K254120 · Subtle Medical, Inc. · QIH · May 15, 2026 · Radiology
Device Facts
Record ID
K254120
Device Name
SubtleHD-CT (1.x)
Applicant
Subtle Medical, Inc.
Product Code
QIH · Radiology
Decision Date
May 15, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K254120 · May 15, 2026
SubtleHD-CT (1.x)
Subtle Medical, Inc.
Retrospectively acquired clinical CT exams
Retrospective clinical CT cases were used in a reader study to assess the diagnostic utility and artifact presence of the SubtleHD-CT enhanced images compared to original images.
SubtleHD-CT Reader Study; Retrospective reader study
270 cases including pediatric and adult subjects (3 to 90 years old) across various anatomies and scanner vendors.; Sample Size: 270 cases; Number of Sites: Multiple institutions
Detectability index (d') >= unenhanced FBP; Noise magnitude p < 0.05; NPS frequency < 0.5 cycles/mm; CT number accuracy +/- 5 HU; Image non-uniformity < 5 HU; TTF < 1.2 lp/cm
Detectability index (d') > unenhanced FBP; Noise magnitude statistically significantly lower; NPS frequency < 0.5 cycles/mm; CT number accuracy within +/- 5 HU; Image non-uniformity < 5 HU; TTF < 1.2 lp/cm
—
—
Reader Study: 270 cases of retrospectively acquired human data
>1 (board-certified radiologists)
Indications for Use
SubtleHD-CT is an AI/ML software intended for the processing and enhancement of CT images regardless of the manufacturer or model of CT scanner.
Device Story
SubtleHD-CT is a deep learning-based image processing SaMD designed to enhance CT images. It operates as background software without a user interface, processing DICOM images received from a compatible medical device data system (MDDS) after acquisition. The software utilizes a convolutional neural network (CNN) to perform image enhancement, specifically targeting noise reduction. Enhanced images are returned to the PACS or other DICOM nodes for clinical review by radiologists. The device is intended to be used in medical facilities; it serves as an adjunctive tool to original reconstructed images, aiming to improve image quality. It is compatible with various CT scanner manufacturers and reconstruction methods (FBP, IR, DLR).
Clinical Evidence
Performance validated via bench testing and a retrospective reader study. Bench testing used ACR 464, Digital CatPhan 600, and Mercury 4.0 phantoms across multiple scanner vendors (GE, Siemens, Canon) to evaluate CT number accuracy, noise magnitude, NPS, and spatial resolution (TTF). Results showed statistically significant noise reduction (p<0.05) and improved detectability index (d') compared to FBP. Reader study included 270 retrospective clinical cases evaluated by board-certified radiologists; results confirmed diagnostic utility (Likert score ≥3) and absence of clinically impactful artifacts (Likert score ≥2).
Technological Characteristics
Software-only device; operates on virtual machines (Linux compatible). Uses a deep learning convolutional neural network (CNN) for image enhancement. Processes DICOM-compliant image data. Designed for primary reconstructed images (FBP). No user interface; configuration via files/environment variables. Standalone deployment model within existing hospital infrastructure.
Indications for Use
Indicated for processing and enhancement of CT images across all anatomies (abdomen, c-spine, cardiac, chest, chest-abdomen-pelvis, head, head & neck, l-spine, lower-extremity, MSK, t-spine, upper-extremity) in pediatric and adult patients (3-90 years old).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
Subtle Medical, Inc.
