Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510825); Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657484); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510826); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657485)
K254149 · Sterilmed, Inc. · OWQ · Jun 29, 2026 · Cardiovascular
Device Facts
Record ID
K254149
Device Name
Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510825); Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657484); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510826); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657485)
Applicant
Sterilmed, Inc.
Product Code
OWQ · Cardiovascular
Decision Date
Jun 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Reprocessed SOUNDSTAR CRYSTAL Ultrasound Catheter is indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO 3 EP Navigation Systems, the catheter provides location information. Refer to the Biosense Webster Compatibility Matrix Insert for compatible CARTO 3 Systems as each catheter is compatible with a specific version of the CARTO 3 System and is not backward compatible with previous versions of the CARTO 3 EP Navigation System. The Reprocessed AcuNav CRYSTAL Ultrasound Catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
Device Story
9F sterile, single-use, disposable ultrasound imaging catheters; reprocessed versions of OEM devices. Distal tip contains ultrasound transducer for 2-D imaging; SOUNDSTAR models include 3-D location sensor for integration with CARTO 3 EP Navigation Systems. Steering mechanism controls image plane orientation via catheter tip rotation and variable deflection. Used by clinicians in cardiac interventional procedures for real-time visualization of anatomy and other devices. Reprocessed maximum one time; tracked through cycle; rejected thereafter. Output is ultrasound image displayed on compatible system; assists physician in procedural guidance. Benefits include visualization of cardiac structures during percutaneous interventions.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including biocompatibility, cleanliness, sterilization, packaging, shelf life, and electrical/mechanical safety testing per ISO 14971, ANSI/AAMI ES 60601-1, ISO 10993-1, ISO 11135, and ISO 11607-1/2.
Technological Characteristics
9F diameter; ultrasound transducer; 3-D location sensor (SOUNDSTAR only); steering mechanism for tip rotation/deflection. Sterilization: Ethylene oxide. Standards: ISO 14971, ANSI/AAMI ES 60601-1, ISO 10993-1, ISO 11135, ISO 11607-1/2. Single-use, reprocessed once.
Indications for Use
Indicated for intra-cardiac and intra-luminal visualization of cardiac/great vessel anatomy, physiology, and other devices in the heart for adult and pediatric patients. Used for imaging guidance during cardiac interventional percutaneous procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{1}

June 29, 2026
Sterilmed, Inc.
Rachel Poltilove
Associate Director, Regulatory Affairs
5010 Cheshire Pkwy. N, Suite 2
Plymouth, Minnesota 55446
Re: K254149
Trade/Device Name: Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510825); Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657484); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510826); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657485)
Regulation Number: 21 CFR 870.1200
Regulation Name: Diagnostic Intravascular Catheter
Regulatory Class: Class II
Product Code: OWQ
Dated: June 7, 2026
Received: June 8, 2026
Dear Rachel Poltilove:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K254149 - Rachel Poltilove
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{3}
K254149 - Rachel Poltilove
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for: **MARCO CANNELLA -S**
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
K254149 - Rachel Poltilove
Page 4
The following devices are included in the scope of this 510(k) premarket notification:
| Device Name | OEM Model Numbers | Sterilmed Model Numbers | Description |
| --- | --- | --- | --- |
| Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter | 11510826 | R11510826 | Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System |
| | 11657485 | R11657485 | Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System |
| Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter | 11510825 | R11510825 | Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System |
| | 11657484 | R11657484 | Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System |
{5}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254149 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510826); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657485); Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510825); Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657484) | | |
| Please provide your Indications for Use below. | | ? |
| The Reprocessed SOUNDSTAR CRYSTAL Ultrasound Catheter is indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO 3 EP Navigation Systems, the catheter provides location information. Refer to the Biosense Webster Compatibility Matrix Insert for compatible CARTO 3 Systems as each catheter is compatible with a specific version of the CARTO 3 System and is not backward compatible with previous versions of the CARTO 3 EP Navigation System. The Reprocessed AcuNav CRYSTAL Ultrasound Catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{6}
