The SAYA 86 Radial Access Guide Catheter is indicated for the introduction of interventional devices into the neurovasculature.
Device Story
SAYA 86 Radial Access Guide Catheter is a sterile, single-use device for neurovascular access; facilitates introduction of interventional devices (e.g., guidewires, therapeutic devices) into neurovasculature. Device comprises guide catheter and dilator; catheter features stainless-steel coils/braids for kink resistance, PTFE inner lumen, and hydrophilic coating on distal segment. Operated by trained physicians in clinical settings. Physician navigates catheter through vasculature; radiopaque marker aids visualization under X-ray. Output is a conduit for interventional tools; enables targeted delivery of devices to neurovascular sites. Benefits include reliable access to neurovasculature for therapeutic interventions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including dimensional verification, simulated use, tensile strength, tip flexibility, kink resistance, torque strength, radiopacity, coating integrity, particulate evaluation, lubricity, burst pressure, infusion flow rate, leak testing, hub compatibility, and corrosion resistance. Biocompatibility testing performed per ISO 10993.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 11, 2026
ASAHI INTECC CO., LTD.
% Ariel Barrett, PhD
Associate Director of Regulatory Affairs
ASAHI INTECC USA, Inc.
22 Executive Park, Suite 110
Irvine, California 92614
Re: K254178
Trade/Device Name: SAYA 86 Radial Access Guide Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP
Dated: April 9, 2026
Received: April 9, 2026
Dear Ariel Barrett:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254178 - Ariel Barrett, PhD
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254178 - Ariel Barrett, PhD
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254178 | |
| Device Name SAYA 86 Radial Access Guide Catheter | |
| Indications for Use (Describe) The SAYA 86 Radial Access Guide Catheter is indicated for the introduction of interventional devices into the neurovasculature. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
ASAHI INTECC CO., LTD.
Page 1 of 8
# 510(k) Summary
(as required by 21 CFR § 807.92)
# ASAH INTECC CO.,LTD.
Global Headquarters and R&D Center
3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan
TEL: +81-561-48-5551 FAX: +81-561-48-5552
http://www.asahi-intecc.co.jp/
# SAYA 86 Radial Access Guide Catheter
510(k) K254178
| Date Prepared: | May 8, 2026 |
| --- | --- |
| Applicant: | ASAHI INTECC CO., LTD. 3-100 Akatsuki-cho, Seto, Aichi 489-0071 Japan |
| Contact: | Dr. Ariel Barrett Associate Director of Regulatory Affairs ASAHI INTECC USA, Inc. 22 Executive Park, Suite 110 Irvine, CA 92614 Tel: 240-429-9335 e-mail: ariel.barrett@asahi-intecc-us.com |
| Device Name: | SAYA 86 Radial Access Guide Catheter |
| Device Classification: | Class II, 21 CFR § 870.1250 |
| Regulation Name: | Percutaneous Catheter |
| Product Code: | QJP- Percutaneous Catheter, Neurovasculature |
| Predicate Device: | Rist™ 079 Radial Access Guide Catheter (K241388) |
| Reference Device: | FUBUKI XF Neurovascular Long Sheath (K213589) |
{5}
ASAHI INTECC CO., LTD.
Page 2 of 8
# Indications For Use
The SAYA 86 Radial Access Guide Catheter is indicated for the introduction of interventional devices into the neurovasculature.
# Device Description
The SAYA 86 Radial Access Guide Catheter is a sterile, single-use guide catheter.
The SAYA 86 Radial Access Guide Catheter is composed of a guide catheter and a dilator. The guide catheter is a single lumen neurovascular catheter designed for introduction of interventional devices, such as guidewires and other therapeutic devices. The guide catheter consists of three sections: (1) a shaft, (2) a protector and (3) a connector. It is comprised of stainless-steel coils and polymeric materials. The distal portion of the shaft consists of a soft tip and a soft tube. The proximal part of the shaft is covered by the protector and the connector is bonded to the proximal end of the shaft. The shaft is made of Austenitic stainless-steel braids as well as Austenitic stainless-steel coils for durability and resistance to kinking.
A hydrophilic coating is applied to the distal segment of the guide catheter. The inner lumen of the shaft consists of a hydrophobic (PTFE) tube. The guide catheter has a 1mm radiopaque marker on the distal part to aid in visualization.
The SAYA 86 Radial Access Guide Catheter is provided sterile, by ethylene oxide sterilization, and is intended for single use only by physicians who have been adequately trained in neurointerventional procedures.
