Irrigation/Urology Sets

K254111 · B.Braun Medical, Inc. · LJH · Sep 10, 2026 · Gastroenterology, Urology

Device Facts

Record IDK254111
Device NameIrrigation/Urology Sets
ApplicantB.Braun Medical, Inc.
Product CodeLJH · Gastroenterology, Urology
Decision DateSep 10, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2

Indications for Use

B. Braun Medical Irrigation Tubing Set is intended for delivery of irrigation solutions from a container to an irrigation site during bladder irrigation or endoscopic procedures such as cystoscopy and transurethral resection procedures.

Device Story

Sterile, single-use, gravity-fed irrigation sets; deliver irrigation solutions from fluid containers to irrigation sites (bladder/endoscopic). Configurations include single, dual, or four-line pathways; vary by tubing dimensions, spike types, flow-control components (clamps), and container compatibility (PIC, flexible bags). Operated by healthcare professionals in hospitals/surgical centers. Connects to indwelling urinary catheters or irrigation devices. Duration of use may exceed 24 hours. Output is controlled gravity flow of irrigation fluid; facilitates surgical/urological procedures by maintaining site visibility and irrigation. Benefits include consistent fluid delivery for procedure maintenance.

Clinical Evidence

Bench testing only. No clinical data. Performance testing included visual inspection, occlusion, pressure integrity (positive/negative), tensile testing, clamp qualification, priming/gravity flow, spike piercing/coring, particulate testing, and biocompatibility (cytotoxicity, sensitization, intracutaneous, systemic toxicity, pyrogen, hemolysis).

Technological Characteristics

Materials: silicone, ABS, LDPE, High Impact Polystyrene, acrylic, PVC, TPE, PTFE. Gravity-fed flow. EO sterilized. Single-use. Configurations: 1-4 lines, various lengths (5-96 in.). Standards: ISO 8536-4, ISO 8536-8, ISO 8536-14, USP <788>, ISO 10993-1.

Indications for Use

Indicated for delivery of irrigation solutions from a container to an irrigation site during bladder irrigation or endoscopic procedures (cystoscopy, transurethral resection) in clinical environments by healthcare professionals.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Submission Summary (Full Text)

