← Product Code [BTT](/productcode/BTT) · K254078

# HVT 2.0 (K254078)

_Vapotherm, Inc. · BTT · Apr 13, 2026 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K254078

## Device Facts

- **Applicant:** Vapotherm, Inc.
- **Product Code:** [BTT](/productcode/BTT.md)
- **Decision Date:** Apr 13, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5450
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The HVT 2.0 system is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing neonate/infant, pediatric and adult patients. The device is intended to be used in hospital and sub-acute facility settings. HVT 2.0 is not for use during field transport. The flow rates may be from 2 to 45 liters per minute (BTPS).

## Device Story

HVT 2.0 delivers high-velocity, warmed, humidified respiratory gases to spontaneously breathing patients via small-bore nasal cannula. System comprises capital unit (internal blower, electronics) and Disposable Patient Circuit (DPC) containing Vapor Transfer Cartridge (VTC) and multi-lumen tubing. Device mixes room air with external oxygen; internal blower drives flow; VTC uses semi-permeable polymer tubing to transfer heat/humidity to gas. Operated by clinicians in hospital/sub-acute settings; user sets flow rate, temperature, and FiO2 via interface. Software monitors sensors to maintain target values; audible/visual alarms ensure essential performance. Closed-loop water path in DPC prevents contamination of main unit. Benefits include precise, comfortable respiratory support; facilitates patient transfer while maintaining therapy. Not for field transport or MRI.

## Clinical Evidence

Bench testing only. Performance validated against ISO 80601-2-74 for humidification, oxygen accuracy, temperature stability, and flow rate accuracy. Biocompatibility assessed per ISO 10993-1 and ISO 18562. Electrical safety/EMC tested per IEC 60601-1 and IEC 60601-1-2. Alarms verified per IEC 60601-1-8. Usability and human factors testing performed.

## Technological Characteristics

Blower-driven respiratory gas humidifier; semi-permeable polymer membrane for heat/moisture transfer. Flow range 2-45 L/min (BTPS); temperature 33-39°C. AC/battery powered (Li-Ion). Small-bore nasal cannula interface. Software-controlled with audible/visual alarms. Complies with ISO 80601-2-74, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, and ISO 14971.

## Regulatory Identification

A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.

## Predicate Devices

- Vapotherm Precision Flow ([K072845](/device/K072845.md))
- Vapotherm HVT 2.0 ([K203357](/device/K203357.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 13, 2026

Vapotherm, Inc.
Michael Webb
Vice President of Regulatory Affairs
100 Domain Dr.
Exeter, New Hampshire 03833

Re: K254078
Trade/Device Name: HVT 2.0
Regulation Number: 21 CFR 868.5450
Regulation Name: Respiratory Gas Humidifier
Regulatory Class: Class II
Product Code: BTT
Dated: December 17, 2025
Received: December 18, 2025

Dear Michael Webb:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254078 - Michael Webb
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K254078 - Michael Webb
Page 3

Sincerely,

Ethan L. Nyberg -S

Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K254078  |   |
|  Device Name HVT 2.0  |   |
|  Indications for Use (Describe) The HVT 2.0 system is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing neonate/infant, pediatric and adult patients. The device is intended to be used in hospital and sub-acute facility settings. HVT 2.0 is not for use during field transport. The flow rates may be from 2 to 45 liters per minute (BTPS).  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

32-1

# 510(k) Summary and Certification

Date: 11-March-2026

Company: Vapotherm, Inc.
100 Domain Drive
Exeter, NH 03833, USA

Official Contact: Michael J. Webb, VP of Regulatory Affairs
Tel: 866-410-9986

Proprietary or Trade Name: HVT 2.0

Common/Usual Name: Respiratory gas humidifier

Classification: Class II
21 CFR 868.5450, Product Code: BTT

Predicate Devices: K072845 Vapotherm Precision Flow (Primary)
K203357 Vapotherm HVT 2.0 (Secondary)

Reason for submission: Align indications with Vapotherm Precision Flow

## Device Description:

The HVT 2.0 System is designed to deliver high velocity respiratory therapy via a small-bore nasal cannula to spontaneously breathing neonate/infant, pediatric and adult patients. HVT 2.0 heats and humidifies respiratory gases to provide comfortable, safe, and effective therapy.

