InSet Reverse Total Shoulder System

K254128 · Shoulder Innovations, Inc. · PHX · Apr 13, 2026 · Orthopedic

Device Facts

Record IDK254128
Device NameInSet Reverse Total Shoulder System
ApplicantShoulder Innovations, Inc.
Product CodePHX · Orthopedic
Decision DateApr 13, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The InSet Reverse Total Shoulder System should be used in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The InSet Reverse Total Shoulder System is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The System can be used to treat the fractures of the humeral head, fractures of the proximal humerus where other methods of treatment are deemed inadequate (with Standard Humeral Stems), and revision of other devices if sufficient bone stock remains. The Glenoid Baseplate is intended for cementless application with the addition of screw fixation. The Humeral Stem may be implanted by press-fit or cement fixation.

Device Story

InSet Reverse Total Shoulder System is a modular shoulder prosthesis; product line extension adding InSet 135 Humeral Stem. Device components include glenoid baseplate and humeral stem; implanted by orthopedic surgeons in clinical settings. Humeral stem provides structural support for humeral head/proximal humerus fractures; glenoid baseplate uses screw fixation. System restores shoulder function and relieves pain in patients with rotator cuff deficiency. Operates via mechanical fixation (press-fit or cement) to replace damaged joint structures. Benefits include pain relief and improved mobility for patients with severe arthropathy or fracture.

Clinical Evidence

No clinical data provided; substantial equivalence demonstrated via non-clinical bench testing, including mechanical fatigue analysis, taper interface justification, sterilization validation, packaging validation, shelf-life testing, and pyrogen testing.

Technological Characteristics

Titanium Alloy (Ti-6Al-4V ELI, ASTM F136) humeral stem with proximal porous coating of Commercially Pure Titanium (CPTi, ASTM F67). Modular design; press-fit or cemented fixation. Includes glenoid baseplate and humeral stem components. Sterilization and packaging validated for clinical use.

