← Product Code [MOS](/productcode/MOS) · K254190

# dS Base 1.5T; dS Base 3.0T; dS Head 1.5T; dS Head 3.0T; dS HeadNeck 1.5T; dS HeadNeck 3.0T (K254190)

_Philips Medical Systems Nederland B.V. · MOS · Jan 14, 2026 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K254190

## Device Facts

- **Applicant:** Philips Medical Systems Nederland B.V.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Jan 14, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Philips Neurovascular MR Coil is intended to be used in conjunction with a Philips 1.5T or 3T Magnetic Resonance Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.

## Device Story

Philips Neurovascular MR Coil family (dS Base, dS Head, dS HeadNeck) consists of phased-array receive-only coils for 1.5T or 3T MRI systems. Coils receive magnetic resonance signals from hydrogen nuclei; signals are amplified and transmitted to the MRI system for tomographic image reconstruction. Used in clinical settings by trained physicians/technicians. dS Base (8-channel) functions independently or with dS Head (7-channel) or dS HeadNeck (8-channel) coils to increase channel count (up to 16-channel) for improved signal-to-noise ratio and workflow. Coils connect to the MR system flex connect socket. Enables diagnostic imaging of head, neck, spine, extremities, and pediatric anatomy. Benefits include optimized image quality and workflow efficiency through modular coil combinations.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-2-33, IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993 series).

## Technological Characteristics

Phased-array receive-only MR coils. Materials: Polycarbonate (Lexan 925) for dS Base; Copolyester (MFX-421) for dS Head/HeadNeck. Connectivity: Flex connect socket to Philips MR systems. Standards: IEC 60601-2-33, IEC 60601-1, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23, IEC 62366-1, ISO 14971.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- dS Base 1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS Head-Neck 1.5T, dS Head-Neck 3.0T ([K123492](/device/K123492.md))

## Reference Devices

- dS Sentinelle Breast 16 Ch Coil ([K213735](/device/K213735.md))
- dS FootAnkle 16 Ch Coils ([K213766](/device/K213766.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 14, 2026

Philips Medical Systems Nederland BV
Ketaki Bendre
Regulatory Affairs Manager
Veenpluis 6
Best, 5684 PC
Netherlands

Re: K254190
Trade/Device Name: dS Base 1.5T; dS Base 3.0T; dS Head 1.5T; dS Head 3.0T; dS HeadNeck 1.5T; dS HeadNeck 3.0T
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: MOS
Dated: December 23, 2025
Received: December 23, 2025

Dear Ketaki Bendre:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254190 - Ketaki Bendre
Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}

K254190 - Ketaki Bendre
Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Michael D.
O'hara -S
Date: 2026.01.14 14:53:37 -05'00' For
Daniel Krainak
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K254190  |   |
|  Device Name dS Base 1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T, dS HeadNeck 3.0T.  |   |
|  Indications for Use (Describe) The Philips Neurovascular MR Coil is intended to be used in conjunction with a Philips 1.5T or 3T Magnetic Resonance Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K254190

# 510(k) Summary

The 510(k) Summary was prepared in accordance with 21 CFR §807.92(c).

Preparation date: January 12, 2026

510(k) Owner: Philips Medical Systems Nederland B.V.
Veenpluis 6
5684 PC Best
The Netherlands
Establishment Registration Number: 3042177665

Contact person: Ketaki Bendre (Primary Contact)
Regulatory Affairs Manager
Philips Medical Systems Nederland B.V.
Email: ketaki.bendre@philips.com

Swapnil Jain (Secondary Contact)
Senior Regulatory Affairs Manager
Philips Medical Systems Nederland B.V.
Email: swapnil.jain@philips.com

Device Trade Name: dS Base 1.5T, dS Base 3, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T, dS HeadNeck 3.0T

Classification: Classification Name: Coil, Magnetic Resonance, Specialty
Regulation Number: 21 CFR 892.1000
Review Panel: Radiology
Device Class: Class II
Product Code: MOS

Predicate Device: Trade name: dS Base 1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS. Head-Neck 1.5T, dS Head-Neck 3.0T
510(k) Clearance: K123492
Manufacturer: Philips Medical Systems Nederland B.V.
Classification Name: Coil, Magnetic Resonance, Specialty
Regulation Number: 21 CFR 892.1000
Review Panel: Radiology
Device Class: Class II
Product Code: MOS

Reference Device: Trade name: dS Sentinelle Breast 16 Ch Coil
510(k) Clearance: K213735
Manufacturer: Philips Medical Systems Nederland B.V.

Reference Device: Trade name: dS FootAnkle 16 Ch Coils

{5}

510(k) Clearance: K213766
Manufacturer: Philips Medical Systems Nederland B.V.

# Device description

A general description of the subject devices, which are part of the Philips Neurovascular MR Coil Family is given below:

The dS Base 1.5T and dS Base 3.0T are lightweight, 8-channel, phased-array coil that are designed to be used independently or in combination with the dS Head or dS HeadNeck coil. The dS Base coil also can be used for extremities and pediatric examinations or can be combined with the dS Posterior coil which is integrated in the patient support of the compatible Magnetic Resonance (MR) Systems for total spine imaging. The dS Base coil can stay on the table for most examinations without exchanging coils, thereby, improving workflow. These coils are connected to the flex connect socket of the compatible MR System.

