Access anti-HBc IgM

K254059 · Beckman Coulter, Inc. · SEI · Mar 13, 2026 · Microbiology

Device Facts

Record IDK254059
Device NameAccess anti-HBc IgM
ApplicantBeckman Coulter, Inc.
Product CodeSEI · Microbiology
Decision DateMar 13, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3173
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K254059 · Mar 13, 2026Access anti-HBc IgMBeckman Coulter, Inc.Retrospective clinical specimens from intended use populationRetrospective clinical specimens were used to evaluate the positive percent agreement (PPA) and negative percent agreement (NPA) of the Access anti-HBc IgM assay compared to FDA-approved reference assays.Retrospective clinical study; Clinical performance evaluation; Diagnostic accuracy

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Multi-center clinical study (retrospective cohort); Retrospective clinical performance evaluationIndividuals with signs and symptoms of hepatitis (including pediatric and adult populations); Sample Size: 2,537 total specimens (includes both prospective and retrospective); Number of Sites: 3FDA approved reference HBc IgM assaysPositive Percent Agreement (PPA) and Negative Percent Agreement (NPA)

Indications for Use

The Access anti-HBc IgM assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in human pediatric (3 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K2) EDTA, tripotassium (K3) EDTA, sodium citrate, acid citrate dextrose (ACD), and citrate phosphate dextrose (CPD)] using the DxI 9000 Access Immunoassay Analyzer. The Access anti-HBc IgM assay results may be used as an aid in the laboratory diagnosis of acute or recent hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis, when used in conjunction with other serological and clinical information. The Access anti-HBc IgM assay is for use on the DxI 9000 Access Immunoassay Analyzer only. This assay is not intended for the screening of blood, plasma, and cell or tissue donors.

Device Story

Access anti-HBc IgM is a two-step enzyme immunoassay for the DxI 9000 Access Immunoassay Analyzer. Input: human serum or plasma samples. Operation: paramagnetic particles coated with anti-human IgM monoclonal antibody capture IgM from the sample; unbound material is washed away. HBc antigen complexed to alkaline phosphatase-conjugated anti-HBc monoclonal antibody is added; conjugate binds to captured IgM. A second wash removes unbound conjugate. Chemiluminescent substrate is added; light output is measured via luminometer. Output: qualitative result (reactive/nonreactive) based on comparison to a stored cutoff value. Used in clinical laboratories by technicians. Results assist physicians in diagnosing acute or recent HBV infection when combined with other clinical/serological data.

Clinical Evidence

Multi-center study of 2,537 specimens (2,281 prospective, 256 retrospective) from pediatric and adult populations (signs/symptoms or increased risk). Primary endpoints: PPA and NPA against a Composite Reference Method (CRM). Overall PPA 50.0% (3/6) and NPA 99.96% (2,274/2,275) in prospective cohort. Retrospective cohort PPA 95.02% (191/201). Bench testing included imprecision (CLSI EP05-A3), interference (CLSI EP07), cross-reactivity, matrix equivalence, and hook effect studies. No significant interference or cross-reactivity observed.

Technological Characteristics

Two-step paramagnetic particle chemiluminescent immunoassay. Reagents: liquid, ready-to-use. Detection: automated luminometer. Sample types: serum, plasma (heparin, EDTA, citrate). Calibration: 56-day frequency. Software: automated qualitative result determination based on stored calibration. Sterilization: not specified.

Indications for Use

Indicated for pediatric (3-21 years) and adult patients with signs/symptoms of hepatitis to aid in the laboratory diagnosis of acute or recent hepatitis B virus (HBV) infection. Not for donor screening.

