K254199 · BLUECORE COMPANY Co., Ltd. · GEX · Sep 24, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254199
Device Name
PicoreMax (PicoreMax)
Applicant
BLUECORE COMPANY Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Submission Summary (Full Text)
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 24, 2026
BLUECORE COMPANY Co., Ltd.
Kwon Bu Gyeong
Official Correspondent
#402-404,34,Yarl 3-Gil, Mulgeum-Eup
Yangsan-Si, Gyeonsangnam-Do
Yangsan, KR 50652
Republic Of Korea
Re: K254199
Trade/Device Name: PicoreMax (PicoreMax)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: September 15, 2026
Received: September 15, 2026
Dear Kwon Bu Gyeong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K254199 - Kwon Bu Gyeong
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K254199 - Kwon Bu Gyeong
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU -S
Digitally signed by YAN FU -S
Date: 2026.09.24 19:41:10
-04'00'
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254199 | ? |
| Please provide the device trade name(s). | | ? |
| PicoreMax (PicoreMax) | | |
| Please provide your Indications for Use below. | | ? |
| The PicoreMax is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology and general plastic surgery as follows. | | |
| Indications for use | | |
| 532 nm Wavelength- Indicated for the removal of tattoos for Fitzpatrick skin types I – III to treat the following tattoo colors : red, yellow and orange.- Indicated for benign pigmented lesions removal for Fitzpatrick skin types I – IV. | | |
| (nominal delivered energy of 585 nm and 650 nm with the Optional Dye Laser Handpiece)- Tattoo removal (light ink: red, sky blue, green)- Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi- Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis- Skin Resurfacing for Acne Scars and Wrinkles- Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 650nm wavelength)- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength) | | |
| 1064 nm Wavelength:- Indicated for the removal of tattoos for all skin types (Fitzpatrick skin types I – VI) to treat the following tattoo colors : black, brown, green, blue and purple.- Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV.- Focus (Fractional) handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I – IV.- Focus (Fractional) handpiece, indicated for the treatment of acne scars in Fitzpatrick Skin Types II – V. | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
# 510(k) summary (510(k) #: K254199)
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
# I. Submitter Information [21 CFR 807.92(a) (1)]
| Applicant/Manufacturer Name | BLUECORE COMPANY Co., Ltd. |
| --- | --- |
| Applicant Address | #402-404, 34, Yari 3-gil, Mulgeum-eup, Yangsan-si, Gyeongsangnam-do, KR 50652, Republic of Korea |
| Contact Person | Bu Gyeong Kwon Regulatory Affairs Manager Tel.: +82 070 4870 2042 Fax: +82 055 366 4319 E-mail: kbg7881@bluecorecompany.com kbg7881@blccp.com |
| Date Prepared | July. 27. 2026 |
# II. Name of device [21 CFR 807.92 (a) (2)]
| Device Trade Name | PicoreMax |
| --- | --- |
| Common Name | PicoreMax Picosecond Nd:YAG Laser System |
| Classification Name | Power Laser Surgical Instrument |
| Regulation Number | 21 CFR 878.4810 |
| Product Code | GEX |
# III. Predicate and Reference Devices [21 CFR 807.92(a) (3)]
| Predicate #1 | K240983 |
| --- | --- |
| Predicate Trade Name | Discovery Pico; Discovery Pico Plus; Discovery Pico Derm |
| Product Code | GEX |
| Predicate #2 | K160488 |
| --- | --- |
| Predicate Trade Name | enlighten III Laser System |
| Product Code | GEX |
| Predicate #3 | K133254 |
| --- | --- |
| Predicate Trade Name | RevLite Q-Switched Nd:YAG Laser System |
| Product Code | GEX |
# IV. Device Description [21 CFR 807.92(a) (4)]
The subject device, PicoreMax, is a picosecond Nd:YAG laser system that uses Nd:YAG as the laser medium to generate laser beams at wavelengths of 532 nm and 1064 nm.
The user can select from various handpieces. Moreover, specific software controls the device functions and allows the user to select through the touchscreen display.
The system is intended to be used in user facilities such as hospitals, physicians' offices and medical spas, and its major components are the main console unit, foot switch, and individual modules.
The device is re-usable and non-sterile; instructions for cleaning its components between uses are provided in the labeling.