The HVT 2.0 System consists of the HVT 2.0 Main Device, also referred to as the capital unit, and a Disposable Patient Circuit (DPC). The HVT 2.0 main unit is portable, self-contained, AC (line voltage) or battery powered with high velocity therapy driven by an internal blower using room air source and supplied with external oxygen. It facilitates patient transfer, while maintaining the levels of heated humidified gas delivered through a small-bore cannula. The unit maintains the integrity of a closed, single-use water path disposable. The main unit contains all the electrical and electronic components and measures and controls water temperature, flow rate and oxygen percentage to reach the user-set target values. The main unit has no water pathway, and the gas pathway contains only gas at room temperature, and consequently does not need internal cleaning or disinfection.

The DPC consists of the Vapor Transfer Cartridge (VTC), Disposable Water Path (DWP) and multi-lumen delivery tubing. The VTC uses semi-permeable small-bore tubing technology to transfer heat and humidity to the inspired gases. Finally, multi-lumen tubing is used to circulate warmed water to provide humidity and body temperature gas from the VTC to the patient. This is the same as the predicate devices.

{5}

32-2

# Indications for Use:

The HVT 2.0 system is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing neonate/infant, pediatric and adult patients. The device is intended to be used in hospital and sub-acute facility settings.

HVT 2.0 is not for use during field transport. The flow rates may be from 2 to 45 liters per minute (BTPS).

# Intended Patient Population:

Neonate/infant, pediatric and adult patients

# Environment of Use:

Hospital and sub-acute facility settings

# Contraindications:

- Not appropriate for patients who are not spontaneously breathing, are unable to protect their airway, or have anatomic or injury induced blockage of the nasal pathway to the nasopharyngeal space.
- Not for treating OSA and snoring.
- HVT 2.0 is not for field transport.
- HVT 2.0 is MRI unsafe. Do not use it in an MR environment.
- Not for use with an Oxygen Concentrator (when treating respiratory distress).

# Basis for Determination of Substantial Equivalence:

The expansion in indications for the proposed device falls within the already cleared indications for the primary predicate. The subject device and predicate have the same intended use, and any minor differences in technological characteristics do not raise different questions of safety and effectiveness based on the performance data. The devices are substantially equivalent.

The following table compares the technological characteristics of the HVT 2.0 subject device to the Vapotherm primary predicate Precision Flow, and where appropriate, to itself as the secondary predicate.

{6}

|   |   | Predicate Devices Comparison Table and Discussion  |   |   |
| --- | --- | --- | --- | --- |
|  Characteristic | Primary Predicate: Precision Flow K072845 | Secondary Predicate: HVT 2.0 K203357 | Subject Device: | Discussion:  |
|  Indicatio ns for Use | Precision Flow® HVNI is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, subacute institutions settings, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula. | The HVT 2.0 system is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing adult and pediatric patients (5 kg and up). The device is intended to be used in hospital, sub-acute facility, and home-use settings. | The HVT 2.0 system is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing neonate/infant, pediatric and adult patients. The device is intended to be used in hospital and sub-acute facility settings. HVT 2.0 is not for use during field transport. The flow rates may be from 2 to 45 liters per minute (BTPS). | Equivalent Patient populations are equivalent to K072845. Flow rate range is a subset of K072845. Environments of use for subject device are a subset of the predicates.  |
|  Patient Populatio n | Neonate, Infant, Pediatric, Adult | Adult and pediatric (5 kg and up) | Neonate, Infant, Pediatric, Adult | Identical to K072845  |
|  Product Codes | BTT | BTT | BTT | Identical  |
|  Environ ment of Use for BTT | Hospital, sub-acute and homecare settings | Hospital, sub-acute and homecare settings | Hospital and sub-acute settings | Equivalent Subject device is a subset of the predicate  |
|  Patient Consciou sness | Spontaneously breathing patients | Spontaneously breathing patients | Spontaneously breathing patients | Identical  |
|  Duration of Use | Disposable can be used for up to 30 days single patient use. | Disposable can be used for up to 30 days single patient use. | Disposable can be used for up to 30 days single patient use. | Identical  |
|  Prescripti ve | Rx Only | Rx Only | Rx Only | Identical  |
|  Patient Interface | Small bore nasal cannula or Trach Adapter | Vapotherm ProSoft small bore nasal cannula or Trach Adapter | Vapotherm ProSoft small bore nasal cannula or Trach Adapter | Equivalent to both predicates  |

