K254024 · Multi Radiance Medical, Inc. · NHN · May 19, 2026 · Physical Medicine
Device Facts
Record ID
K254024
Device Name
MR5 Activ Pro
Applicant
Multi Radiance Medical, Inc.
Product Code
NHN · Physical Medicine
Decision Date
May 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MR5 ACTIV PRO is indicated for adjunctive use in the temporary relief of pain associated with Lateral Epicondylitis.
Device Story
Handheld, battery-powered, non-invasive light-based therapy device; utilizes 905 nm super-pulsed laser diode, 850 nm infrared LEDs, and 660 nm red LEDs. Operates via fixed 4 cm² aperture to deliver localized optical energy for non-heating pain relief. Used in clinical environments by trained healthcare professionals. Device software controls pulse repetition frequency, duration, and synchronized emitter operation. Healthcare providers apply device to affected lateral epicondyle; output provides temporary pain relief, potentially reducing reliance on other pain management therapies. System is cordless, rechargeable, and designed for repeatable treatment delivery.
Clinical Evidence
Prospective, 1:1 randomized, triple-blind, placebo-controlled multi-center study (n=50). Subjects (adults with lateral epicondylitis) received 6 treatments over 3 weeks. Primary endpoint: proportion of subjects achieving ≥30% reduction in VAS pain score. Study met primary endpoint with larger proportion of successes in active group vs placebo. Safety: 2 mild adverse events (heat/biting sensation) reported in active group, both resolved upon treatment cessation.
Technological Characteristics
Handheld, battery-powered (Li-polymer, 4 Ah). Optical sources: 905 nm SPL, 850 nm IR LED, 660 nm red LED. Fixed 4 cm² aperture. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 14971, IEC TR 60601-4-2, IEC 60825-1, ISO 10993-1. Software-controlled optical delivery parameters. Class I laser. Non-heating mechanism.
Indications for Use
Indicated for adjunctive use in the temporary relief of pain associated with Lateral Epicondylitis in adults aged 22 years and older.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 19, 2026
Multi Radiance Medical, Inc.
Glenn Persello
Regulatory Affairs Manager
6521 Davis Industrial Pkwy.
Solon, Ohio 44139
Re: K254024/S001
Trade/Device Name: MR5 Activ Pro
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: NHN
Dated: April 17, 2026
Received: April 17, 2026
Dear Glenn Persello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254024 - Glenn Persello
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254024 - Glenn Persello
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHUN XU -S
For Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254024/S001 | |
| Device Name | |
| MR5 ACTIV PRO | |
| Indications for Use (Describe) The MR5 ACTIV PRO is indicated for adjunctive use in the temporary relief of pain associated with Lateral Epicondylitis. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
1
# 510(k) Summary
| Item | Information |
| --- | --- |
| Application Type | Traditional |
| Applicant | Multi Radiance Medical |
| Address | 6521 Davis Industrial Parkway, Solon, OH 44139 |
| Contact Person | Glenn Persello |
| Contact Information | gpersello@multiradiance.com • 440-542-0761 |
| Preparation Date | May 19, 2026 |
| Device Trade Name | MR5 Activ Pro |
| Common Name | Infrared Lamp |
| Classification Name | Powered Light-Based Laser Non-Thermal Instrument with Non-Heating Effect for Adjunctive Use in Pain Therapy |
| Regulation Number | 890.5500 |
| Product Code | NHN |
| Regulatory Class | Class II |
| Legally Marketed Predicate Device | FibroLux — K212189 |
## Device Description:
The MR5 ACTIV Pro is a handheld, portable, reusable, battery-powered, non-invasive, non-sterile light-based therapy medical device that utilizes a single laser diode and multiple light emitting diodes (LEDs) to provide therapeutic infrared and visible red-light optical output through the device emitter aperture. The device platform includes optical sources at 905 nm, 850 nm, and 660 nm. For the Lateral Epicondylitis treatment mode that is the subject of this submission, the active optical outputs consist of a 905 ± 10 nm super-pulsed laser diode, 850 ± 15 nm infrared LEDs, and 660 ± 10 nm red LEDs. The device is designed to deliver localized optical energy through a fixed 4 cm² treatment aperture using predefined operating parameters consistent with a non-heating mechanism of action under product code NHN. The MR5 ACTIV Pro is a cordless, self-contained system powered by a rechargeable lithium-polymer battery and is intended for use in a clinical environment by trained healthcare professionals. The device incorporates software-controlled optical delivery parameters that regulate treatment duration, pulse repetition frequency, and synchronized emitter operation to ensure repeatable treatment delivery under labeled conditions.