Regulatory Classification

Identification

A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include: (i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors. (ii) A detailed explanation of the principles of operation and procedures for performing the assay. (iii) A detailed explanation of the interpretation of results. (iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate: (A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected. (B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection. (C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results. (D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures. (E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay. (F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay. (v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include: (A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate ( *e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate ( *e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following: (i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration. (ii) Documentation and characterization ( *e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance. (iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability. (v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range. (vi) All stability protocols, including acceptance criteria. (vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays. (viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate. (ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses. (x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review. (xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas. (3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply: (i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate. (ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K254059 B Applicant Beckman Coulter, Inc. C Proprietary and Established Names Access anti-HBc IgM D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | SEI | Class II | 21 CFR 866.3173 Hepatitis B Virus (HBV) Antibody Assay | Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain clearance for a new device, Access anti-HBc IgM. B Measurand: Hepatitis B virus core antigen IgM antibody (anti-HBc IgM) C Type of Test: The Access anti-HBc IgM assay is a chemiluminescent 2-step enzyme immunoassay (CLIA). ## III Intended Use/Indications for Use: Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} A Intended Use(s): See Indications for Use below. B Indication(s) for Use: Access anti-HBc IgM assay: The Access anti-HBc IgM assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in human pediatric (3 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K2) EDTA, tripotassium (K3) EDTA, sodium citrate, acid citrate dextrose (ACD), and citrate phosphate dextrose (CPD)] using the DxI 9000 Access Immunoassay Analyzer. The Access anti-HBc IgM assay results may be used as an aid in the laboratory diagnosis of acute or recent hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis, when used in conjunction with other serological and clinical information. The Access anti-HBc IgM assay is for use on the DxI 9000 Access Immunoassay Analyzer only. This assay is not intended for the screening of blood, plasma, and cell or tissue donors. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: DxI 9000 Access Immunoassay Analyzer IV Device/System Characteristics: A Device Description: Access anti-HBc IgM The Access anti-HBc IgM assay is a two-step enzyme immunoassay for qualitative detection of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in serum or plasma samples using Dxl 9000 Access Immunoassay Analyzer. The Access anti-HBc IgM assay requires Access anti-HBc IgM reagent pack, Access anti-HBc IgM Calibrator (C1) and Access anti-HBc IgM Quality Controls (QC1-QC2). B Principle of Operation: Paramagnetic particles coated with anti-human IgM monoclonal antibody and prediluted sample are added to a reaction vessel. After incubation, material bound to the solid phase are held in a magnetic field while unbound materials are washed away. HBc antigen complexed to anti-HBc monoclonal antibody alkaline phosphatase conjugate is added and the conjugate binds to the IgM antibodies captured on magnetic particles. A second separation and wash step remove unbound conjugate. A chemiluminescent substrate is then added to the vessel and light generated by the reaction is measured with a luminometer. The light production is compared to the cutoff value defined Kxxxxxx - Page 2 of 18 {3} during calibration of the instrument. The qualitative assessment is automatically determined from a stored calibration. ## Interpretation of Results: Test results are determined automatically by the system software. Results (Signal/Cutoff (S/CO)) are reported as reactive or nonreactive as follows | Result (S/CO) | Interpretation | Reporting Instructions | | --- | --- | --- | | < 1.00 | Nonreactive | Report results as nonreactive for anti-HBc IgM antibodies | | ≥ 1.00 | Reactive | Report results as reactive for anti-HBc IgM antibodies | ## C Instrument Description Information: 1. Instrument Name: DxI 9000 Access Immunoassay Analyzer 2. Specimen Identification: Specimen identification can be done by a bar code or by manual entry. Bar code labels are read by the analyzer or when ordered by the LIS or middleware. Orders that are entered manually include the rack ID and position, so bar code labels are not necessary for sample identification. 