{7}

|   |   | Predicate Devices Comparison Table and Discussion  |   |   |
| --- | --- | --- | --- | --- |
|  Characteristic | Primary Predicate: Precision Flow K072845 | Secondary Predicate: HVT 2.0 K203357 | Subject Device: | Discussion:  |
|  Flow Range | 1 to 40 L/Min (SLPM), converted as: 1.1 to 45 L/Min (BTPS) per ISO 80601-2-74 | 3-45 L/Min (BTPS) | 2-45 L/Min (BTPS) | Equivalent Subject Device flow range falls within cleared predicate K072845  |
|  Patient contacting materials | Externally communicating, tissue, permanent duration | Externally communicating, tissue, permanent duration | Externally communicating, tissue, permanent duration | Identical  |
|  Temperature Range | 33°C to 39°C | 33°C to 39°C | 33°C to 39°C | Identical  |
|  Temperature Alarm | Alarms if 2°C above set point. | Alarms if 2°C above set point. | Alarms if 2°C above set point. | Identical  |
|  Temperature Accuracy | ±2 °C | ±2 °C | ±2 °C | Identical  |
|  FiO2 Range | 0.21 to 1.00 | 0.21 to 1.00 | 0.21 to 1.00 | Identical  |
|  Oxygen Source | Wall or tank | Wall, tank, or Oxygen Concentrator | Wall, tank, or Oxygen Concentrator | Identical to K203357  |
|  Oxygen Accuracy | ±2% | ±2% (wall or tank) ±4% Oxygen Concentrator | ±2% (wall or tank) ±4% Oxygen Concentrator | Identical to K203357  |
|  Operating Principle | Air flow via a compressor and humidification via semi-permeable polymer technology which allows for delivery of entrained, humidified gases at constant flow to the patient | Air flow via a blower and humidification via semi-permeable polymer technology which allows for delivery of entrained, humidified gases at constant flow to the patient | Air flow via a blower and humidification via semi-permeable polymer technology which allows for delivery of entrained, humidified gases at constant flow to the patient | Identical to K203357  |
|  Alarms | Extensive audible and visual alarms to ensure essential performance of the device is maintained. | Extensive audible and visual alarms to ensure essential performance of the device is maintained. | Extensive audible and visual alarms to ensure essential performance of the device is maintained. | Identical to K203357  |
|  User interface settings | User sets flow rate, temperature, and oxygen fraction F_{i}O_{2} | User sets flow rate, temperature, and oxygen fraction F_{i}O_{2} | User sets flow rate, temperature, and oxygen fraction F_{i}O_{2} | Identical to K203357  |
|  Control | Software control | Software control | Software control | Identical  |
|  Modes of operation | Sleep, Standby and Run Modes | Sleep, Standby and Run Modes | Sleep, Standby and Run Modes | Identical  |
|  Gas/Air input | Standard DISS non-interchangeable fitting for oxygen, room air | Standard DISS non-interchangeable | Standard DISS non-interchangeable | Identical to K203357  |