Adam Heroux
Quality Engineering and Regulatory Affairs Manager
4190 Grove St
Denver, Colorado 80211
Re: K254120
Trade/Device Name: SubtleHD-CT (1.x)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: April 11, 2026
Received: April 13, 2026
Dear Adam Heroux:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K254120 - Adam Heroux
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K254120 - Adam Heroux
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254120 | ? |
| Please provide the device trade name(s). | | ? |
| SubtleHD-CT (1.x) | | |
| Please provide your Indications for Use below. | | ? |
| SubtleHD-CT is an AI/ML software intended for the processing and enhancement of CT images regardless of the manufacturer or model of CT scanner. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
K254120
# SubtleHD-CT 510(k) Summary
Table 1. Contact Details and Device Name
| Date Summary Prepared: | 10 April 2026 |
| --- | --- |
| Contact Details: | |
| Company Name: | Subtle Medical, Inc. |
| Company Address: | 883 Santa Cruz Ave, Suite 205 Menlo Park, CA 94025 United States |
| Applicant Name: | Mr Ajit Shankaranarayanan |
| Application Telephone: | 650-448-4285 |
| Applicant Email: | Ajit@subtlemedical.com |
| Correspondent Name: | Highland Biomedical |
| Correspondent Address: | 4190 Grove St Denver, CO 80211 |
| Correspondent Contact: | Mr. Adam Heroux |
| Correspondent Contact Telephone: | 617-823-0515 |
| Correspondent Contact Email: | regulatory@subtlemedical.com |
| Device Name: | |
| Device Trade Name | SubtleHD-CT (1.x) |
| Common Name: | Medical image management and processing system |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Number: | 892.2050 |
| Product Code: | QIH |
| Device Class: | Class II |
| Legally Marketed Predicate Devices: | Primary Predicate #: K212074 Primary Predicate Name: ClariCT.AI Primary Predicate Legal Manufacturer: Claripi |
Device Description Summary
{6}
K254120
SubtleHD-CT is an image processing Software as a Medical Device (SaMD) consisting of a deep learning artificial intelligence/machine learning (AI/ML) to enhance medical images acquired by computed tomography (CT) scanners. It is post-processing software that does not directly interact with the scanner. Once a CT scan is acquired, a technologist sends the study from the scanner to a compatible medical device data system (MDDS) via the DICOM protocol. The compatible MDDS, then, makes the images available to SubtleHD-CT for processing. The study is enhanced by SubtleHD-CT and sent back to the picture archiving and communication system (PACS) or other DICOM node by the compatible MDDS for clinical review. Please note that interpretation of SubtleHD-CT's output should be adjunctive to the original reconstructed image. Because the software runs in the background, it has no user interface. Configuration is specified as configuration files and OS environment variables.
The SubtleHD-CT software implements an image enhancement algorithm using a convolutional neural network. Input images are enhanced by running through the CNN that consists of cascaded layers to perform numerical operations, including convolutional filtering, linear combination, and nonlinearity operations.
# Intended Use / Indications for Use
SubtleHD-CT is an AI/ML software intended for the processing and enhancement of CT images regardless of the manufacturer or model of CT scanner.
# Intended Use / Indications for Use Comparison
SubtleHD-CT and its predicate are both intended to enhance CT images regardless of the manufacturer or model of CT scanner.
# Technological Comparison
SubtleHD-CT and its predicate are used for image enhancement. They operate on DICOM files, enhance the images, and send the enhanced images to any desired destination. The receipt of original DICOM image files and delivery of enhanced images as DICOM files depends on other software systems. Both the subject and predicate device use pre-trained deep learning models. The software performs image enhancement. The following table provides a detailed description of the technological characteristics of the subject and predicate device.
SubtleHD-CT v1.x is classified as Class II with product code QIH through the FDA using ClariCT.AI as predicate.