3. Specimen Sampling and Handling: No specific requirements for specimen handling for the analyzer are needed. Each specific reagent package insert contains sample handling instructions. 4. Calibration: The Access anti-HBc IgM Calibrator, provided at one level, is used to establish calibration for the Access anti-HBc IgM assay on the DxI 9000 Access Immunoassay Analyzer. An active calibration is required for all tests. For the Access anti-HBc IgM assay, calibration is required every 56 days. 5. Quality Control: The Access anti-HBc IgM QC kit contains QC1 (negative for anti-HBc IgM) and QC2 (positive for anti-HBc IgM). ## V Substantial Equivalence Information: A Predicate Device Name(s): ARCHITECT CORE-M Kxxxxxx - Page 3 of 18 {4} B Predicate 510(k) Number(s): P060035 # C Comparison with Predicate(s): | Device & Predicate Device(s): | Candidate Device K254059 | Predicate Device P060035 | | --- | --- | --- | | Device Trade Name | Access anti-HBc IgM | ARCHITECT CORE-M | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | The Access anti-HBc IgM assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in human pediatric (2 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K2) EDTA, tripotassium (K3) EDTA, sodium citrate, acid citrate dextrose (ACD), and citrate phosphate dextrose (CPD)] using the DxI 9000 Access Immunoassay Analyzer. The Access anti-HBc IgM assay results may be used as an aid in the laboratory diagnosis of acute or recent hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis, when used in conjunction with other serological and clinical information. The Access anti-HBc IgM assay is for use on the DxI 9000 Access Immunoassay Analyzer only. This assay is not intended for the screening of blood, plasma, and cell or tissue donors. | The ARCHITECT CORE-M assay is a chemiluminescent microparticle immunoassay (CMIA) used for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) in human adult and pediatric serum or plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum on the ARCHITECT i System. The ARCHITECT CORE-M assay is to be used as an aid in the diagnosis of acute or recent hepatitis B virus (HBV) infection in conjunction with other laboratory results and clinical information. | | Calibrator Intended Use | The Access anti-HBc IgM Calibrator is intended to calibrate the Access anti-HBc IgM assay for the in vitro qualitative detection of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in human serum and plasma using the DxI 9000 Access Immunoassay Analyzer. | The ARCHITECT CORE-M Calibrators are used for the calibration of the ARCHITECT i System when the system is used for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) using the ARCHITECT CORE-M Reagent Kit. The performance of the ARCHITECT CORE-M Calibrators has not been established with any other IgM anti-HBc assays. | Kxxxxxx - Page 4 of 18 {5} Kxxxxxx - Page 5 of 18 | | | | | --- | --- | --- | | QC Intended Use | The Access anti-HBc IgM QC is intended for monitoring system performance of the Access anti-HBc IgM assay. The Access anti-HBc IgM QC is for use on the DxI 9000 Access Immunoassay Analyzer. | The ARCHITECT CORE-M Controls are used for monitoring the performance of the ARCHITECT i System when used for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) in human adult serum and plasma when using the ARCHITECT CORE-M Reagent Kit. The performance of the ARCHITECT CORE-M Controls has not been established with any other IgM anti-HBc assays. | | Analyte Measured | Same | Anti-HBc IgM antibodies | | Assay Type | Same | Qualitative | | Detection Method | Automated, Chemiluminescence immunoassay | Automated, Chemiluminescent microparticle immunoassay (CMIA) | | Reagent, Calibrator and QC format | Same | Liquid, ready-to-use | | Control(s) | Same | 2 levels Negative Positive | | Sample Type | Same | Serum and Plasma | | | | | | **General Device Characteristic Differences** | | | | Compatible Anticoagulants | Serum, Serum separator tube, Plasma [Lithium Heparin, Lithium Heparin separator tube, Dipotassium (K2) EDTA, Tripotassium (K3) EDTA Sodium Citrate, Acid Citrate Dextrose (ACD),Citrate Phosphate Dextrose (CPD)] | Human serum, plasma (dipotassium EDTA, Lithium Heparin, Sodium Heparin) | | Sample Volume | 10 μL+ dead volume | 64 μL + dead volume | | Instrument | DxI 9000 Access Immunoassay Analyzer | ARCHITECT i Systems | | Traceability/Standardization | Calibrator is traceable to the manufacturer’s working calibrator. | Calibrator 2 is traceable to the Reference Standard of the Paul Ehrlich Institute, HBc Reference serum IgM 84 (IgM anti-HBc). | | Time to Result | ~ 29 minutes | ~ 30 minutes | | Calibration Frequency | 56 days | 30 days | {6} | Purpose | Aid in the laboratory diagnosis of acute or recent hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis, when used in conjunction with other serological and clinical information | Aid in the diagnosis of acute or recent hepatitis B virus (HBV) infection in conjunction with other laboratory results and clinical information | | --- | --- | --- | # VI Standards/Guidance Documents Referenced: 1. EP05-A03: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition 2. EP07-3rd: Interference Testing in Clinical Chemistry - Third Edition 3. EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition 4. EP21-A2: Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Approved Guideline - Second Edition 5. EP25: Evaluation of Stability of In vitro Medical Laboratory Test Reagents - Second Edition 6. EP37: Supplemental Tables for Interference Testing in Clinical Chemistry - First Edition 7. GP44-A4: Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline-Fourth Edition 8. EP12-A02: User protocol evaluation of qualitative test performance 9. ISO 14971: Medical devices - Application of risk management of medical devices - Third Edition; 2019. 