Common or Usual name | Power Wheelchair | Electric Wheelchair | -- |
| Model(s) | ZS-EW8026 | ZH-W001,ZH-W002,ZH-W003 | -- |
| Classification | Class II | Class II | Same |
| Classification regulation | 21 CFR890.3860 | 21 CFR890.3860 | Same |
| Product code | ITI | ITI | Same |
| Indications for use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | S.E. |
| Intended user | disabled or elderly person limited to a seated position. | disabled or elderly person limited to a seated position. | S.E. |
| Use condition | indoor and outdoor use | indoor and outdoor use | S.E |
| Number of wheels | 6,including two front wheels, two rear wheels,two anti-tips | 6,including two front wheels, two rear Wheels,two anti-tips | S.E |
{7}
| Function of wheels | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | S.E |
| --- | --- | --- | --- |
| Movement control method | By Joystick control | By Joystick control | S.E |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | S.E |
| Brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | S.E |
| Braking distance | 0.7m | ZH-W001:≤1.2m
ZH-W002:≤1.0m
ZH-W003:≤1.0m | Minor difference on braking distance will not cause different performance. Shorter distance for braking will be more safety. |
| Maximum safe operational incline degree | 10° | 10° | S.E |
| Main frame material | Aluminium alloy | ZH-W001:aluminium alloy
ZH-W002:aluminium alloy
ZH-W003:carbon steel | S.E |
| Armrest | PUR | PUR | S.E |
| Back cushion | Nylon | Nylon | S.E |
{8}
| Seat cushion | Nylon | Nylon | S.E |
| --- | --- | --- | --- |
| Overall length | 990mm | ZH-W001:1130mm
ZH-W002:1020mm
ZH-W003:1020mm | Minor difference on wheelchair dimension
will not cause different performance. All safety and performance have been validated with the maximum rated weight dummy. |
| Overall width | 650mm | ZH-W001:660mm
ZH-W002:600mm
ZH-W003:600mm | |
| Folded Dimension (length*width*height) | 780*660*770mm | ZH-W001:770mm*650mm*380mm
ZH-W002:760mm*600mm*330mm
ZH-W003:760mm*600mm*330mm | |
| Front wheel size/type | 8"/PU Solid tire | ZH-W001:200mm*50mm
ZH-W002:190mm*30mm
ZH-W003:177mm*40mm | Minor difference on dimension of wheels
will not cause different performance. |
| Rear wheel size/type | 12" PU pneumatic tyre | ZH-W001:317mm*57mm
ZH-W002:320mm*35mm
ZH-W003:272mm*40mm | |
| Max speed forward | Up to 6 km/h (1.6 m/s), adjustable | ZH-W001:1.8m/s
ZH-W002:1.6m/s
ZH-W003:1.6m/s | S. E
Same as ZH-W002 |
| Max speed backward | 0.6m/s | 0.5m/s | higher max. backward speed will not impact overall safety |
| Max loading weight | 150kg | 120kg (265lbs) | Difference on loading weight will not cause different performance.
All safety and performance have been validated with the maximum loading weight. |
{9}
| Maximum distance of travel on the fully charged battery | 15.1 km | ZH-W001:14.3 km
ZH-W002:13.8 km
ZH-W003: 13.8km | It is caused by the size of the wheel, will not cause different performance, The further away the better. |
| --- | --- | --- | --- |
| Battery | Li-ion battery pack; 24 VDC 20Ah | li-ion battery pack 24V,12Ah, | the battery capacity will impact the travel distance, which will not cause new safety and effectiveness concerns raised. |
| Battery charger | Off-board charger
Input: 110-240V, 50/60Hz, 1.5A
Output: 29.4 Vdc, 2A; | Off-board charger
Input: 100-240V, 50/60Hz, 1.5A,
Output: 24 Vdc, 2A; | S.E |
| Motor | Brushless DC motor; 24VDC; 250W; 2pcs | Brushless DC motor,
24V,250W, 2pcs | Minor difference on motor power. Both meet the require of ISO 7176-14 standard. It will not cause new safety and effectiveness concerns. |
| Electronic controller | 24V,35A | Brushless dual-drive rocker controller;40A | Similar controller is used, both the control system, including the joystick controller, the electromagnetic brakes and the user interface are similar. The joystick controls the directions and |
{10}
| | | | when the joystick is released, the powered wheelchair will slow down to stop and the brakes will automatically re-engage.
The controller also provides the battery status displaying and abnormal condition displaying.