{1} ![LOGO](U.S. Food & Drug Administration) September 10, 2026 B.Braun Medical, Inc. Rachel Pappas Regulatory Affairs Specialist 3773 Corporate Pkwy. Center Valley, Pennsylvania 18034 Re: K254111 Trade/Device Name: Irrigation/Urology Sets Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and accessories Regulatory Class: II Product Code: LJH Dated: August 10, 2026 Received: August 11, 2026 Dear Rachel Pappas: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K254111 - Rachel Pappas Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {3} K254111 - Rachel Pappas Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ANDREW M. FALES -S for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K254111 Device Name Irrigation/Urology Sets Indications for Use (Describe) B. Braun Medical Irrigation Tubing Set is intended for delivery of irrigation solutions from a container to an irrigation site during bladder irrigation or endoscopic procedures such as cystoscopy and transurethral resection procedures. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets K254111 Page 1 of 11 ## Irrigation/Urology Sets ### 510(k) Summary This summary of 510(k) information is being provided in accordance with 21 CFR 807.92. | Submitter Information | | | --- | --- | | Name | B. Braun Medical Inc. | | Address | 824 12th Avenue, Bethlehem, PA 18018 | | Telephone number | (610) 266-0500 | | Primary contact Title e-mail Telephone number | Rachel Pappas Regulatory Specialist, Regulatory Affairs rachel.pappas@bbraunusa.com (610) 596-2673 | | Secondary contact Title e-mail Telephone number | Tracy Larish Manager II, Regulatory Affairs tracy.larish@bbraunusa.com 484-375-9064 | | Date prepared | September 9, 2026 | | Device | | | Device Name / Proprietary Name | Irrigation/ Urology Set | | Common or Usual Name | Irrigation Sets and Accessories | | Regulation Name | Urological catheter and accessories | | Regulation Number | 21 CFR 876.5130 | | Device Class | Class II | | Product Code(s) | LJH: System, Irrigation, Urological | | Review Panel | Gastroenterology & Urology | | Legally marketed device to which equivalence is claimed (predicate) | | | Device Name / 510(k) | TUR, Arthroscopic, Cystoscopy Irrigation Sets / K960787 | ### Device Description The proposed B. Braun Irrigation/Urology Sets are sterile, single-use devices intended for the delivery of irrigation solutions from a fluid container to an irrigation site. The devices operate by gravity-fed flow and are available in configurations designed for bladder irrigation and endoscopic procedures. The proposed device family consists of eight configurations that vary in the number of irrigation lines, tubing configuration and dimensions, irrigation-container compatibility, spike configuration, flow-control components, connectors, and other configuration-specific design features. Depending on the configuration, the sets provide single-, dual- or four-line irrigation pathways and are designed for connection to the applicable irrigation fluid container. The proposed B. Braun Irrigation/Urology Sets are intended for delivery of irrigation solutions from a container to an irrigation site during bladder irrigation. {6} K254111 Page 2 of 11 ***B. Braun Medical Inc.*** *510(k) Premarket Notification* *Irrigation/ Urology Sets* Depending on the configuration, the devices are designed for connection to B. Braun PIC™ Plastic Irrigation Containers, flexible IV bags, or applicable spike-compatible irrigation containers. The sets may incorporate single or multiple irrigation lines, with up to four inlet pathways. The patient-contacting end of the set connects to the applicable irrigation device or catheter. When used for continuous bladder irrigation via an indwelling urinary catheter, the duration of use may exceed 24 hours. The devices are intended for use by healthcare professionals in clinical environments, including hospitals and surgical centers. The proposed B. Braun Irrigation/Urology Sets have the same intended use and fundamental operating principle as the predicate device. Both device families are sterile, single-use irrigation sets that deliver irrigation solutions from a fluid container to an irrigation site through a tubing pathway. The proposed configurations incorporate technological characteristics appropriate to their respective irrigation applications, including differences in line count, tubing dimensions, irrigation-container compatibility, spike configuration, flow-control components, and connectors. Functional and performance testing demonstrates that the proposed devices perform as intended. The identified technological differences do not raise different questions of safety or effectiveness. Therefore, the proposed B. Braun Irrigation/Urology Sets are substantially equivalent to the Baxter TUR, Arthroscopic, Cystoscopy Irrigation Sets cleared under K960787. Refer to Table 1 for a side-by-side comparison of the proposed and predicate devices. {7} B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets K254111 Page 3 of 11 Table 1: Substantial Equivalence - Comparison of Proposed Device with Predicate Device | Characteristic | Proposed Device- B. Braun Irrigation/Urology Sets | Predicate Device- Baxter Irrigation Sets, K960787 | Difference/ Substantial Equivalence Assessment | | --- | --- | --- | --- | | Description | Sterile, single-use, non-pyrogenic irrigation sets and accessories intended to deliver irrigation solutions from a fluid container to an irrigation site. The device family includes single-, dual-, and four-line configurations with configuration-specific tubing dimensions, irrigation-container compatibility, spike configurations, flow-control