10. ISO 15223-1: Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements - Fourth Edition; 2021. 11. ISO 17511: In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators trueness control materials and human samples - second edition; 2020. # VII Performance Characteristics (if/when applicable): # A Analytical Performance: 1. Precision/Reproducibility: # A. Within-Laboratory precision: Within-laboratory precision was evaluated for twenty days on three different DxI 9000 Access Immunoassay Analyzer at one internal site. An eight-member panel containing four serum samples (S1-S4) and four plasma samples (P1-P4) were run at two runs per day with three replicates per sample using three reagent lots of Access anti-HBc IgM and two lots of Access anti-HBc IgM calibrators. Calibration was conducted at the beginning of the study. Table 1 presents the results for within-laboratory precision study. Table 1. Within-Laboratory precision results of the Access anti-HBc IgM Kxxxxxx - Page 6 of 18 {7} | Sample | N | Mean (S/CO) | Repeatability (Within-Run) | | Between-Run | | Between-Day | | Between- Reagent Lot | | Between- Instrument | | Within Laboratory | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV | | QC1 | 2,160 | 0.01 | 0.000 | N/A | 0.001 | N/A | 0.001 | N/A | 0.001 | N/A | 0.000 | N/A | 0.002 | N/A | | QC2 | 2,160 | 2.84 | 0.062 | 2.2 | 0.052 | 1.8 | 0.039 | 1.4 | 0.047 | 1.7 | 0.022 | 0.8 | 0.104 | 3.7 | | S1 | 2,160 | 0.01 | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.001 | N/A | 0.000 | N/A | 0.001 | N/A | | S2 | 2,160 | 0.73 | 0.018 | 2.5 | 0.018 | 2.5 | 0.008 | 1.1 | 0.018 | 2.5 | 0.006 | 0.9 | 0.033 | 4.5 | | S3 | 2,160 | 1.13 | 0.028 | 2.5 | 0.033 | 2.9 | 0.015 | 1.3 | 0.036 | 3.1 | 0.01 | 0.9 | 0.059 | 5.2 | | S4 | 2,160 | 4.45 | 0.102 | 2.3 | 0.097 | 2.2 | 0.063 | 1.4 | 0.092 | 2.1 | 0.034 | 0.8 | 0.183 | 4.1 | | P1 | 2,160 | 0.01 | 0.001 | N/A | 0.000 | N/A | 0.000 | N/A | 0.001 | N/A | 0.000 | N/A | 0.001 | N/A | | P2 | 2,160 | 0.71 | 0.019 | 2.7 | 0.015 | 2.1 | 0.014 | 2 | 0.018 | 2.5 | 0.004 | 0.5 | 0.034 | 4.7 | | P3 | 2,160 | 1.08 | 0.03 | 2.8 | 0.018 | 1.7 | 0.031 | 2.9 | 0.038 | 3.6 | 0.008 | 0.8 | 0.061 | 5.7 | | P4 | 2,160 | 3.84 | 0.099 | 2.6 | 0.054 | 1.4 | 0.085 | 2.2 | 0.11 | 2.9 | 0.047 | 1.2 | 0.185 | 4.8 | Note: %CV are not meaningful when S/CO approaches zero. Results are noted as N/A. P = Plasma, S = Serum, N/A = Not applicable, N = number of replicates, QC = Quality control Within-laboratory precision of the Access anti-HB IgM assay on the Automation Line DxA 5000 total laboratory automation system coupled to the DxI 9000 Access Immunoassay Analyzer was evaluated at one internal site on one DxI 9000 Analyzer. The same panel of samples, used in the within-laboratory precision study, were tested in 3 replicates per run, 2 runs per day, for 20 days, using 1 reagent lot and 1 calibrator lot. The study results are summarized in Table 2. Table 2. Within-Laboratory precision results with Automation of the Access anti-HBc IgM | Sample | N | Mean (S/CO) | Repeatability (Within-run) | | Between- Run | | Between- Day | | Within- Laboratory (total) | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | | QC1 | 120 | 0.00 | 0.000 | NA | 0.000 | NA | 0.000 | NA | 0.000 | NA | | QC2 | 120 | 2.83 | 0.073 | 2.6% | 0.046 | 1.6% | 0.067 | 2.4% | 0.109 | 3.9% | | S1 | 120 | 0.01 | 0.004 | N/A | 0.002 | N/A | 0.003 | N/A | 0.005 | N/A | | S2 | 120 | 0.72 | 0.023 | 3.3% | 0.013 | 1.8% | 0.023 | 3.1% | 0.035 | 4.9% | | S3 | 120 | 1.14 | 0.039 | 3.4% | 0.029 | 2.5% | 0.037 | 3.3% | 0.062 | 5.4% | | S4 | 120 | 4.46 | 0.134 | 3.0% | 0.085 | 1.9% | 0.114 | 2.6% | 0.195 | 4.4% | | P1 | 120 | 0.01 | 0.000 | 0.0% | 0.000 | 0.0% | 0.000 | 0.0% | 0.000 | 0.0% | | P2 | 120 | 0.70 | 0.022 | 3.2% | 0.017 | 2.4% | 0.023 | 3.3% | 0.036 | 5.2% | | P3 | 120 | 1.06 | 0.029 | 2.8% | 0.032 | 3.0% | 0.021 | 2.0% | 0.048 | 4.5% | | P4 | 120 | 3.81 | 0.189 | 4.9% | 0.000 | 0.0% | 0.122 | 3.2% | 0.225 | 5.9% | Note: %CV are not meaningful when S/CO approaches zero. Results are noted as N/A. P = Plasma, S = Serum, N/A = Not applicable, N = number of replicates, QC = Quality control ## B. Reproducibility: Reproducibility of the Access anti-HBc IgM on the DxI 9000 Access Immunoassay analyzer was evaluated by running a five-day multi-site reproducibility study. A panel of samples, serum (S1-S6) and plasma (P1-P6), was assayed at three clinical sites, on three instruments Kxxxxxx - Page 7 of 18 {8} (one per site). Each panel member was assayed in replicates of three at two separate times (runs) per day. The results are summarized in table 3. Table 3. Reproducibility study results of the Access anti-HBc IgM | Sample | N | Mean (S/CO) | Between-Site | Between-Day | Between-Run | Repeatability | | Reproducibility | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | (Within-Run) | | | | | | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | | | | (S/CO) | | (S/CO) | | (S/CO) | | (S/CO) | | (S/CO) | | | S1 | 90 | 0.01 | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | | S2 | 90 | 0.32 | 0.011 | 3.6 | 0.000 | 0.000 | 0.003 | 1 | 0.009 | 2.7 | 0.015 | 4.5 | | S3 | 90 | 0.55 | 0.017 | 3 | 0.000 | 0.000 | 0.015 | 2.8 | 0.015 | 