## Indications for use:
The MR5 ACTIV PRO is indicated for adjunctive use in the temporary relief of pain associated with Lateral Epicondylitis.
{5}
2
# Intended Population:
The MR5 ACTIV PRO is intended for use in adults aged 22 years and older.
# Technology Description
The MR5 ACTIV PRO utilizes a light-based therapy technology consisting of a 905 nm super-pulsed laser (SPL) combined with pulsed red (660 nm) and pulsed infrared (850 nm) light-emitting diodes (LEDs). Optical energy is delivered through a fixed 4 cm² treatment aperture using nanosecond super-pulsed laser emission together with electronically pulsed LED sources operating at defined duty cycles and time-averaged optical output levels. The device is configured to provide localized treatment for the temporary adjunctive relief of pain associated with Lateral Epicondylitis. Device operating parameters are controlled to maintain low duty-cycle optical exposure and safe skin temperature profiles, as demonstrated through performance and thermal testing described in this submission.
510(k) Summary Table — Technical Characteristics Comparison (Part 1)
| Technological Characteristic | Subject Device: MR5 ACTIV PRO (Lateral Epicondylitis Mode) | Predicate Device: FibroLux (K212189) | Equivalence Discussion |
| --- | --- | --- | --- |
| Wavelength (nm) | • Laser: 905 ±10 nm • Red LED: 660 ±10 nm • IR LED: 850 ±15 nm | Not publicly available | Similar therapeutic wavelength ranges. |
| Number of Diodes | • Laser: 1 • Red LED: 3 • IR LED: 3 | Not publicly available | Differences in diode count reflect device design and treatment geometry, with the subject device delivering energy through a smaller 4 cm² treatment aperture compared to the 30 cm² predicate aperture. Clinical performance data demonstrates that the device does not raise effectiveness concerns compared to the predicate device. |
| Peak Power (per diode) | • SPL: 50 W • Red LED: 133 mW • IR LED: 167 mW | Not publicly available | Peak laser power per SPL diode is identical between devices (50 W). Reported peak power values for the red and infrared LEDs represent characterization of instantaneous output during the active (ON) phase of pulsed operation. Thermal characterization testing and clinical performance data confirm that the device can be used as safely and effectively as the predicate device under the intended use conditions. |
| Average Power (per diode) | • SPL: 1.0875 mW • Red LED: 66.7 mW • IR LED: 83.3 mW | Not publicly available | SPL average power per diode differs due to pulse duration (87 ns vs 68 ns) while peak SPL power is identical at 50 W. The difference is small and remains in the milliwatt range. Total SPL average power is lower for the subject device because the subject has 1 SPL diode. and the predicate has 4 SPL diodes. |
{6}
# 510(k) Summary Table — Technical Characteristics Comparison (Part 2)
| Technological Characteristic | Subject Device: MRS ACTIV PRO (Lateral Epicondylitis Mode) | Predicate Device: FibroLux | Equivalence Discussion |
| --- | --- | --- | --- |
| Pulse Repetition / Frequency | ·SPL: 250 Hz ·Red LED: 2 Hz ·IR LED: 250 Hz | Not publicly available | Both devices use pulsed optical output with identical red LED frequency. The lower SPL and IR LED frequencies in the subject device do not raise effectiveness concerns compared to the predicate device because the subject device's operating configuration delivers clinically relevant optical dose to localized lateral epicondylitis targets through a smaller 4 cm² aperture. Clinical performance data and bench testing demonstrates that the device does not raise increased safety or effectiveness concerns compared to the predicate device. |