components, and connectors. The devices are EO sterilized and latex-free. | Baxter irrigation sets are sterile, single-use devices intended for delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures (cystoscopy, TUR, arthroscopy). The submission focused on a material change in the distal tubing segment from natural rubber latex to SEBS. | Similar. Both device families provide sterile, single-use, gravity-fed delivery of irrigation solutions. Configuration-specific differences in line count, tubing dimensions, container compatibility, spike configuration, and other components are addressed through design and performance evaluation. | | Intended use | Delivery of irrigation solutions from a container to an irrigation site during bladder irrigation or endoscopic procedures such as cystoscopy and transurethral resection (TUR) procedures. | Delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cystoscopy, transurethral resection (TUR), and arthroscopic procedures. | Similar. However, the proposed is not indicated for arthroscopic procedures. | {8} B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets K254111 Page 4 of 11 | Characteristic | Proposed Device- B. Braun Irrigation/Urology Sets | Predicate Device- Baxter Irrigation Sets, K960787 | Difference/ Substantial Equivalence Assessment | | --- | --- | --- | --- | | Indications for Use | B. Braun Medical Irrigation Tubing Set is intended for delivery of irrigation solutions from a container to an irrigation site during bladder irrigation, or endoscopic procedures such as cystoscopy and transurethral resection procedures. | Baxter's irrigation sets are intended for delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cystoscopy, transurethral resection (TUR), and arthroscopic procedures. | Similar. The proposed devices are not indicated for arthroscopic procedures. | | Design/Components | Depending on the configuration, the proposed devices incorporate irrigation spikes, drip chambers, single or multiple irrigation lines, roller or pinch clamps, Y-sites or multi-port connectors, and Luer or tapered connectors. Configurations differ in tubing dimensions, irrigation-container compatibility, spike configuration, flow-control components, and other configuration-specific design features. | Includes irrigation-container access components, tubing, clamps, connectors, and associated fluid-path components. K960787 identifies a change in the distal tubing segment from natural rubber latex to SEBS while maintaining the overall device design and intended use. | Similar. Component implementation and configuration-specific characteristics differ; however, both devices use comparable components to establish and control a gravity-fed irrigation fluid pathway. | {9} B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets K254111 Page 5 of 11 | Characteristic | Proposed Device- B. Braun Irrigation/Urology Sets | Predicate Device- Baxter Irrigation Sets, K960787 | Difference/ Substantial Equivalence Assessment | | --- | --- | --- | --- | | Material Composition | The material composition of the components of B. Braun Irrigation/Urology sets are commonly used materials for medical devices such as silicone, ABS, LDPE, High Impact Polystyrene, acrylic, PVC, TPE, PTFE, etc. | SEBS material used in place of latex; other materials not specified. | Similar. The proposed and predicate devices use medical-grade polymeric materials appropriate for their intended function. Differences in material composition do not alter the intended use or fundamental operating principle of the devices. The proposed device materials have been evaluated for biocompatibility. | | Sterilization Process | Ethylene Oxide (EtO) | Ethylene Oxide (EtO), as identified in the submitted predicate comparison/supporting information | Same | | Prescription | Yes | Yes | Same | {10} B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets K254111 Page 6 of 11 Table 2: Configuration-Level Technological Comparison | Proposed Configuration | Proposed Technical Specifications / Features | Predicate (K960787) / Public Baxter Information | Difference | Performance | SE Rationale | | --- | --- | --- | --- | --- | --- | | 1- Two-Line PIC™, 94 in. | 2 fluid paths; overall set 94 in.; two Urology Unispikes X59-952; two 16-in. inlet tubing segments and 66-in. main tubing segment; PIC™ Plastic Irrigation Container compatibility. Tubing ID: 0.281 in. nominal ID (X55-389 principal tubing); 0.211 in. ID (X55-393 transition tubing); 0.194 in. ID (A7208140 distal TPE tubing). | K960787: irrigation-set family; configuration dimensions not enumerated. Current Baxter catalog: Y-Type TUR Series 2C4005, approx. 78 in., for Baxter UROMATIC/ARTHROMATIC plastic irrigation containers. Numeric tubing ID and spike venting are not publicly specified. | Dual-line/Y-type concept and rigid plastic irrigation-container use are similar; overall length differs. | Functional gravity flow; spike piercing/coring; joint integrity; clamp performance; associated-device compatibility, as applicable. | Differences in length and component implementation do not change gravity-fed irrigation principle. Proposed-device verification demonstrates functional flow, connection integrity, and spike performance. | | 2 -TUR/CYSTO, 79 in. | 1 fluid path; 79-in. set; one non-vented spike S5069200 with disk guard; 72-in. primary tubing; large-bore configuration; universal spike-compatible irrigation source. Tubing ID: 0.281 in. nominal ID (X55-389 principal tubing); 0.211 in. ID (X55-393 transition tubing); 0.194 in. ID (A7208140 distal TPE tubing). | K960787: TUR/cystoscopy irrigation-set family; detailed dimensions