{6}
PicoreMax Laser System consists of:
- System console (contains the system software, power supply, and various other electronic and mechanical parts)
Operator control panel with touch-screen technology (GUI)
- Handpiece (Zoom / Collimated / Recurve S, D, SS, E / Focus 7/ Gold/ Ruby)
- Footswitch and accessories
## V. Indications for Use [21 CFR 807.92(a) (5)]
The PicoreMax is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology and general plastic surgery as follows.
### 532 nm
Indicated for the removal of tattoos for Fitzpatrick skin types I – III to treat the following tattoo colors : red, yellow and orange.
Indicated for benign pigmented lesions removal for Fitzpatrick skin types I – IV.
### (nominal delivered energy of 585 nm and 650 nm with the Optional Dye Laser Handpiece)
Tattoo removal (light ink: red, sky blue, green)
Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi
Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis
Skin Resurfacing for Acne Scars and Wrinkles
Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 650nm wavelength)
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
### 1064 nm
Indicated for the removal of tattoos for all skin types (Fitzpatrick skin types I – VI) to treat the following tattoo colors : black, brown, green, blue and purple.
Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV.
Focus (Fractional) handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I – IV.
Focus (Fractional) handpiece, indicated for the treatment of acne scars in Fitzpatrick Skin Types II – V.
{7}
## VI. Summary of technological characteristics of the device compared to the predicate [21 CFR 807.92(a)(6)]
The subject PicoreMax system shares with its predicate devices, Discovery Pico Family (K240983), Enlighten III Laser System (K160488), and RevLite (K133254), the same underlying technology and presents similar technological characteristics. All systems use laser energy delivered to the skin surface via applicators.
| | Subject Device | Predicate Device |
| --- | --- | --- |
| Product Name | PicoreMax | Discovery Pico; Discovery Pico Plus; Discovery Pico Derm |
| 510(k) Number | - | K240983 |
| Manufacturer | BLUECORE COMPANY Co., Ltd. | Quanta System Spa |
| Laser Medium | Picosecond Nd:YAG | Picosecond Nd:YAG |
| Wavelength | 1064 / 532 [nm] | 1064 / 532 [nm] |
| Pulse Energy (mJ) | 1064 nm: Max.800 [mJ] 532 nm: Max. 300 [mJ] | 1064 nm: Max. 800 [mJ] 532 nm: Max. 300 [mJ] |
| Fluence (J/cm²) | 1064 nm: Zoom Handpiece: Max. 5.6 J/cm² Recurve S Handpiece: Max. 2.2 J./ cm² Recurve D Handpiece: Max. 2.2 J/cm² Recurve SS Handpiece: Max. 2.2 J/cm² Collimated Handpiece: Max. 2.1 J/cm² 532 nm: Zoom Handpiece: Max. 2.8 J/cm² Recurve E Handpiece: Max. 1.0 J/cm² | 1064 nm: Max. 5.6 J/cm² 532 nm: Max. 2.8 J/cm² |
| Pulse Width (Pulse Duration) | 1064 nm: 450 [ps] 532 nm: 380 [ps] | 1064 nm: 450 – 800 [ps] 532 nm: 370 – 600 [ps] |
| Frequency | Max.10 [Hz] | Max.10 [Hz] |
| Spot Size | Zoom H/P: 2 – 10 [mm] Collimated H/P: 7 [mm] Recurve (Square): 6 [mm] | Round: 2 – 12 [mm] Square: 2,3,4,5,7 [mm] |
| Indication for use | The PicoreMax is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows. 532 nm - Indicated for the removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. - Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. 1064 nm - Indicated for the removal of tattoos for all skin types (Fitzpatrick skin types I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. - Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. - Focus (Fractional) handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV. - Focus (Fractional) handpiece, indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V. | The Discovery Pico Family is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows. 532 nm - Indicated for the removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. - Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. - Only with fractional handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV. 1064 nm - Indicated for the removal of tattoos for all skin types (Fitzpatrick skin types I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. - Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. - Only with fractional handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV. Only with fractional handpiece, indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V. |
