The hemoclip is indicated for endoscopic clip placement within the digestive tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1. Endoscopic marking, 2. Hemostasis for: - Mucosal/sub-mucosal defects <3cm - Bleeding ulcers - Arteries<2mm - Polyps<1.5cm in diameter - Diverticula in the colon 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel. 4. As a supplementary method, closure of GI tract luminal perforations<20mm that can be treated conservatively.
Device Story
Hemoclip is a single-use, EO-sterilized endoscopic device consisting of a clip and a release mechanism (torque wire, spring tube, handle). Operated by a physician via a flexible endoscope; the device is inserted through the working channel. For 511 series, a swinging mechanism allows clips to adjust to tissue contact via a pin-fin tube. The operator uses the sliding handle to deploy and close the clip on target tissue. The device provides mechanical pressure for hemostasis, marking, anchoring, or closure of perforations. It is not MRI compatible.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993), sterility (ISO 11135), and performance testing for swinging functionality and claw angles.
Technological Characteristics
Materials: PE, stainless steel. Principle: Mechanical clip deployment via flexible endoscope. Features: 511 series includes swinging mechanism; 543 series is standard. Dimensions: Claw angles 90°, 135°. Sterilization: Ethylene Oxide (EO). Single-use. MR Unsafe.
Indications for Use
Indicated for adult patients requiring endoscopic clip placement in the digestive tract for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula), anchoring jejunal feeding tubes, or closure of GI luminal perforations <20mm.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
August 10, 2026
Hangzhou AGS MedTech Co., Ltd.
Yangping Fu
RA
Bldg. 2, 5 And 7, #389 Xingzhong Rd., Linping District
Hangzhou, Zhejiang 311103
China
Re: K254069
Trade/Device Name: Hemoclip (511 series, 543 series)
Regulation Number: 21 CFR 876.4400
Regulation Name: Hemorrhoidal Ligator
Regulatory Class: Class II
Product Code: PKL
Dated: July 10, 2026
Received: July 10, 2026
Dear Yangping Fu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254069 - Yangping Fu
2
Page
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254069 - Yangping Fu
3
Page
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
SIVAKAMI VENKATACHALAM -S
for
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K254069
Device Name
Hemoclip (511 series, 543 series)
Indications for Use (Describe)
The hemoclip is indicated for endoscopic clip placement within the digestive tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of :
1. Endoscopic marking,
2. Hemostasis for:
- Mucosal/sub-mucosal defects <3cm
- Bleeding ulcers
- Arteries<2mm
- Polyps<1.5cm in diameter
- Diverticula in the colon
3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
4. As a supplementary method, closure of GI tract luminal perforations<20mm that can be treated conservatively.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
AGS
CH2.03 510(k) Summary
Hemoclip
K254069
Page 1 of 7
### 510(k) Summary
We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.
CH2.03.1 Submitter
| Submitted by/Owner: | Hangzhou AGS MedTech Co., Ltd. Building 2, 5 and 7, No.389 Xingzhong Road, Linping District, 311103 Hangzhou, Zhejiang, PEOPLE'S REPUBLIC OF CHINA |
| --- | --- |
| Establishment Registration Number: | 3010288205 |
| Registration Status: | Active |
| Contact Person: | Yanping Fu Building 2, 5 and 7, No.389 Xingzhong Road, Linping District, 311103 Hangzhou, Zhejiang, PEOPLE'S REPUBLIC OF CHINA Phone: +86-0571-87671223 Fax: +86-0571-87671225 Email: fuyp@bioags.com |
| Date Prepared: | July 6, 2026 |
CH2.03.2 Proposed Device
| Trade Name: | / |
| --- | --- |
| Device Name: | Hemoclip (511 series, 543 series) |
| Common Name: | Hemoclip |
| Regulation class: | Class II |
| Regulation Number: | 876.4400 |
| Regulation Description: | Hemorrhoidal ligator. |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | PKL |
| Common Name: | Hemostatic Metal Clip For The GI Tract |
CH2.03.3 Predicate Device
| Trade Name: | Lockado™ Repositionable Hemostasis Clip |
| --- | --- |
| Device Name: | Lockado™ Repositionable Hemostasis Clip |
| Common Name: | Hemostasis Clip |
| 510(k) Number: | K202333 |
| Regulation class: | Class II |
| Regulation Number: | 876.4400 |
| Regulation Description: | Hemorrhoidal ligator. |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | PKL |
CH2.03-2
{5}
AGS
CH2.03 510(k) Summary
Hemoclip
K254069
Page 2 of 7
| Product Code Name: | Hemostatic Metal Clip For The Gi Tract |
| --- | --- |
### CH2.03.4 Device Description
Hemoclip consists of Clip part and Release part.
For 511 series, clip part consists of Clip, Frap tube and Pin-fin tube. Release part consists of Torque wire, Spring end, Plastic coated spring tube, Sliding handle and Handle assembly. EO Sterilization and use for single use only.
For 543 series, clip part consists of Clip and Frap tube. Release part consists of Spring end, Plastic coated spring tube, Sliding handle and Handle assembly. EO Sterilization and use for single use only.
### CH2.03.5 Indication for use statement
The hemoclip is indicated for endoscopic clip placement within the digestive tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of:
1. Endoscopic marking,
2. Hemostasis for:
- Mucosal/sub-mucosal defects <3cm
- Bleeding ulcers
- Arteries<2mm
- Polyps<1.5cm in diameter
- Diverticula in the colon
3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
4. As a supplementary method, closure of GI tract luminal perforations<20mm that can be treated conservatively.
### CH2.03.6 A description of the accessories
For this proposed device, no other medical devices and no other products are provided with it as accessories.