Table 2. Comparison of Technological Characteristics
| Comparison | SubtleHD-CT (Subject Device) | ClariCT.AI (K212074) | Noted Differences |
| --- | --- | --- | --- |
| Product Code | QIH | LLZ | Substantially Equivalent |
| Modalities | CT | Same | Same |
{7}
K254120
| Comparison | SubtleHD-CT (Subject Device) | ClariCT.AI (K212074) | Noted Differences |
| --- | --- | --- | --- |
| Intended Use / Indications for Use | SubtleHD-CT is an AI/ML software intended for the processing and enhancement of CT images regardless of the manufacturer or model of CT scanner. | ClariCT.AI is a software device intended for networking, communication, processing and enhancement of CT images in DICOM format regardless of the manufacturer of the CT scanner or model. | Substantially Equivalent |
| Product Description | SubtleHD-CT is an AI/ML image processing Software as a Medical Device (SaMD) consisting of a deep learning artificial intelligence/machine learning (AI/ML) to enhance medical images acquired by computer tomography (CT) scanners. SubtleHD-CT can enhance CT images of the abdomen, c-spine, cardiac, chest, chest-abdomen-pelvis (CAP), head, head & neck, l-spine, lower-extremity, MSK, t-spine, and upper-extremity acquired on a wide range of CT doses. | ClariCT.AI software is intended for denoise processing and enhancement of CT DICOM images when higher image quality and/or lower dose acquisitions are desired. ClariCT.AI software can be used to reduce noises in CT images of the head, chest, heart, and abdomen, in particular in CT images with a lower radiation dose. ClariCT.AI may also improve the image quality of low-dose nondiagnostic Filtered Back Projection images as well as Iterative Reconstruction images. | Substantially Equivalent |
| Workflow | The software operates on DICOM files, enhances the images, and sends the enhanced images to any desired destination with an AE Title (e.g., PACS, CT device, workstation, and more). Enhanced images coexist with the original images. | ClariCT.AI can seamlessly integrate with PACS and the existing user site infrastructure using features provided by syngo.via. The recommended setting is to automate the whole workflow from taking an exam using a CT scanner to viewing the denoised image in the radiologist's reading environment. | Substantially Equivalent |
3 of 10
{8}
K254120
| Comparison | SubtleHD-CT (Subject Device) | ClariCT.AI (K212074) | Noted Differences |
| --- | --- | --- | --- |
| Physical Characteristics | Software package Operates on a virtual machine | Window or Linux Operating System. PC Hardware supported graphics card or equivalent. Docker images in a virtual machine. | Substantially Equivalent |
| Intended User | Radiologists | Same | Same |
| Intended Location | Medical facility (hospitals, clinics, imaging center, etc.) | Same | Same |
| Operating System / Computer | Linux Compatible; PC or Mac | Can be integrated into the AI Marketplace system that distributes Docker images in a virtual machine running in a Linux environment. | Substantially Equivalent |
| Rx or OTC | Rx | Same | Same |
| User Interface | Runs in the background (no user interact) | GUI mode and Background mode. MixView to view original and denoised images side-by-side for easy image quality comparison. | Both have background modes, ClariCT.AI has an optional user interface. Substantially Equivalent |
| DICOM Standard Compliance | The software processes DICOM-compliant image data. | Same | Same |
| Image Enhancement Algorithm Description | Image enhancement is performed with the use of pre-trained deep learning models. | Noise reduction is performed with the use of pre-trained deep learning models. | Substantially Equivalent |
| Model Architecture | One model is used for image enhancement of CT. | Pre-trained deep learning models. | Substantially Equivalent |
Non-Clinical and/or Clinical Tests Summary & Conclusions
{9}
K254120
Subtle Medical conducted the following performance testing:
- Software Verification and Validation testing (unit, integration, and system testing) to demonstrate that software requirements are implemented. These tests PASSED.
- Performance Validation testing utilizing ACR Phantom, Digital CatPhan Phantom, and Mercury 4.0 Phantom to demonstrate the software enhanced image quality. These tests PASSED.
- A Reader Study utilizing retrospective clinical data to demonstrate the software enhanced image quality in CT images. These tests PASSED.
## Performance Phantom Characteristics:
The performance validation utilized industry standard phantoms:
- CT ACR phantom: used by the proposed predicate device.
- Digital CatPhan Phantom 600: from the FDA's Division of Imaging, Diagnostics, and Software Reliability (DIDSR).
- Mercury 4.0 Phantom
## Reader Study Dataset Characteristics:
To represent the patient population and use of CT in the field, the SubtleHD-CT performance and reader study validation test dataset consists of:
- Split between all anatomies (abdomen, c-spine, cardiac, chest, chest-abdomen-pelvis (CAP), head, head & neck, l-spine, lower-extremity, MSK, t-spine, upper-extremity).
- Canon, GE Medical Systems, Hitachi, Philips, and Siemens scanner vendors.
- Pediatric (adolescent, children) and adult subjects (ranging from 3 to 90 years old).
- Even distribution of subject sex.
- Contrasted and non-contrasted images.