Indications for Use

Indicated for patients with grossly deficient rotator cuff and severe arthropathy, or failed prior shoulder replacement. Requires functional deltoid muscle and suitable anatomy. Indicated for primary, fracture, or revision total shoulder replacement for pain/disability relief. Includes treatment of humeral head and proximal humerus fractures.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 13, 2026 Shoulder Innovations, Inc. Mark Hanes Senior Technical Director 1535 Steele Ave. SW Suite B Grand Rapids, Michigan 49507 Re: K254128 Trade/Device Name: InSet Reverse Total Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, MBF Dated: March 5, 2026 Received: March 6, 2026 Dear Mark Hanes: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K254128 - Mark Hanes Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K254128 - Mark Hanes Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, FARZANA SHARMIN -S Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254128 | ? | | Please provide the device trade name(s). | | ? | | InSet Reverse Total Shoulder System | | | | Please provide your Indications for Use below. | | ? | | Reverse Total Shoulder: The InSet Reverse Total Shoulder System should be used in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The InSet Reverse Total Shoulder System is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The System can be used to treat the fractures of the humeral head, fractures of the proximal humerus where other methods of treatment are deemed inadequate (with Standard Humeral Stems), and revision of other devices if sufficient bone stock remains. The Glenoid Baseplate is intended for cementless application with the addition of screw fixation. The Humeral Stem may be implanted by press-fit or cement fixation. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K254128- Page 1 # 510(k) Summary Date Prepared: December 20, 2025 Submitter: Shoulder Innovations, Inc. 1535 Steele Ave SW, Suite B Grand Rapids, MI 49507 Contact: Mark D. Hanes, Ph.D Sr. Technical Director Shoulder Innovations, Inc. (574) 575-0903 Mark.Hanes@GenesisInnovationGroup.com Proprietary Name: InSet Reverse Total Shoulder System Common Name: Shoulder Prosthesis, Reverse Configuration Classification Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulation Number: 21 CFR 888.3660 Classification Code: PHX, MBF Review Panel: Orthopedic Predicate Devices: Primary Predicate K252221 – InSet Reverse Total Shoulder System Secondary Predicate K212737 – INHANCE Shoulder System ## Device Description: This submission represents a product line extension to add the InSet 135 Humeral Stem to the InSet Reverse Total Shoulder System, previously cleared under K252221. No modifications are proposed to the cleared InSet Reverse Total Shoulder System components as part of this submission. The purpose of this 510(k) submission is to obtain FDA clearance for a product line extension adding the InSet 135 Humeral Stem (“I135” or “subject device”) and its associated stem-specific instruments to the InSet Reverse Total Shoulder System, and to expand the Indications for Use to include treatment of fractures of the proximal humerus. This fracture indication applies only when the InSet Reverse Total Shoulder System is used in combination with the InSet 135 Humeral Stem. InSet 135 Humeral Stems are collarless and manufactured from Titanium Alloy (Ti-6Al-4V Extra Low Interstitial, (ELI) per ASTM F136) with a proximal porous coating of Commercially Pure Titanium (CPTi) per ASTM F67, with fins to provide rotational stability; identical to the InSet 95 Humeral Stem (primary predicate device). Page 1 of 3 {5} SHOULDER K254128- Page 2 ## Indications for Use: **Reverse Total Shoulder:** The Inset Reverse Total Shoulder System should be used in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The Inset Reverse Total Shoulder System is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The System can be used to treat the fractures of the humeral head, fractures of the proximal humerus where other methods of treatment are deemed inadequate (with Standard Humeral Stems), and revision of other devices if sufficient bone stock remains. The Glenoid Baseplate is intended for cementless application with the addition of screw fixation. The Humeral Stem may be implanted by press-fit or cement fixation. ## Comparison of Technological Characteristics: The InSet Reverse Total Shoulder System with InSet 135 Humeral Stem is substantially equivalent to the InSet Reverse Total Shoulder System with InSet 95 Humeral Stem (K252221) with respect to intended use, fundamental scientific technology, overall system design, materials, modularity, fixation methods, and sterilization. The InSet 135 Humeral Stem represents a length variant of the previously cleared humeral stem platform and maintains identical design features, material composition, surface characteristics, and taper interface. The proposed expansion of the indications for use to include treatment of fractures of the proximal humerus is supported by the legally marketed INHANCE™ Reverse Shoulder System (K212737), which is cleared under the same regulation and product code. Differences related to stem length, packaging configuration, shelf life, and manufacturing and sterilization sites were evaluated through appropriate non-clinical testing, including mechanical performance testing, cadaver laboratory evaluation, sterilization validation, packaging validation, and shelf-life testing, and were shown not to raise different questions of safety or effectiveness. ## Non-Clinical Testing Summary: Non-clinical testing was conducted to demonstrate that the InSet Reverse Total Shoulder System with InSet 135 Humeral Stem performs as intended and is substantially equivalent to the predicate devices. Testing included mechanical evaluation (worst-case fatigue analysis) and taper interface justification. Additional testing, including sterilization validation, packaging validation, shelf-life testing, and pyrogen testing, demonstrated that the subject device meets applicable performance requirements and does not raise different questions of safety or effectiveness. ## Clinical Testing Summary: Clinical testing was not necessary to demonstrate substantial equivalence of the InSet Reverse Total Shoulder System InSet 95 Humeral Stem to the predicate device. Page 2 of 3 {6} SHOULDER K254128- Page 3 ## Overall Conclusion: The InSet Reverse Total Shoulder System with InSet 135 Humeral Stem is substantially equivalent to the legally marketed InSet Reverse Total Shoulder System with InSet 95 Humeral Stem (K252221) with respect to intended use, fundamental scientific technology, design, materials, fixation methods, and sterilization. The proposed expansion of the indications for use to include treatment of fractures of the proximal humerus is supported by the legally marketed INHANCE™ Reverse Shoulder System (K212737), which is cleared under the same regulation and product code. Differences between the subject and predicate devices were evaluated through appropriate non-clinical testing and do not raise different questions of safety or effectiveness; therefore, the subject device is substantially equivalent to the identified predicate device. Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...