Both of the control systems are evaluated according to standard ISO 7176-14:2008 and software validation requirement and there are no new safety and effectiveness concerns due to the difference. |
| --- | --- | --- | --- |
| Turning Radius | 900 mm | ZH-W001:938mm
ZH-W002:875mm
ZH-W003:875mm | Minor difference on turning radius will not cause new safety and effectiveness concerns. |
| Maximum obstacle climbing | 25 mm | 20mm | Minor difference on obstacle climbing will not cause new safety and effectiveness concerns. |
{11}
Table 2 safety comparison
| Item | Subject device | Predicate Device | Results |
| --- | --- | --- | --- |
| Biocompatibility | The biocompatibility of the subject device is based on the use of low-risk biocompatible materials in Attachment G of FDA's 2023 Biocompatibility Guidance. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | S.E. |
| EMC | ISO7176-21, IEC 60601-1-2, IEC 60601-4-2 | ISO7176-21 and IEC 60601-1-2 | S.E. |
| Performance | ISO7176 series | ISO7176 series | S.E. |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. |
| Item | Subject device | Predicate Device | Results |
| --- | --- | --- | --- |
| ISO7176-1 | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | S.E. |
| ISO7176-2 | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | S.E. |
| ISO7176-3 | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | S.E. |
| ISO7176-4 | The theoretical distance range has been | The theoretical distance range has been determined | S.E. |
{12}
| | determined after the testing according to the ISO 7176-4, and test results meet its design specification. | after the testing according to the ISO 7176-4, and test results meet its design specification. | |
| --- | --- | --- | --- |
| ISO7176-5 | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | S.E. |
| ISO7176-6 | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | S.E. |
| ISO7176-7 | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7. | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7. | S.E. |
| ISO7176-8 | All test results meet the requirements in Clause 4 of ISO 7176-8. | All test results meet the requirements in Clause 4 of ISO 7176-8. | S.E. |
| ISO7176-9 | The test results shown that the device under tests could continue to function according to manufacturer’s specification after being subjected to each of the testsspecified in Clause 8 of ISO 7176-9. | The test results shown that the device under tests could continue to function according to manufacturer’s specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9. | S.E. |
| ISO7176-10 | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | S.E. |
| ISO7176-11 | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11. | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11. | S.E. |
{13}
| ISO7176-13 | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved. | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved. | S.E. |
| --- | --- | --- | --- |
| ISO7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14. | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14. | S.E. |
| ISO7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO7176-15. | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15. | S.E. |
| ISO16840-10:2021 | The performance of resistance to ignition meet the requirements of ISO 16840-10:2021. | The performance of resistance to ignition meet the requirements of ISO 16840-10:2021. | S.E. |
| ISO 7176-21 | The EMC performance results meet the requirements of ISO 7176-21and IEC60601-1-2. | The EMC performance results meet the requirements of ISO 7176-21. | S.E. |
| ISO7176-25 | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25. | The performance of batteries and charger of device meet the Requirements in Clause 5and 6 of ISO 7176 -25. | S.E. |
2. Substantial Equivalence Discussion
The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3,
{14}
ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-21, ISO 7176-25, and FDA guidance Submission for Power Wheelchair.
The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/control criteria following above mentioned standards. And the test results show that the subject product is substantially equivalent to the predicate device in performance.
The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability (tipping angle), The Dynamic stability (Safe Gradient Maximum Gradient), Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to ignition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers.
The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.
## 3. Product Performance
The following performance data were provided to verify that the proposed device met all design specifications and provided support of the substantial equivalence determination.
{15}
> Risk Analysis developed in accordance with ISO 14971: 2019.
> Software validation
> ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
> ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
> ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
> ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
> ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
> ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
> ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
> ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
> ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
> ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
> ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
> ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
{16}
> ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
> ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
> ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method
> ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
> ISO 7176-22: 2014 Wheelchairs-Part 22: Set-up procedures
> ISO 7176-25: 2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs.
> Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 and IEC TR 60601-4
> ANSI C63.18-2014 American National Standard Recommended Practice for an On-Site, Ad-Hoc Test Method for Estimating Electromagnetic Immunity of Medical Devices to Radiated Radio-Frequency (RF) Emissions from RF Transmitters
> IEC 62133-2:2017+A1:2021 Secondary cells and batteries containing alkaline or other non- acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
## Biocompatibility of patient-contacting material
{17}
The biocompatibility of the subject device is based on the use of low-risk biocompatible materials in Attachment G of FDA's 2023 Biocompatibility Guidance.
## 4. Summary of clinical testing
No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate device.
## 5. Substantially Equivalency Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices, K250158 Electric Wheelchair from Zhejiang Cleisman Industry and Trade Co.,LTD.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.