- Various reconstruction methods (FBP, IR, DLR).
- Various slice thicknesses: between 0.5 mm and 15 mm
- Various FOV: between 113mm and 1246 mm.
- Axial, coronal, sagittal, and oblique orientations.
- Various imaging acquisition parameters (XRay Tube Current, Exposure, kVP, Pixel Size).
- 64 unique kernels representing various reconstructions (soft, standard, and sharp) performed for the wide range of clinical protocols in different scanner manufacturers.
Summary of CTDIvol dose levels represented in the clinical cases used in the reader study
| CT Protocol | Minimum Dose (mGy) | Average Dose (mGy) | Maximum Dose (mGy) |
| --- | --- | --- | --- |
| Abdomen | 2.75 | 12.94 | 45.50 |
| C-Spine | 2.10 | 11.29 | 28.68 |
| Cardiac | 2.10 | 15.99 | 55.28 |
| Chest | 0.90 | 5.47 | 15.07 |
{10}
K254120
| Chest-Abd-Pelvis | 2.86 | 11.25 | 27.84 |
| --- | --- | --- | --- |
| Head | 13.37 | 38.47 | 68.50 |
| Head-Neck | 1.35 | 10.70 | 39.06 |
| L-Spine | 8.87 | 19.78 | 32.78 |
| Lower-Extremity | 3.73 | 10.05 | 18.88 |
| MSK | 15.50 | 15.50 | 15.50 |
| T-Spine | 4.11 | 13.75 | 29.72 |
| Upper-Extremity | 0.93 | 10.75 | 21.03 |
| Min/Avg/Max | 0.90 | 15.38 | 68.50 |
Various clinical conditions are present in the performance dataset, as per reports accompanying the images from the sources, including: chronic sinusitis, trauma, stroke, head injury, osteosarcoma, dissection, pain, fracture, mass/tumor, aneurysm, calcium scoring, coronary CTA, etc. Some clinical conditions were not reported with the data. This selection criteria represents a well characterized clinically-relevant reference dataset.
To show the performance of the device was not hindered by site variability, in the validation dataset, data was selected from sources not included in the training dataset. The majority of performance data comes from sources in the United States. Independence between the training and test datasets was ensured by using non-overlapping datasets, with no shared subjects, studies, or images between training and testing.
## Performance Validation:
The quantitative performance validation for SubtleHD-CT was conducted exclusively using phantom data. Specifically, testing utilized the CT ACR 464 phantom, the Digital CatPhan 600 phantom, and the Mercury 4.0 phantom. As part of the quantitative evaluation: CT Number Accuracy, Image Noise (Magnitude), Image Non-Uniformity (NU), Spatial Resolution (Task Transfer Function - TTF), Noise Texture (Noise Power Spectrum - NPS), Detectability Index (d') were characterized when compared to traditional FBP reconstructions. Where applicable, testing relied on simulated targets rather than clinical anomalies to estimate task-based clinical utility, such as how well a signal or lesion can be detected given the image resolution and noise properties.
SubtleHD-CT's quantitative performance was validated on these phantoms for CT images reconstructed via standard Filtered Back Projection (FBP), acquired using standard-dose or higher-dose protocols.
The software is engineered to complement the characteristics of FBP-based imaging; use with Iterative Reconstruction (IR) or Deep Learning (DL)-based reconstructions may result in unknown performance. Please refer to Table 4 for more details.
SubtleHD-CT has been validated for use on primary reconstructed images. Performance has not been specifically characterized on images that have undergone advanced third-party
6 of 10
{11}
K254120
post-processing or specialized non-standard filtering. It is recommended to apply SubtleHD-CT enhancement to the direct output of the CT scanner's primary reconstructed images.
For the assessment of task-based image quality on the ACR phantom, Digital CatPhan and Mercury 4.0 phantoms, simulated lesions of varying diameters and contrasts were used to mimic low-contrast detectability across different tasks. Table 3 summarizes the target characteristics of the tasks performed.