2.7 | 0.027 | 4.9 | | S4 | 90 | 0.76 | 0.024 | 3.2 | 0.003 | 0.4 | 0.009 | 1.2 | 0.018 | 2.4 | 0.032 | 4.2 | | S5 | 90 | 1.19 | 0.032 | 2.6 | 0.000 | 0.000 | 0.027 | 2.2 | 0.03 | 2.5 | 0.051 | 4.3 | | S6 | 90 | 4.52 | 0.087 | 1.9 | 0.049 | 1.1 | 0.068 | 1.5 | 0.108 | 2.4 | 0.162 | 3.6 | | P1 | 90 | 0.01 | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | | P2 | 90 | 0.33 | 0.009 | 2.7 | 0.004 | 1.1 | 0.007 | 2 | 0.008 | 2.5 | 0.014 | 4.3 | | P3 | 90 | 0.53 | 0.017 | 3.2 | 0.000 | 0.000 | 0.012 | 2.2 | 0.013 | 2.6 | 0.025 | 4.7 | | P4 | 90 | 0.74 | 0.021 | 2.8 | 0.000 | 0.000 | 0.013 | 1.8 | 0.019 | 2.6 | 0.031 | 4.2 | | P5 | 90 | 1.12 | 0.035 | 3.1 | 0.000 | 0.000 | 0.029 | 2.6 | 0.024 | 2.1 | 0.052 | 4.6 | | P6 | 90 | 3.95 | 0.073 | 1.8 | 0.042 | 1.1 | 0.081 | 2.1 | 0.097 | 2.5 | 0.152 | 3.8 | Note: %CV are not meaningful when S/CO approaches zero. Results are noted as N/A. P= Plasma, S= Serum, N/A= Not applicable, N=number of replicates # 2. Linearity: Not applicable as this is a qualitative assay # 3. Analytical Specificity/Interference: # A. Interference study: Interference study was conducted to evaluate the susceptibility of the Access anti-HBc IgM assay to potentially interfering substances using negative, low positive and moderately positive samples. Potential endogenous, and exogenous interferents were tested. No interference was observed for any of the interferents tested at concentrations indicated in Table 4. Table 4. Interference study results of the Access anti-HBc IgM | | Potential Interferents | Highest Concentration tested | | --- | --- | --- | | Endogenous Interferents | Hemoglobin | 1,000 mg/dL | | | Total Protein | 15 g/dL | | | Bilirubin Conjugated | 40 mg/dL | | | Bilirubin Unconjugated | 40 mg/dL | | | Biotin | 3,510 ng/mL | | | Cholesterol | 400 mg/dL | | | Triglycerides (Intralipid) | 37 mmol/L (3,854 mg/dL) | Kxxxxxx - Page 8 of 18 {9} # B. Cross reactivity: Potential cross-reactivity with the Access anti-HBc IgM assay was determined by testing specimens with antibodies to other microorganisms or with medical conditions unrelated to HBV infections. No cross reactivity was observed. The cross-reactants tested and study results are summarized in Table 5. Table 5. Cross reactivity study results for the Access anti-HBc IgM | Cross reactants | Number of Samples Tested | Number of Reactive Samples | Number of Non-Reactive Samples | | --- | --- | --- | --- | | Pathogenic Agents and Related Diseases | | | | | Epstein-Barr Virus (EBV) | 10 | 0 | 10 | | Cytomegalovirus (CMV) | 10 | 0 | 10 | | Herpes Simplex Virus (HSV1/2) | 10 | 0 | 10 | | Toxoplasmosis | 10 | 0 | 10 | | Human Immunodeficiency Virus (HIV) | 10 | 0 | 10 | | Hepatitis A Virus (HAV) | 10 | 0 | 10 | | Hepatitis C Virus (HCV) | 10 | 0 | 10 | | Hepatitis E Virus (HEV) | 10 | 0 | 10 | | Flavivirus (Dengue) | 10 | 0 | 10 | | Flavivirus (West Nile) | 10 | 0 | 10 | | Flavivirus (Zika) | 10 | 0 | 10 | | Measles (Rubeola) | 10 | 0 | 10 | | Mumps | 10 | 0 | 10 | | Rubella | 10 | 0 | 10 | | Varicella Zoster | 10 | 0 | 10 | | Syphilis | 10 | 0 | 10 | | Other Related Diseases | | | | | Alcohol Liver Disease | 10 | 0 | 10 | | Primary Biliary Cirrhosis | 10 | 0 | 10 | | Hepatitis B Virus (HBV) | 10 | 0 | 10 | | Hepatitis C Virus (HCV) | 10 | 0 | 10 | | Hepatitis E Virus (HEV) | 10 | 0 | 10 | | Hepatitis F Virus (Zika) | 10 | 0 | 10 | | Mumps | 10 | 0 | 10 | | Rubella | 10 | 0 | 10 | | Varicella Zoster | 10 | 0 | 10 | Kxxxxxx - Page 9 of 18 {10} | Vaccination Potential Interference | | | | | --- | --- | --- | --- | | Influenza Post Vaccination | 10 | 0 | 10 | | Auto-Immune Diseases (Heterophile) | | | | | HAMA | 10 | 0 | 10 | | Anti-Nuclear Antibody (ANA) | 10 | 0 | 10 | | Rheumatoid Factor | 10 | 0 | 10 | | Systemic Lupus Erythematosus (SLE) | 10 | 0 | 10 | | Multiple Myeloma | 10 | 0 | 10 | | Cross-Reactivity Related to Design | | | | | Anti-E. coli or E. coli Infection | 10 | 0 | 10 | | Pregnant Women | | | | | Pregnancy Multipara | 10 | 0 | 10 | | Pregnancy First Trimester | 10 | 0 | 10 | | Pregnancy Second Trimester | 10 | 0 | 10 | | Pregnancy Third Trimester | 10 | 0 | 10 | | Other | | | | | Transplant Recipient | 10 | 0 | 10 | | Dialysis Patients | 10 | 0 | 10 | | Hemophilia / CIDTTing Factor Deficiency | 10 | 0 | 10 | # C. IgM class specificity study: A study was performed to demonstrate that the Access anti-HBc IgM assay specifically detects IgM not IgG and does not generate false results in the presence of IgG in patients' samples. The IgM class specificity of the Access anti-HBc IgM assay was evaluated by depleting IgM molecules present in the samples by treating with dithiothreitol (DTT). Ten anti-HBc IgM known positive samples were tested. All HBc IgM reactive samples treated with DTT showed decreased signal while untreated samples remained unaffected, indicating the Access HBc IgM assay is specifically detects HBc IgM. Results are presented in table 6. Table 6. IgM class specificity study results of the Access anti-HBc IgM. | Sample | Condition | Reference anti-HBc Total | | Access anti-HBc IgM | | Reference anti-HBc IgM | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Signal S/CO | Results | Signal S/CO | Status | Signal S/CO | Results | | Positive Control | NEAT | 1.38 | R | 8.62 | R | 9.72 | R | | | Untreated -PBS) | 1.42 | R | 8.85 | R | 9.30 | R | | | DTT treated | 0.07 | NR | 0.01 | NR | 0.54 | NR | | Negative Control | NEAT | 7.50 | R | 0.18 | NR | 0.12 | NR | | | Untreated (PBS) | 7.32 | R | 0.16 | NR | 0.11 | NR | Kxxxxxx - Page 10 