The dS Head 1.5T and dS Head 3.0T are 7-channel receive only coils, which are designed to be used in combination with the dS Base (bottom) coil to create a 15-channel coil with or without the dS Posterior coil which is integrated in the patient support of the compatible MR Systems for total spine imaging. The coil is most frequently used for imaging the head/brain and its vasculature. The coil can also be used for extremities and pediatric examinations. These coils are available in both 1.5T and 3.0T and are connected to the flex connect socket of the compatible MR System.

The dS HeadNeck 1.5T and dS HeadNeck 3.0T are 8-channel receive only coils, which are designed to be used in combination with the dS Base (bottom) coil to create a 16-channel coil with or without the dS Posterior coil which is integrated in the patient support of the compatible MR Systems for total spine imaging. The coil is used for brain, head, neck, cervical spine, and neurovascular imaging. The coil also can be used for extremities and pediatric examinations. These coils are available in both 1.5T and 3.0T and are connected to the flex connect socket of the compatible MR System

The dS HeadNeck coil can be used either alone or in combination with the dS Anterior coil and dS Posterior coil. When used in combination with the dS Anterior coil and dS Posterior coil for Whole Body exams this is called a 'Whole Body solution.

# Indications for use

The indications for Use statement provided below for the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T is similar to the predicate device dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS Head-Neck 1.5T and dS Head-Neck 3.0T (K123492). While the new indication for use statement of the subject device is modified to make the wording specific for the subject device coils. The difference in wording in the Indications for Use statement do not raise any new safety and effectiveness questions, both devices have the same intended use and are intended to be used in conjunction with a MR Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.

{6}

The Philips Neurovascular MR Coil is intended to be used in conjunction with a Philips 1.5T or 3T Magnetic Resonance Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.

# Design Features/Fundamental Scientific Technology:

Same as the predicate device, the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T is based on the principle that the coil receives magnetic resonance signals generated in hydrogen nuclei (protons) in the human body while blocking the radio frequency magnetic field applied by the MRI system at specified timings. The received signal is amplified and transmitted to the MRI system, where it is processed into tomographic images by the computer

The principal technological components (PCBs, system cable, baluns and positioning/comfort pads) of the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T is unchanged compared to predicate device.

The following technological differences exist between the subject and predicate devices:

- The subject device housing material has changed from polyester (Trirex) in predicate device to polycarbonate (Lexan 925) for dS Base coils.
- The subject device housing material has changed from polyester (Trirex) in predicate device to copolyester (MFX-421) for dS Head /dS HeadNeck coils

# Summary of Non-Clinical Performance Data

The Non-clinical performance testing has been performed on the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T and demonstrates compliance with following international and FDA-recognized consensus standards

|  Recognition Number | Standard Number and Date | Standard Name  |
| --- | --- | --- |
|  12-347 | IEC 60601-2-33 Edition 4.0 2022-08 | Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis  |
|  19-46 | ANSI / AAMI ES60601-1:2005/(R)2012 and A1:2012 | Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2006, MOD).  |
|  19-36 | IEC 60601-1-2:2014 [Including AMD 1:2021] | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]  |

{7}

|  2-258 | ISO 10993-1:2018 | ISO 10993-1:2018 - Biological evaluation of medical devices. Part 1: Evaluation and testing within a risk management process  |
| --- | --- | --- |
|  2-291 | 10993-23 First edition 2021-01 | Biological evaluation of medical devices - Part 23: Tests for irritation  |
|  2-245 | 10993-5 Third edition 2009-06-01 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity  |
|  2-296 | 10993-10 Fourth edition 2021-11 | Biological evaluation of medical devices - Part 10: Tests for skin sensitization  |
|  5-129 | ANSI AAMI IEC 62366-1:2015+AMD1:2020 (Consolidated Text) | Medical devices Part 1: Application of usability engineering to medical devices including Amendment 1  |
|  5-125 | ANSI AAMI ISO 14971: 2019 | Medical devices – Application of risk management to medical devices.  |
|  5-134 | ISO 15223-1:2021, 15223-1 Fourth Edition 2021-07 | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements  |
|  5-126 | ISTA 3A 2018 | Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less  |
|  14-579 | 17664-2 First edition 2021-02 | Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices.  |

Non-Clinical verification and validation tests have been performed with regards to the intended use, the technical claims, the requirement specifications and the risk management results.

The verification and/or validation test results demonstrate that the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T meet the acceptance criteria and are adequate for the intended use.

The risk management activities show that all risks are sufficiently mitigated, that no new risks are introduced, and that the overall residual risks are acceptable.

Therefore, the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T are substantially equivalent to the legally marketed predicate devices in terms of safety and effectiveness.

## Summary of Clinical Data:

With the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T, the indications for use remain similar and there were no technological characteristics relative to the predicate device that would require clinical testing.

{8}

# Conclusion:

The subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0T are substantially equivalent to the legally marketed predicate device dS Base1.5T, dS Bae 3.0T, dS Head 1.5T, dS Head 3.0T, dS Head-Neck 1.5T and dS Head-Neck 3.0T (K123492) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, substantial equivalence is demonstrated with non-clinical performance (verification and validation) tests, which complied with the requirements specified in the international and FDA-recognized consensus standards and device-specific guidance.

The results of these tests demonstrate that the subject dS Base1.5T, dS Base 3.0T, dS Head 1.5T, dS Head 3.0T, dS HeadNeck 1.5T and dS HeadNeck 3.0 meet the acceptance criteria and are adequate for the intended use.

---

**Source:** [https://fda.innolitics.com/device/K254190](https://fda.innolitics.com/device/K254190)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