{6}
# Comparison with Predicate/Reference Devices
The SAYA 86 Radial Access Guide Catheter has the same intended use and operating principles, and similar technological characteristics compared to the predicate device. While differences exist between the SAYA 86 Radial Access Guide Catheter and predicate device, with respect to the effective length, dimensions, tip shape and materials, those differences do not raise different or new questions of safety and effectiveness.
| Device | Subject Device SAYA 86 Radial Access Guide Catheter | Predicate Device Rist™ 079 Radial Access Guide Catheter | Reference Device FUBUKI XF Neurovascular Long Sheath |
| --- | --- | --- | --- |
| 510(k) Number | K254178 | K241388 | K213589 |
| Manufacturer | ASAHI INTECC CO., LTD. | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular | ASAHI INTECC CO., LTD. |
| Regulation Number | 21 CFR 870.1250 | 21 CFR 870.1250 | 21 CFR 870.1250 |
| Regulation Name | Percutaneous Catheter | Percutaneous Catheter | Percutaneous Catheter |
| Regulatory Class | II | II | II |
| Product Code | QJP | QJP, DQY | QJP, DQY |
| Indications for Use | The SAYA 86 Radial Access Guide Catheter is indicated for the introduction of interventional devices into the neurovasculature. | The Rist™ 079 Radial Access Guide Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | This product is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. This product is also intended to be used for injection of contrast media. This product is intended for use only in the neurovasculature. |
ASAHI INTECC CO., LTD.
Page 3 of 8
{7}
| Device | Subject Device SAYA 86 Radial Access Guide Catheter | Predicate Device Rist™ 079 Radial Access Guide Catheter | Reference Device FUBUKI XF Neurovascular Long Sheath |
| --- | --- | --- | --- |
| Guide Catheter | | | |
| Effective Length | 80cm, 90cm, 95cm, 100cm,105cm,110cm | 95cm, 100cm, 105cm | 80cm, 90cm, 100cm, 110cm |
| Tip Shape | Straight, Angle | Straight | Straight, Angle |
| Outer Diameter | 2.46mm (0.097", 7Fr) | 2.37mm (0.093", 7Fr) | 2.70mm (0.106", 8Fr) |
| Inner Diameter | 2.19mm (0.086") | 2.00mm (0.079") | 2.28mm (0.090") |
| Lubricious Coating/Length | Hydrophilic 20cm | Hydrophilic 25cm | Hydrophilic 8cm |
| Dilator | | | |
| Effective Length | 94.5cm, 104.5cm, 109.5cm, 114.5cm, 119.5cm, 124.5cm | Not available | 1020mm, 1120mm, 1220mm |
| Outer Diameter | 2.15mm (0.085") | Not available | 2.24mm (0.088") |
| Inner Diameter | Distal: 0.91mm (0.036") Proximal: 1.10mm (0.043") | Not available | Distal: 0.91mm (0.036") Proximal: 1.10mm (0.043") |
| Other Information | | | |
| Accessories | None | Hemostasis Valve | Rotating Hemostasis Valve |
| Sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Shelf Life | 3 years | 3 years | 3 years |
| Number of Uses | Single Use | Single Use | Single Use |
ASAHI INTECC CO., LTD.
{8}
# Non-Clinical Performance Testing
The following performance testing for the SAYA 86 Radial Access Guide Catheter was conducted to support the substantial equivalence determination. The SAYA 86 Radial Access Guide Catheter met all acceptance criteria and performed similarly to the predicate and/or reference devices.
Testing was performed per FDA guidance “Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions” and the recommendations provided in ISO 10555-1 “Intravascular Catheters – Sterile and Single-Use Intravascular Catheters – Part 1: General Requirements”.
| Guide Catheter | | |
| --- | --- | --- |
| Test | Test Method Summary | Results/Conclusion |
| Dimensional Verification | The device dimensions were measured to confirm they meet design requirements. | All samples met the acceptance criteria. |
| Simulated Use | The device was used in simulated anatomical model to assess overall performance. | All samples met the acceptance criteria. |
| Tensile Strength (Catheter Bond Strength) | Tensile force was applied to joints to assess mechanical strength. | All samples met the acceptance criteria. |
| Tip Pull | Tensile force was applied to tip joint to assess mechanical strength. | All samples met the acceptance criteria. |
| Tip Flexibility | Stiffness of the device tip was assessed. | All samples met the acceptance criteria. |
| Kink Resistance | The device was bent to assess resistance to kinking. | All samples met the acceptance criteria. |
| Torque Strength | The device was rotated in simulated anatomical mode to assess torque durability. | All samples met the acceptance criteria. |
| Radiopacity | The device was evaluated for visibility under X-ray imaging. | All samples met the acceptance criteria. |
| Coating Integrity/Particulate Evaluation | The device was tracked in a simulated anatomical model to evaluate coating integrity and particulate generation. | The results were comparable to a similar cleared device. |
| Lubricity | The lubricity of the coating was assessed. | The results were comparable to a similar cleared device. |
| Catheter Body Burst Pressure | The device was pressurized until failure to assess burst strength. | All samples met the acceptance criteria. |
ASAHI INTECC CO., LTD.