Table 3: Simulated Target Characteristics for ACR 464, Digital CatPhan 600 and Mercury 4.0 Phantoms
| Phantom | Target type | Target sizes | Target Contrasts / Materials |
| --- | --- | --- | --- |
| ACR 464 | Simulated circular lesions in varying backgrounds | 3mm, 4mm, 5mm, 6mm, 7mm | Materials: Air, Polystyrene, Acrylic, Bone. Each material was evaluated at 5 varying contrasts. |
| Digital CatPhan 600 | Simulated circular lesions | 3mm, 5mm, 7mm, 10mm | 14 HU, 7 HU, 5 HU, 3 HU (respectively) |
| Mercury 4.0 | Simulated circular lesion in varying backgrounds | 1mm | Air, Water, Bone, Polystyrene, Iodine |
Phantom Testing Details: To characterize the performance of SubtleHD-CT, phantoms were tested across multiple vendor scanners (GE, Siemens, and Canon) using a wide variety of scanning parameters. Across the full dataset, testing encompassed a wide operational range including slice thicknesses from 0.28 to 5 mm, reconstruction fields of view (FOV) from 144 to 500 mm, CTDIvol of 0.2 to 20.6 mGy and pixel size from 0.28 to 0.98 mm. Furthermore, the Mercury 4.0 phantom consists of 5 cylindrical modules composed of Polyethylene, varying in diameter from 16, 21, 26, 31, and 36 cm to represent varying patient sizes. Table 4 summarizes the key acquisition parameters.
Table 4: Evaluated Acquisition Parameters for ACR 464, Digital CatPhan 600 and Mercury 4.0 Phantom scans
{12}
K254120
| Phantom | Pixel size (mm) | FOV (mm) | Matrix Size | Slice Thickness (mm) | FBP Kernel | CTDivol (mGy) | kVp (kV) | X-ray Tube Current (mA) | Vendor |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| ACR 464 | 0.488 | 250 | 512 x 512 | 1 | B30f | 3.38 | 120 | 60 | Siemens |
| | 0.488 | 250 | 512 x 512 | 1 | D45f | 3.38 | 120 | 60 | Siemens |
| | 0.488 | 250 | 512 x 512 | 3 | B30f | 3.38 | 120 | 60 | Siemens |
| | 0.488 | 250 | 512 x 512 | 3 | D45f | 3.38 | 120 | 60 | Siemens |
| Digital CatPhan 600 (13 cm) | 0.281 | 144 | 512 x 512 | 0.28 | fbp D45 | / | 120 | 170 | Siemens(simula-te d) |
| | 0.281 | 144 | 512 x 512 | 0.28 | fbp D45 | / | 120 | 170 | Siemens(simula-te d) |
| Mercury 4.0 | 0.586 | 300 | 512 x 512 | 5 | STAND ARD | / | 100 | 205 | GE |
| | 0.586 | 300 | 512 x 512 | 5 | STAND ARD | / | 120 | 130 | GE |
| | 0.586 | 300 | 512 x 512 | 5 | STAND ARD | / | 140 | 90 | GE |
| | 0.586 | 300 | 512 x 512 | 5 | STAND ARD | / | 80 | 380 | GE |
| | 0.977 | 500 | 512 x 512 | 4 | Br38s | 6 | 70 | 494 | Siemens |
| | 0.977 | 500 | 512 x 512 | 4 | Br38s | 12.65 | 80 | 646 | Siemens |
| | 0.977 | 500 | 512 x 512 | 4 | Br38s | 19.72 | 100 | 475 | Siemens |
| | 0.977 | 500 | 512 x 512 | 4 | Br38s | 20.25 | 120 | 285 | Siemens |
| | 0.977 | 500 | 512 x 512 | 4 | Br38s | 20.6 | 140 | 190 | Siemens |
| | 0.702 | 360 | 512 x 512 | 1 | Body+ BHC | 0.4 | 80 | 209 | Canon |
| | 0.977 | 500 | 512 x 512 | 1 | Body | 0.2 | 120 | 40 | Canon |
The following are the endpoints, acceptance criteria, results, and conclusions from the SubtleHD-CT performance validation study:
{13}
K254120
Table 5. Performance Validation Summary
| Endpoint | Acceptance Criteria | Results / Conclusions |
| --- | --- | --- |
| Detectability Index (d') Primary Endpoint | The detectability index (d') value of the SubtleHD-CT enhanced outputs must be equal to or greater than the corresponding value obtained from the unenhanced filtered back projection (FBP) inputs across all evaluated clinical tasks. | PASS: For SubtleHD-CT as compared to the unenhanced FBP, across all phantoms and all evaluated clinical tasks the SubtleHD-CT detectability index (d') was greater than the unenhanced FBP. |