| With Fractional handpiece | | |
| Spot overall Dimension | Focus 7 (mm) | Round 8 (mm) |
| Wavelength (nm) | 1064 | 1064 |
| Energy(J) Max | Focus 7: 0.283 | 1064 (ps): 0.73 |
| Fluence /Dot (J/cm²) | Focus 7: 18.39 | 1064 (ps): 18.4 |
| Dot Diameter (um) | Focus 7: 200 | 1064 (ps): 200 |
| Coverage (%) | Focus 7: 4 | 1064 (ps): 4 |
{8}
| | Subject Device | Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| Product Name | PicoreMax | enlighten III Laser System | RevLite Q-Switched Nd:YAG Laser System |
| 510(k) Number | - | K160488 | K133254 |
| Manufacturer | BLUECORE COMPANY Co., Ltd. | Cutera, Inc | Cynosure |
| Laser Medium | Picosecond Nd:YAG | Q-switched Nd:YAG laser | Q-switched Nd:YAG laser |
| Wavelength | 585 / 650 [nm] | 670 [nm] | 532 [nm] 585 [nm] (with optional dye handpiece) 650 [nm] (with optional dye handpiece) |
| Pulse Energy (mJ) | 585 nm: Max.200 [mJ] 650 nm: Max. 125 [mJ] | 670 nm: Max. 125 [mJ] | 1.6 J |
| Fluence (J/cm²) | 585 nm: Max. 1.02 J/cm² 650 nm: Max. 1.77 J/cm² | 670 nm: Max. 4.0 J/cm² | 532 nm: Max. 1- 8 J/cm² |
| Pulse Width (Pulse Duration) | 585 nm / 650 nm: 380 [ps] | 670 nm: 660 [ps] | 532 nm: 7-20 [ns] |
| Frequency | Max.2 [Hz] | Max.10 [Hz] or single shot | Max. 10 [Hz] or single shot |
| Spot Size | 585nm: 5 [mm] 650nm: 3 [mm] | 670 nm: 2, 3, 4, 5, 6 [mm] | 532 nm: 2 - 8.5 [mm] |
| Indication for use | The PicoreMax is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. 532nm (nominal delivered energy of 585 nm and 650 nm with the Optional Dye Laser Handpiece) Tattoo removal (light ink: red, sky blue, green) Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis Skin Resurfacing for Acne Scars and Wrinkles Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 650nm wavelength) Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength) | The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. 670 nm The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III. | The RevLite Q-Switched Nd:YAG Laser System consists of an electrically powered Console, in which laser energy produced within the system is delivered to the tissues by means of an articulated arm, Handpiece Adaptor and specially designed handpiece. The user activates laser emission by means of a footswitch. 532 nm (nominal delivered energy of 585 nm and 650 nm with the Optional Multilite Dye Laser Handpiece) Tattoo removal (light ink: red, sky blue, green) Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis Skin Resurfacing for Acne Scars and Wrinkles Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 650nm wavelength) Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength) |
{9}
### VII. Nonclinical Performance Testing [21 CFR 807.92(b)(1)]
IEC 60601-1 2020-08: Medical Electrical Equipment – Part 1: General Requirements for safety and essential performance
IEC 60601-1-2 2020-09: Medical Electrical Equipment 1-2 General Requirements for basic safety and essential performance
IEC 60601-2-22 – 2019-11: Medical Electrical Equipment – Part 2-22 – Particular Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
IEC 60825-1, 2007-003: Safety of laser products - Part 1: Equipment classification, and requirements IEC 10993-5, 2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity IEC 10993-10, 2021 Biological evaluation of medical devices – Part 10: Tests for skin sensitization IEC 10993-23, 2021 Biological evaluation of medical devices – Part 23: Tests for irritation
In addition, software verification and validation testing was performed to the requirements of IEC 62304 and biocompatibility conformance to FDA standards was established.
### VIII. Clinical Data [21 CFR 807.92(b) (2)]
Based on the similarities in the safety and effectiveness profiles of the subject, reference, and primary predicate devices, no animal or clinical studies were deemed needed to support this submission.
### IX. Conclusions Safety and Effectiveness SE [21 CFR 807.92(b) (3)]
The PicoreMax is as safe and effective as the predicate devices.
The proposed PicoreMax has the same intended use and indications, similar technological characteristics, and the same principle of operation as the predicate devices.
Thus, the PicoreMax is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.