Page 5 of 8
{9}
| Guide Catheter | | |
| --- | --- | --- |
| Test | Test Method Summary | Results/Conclusion |
| Infusion Flow Rate | The flow rate through the device was measured. | All samples met the acceptance criteria. |
| Leak (Liquid) | The device was pressurized with liquid and examined for leaks. | All samples met the acceptance criteria. |
| Leak (Air) | The hub was aspirated and examined for air ingress. | All samples met the acceptance criteria. |
| Power Injection Burst Pressure | The device was tested for performance during high-pressure injection. | All samples met the acceptance criteria. |
| Hub Compatibility | The connector was tested per ISO 80369. | All samples met the acceptance criteria. |
| Surface/Distal Tip | The device surface and tip were examined for cleanliness and absence of defects. | All samples met the acceptance criteria. |
| Corrosion Resistance | The device was visually examined for any sign of corrosion after immersion into an aqueous solution of sodium chloride to evaluate for occurrence of corrosion. | All samples met the acceptance criteria. |
| Dilator | | |
| Test | Test Method Summary | Results/Conclusion |
| Dimensional Verification | The device dimensions were measured to confirm they meet design requirements. | All samples met the acceptance criteria. |
| Tensile Strength | Tensile force was applied to joints to assess mechanical strength. | All samples met the acceptance criteria. |
| Radiopacity | The device was evaluated for visibility under X-ray imaging. | All samples met the acceptance criteria. |
| Surface | The device surface was examined for cleanliness and absence of defects. | All samples met the acceptance criteria. |
| Distal Tip | The device tip was examined for appearance. | All samples met the acceptance criteria. |
ASAHI INTECC CO., LTD.
Page 6 of 8
{10}
# Biocompatibility
The guide catheter of the SAYA 86 Radial Access Guide Catheter was tested for the following items per ISO 10993 series and FDA guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". All testing performed met the requirements as specified within the applicable standard. The biocompatibility of the dilator was supported by the biocompatibility testing results of the reference device.
| Test | Test Method Summary | Results/Conclusion |
| --- | --- | --- |
| Cytotoxicity L929 MEM Elution Test | Determine the potential cytotoxicity of a mammalian cell culture (L929) in response to the test article extract. | Non-cytotoxic |
| Sensitization Kligman Maximization Test | Allergenic or sensitizing potential of the device was evaluated using polar and non-polar extracts in a guinea pig maximization test. | Non-sensitizing |
| Irritation or Intracutaneous Reactivity Intracutaneous Injection Test | Potential irritation effect of the extract of the device as a result of intracutaneous injection of polar and non-polar extracts was tested. | Non-irritant |
| Acute Systemic Toxicity Systemic Injection Test | Determine the potential toxic effects of the test article extract as a result of a single-dose systemic injection of polar and non-polar extracts in mice. | Non-toxic |
| Material Mediated Pyrogenicity Rabbit Pyrogen Test | Determine the potential presence of material-mediated pyrogen. | Non-pyrogenic |
| Hemocompatibility Rabbit Blood Hemolysis Test | Determine the potential hemolytic activity, via the induction of increased levels of free plasma hemoglobin in rabbit blood, in response to the test article (direct) and its extract (indirect). | Non-hemolytic |
| Hemocompatibility SC5b-9 Complement Activation Test (Direct Contact) | Human plasma was exposed to the device (direct contact) to determine the potential activation of the SC5b-9 complement system. | Non-activator |
| Hemocompatibility Partial Thromboplastin Time Test (PTT) (Direct Contact) | Human plasma was exposed to the device (direct contact) to assess any effect on the intrinsic coagulation pathway by measuring clotting time. | Non-activator |
| Hemocompatibility Platelet and Leukocyte Count Test (PLC) (Direct Contact) | Compared and evaluated the thrombogenicity properties of direct blood contacting components of the SAYA 86 Radial Access Guide Catheter and predicate device in vitro. | Test article was not considered to have an effect on platelet and leukocyte concentrations. |
| Comparative Surface Assessment | Observation at 50x magnification and scanning electron microscopy. | Test article was similar to the predicate device. |
ASAHI INTECC CO., LTD.
Page 7 of 8
{11}
ASAHI INTECC CO., LTD.
Page 8 of 8
## Sterilization and Shelf Life
The SAYA 86 Radial Access Guide Catheter is sterilized using ethylene oxide. The sterilization cycle was verified to ensure a sterility assurance level (SAL) of 10⁻⁶ in accordance with ISO 11135:2014, “Sterilization of health-care products – Ethylene oxide – Requirements for the development, validation, and routine control of a sterilization process for medical devices”. The packaging integrity and performance testing confirmed that the SAYA 86 Radial Access Guide Catheter remains functional throughout its shelf-life.
## Conclusion
The SAYA 86 Radial Access Guide Catheter and the predicate device share the same intended use, operating principles, and similar technological characteristics. Differences in technological characteristics between the SAYA 86 Radial Access Guide Catheter and predicate device do not raise new or different questions of safety and effectiveness. Performance data demonstrate that the device functions as intended. Therefore, the SAYA 86 Radial Access Guide Catheter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.