not enumerated. Current Baxter catalog includes Straight-Type TUR Series 2C4006 (22 in.) and Cysto/Bladder 2C4040 (81 in., straight). Numeric tubing ID and spike venting are not publicly specified. | Same clinical technology; line arrangement and available lengths vary across predicate-family catalog examples. | Functional gravity flow; non-vented spike piercing/coring; joint integrity; clamp performance; associated-device compatibility. | Single-line configuration remains a gravity-fed irrigation pathway. Verified flow, spike, joint, and connection performance support that dimensional/component differences do not raise different questions of safety or effectiveness. | {11} K254111 Page 7 of 11 B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets | Proposed Configuration | Proposed Technical Specifications / Features | Predicate (K960787) / Public Baxter Information | Difference | Performance | SE Rationale | | --- | --- | --- | --- | --- | --- | | 3- Two-Line Flexible Bag, 94 in. | 2 fluid paths; 94-in. set; two non-vented spikes S5069200; two 16-in. inlet tubing segments and 66-in. main tubing segment; flexible IV bag compatibility. Tubing ID: 0.281 in. nominal ID (X55-389 principal tubing); 0.211 in. ID (X55-393 transition tubing); 0.194 in. ID (A7208140 distal TPE tubing). | K960787: irrigation-set family; detailed configuration not enumerated. Current Baxter catalog includes Y-Type TUR Series 2C4005 (78 in.) for Baxter plastic irrigation containers. Numeric tubing ID and spike venting are not publicly specified. | Dual-line/Y-type arrangement is similar; proposed container interface and overall length are configuration-specific. | Functional gravity flow; non-vented spike piercing/coring; joint integrity; clamp performance; associated-device compatibility. | Container-interface and length differences do not change the fundamental irrigation function. Spike and functional testing demonstrate acceptable connection and flow performance. | | 4- Four-Bag Flexible Bag, 96 in. | 4 fluid paths; 96-in. set; four non-vented spikes S5069200; four 19-in. inlet tubing segments and 66-in. main tubing segment; four-line manifold; flexible IV bag compatibility. Tubing ID: 0.281 in. nominal ID (X55-389 principal tubing); 0.211 in. ID (X55-393 transition tubing); 0.194 in. ID (A7208140 distal TPE tubing). | FDA Recall Z-1684-2013 identifies Baxter Product Code 2C4031, Four Lead Arthroscopic Irrigation Set, as a device associated with 510(k) K960787. The FDA record identifies a four-lead configuration, approximate length of 99 in. (2.5 m), for use with ARTHROMATIC Plastic Containers. Numeric predicate tubing ID and spike venting are not publicly specified. | Both devices incorporate a four-line/four-lead irrigation arrangement and have similar overall lengths (96 in. proposed versus approximately 99 in. predicate). The container interface and component implementation are manufacturer specific. Predicate tubing ID and spike venting are not publicly specified. | Functional gravity flow; non-vented spike piercing/coring; joint integrity; clamp performance; particulate testing; associated-device compatibility, as applicable. | The proposed and predicate configurations use four inlet pathways to provide gravity-fed irrigation from multiple irrigation containers. The approximately 3-in. difference in overall length and manufacturer-specific component implementation do not alter the fundamental operating principle. Proposed-device functional, mechanical, spike, and compatibility testing demonstrates that the four-line configuration performs as intended and that the identified differences do not raise different questions of safety or effectiveness. | {12} K254111 Page 8 of 11 B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets | Proposed Configuration | Proposed Technical Specifications / Features | Predicate (K960787) / Public Baxter Information | Difference | Performance | SE Rationale | | --- | --- | --- | --- | --- | --- | | 5- Single-Line PIC™, 79 in. | 1 fluid path; 79-in. set; one Urology Unispike X59-952; 10-in. proximal tubing and 62-in. main tubing segment; PIC™ Plastic Irrigation Container compatibility. Tubing ID: 0.187 in. nominal ID (X55-384 principal tubing); 0.194 in. ID (A7208140 distal TPE tubing). | K960787: irrigation-set family; detailed configuration not enumerated. Current Baxter catalog includes straight irrigation configurations, including 2C4040 at 81 in. and 2C4006 at 22 in., for irrigation use. Numeric tubing ID and spike venting not publicly specified. | Single/straight irrigation pathway is similar; length and container interface are configuration specific. | Spike piercing/coring (PIC spike); joint integrity; clamp performance; associated-device compatibility, as applicable. | Differences in length and spike/container interface accommodate the intended irrigation source without changing the device's fundamental mode of action. | | 6- Single-Line PIC™, 80 in. | 1 fluid path; 80-in. set; one Urology Unispike X59-952; 72-in. primary tubing; PIC™ Plastic Irrigation Container compatibility. Tubing ID: 0.187 in. nominal ID (X55-384 principal tubing); 0.194 in. ID (A7208140 distal TPE tubing) | K960787: irrigation-set family; detailed configuration not enumerated. Current Baxter catalog includes Cysto/Bladder 2C4040 at 81 in. (straight). Numeric tubing ID and spike venting are not publicly specified. | Very similar overall length to a current Baxter straight irrigation-set example; container interface differs by manufacturer system. | Spike piercing/coring (PIC spike); joint integrity; clamp performance; associated-device compatibility, as applicable. | The