In clinical use, the distal-end of the product needs to enter the digestive tract through the working channel of the straight flexible endoscope, though there's no physical connection.
The distal-end of the 5431 series needs to enter the digestive tract through the working channel of the straight or side viewing flexible endoscope, though there's no physical connection.
And our proposed device Hemoclip have been designed to be used individually.
### CH2.03.7 Comparison of Technology Characteristics
Our proposed device Hemoclip is substantially equivalent to the predicate devices.
The differences between the Hemoclip and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below:
CH2.03-3
{6}
AGS
K254069
Page 3 of 7
CH2.03 510(k) Summary
Hemoclip
Table CH2.03.7 Comparison of technical characteristics
| Item | | Proposed device Hangzhou AGS MedTech Co., Ltd. | Predicate device (K202333) Micro-Tech (Nanjing) Co., Ltd. | Comparison |
| --- | --- | --- | --- | --- |
| Clinical | Intended use | The clip is compatible with endoscope, which is indicated for clip placement within digestive tract for the purpose of mechanical pressure treatment of bleeding of small arteries and pulsation. The device is intended for single use. | / | Different. The predicate device only indicates "Indications for use" in the IFU. The intended use of the proposed device is same as the reference device (K252271). |
| Technical | Principle of operation | Principle of Swinging performance(only for 5111 series) When one of the clips comes into contact with the tissue, both clips will swing around the pin-fin tube under the action of an external force. The clip part is attached to the tissue by swinging. Then, by pulling the sliding handle, the clip component completes the closure of the wound in the swinging position. | / | Different. Proposed device has the swinging principle while the predicate device doesn't have. We conduct the Bench test to confirm the effectiveness and safety of swinging performance. |
| | Claw angle | 90°, 135° | Unknown | Different. Our already cleared 510k K211787 has claw angle 90° and 135°. Claw angle test been |
CH2.03-4
{7}
AGS
K254069
Page 4 of 7
CH2.03 510(k) Summary
Hemoclip
| Item | Proposed device Hangzhou AGS MedTech Co., Ltd. | Predicate device (K202333) Micro-Tech (Nanjing) Co., Ltd. | Comparison |
| --- | --- | --- | --- |
| | | | done for proposed device. Please refer to CH3.08. |
| Performance | Swinging performance(only for 5111 series) | / | Different. Proposed device has Swinging performance. We conduct the Bench test to confirm the effectiveness and safety for swinging performance. |
| MRI Conditional | MR Unsafe | MRI Safety Information MR Conditional A person with the Lockado™ Repositionable Hemostasis Clip may be safely scanned under the following conditions. Failure to follow these conditions may result in injury. | Different. The proposed device can't be used in MRI environments and we have clarified in the IFU, Label and Patient implant card. Therefore, its safety risks are lower. Please refer to CH5.02_Package Labeling and CH5.03_Instruction for |
| | | Device Name Lockado™ Repositionable Hemostasis Clip | |
| | | Static Magnetic Field Strength (B0) 1.5 T or 3.0 T | |
CH2.03-5
{8}
[Non-Text]
AGS
K254069
Page 5 of 7
CH2.03 510(k) Summary
Hemoclip
| Item | | Proposed deviceHangzhou AGS MedTech Co., Ltd. | Predicate device (K202333)Micro-Tech (Nanjing) Co., Ltd. | | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | Maximum Spatial Field Gradient | 40 T/m or 4000 gause/cm | use |
| | | | RF Excitation | Circularly Polarized (CP) | |
| | | | RF Transmit Coil Type | Volume RF body coil | |
| | | | Operating Mode | Normal Operating Mode | |
| | | | Maximum Whole-Body SAR | 2 W/kg (normal operating mode) | |
| | | | Maximum Head SAR | 3.2 W/kg (first level control mode) | |
| | | | Scan Duration | 2 W/kg whole-body average SAR for 60 minutes of continuous RF with less than 2 degrees temperature rise | |
| | | | MR Image Artifact | The pressure of this implant may produce an image artifact at 25mm away from the device. | |
| | | | Precaution: It is recommended that healthcare providers distribute patient implant cards with the name of the clip and date it was placed. | | |
| Biological | Main Patient contact Material | PE, stainless steel | PE, stainless steel | | Different.The composition of the predicate device's |
CH2.03-6
{9}
AGS
K254069
Page 6 of 7
CH2.03 510(k) Summary
Hemoclip
| Item | Proposed device Hangzhou AGS MedTech Co., Ltd. | Predicate device (K202333) Micro-Tech (Nanjing) Co., Ltd. | Comparison |
| --- | --- | --- | --- |
| | | | stainless steel and polymer are unknown. Comparison of biocompatibility equivalence have been done for the proposed device. Biological risks are acceptable. Please refer to CH3.05.06. |
| Shelf Life | 3 years | 2 years | Different. Shelf life verification have been done for the proposed device. Please refer to CH3.07. |
### CH2.03.8 Applicable Guidance Document
NA
### CH2.03.9 Performance Data
The Hemoclip meets all design specifications and medical device standards for biocompatibility (ISO 10993) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device.
CH2.03-7
{10}
[LOGO]
AGS
K254069
Page 7 of 7
CH2.03 510(k) Summary
Hemoclip
### CH2.03.10 Clinical Test
No Clinical test is included in this submission.
### CH2.03.11 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS MedTech Co., Ltd has demonstrated that proposed device Hemoclip is substantially equivalent to the predicate devices.
CH2.03-8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.