| Image Noise (Magnitude) Primary Endpoint | The noise magnitude of the SubtleHD-CT enhanced outputs shall be statistically significantly lower, p-value < 0.05 level with a Student's t-test, than that of the unenhanced filtered back projection (FBP) inputs. | PASS: Across all phantoms the image noise magnitude was statistically significantly lower, p-value < 0.05 level with a Student's t-test, than that of the unenhanced filtered back projection (FBP) inputs. |
| Noise Texture (Noise Power Spectrum - NPS) Secondary Endpoint | The NPS average frequency (f_{av}) and peak frequency (f_{peak}) of the SubtleHD-CT outputs shall be < 0.5 cycles/mm relative to unenhanced filtered back projection (FBP) inputs. | PASS: For SubtleHD-CT as compared to the unenhanced FBP, across all phantoms, scanner parameters, and diameters the NPS average frequency (fav) and peak frequency (fpeak) for the SubtleHD-CT outputs was < 0.5 cycles/mm relative to the unenhanced filtered back projection (FBP) inputs. |
| CT Number Accuracy Secondary Endpoint | The CT number accuracy shall remain within ± 5 houndsfield units (HU) of the unenhanced filtered back projection (FBP) inputs. | PASS: For SubtleHD-CT as compared to the unenhanced FBP, the CT number accuracy was within ± 5 houndsfield units (HU) of the unenhanced filtered back projection (FBP) inputs. |
| Image Non-Uniformity (NU) Secondary Endpoint | The difference between the mean CT value of each peripheral ROI and the center ROI shall be < 5 houndsfield units (HU). | PASS: The mean CT value of each peripheral ROI and the center ROI for both the Digital CatPhan Phantom and ACR Phantom remains within 5 houndsfield units. |
9 of 10
{14}
K254120
| Endpoint | Acceptance Criteria | Results / Conclusions |
| --- | --- | --- |
| Spatial Resolution (Target Transfer Function - TTF) Secondary Endpoint | The TTF curve at 50% (f_{50}) and 10% (f_{10}) of the TTF shall be < 1.2 lp/cm of the unenhanced filtered back project (FBP) inputs. | PASS: Across all ROIs for all the phantoms, the range of the TTF curve at 50% (f50) and 10% (f10) of the TTF was < 1.2 lp/cm of the unenhanced filtered back project (FBP) inputs. |
## Reader Study:
This study utilized 270 cases of retrospectively acquired human data obtained as part of research studies or clinical exams from various institutions. The datasets were presented side-by-side with original and SubtleHD-CT enhanced images.
Board-certified radiologists evaluated SubtleHD-CT enhanced images for diagnostic utility using a 5-point Likert scale and artifact presence using a 3-point Likert scale.
The following are the endpoints, acceptance criteria, results, and conclusions from the SubtleHD-CT Reader Study:
Table 6. Reader Study Summary
| Endpoint | Acceptance Criteria | Results / Conclusions |
| --- | --- | --- |
| Diagnostic Utility Endpoint | The diagnostic utility of the SubtleHD-CT images is the same or better than the original images. | PASS: The average scores across all readers of the SubtleHD-CT enhanced images are an average of 3 or better (enhanced images are the same or better than original). |
| Artifact Presence Endpoint | SubtleHD-CT does not introduce artifacts that could impact diagnosis in any image. | PASS: SubtleHD-CT enhanced images have an average of 2 or better (SubtleHD-CT does not introduce artifacts that impact diagnosis in any image). |
## Conclusion:
These results demonstrate that SubtleHD-CT is substantially equivalent to the predicate devices.
10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.