of 18 {11} | | DTT treated | 8.11 | R | 0.12 | NR | 0.30 | NR | | --- | --- | --- | --- | --- | --- | --- | --- | | Sample 1 | NEAT | 5.41 | R | 12.69 | R | 7.20 | R | | | Untreated (PBS) | 5.38 | R | 12.25 | R | 7.38 | R | | | DTT treated | 5.62 | R | 0.09 | NR | 0.12 | NR | | Sample 2 | NEAT | 5.44 | R | 10.07 | R | 4.14 | R | | | Untreated (PBS) | 5.52 | R | 10.06 | R | 4.26 | R | | | (treated | 5.85 | R | 0.06 | NR | 0.11 | NR | | Sample 3 | NEAT | 6.01 | R | 9.34 | R | 5.46 | R | | | Untreated (PBS) | 5.94 | R | 8.99 | R | 3.77 | R | | | DTT treated | 6.03 | R | 0.13 | NR | 0.38 | NR | | Sample 4 | NEAT | 3.82 | R | 12.39 | R | 5.70 | R | | | Untreated (PBS) | 3.83 | R | 12.42 | R | 6.02 | R | | | DTT treated | 3.73 | R | 0.16 | NR | 0.16 | NR | | Sample 5 | NEAT | 2.72 | R | 6.80 | R | 4.95 | R | | | Untreated (PBS) | 2.67 | R | 6.50 | R | 4.95 | R | | | DTT treated | 0.71 | NR | 0.07 | NR | 0.11 | NR | | Sample 6 | NEAT | 6.10 | R | 16.15 | R | 15.49 | R | | | Untreated <PBS) | 6.14 | R | 15.64 | R | 14.06 | R | | | DTT treated | 6.21 | R | 2.62 | R | 3.27 | R | | Sample 7 | NEAT | 4.06 | R | 3.34 | R | 2.40 | R | | | Untreated <PBS) | 4.17 | R | 3.25 | R | 2.13 | R | | | DTT treated | 4.31 | R | 0.03 | NR | 0.07 | NR | | Sample 8 | NEAT | 1.35 | R | 5.69 | R | 2.43 | R | | | Untreated (PBS) | 1.45 | R | 6.23 | R | 2.49 | R | | | DTT treated | 2.75 | R | 0.01 | NR | 0.05 | NR | | Sample 9 | NEAT | 5.72 | R | 6.57 | R | 4.93 | R | | | Untreated (PBS) | 5.83 | R | 6.40 | R | 4.79 | R | | | DTT treated | 6.29 | R | 0.06 | NR | 0.20 | NR | | Sample 10 | NEAT | 6.09 | R | 6.98 | R | 4.76 | R | | | Untreated (PBS) | 6.02 | R | 8.07 | R | 5.68 | R | | | DTT treated | 5.91 | R | 0.08 | NR | 0.36 | NR | # D. Hook effect: A study was conducted to characterize the performance of the Access anti-HBc IgM assay with specimens containing high levels of anti-HBc IgM that have the potential to cause a high dose hook effect. The data demonstrated that the Access anti-HBc IgM assay is not susceptible to interference from specimens with high levels HBc IgM. # 4. Detection Limit and Assay Reportable Range: # A. Seroconversion: The ability of the Access anti-HBc IgM assay to detect anti-HBc IgM was evaluated by testing six anti-HBc IgM seroconversion panels. For each seroconversion panel, the difference in the Kxxxxxx - Page 11 of 18 {12} days from the initial draw to the first reactive result was calculated for all panels between the Access anti-HBc IgM assay and the reference assay. Equivalent detection with no difference in bleed number was observed in three of the six panels, and earlier detection by the Access anti-HBc IgM assay was observed in three panels. The results are summarized in table 7. Table 7. Seroconversion study results of the Access HBc IgM. | Panel ID | First anti-HBc IgM Positive Result from Initial Draw Date | | Access anti-HBc IgM assay vs Reference assay | | --- | --- | --- | --- | | | Access anti-HBc IgM assay (days) | Reference assay (days) | Difference in bleed number for the first reactive bleed* | | HBV-6281 | 43 | 43 | 0 | | HBV-9092 | 85 | 92 | -1 | | HBV-9093 | 49 | 49 | 0 | | HBV-001 | 36 | 43 | -1 | | HBV-002 | 59 | 59 | 0 | | HBV-004 | 71 | 76 | -1 | *The difference in bleed number is compared to the reference assay. For example, -1 indicates that the reference assay required 1 additional bleed before reactivity was determined compared to the Access anti-HBc IgM assay. # 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): # A. Reagent pack, calibrators and control stability: Four serum samples (2 samples each for HBc IgM negatives and positives) were evaluated at different storage conditions at different timepoints using three reagent lots for stability claims. Each time point signal values were compared to the baseline signal values (time 0) to assess degradation status of the reagents, controls and calibrators. The timepoint before the last tested timepoint at which all three lots of reagents meet the acceptance criteria (within $+/- 0.100$ S/CO from baseline for negative samples, within $+/- 10.0\%$ relative signal change from baseline for positive samples) was considered for the stability claims. Table 8 shows the stability results and claims of the Access anti-HBc IgM reagents. Table 8. Reagent pack, calibrator and quality control stability results of the Access anti-HBc IgM | Reagent | Storage | Stability | Claimed stability | | --- | --- | --- | --- | | Access anti-HBc IgM Calibration curve | on the instrument | 56 days | 56 days | | Access anti-HBc IgM Calibrator open vial | (2-10°C) | 181 days | 14 days | | Access anti-HBc IgM Calibrator shelf life | (2-10°C) | 365 days | As stated in the package insert | | Access anti-HBc IgM Quality control open vial | (2-10°C) | 90 days | 90 days | | Access anti-HBc IgM Quality control shelf life | (2-10°C) | 540 Days | As stated in the package insert | | Access anti-HBc IgM Reagent pack | (2-10°C) | 366 days | As stated in the package insert | | Access anti-HBc IgM Open reagent | (2-10°C) | 56 days | 56 days | # B. Specimen stability: Kxxxxxx - Page 12 of 18 {13} A study was conducted to evaluate the specimen storage temperatures and number of freeze/thaw cycles for each blood collection tube type acceptable for use with the Access anti-HBc IgM assay. This study evaluated sample stability at room temperature (20 - 25°C), refrigerated (2 - 8°C), and frozen (≤ -18°C). Sample stability after multiple freeze/thaw cycles was also evaluated. Low negative, high