{8}

|   |   | Predicate Devices Comparison Table and Discussion  |   |   |
| --- | --- | --- | --- | --- |
|  Characteristic | Primary Predicate: Precision Flow K072845 | Secondary Predicate: HVT 2.0 K203357 | Subject Device: | Discussion:  |
|   | via wall source or compressor | fitting for oxygen, room air via blower | fitting for oxygen, room air via blower |   |
|  Humidification | Via semi-permeable small-bore tubing | Via semi-permeable small-bore tubing | Via semi-permeable small-bore tubing | Identical to K203357, Complies with ISO 80601-2-74  |
|  Power Requirements | 100-240 VAC, 50-60Hz
Back-up power: 4.8V nickel-metal hydride battery pack

Back-up power with Vapotherm Transfer Unit (VTU): Medipower™ unit, one hour limited use | 110 to 240VAC, 50 to 60Hz
Back-up power (Safety Battery): Li-Ion Battery 14.4V, 2900mAh, 99.Wh.

Back-up power with Transfer Battery: Li-Ion Battery Pack consisting of 2X (14.4V, 2900mAh, 99.4Wh), one hour limited use | 110 to 240VAC, 50 to 60Hz
Back-up power (Safety Battery): Li-Ion Battery 14.4V, 2900mAh, 99.Wh.

Back-up power with Transfer Battery: Li-Ion Battery Pack consisting of 2X (14.4V, 2900mAh, 99.4Wh), one hour limited use | Identical to K203357  |

From the comparisons above, the subject and predicate devices have the same intended use. Any minor differences do not raise different questions of safety or effectiveness.

## Non-clinical performance testing

The subject device was designed and validated in accordance with the application recommendations of FDA guidance documents and requirements of international standards including:

## Biocompatibility / Materials:

Biocompatibility was assessed in accordance with FDA guidance, Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", including the ISO 18562 series Biocompatibility evaluation of breathing gas pathways in healthcare applications. Testing of the patient-contacting parts of the HVT 2.0 System demonstrates an appropriate biocompatibility profile for the device.

## Electrical Safety, Alarms, RFID:

Electrical safety and electromagnetic compatibility testing was conducted in accordance with IEC 60601-1:2005 +AMD1:2012, +AMD2:2020, and IEC 60601-1-2: 2014, +A1:2020 to demonstrate the basic safety, essential performance and emissions and immunity characteristics of the device. The testing demonstrated the appropriate electrical safety and electromagnetic

32-5

{9}

compatibility profile for the device. Alarms testing was performed compliant with IEC 60601-1-8:2006 +AMD1:2012, +AMD2:2020. RFID testing was successfully completed in accordance with AIM 7351731.

## Software Verification, Validation Testing, and Hazard Analysis

Software verification and validation testing was conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”, as well as “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”. Hazard Analysis/Risk Management activities for the device comply with ISO 14971:2019 Medical Devices – Application of risk management to medical devices.

## Bench / Performance Testing –

Comparative performance testing was performed to demonstrate substantial equivalence to the predicate device. Given the HVT 2.0 was previously cleared under K203357 and K221318, prior performance testing continues to apply and includes:

- Oxygen accuracy/Blender Performance
- Temperature accuracy/Thermal Stability
- Flow rate accuracy
- Humidification output ISO 80601-2-74
- Patient contacting surface temperature/Thermal Safety
- Volume of liquid exiting the humidification chamber outlet
- Continuous use (Use Lifespan)
- Operating environment
- Nurse call compatibility
- Usability/Human Factors
- Cleaning Validation (for hardware)

## Substantial Equivalence Conclusion

The intended use of the subject and predicate devices are the same and the technological characteristics are similar or equivalent. Any minor differences do not raise different questions of safety and effectiveness. The testing performed and documentation provided demonstrates that the subject device is substantially equivalent to the predicate devices.

32-6

---

**Source:** [https://fda.innolitics.com/device/K254078](https://fda.innolitics.com/device/K254078)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