approximately comparable length and common gravity-fed straight-path architecture support technological similarity; proposed-device testing confirms performance. | {13} B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets K254111 Page 9 of 11 | Proposed Configuration | Proposed Technical Specifications / Features | Predicate (K960787) / Public Baxter Information | Difference | Performance | SE Rationale | | --- | --- | --- | --- | --- | --- | | 7- Single-Line Flexible Bag, 82 in. | 1 fluid path; 82-in. set; one non-vented spike S5069200 with disk guard; 10-in. proximal and 62-in. main tubing segments; flexible IV bag compatibility. Tubing ID: 0.187 in. nominal ID (X55-384 principal tubing); 0.194 in. ID (A7208140 distal TPE tubing). | K960787: irrigation-set family; detailed configuration not enumerated. Current Baxter catalog includes Cysto/Bladder 2C4040 at 81 in. (straight). Numeric tubing ID and spike venting are not publicly specified. | Straight/single-line configuration and overall length are similar; container interface/component details may differ. | Functional gravity flow; non-vented spike piercing/coring; joint integrity; clamp performance; particulate testing; associated-device compatibility, as applicable. | The 82-in. proposed configuration is technologically similar in overall architecture and length to a current Baxter straight irrigation-set example. Testing confirms acceptable flow and mechanical performance. | | 8- Spike Adapter, 5 in. | Accessory; nominal set length 5 in.; Urology Unispike X59-952; tubing X55-384-625 nominal 1.25 in.; spike connector assembly X71-818 with breather; connects with PIC™ Plastic Irrigation Container / flexible IV bag. Tubing ID: 0.187 in. nominal ID (X55-384 tubing). | K960787/public Baxter irrigation-set information does not identify a directly comparable stand-alone spike adapter configuration. | Configuration-specific accessory; no direct comparable public predicate configuration identified. | Spike piercing/coring; joint integrity; associated device compatibility; other applicable component verification. | The spike adapter uses irrigation-container access and fluid-path components consistent with the proposed device family. Applicable performance testing supports container compatibility and functional performance. The adapter does not alter the intended use or fundamental operating principle and does not raise different questions of safety or effectiveness | {14} K254111 Page 10 of 11 B. Braun Medical Inc. 510(k) Premarket Notification Irrigation/ Urology Sets ## Summary of Non-Clinical Testing Performance testing was conducted to evaluate the functional and mechanical performance of the proposed Irrigation/Urology Sets. Testing included, as applicable, visual inspection, occlusion, positive and negative pressure integrity, static and dynamic tensile testing, clamp qualification, priming and gravity flow, spike piercing/coring, associated-device compatibility, particulate testing, and other applicable functional evaluations. Testing was conducted in accordance with applicable standards and established procedures and met the predetermined acceptance criteria: | Biocompatibility | Biocompatibility evaluation in accordance with ISO 10993-1: 2018 “Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process”. Tests included cytotoxicity, sensitization, intracutaneous, Acute systemic toxicity, subacute toxicity, material mediated pyrogen, and hemolysis. | | --- | --- | | Device Performance | Performance testing was conducted in accordance with: • ISO 8536-4 Infusion Equipment for Medical Use – Part 4: Infusion Sets for Single Use, Gravity Feed • ISO 8536-8 — Infusion equipment for medical use — Part 8: Infusion sets for single use with pressure infusion apparatus. • ISO 8536-14 Clamps and Flow Regulators for Transfusion and Infusion Equipment without Fluid Contact • USP <788> Particulate Matter in Injections. Tests included: visual inspection, occlusion, positive air pressure, positive water pressure, negative pressure, static tensile, dynamic tensile, air tightness, fluid tightness, cyclic opening and closing, clamp retention, air shut-off, fluid shut-off, flow consistency, particulate, and chemical testing. The proposed devices are EO-sterilized, labeled for single use, and provided in either Tyvek pouch or blister tray primary packaging with secondary corrugated carton. The shelf life of the proposed devices is 5 years, validated through real-time and accelerated aging studies. The validation program included packaging process validation, simulated distribution, accelerated aging, and real-time aging, as applicable to each sterile barrier system. Package integrity and seal performance were evaluated using applicable recognized consensus standards and controlled internal procedures. | {15} K254111 Page 11 of 11 **B. Braun Medical Inc.** 510(k) Premarket Notification Irrigation/ Urology Sets # **Conclusions** The proposed B. Braun Irrigation/Urology Sets and the predicate device have the similar intended use and the same fundamental operating principle. Differences in technological characteristics, including configuration-specific differences in line count, tubing dimensions, irrigation-container compatibility, spike configuration, and associated components, do not raise different questions of safety or effectiveness. Performance testing demonstrates that the proposed devices perform as intended. Therefore, the proposed B. Braun Irrigation/Urology Sets are substantially equivalent to the Baxter TUR, Arthroscopic, Cystoscopy Irrigation Sets cleared under K960787.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...