negative, low positive and moderate positive samples were tested. Table 9 shows the summary of the results and specimen stability claims. Table 9. Sample stability results of the Access anti-HBc IgM. | Specimen Stability | | | | | | | --- | --- | --- | --- | --- | --- | | Specimen | Type | 20 to 25°C (hours) | 2 to 8°C (days) | -20°C or colder (days) | Number of freeze thaw | | Serum | Serum and Serum separator tube | 72 | 7 | 30 | Maximum 5 freeze thaws | | Plasma | Lithium Heparin Lithium Heparin Separator Tube Dipotassium (K₂) EDTA Tripotassium (K₃) EDTA Sodium Citrate Acid Citrate Dextrose (ACD) Citrate Phosphate Dextrose (CPD) | | | | | C. Fresh versus frozen samples study: This study evaluated equivalency between fresh (never frozen) serum samples and samples frozen at ≤ -18°C using the Access Anti-HBc IgM assay on the DxI 9000 Access Immunoassay Analyzer. Total of fifty samples including different levels of anti-HBc IgM (ten low negatives, twenty-seven low positives, and thirteen moderate positives) were tested. The regression analysis demonstrated a slope of 1.00 indicating the fresh sample is equivalent to frozen samples in detection of anti-HBc IgM analyte. 6. Assay Cut-Off: Beckman evaluated the Access HBc IgM assay cut-off by testing one hundred HBc IgM negatives and fifty HBc IgM positives from intended use population. The results of this study established cut-off of the Access HBc IgM as 1.00 S/CO. B Comparison Studies: 1. Method Comparison with Predicate Device: See Section C. Clinical Studies. 2. Matrix Comparison: To determine whether the claimed matrices have equivalent performance on the Access anti-HBc IgM assay on the DxI 9000 Access Immunoassay Analyzer, matched donor sets Kxxxxxx - Page 13 of 18 {14} consisting of nine specimen types (serum with or without Gel, and plasma [lithium heparin with Gel, lithium heparin without Gel, dipotassium (K2) EDTA, tripotassium (K3) EDTA, sodium citrate, acid citrate dextrose (ACD) and citrate phosphate dextrose (CPD)] were evaluated. Serum (without gel) served as the reference tube type. Regression analysis showed equivalency between reference matrix and intended matrices. Table 10 presents the results of matrix comparison study results. Table 10. Matrix comparison study results of the Access anti-HBc IgM | Sample Type | Slope (Bootstrap 95% CI) | Correlation Coefficient (r) | | --- | --- | --- | | Serum with Gel | 1.01 (1.00 – 1.02) | 1.00 | | Plasma Lithium Heparin without Gel | 1.02 (1.02 – 1.03) | 1.00 | | Plasma Lithium Heparin with Gel | 1.01 (0.99 – 1.02) | 1.00 | | Plasma K2 EDTA | 1.01 (1.00 – 1.02) | 1.00 | | Plasma K3 EDTA | 1.03 (1.01 – 1.03) | 1.00 | | Plasma Sodium Citrate | 1.10 (1.08 – 1.10) | 1.00 | | Plasma Acid Citrate Dextrose | 1.12 (1.08 – 1.16) | 1.00 | | Plasma CPD | 1.12 (1.10 – 1.14) | 1.00 | # C Clinical Studies: A multi-center study was conducted using the Access anti-HBc IgM assay with the DxI 9000 Access Immunoassay Analyzer at three clinical testing sites in the United States. A total of 2,537 specimens including prospective and retrospective specimens from intended use population were tested with the Access anti-HBc IgM assay and FDA approved reference HBc IgM assays using composite comparators method. The HBV classification was determined by serological assessment for all prospective and retrospective specimens based on the reactivity patterns of six (6) HBV serological markers (HBsAg, HBeAg, anti-HBc IgM, total anti-HBc, anti-HBe and anti-HBS). # Comparison results of prospective and retrospective population: The positive percent agreement (PPA) and negative percent agreement (NPA) and results with $95\%$ confidence intervals (CI) for the prospective and retrospective populations are presented in tables 11-12. Table 11. Access anti-HBc IgM assay PPA and NPA for prospective population by HBV Classification | HBV Classification | Total | PPA | | NPA | | | --- | --- | --- | --- | --- | --- | | | | % (n/N) | 95% CI | % (n/N) | 95% CI | | Acute | 4 | 100.0 (2/2) | 34.2-100.0 | 100.0 (2/2) | 34.2-100.0 | | Chronic | 20 | N/A (0/0) | N/A | 95.00 (19/20) | 76.4-99.1 | Kxxxxxx - Page 14 of 18 {15} Table 12. Access anti-HBc IgM retrospective population PPA by HBV Classification | HBV Classification | PPA | | | | --- | --- | --- | --- | | | % (n/N) | 95% CI | | | Acute | 98.08 (51/52) | 89.9-99.7 | | | Acute (Late) | 100.0 (85/85) | 95.7-100.0 | | | Chronic | 100.0 (5/5) | 56.6-100.0 | | | Early Recovery | 90.38 (47/52)* | 79.4-95.8 | | | HBV Vaccine Response | N/A (0/0) | N/A | | | Not Previously Infected | N/A (0/0) | N/A | | | Recovered or Immune due to Natural Infection | N/A (0/0) | N/A | | | Recovery | N/A (0/0) | N/A | | | Uninterpretable | 42.86 (3/7) | 15.8-75.0 | | * All 4 early recovery samples had result values close to cutoff of candidate and comparator devices; 2 of the discordant samples produced results in concordance with the Access anti-HBc IgM upon re-testing (re-test results not used in performance calculations). ** One subject classified using four-marker classification table. Kxxxxxx - Page 15 of 18 {16} Total 95.02 (191/201) 91.1-97.3 *All early recovery samples discordant with the Access anti-HBc IgM had result value close to cutoff values of candidate and comparator devices. # Comparison of Results for Pregnant Women One hundred and seventy-one serum samples were prospectively collected from an increased risk and/or signs and symptoms U.S. pregnant population. This population included 21 pregnant subjects of pediatric age (range 18 - 21 years). There were 71 subjects in the first trimester, 67 subjects in the second trimester and 33 subjects in the third trimester in this study. The results of the Access anti-HBc IgM testing at all sites combined are presented in table 13. Table 13. Percent Agreement between Access anti-HBc IgM and Comparator Final interpretation by HBV Classification: Pregnant Population. | HBV Classification | PPA | | NPA | | | --- | --- | --- | --- | --- | | | % (n/N) | 95% CI | % (n/N) | 95% CI | | Chronic | 0 (0/0) | NA | 100.0 (1/1) | 20.65-100.0 | | HBV Vaccine Response | 0 (0/0) | NA | 100.0 (77/77) | 95.25-100.0 | | Not Previously Infected | 0 (0/0) | NA | 100.0 (82/82) | 95.52-100.0 | | Possible HBV Vaccination Response | 0 (0/0) | NA | 100.0 (9/9) | 70.09-100.0 | | Recovery | 0 (0/0) | NA | 100.0 (2/2) | 34.24-100.0 | | Total | 0 | NA | 100.0 (171/171) | 97.80-100.0 | Because no HBc IgM positive samples were identified in pregnant cohort, a spiking study was conducted to evaluate the Access Anti-HBC IgM performance with HBc IgM spiked pregnant samples. A total of thirty-one pregnant and control adult serum samples were spiked with unique native anti-HBc IgM positive sample. The results revealed that there were no significant differences in spiked pregnant samples versus control samples. # Comparison of Results for pediatric population Prospectively collected one hundred and fifty-five pediatric subjects were tested by the candidate and comparator devices. Results are presented in table 14. Table 14. Percent Agreement between Access anti-HBc IgM and Comparator Final Interpretation by HBV Classification: Pediatric Population. | HBV Classification | Total | PPA | | NPA | | | --- | --- | --- | --- | --- | --- | | | | % (n/N) | 95% CI | % (n/N) | 95% CI | | Early Recovery | 1 | 0.00 (0/1) | (0.0-79.3) | N/A | N/A | | HBV Vaccine Response | 53 | N/A | N/A | 100.0 (53/53) | (93.2-100.0) | | Not Previously Infected | 95 | N/A | N/A | 100.0 (95/95) | (96.1-100.0) | Kxxxxxx - Page 16 of 18 {17} Supplemental retrospective pediatric study: Additional Anti-HBc IgM known positive samples from pediatric population were obtained from vendors. Table 15 shows results for retrospective samples by HBV classification. Table 15. Access anti-HBc IgM retrospective samples: pediatric cohort PPA and NPA by HBV Classification | HBV Classification | Total | PPA | | | --- | --- | --- | --- | | | | % (n/N) | 95% CI | | Acute | 2 | 100.0 (2/2) | 34.2-100.0 | | Acute (Late) | 2 | 100.0 (2/2) | 34.2-100.0 | | Early Recovery | 4 | 100.00 (4/4) | 51.0-100.0 | | HBV Vaccine Response | 0 | N/A (0/0) | N/A | | Not Previously Infected | 0 | N/A (0/0) | N/A | | Possible HBV Vaccination Response | 0 | N/A (0/0) | N/A | | Recovered or Immune Due to Natural Infection | 0 | N/A (0/0) | N/A | | Total | 8 | 100 (8/8) | 67.6-100.0 | Because insufficient HBc IgM positive samples were identified in pediatric cohorts, a spiking study was conducted to evaluate the results when HBc IgM spiked pediatric samples tested with the Access anti-HBc IgM assay. A total of thirty-one pediatric (3-21 years) and control adult serum samples were spiked with unique native anti-HBc IgM positive sample. The results revealed that there were no significant differences in results of spiked pediatric samples versus control samples. # D Expected Values/Reference Range: The Access anti-HBc IgM results for the prospective population for all clinical trial sites combined by age group and gender are summarized in table 16. Table 16. Expected Results of Access HBc IgM by Age and Gender. | Age Range (years) | Sex | Access anti-HBc IgM | | | | Total | | --- | --- | --- | --- | --- | --- | --- | | | | Nonreactive | | Reactive | | | | | | N | % | N | % | | | 9-12 | Female | 6 | 0.26 | 0 | 0.00 | 6 | | | Male | 11 | 0.48 | 0 | 0.00 | 11 | | 13-18 | Female | 34 | 1.49 | 0 | 0.00 | 34 | | | Male | 24 | 1.05 | 0 | 0.00 | 24 | Kxxxxxx - Page 17 of 18 {18} | 19-21 | Female | 73 | 3.20 | 0 | 0.00 | 73 | | --- | --- | --- | --- | --- | --- | --- | | | Male | 28 | 1.23 | 0 | 0.00 | 28 | | 22-29 | Female | 301 | 13.20 | 0 | 0.00 | 301 | | | Male | 108 | 4.73 | 0 | 0.00 | 108 | | 30-39 | Female | 276 | 12.10 | 0 | 0.00 | 276 | | | Male | 129 | 5.66 | 0 | 0.00 | 129 | | 40-49 | Female | 190 | 8.33 | 0 | 0.00 | 190 | | | Male | 150 | 6.58 | 1 | 0.04 | 151 | | 50-59 | Female | 253 | 11.09 | 0 | 0.00 | 253 | | | Male | 225 | 9.86 | 1 | 0.04 | 226 | | 60-69 | Female | 153 | 6.71 | 0 | 0.00 | 153 | | | Male | 178 | 7.80 | 2 | 0.09 | 180 | | 70-79 | Female | 52 | 2.28 | 0 | 0.00 | 52 | | | Male | 55 | 2.41 | 0 | 0.00 | 55 | | 80-89 | Female | 9 | 0.39 | 0 | 0.00 | 9 | | | Male | 16 | 0.70 | 0 | 0.00 | 16 | | 90+ | Female | 1 | 0.04 | 0 | 0.00 | 1 | | | Male | 5 | 0.22 | 0 | 0.00 | 5 | ## E Other Supportive Instrument Performance Characteristics Data: **Carryover study:** A Carryover study was performed to determine the Access anti-HBc IgM carryover effect by running negative and high positive samples at varying sequences. This study evaluated 1. reagent carryover and, 2. sample carryover. Results demonstrated no carryover with both studies. **EMC:** Electrical safety and electromagnetic compatibility (EMC) testing were performed, and the system was found to be acceptable. **Software and cybersecurity:** Software and cybersecurity documentation was reviewed and found to be acceptable. ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. Kxxxxxx - Page 18 of 18
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...