# 510(k) Summary for K254149
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| Date Prepared: | June 5, 2026 |
| --- | --- |
| Submitter and Manufacturer: | Sterilmed, Inc. 5010 Cheshire Parkway N, Suite 2 Plymouth, MN 55446 www.sterilmed.com |
| Manufacturing Facility Address: | 11400 73rd Avenue North Maple Grove, MN 55369 |
| Official Correspondent: | Rachel Poltilove, PhD Associate Director, Regulatory Affairs Sterilmed, Inc. Tel: 301-213-7224 Email: rpoltilo@its.jnj.com |
| Trade Name: | Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter |
| Classification Name: | Reprocessed Diagnostic Ultrasound Catheter |
| Common Name: | Diagnostic Intravascular Catheter |
| Device Classification: | Class II, 21 CFR 870.1200 |
| Product Code: | OWQ |
| Primary Predicate Device: | Biosense Webster SOUNDSTAR CRYSTAL™ Ultrasound Catheter (K240050) |
| Secondary Predicate Devices: | Siemens AcuNav CRYSTAL™ Ultrasound Catheter (K233270) |
| Device Description: | The Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter is a 9F sterile, single patient use, disposable imaging catheter. The distal end of the catheter has an ultrasound transducer providing 2-D imaging and a 3-D location sensor providing location information to compatible Carto® 3 EP Navigation Systems with ultrasound capability. A steering mechanism controls the image plane orientation by rotating both the catheter tip and the variable deflection. The Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter is a 9F sterile, single patient use, disposable imaging catheter. The distal end of the catheter has an ultrasound transducer providing 2-D imaging ultrasound capability. A steering mechanism controls the image plane orientation |
K254149, page 1 of 3
{7}
| | by rotating both the catheter tip and the variable deflection. For ultrasound purposes, the Reprocessed AcuNav CRYSTAL™ and Reprocessed SOUNDSTAR CRYSTAL™ catheters are identical. The Reprocessed SOUNDSTAR CRYSTAL™ and AcuNav CRYSTAL™ Ultrasound Catheters are reprocessed no more than one time. Each device is marked and tracked through the reprocessing cycle. After the device has been reprocessed once, it is rejected from further reprocessing. Sterilmed does not reprocess catheters that have previously been reprocessed by other 3^{rd} party reprocessors. | |
| --- | --- | --- |
| **Model Numbers:** | R11510826 | Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System |
| | R11657485 | Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System |
| | R11510825 | Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System |
| | R11657484 | Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System |
| **Indications For Use:** | The Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter is indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO 3 EP Navigation Systems, the catheter provides location information. Refer to the Biosense Webster Compatibility Matrix Insert for compatible CARTO 3 Systems as each catheter is compatible with a specific version of the CARTO 3 System and is not backward compatible with previous versions of the CARTO 3 EP Navigation System. The Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. | |
| **Technological Characteristics:** | The technological characteristics of the Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter including claims, clinical applications, patient population, performance specifications, and method of operation, are identical to the Biosense Webster SOUNDSTAR CRYSTAL™ Ultrasound Catheter. The technological characteristics of the Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter including claims, clinical applications, patient population, performance specifications, and method of operation, are | |
K254149, page 2 of 3
{8}
| | identical to the Siemens AcuNav CRYSTAL™ Ultrasound Catheter. |
| --- | --- |
| **Summary of Non-Clinical Tests Conducted:** | The Reprocessed SOUNDSTAR CRYSTAL™ and AcuNav CRYSTAL™ Ultrasound Catheters have been evaluated for biocompatibility, cleanliness, sterilization, packaging, shelf life, and electrical, electromagnetic and mechanical safety, and have been found to conform with applicable medical device safety standards. The devices comply with the following voluntary standards: • ISO 14971, Application of Risk Management to Medical Devices • ANSI/AAMI ES 60601-1, Safety Requirements for Medical Equipment • ISO 10993-1, Biocompatibility • ISO 11135, Sterilization of Health-Care Products - Ethylene oxide • ISO 11607-1 and ISO 11607-2, Packaging for Terminally Sterilized Medical Devices Performance testing shows the Reprocessed SOUNDSTAR CRYSTAL™ and Reprocessed AcuNav CRYSTAL™ Ultrasound Catheters perform as originally intended. |
| **Conclusion:** | The Reprocessed SOUNDSTAR CRYSTAL™ and Reprocessed AcuNav CRYSTAL™ Ultrasound Catheters are substantially equivalent to their respective currently cleared predicate devices based on the successful completion of nonclinical bench testing as well as the technological comparison of identical principles of design, operation, and indications for use. |
K254149, page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.