| Duty Cycle | ·SPL: 0.00218% ·Red LED: 50% ·IR LED: 50% | Not publicly available | Both devices use extremely low SPL duty cycles and identical 50% LED duty cycles. The lower SPL duty cycle in the subject device does not raise effectiveness concerns because the complete operating configuration, including wavelength, average power, delivered energy, treatment time, treatment geometry, and LED contribution, was clinically validated for the temporary pain relief associated with lateral epicondylitis and bench thermal testing. Thermal bench testing data was submitted to demonstrate non-heating operation. |
| Spot Size(s) | ·SPL: 0.44 cm² ·Red LED: 4 cm² (aperture-based emission area) ·IR LED 4 cm² | Not publicly available | SPL spot size per diode is equivalent. LED emission is divergent and fills the treatment aperture; therefore, LED spot size is appropriately represented by the treatment aperture area (4 cm² for the subject device and 30 cm² for the predicate device). |
{7}
| Laser Output Mode | Pulsed | Not publicly available | Same pulsed operating principle. |
| --- | --- | --- | --- |
| Laser Pulse Shape | Gaussian | Not publicly available | Same. |
| Laser Beam Profile | Elliptical (~9" × 25" divergence) | Not publicly available | Same optical beam geometry. |
| Aperture Size (Treatment Window) | 4 cm² | Not publicly available | Subject designed for localized treatment through a smaller treatment aperture; predicate designed for distributed coverage across a larger treatment aperture. |
| Irradiance | •SPL: 0.00027 W/cm² •Red: 0.050 W/cm² •IR: 0.0625 W/cm² •Total: 0.113 W/cm² | Not publicly available | Higher calculated irradiance reflects localized delivery through a smaller 4 cm² treatment aperture compared to the 30 cm² predicate aperture. Total delivered energy remains comparable to or lower than the predicate device under indicated conditions. Clinical data and supplemental thermal characterization testing across Fitzpatrick Skin Types I–VI demonstrate safe operation under labeled conditions, with no adverse thermal events, burns, erythema, or tissue injury observed. Increased irradiance values reflect localized delivery through a smaller 4 cm² treatment aperture and do not result in increased thermal risk. |
| Total Fluence (per site)** | 60 s exposure: •SPL 0.148 J/cm² •Red 3.0 J/cm² •IR 3.75 J/cm² •Total aperture-based fluence 6.77 J/cm² | Not publicly available | Higher red, infrared, and total aperture-based fluence values reflect localized delivery through the 4 cm² treatment aperture, while SPL fluence remains lower than the predicate device. Total delivered energy per treatment site is lower for the subject device compared to the predicate under indicated conditions. Clinical data and supplemental thermal characterization testing demonstrate non-heating operation. |
| Total Energy Delivered (per site) | 27.065 J/site | Not publicly available | Although the subject device delivers lower total energy per site than the predicate, the MR5 ACTIV PRO is intended for localized treatment of discrete lateral epicondylitis targets using a smaller 4 cm² treatment aperture and 60-second exposure. Clinical performance data submitted in this application demonstrates temporary pain relief under the subject device operating configuration. Therefore, the lower total energy per site does not raise increased effectiveness concerns compared to the predicate device. |
{8}
| Technological Characteristic | Subject Device: MR5 ACTIV PRO (Lateral Epicondylitis Mode) | Predicate Device: FibroLux (K212189) | Equivalence Discussion |
| --- | --- | --- | --- |
| Power Source | Li-polymer battery | Not publicly available | Same power source configuration. |
| Battery Capacity | 4 Ah | Not publicly available | Same battery capacity. |
| Continuous Operation Time | 6 hours | Not publicly available | Similar operating duration; no impact on safety or performance. |
| Battery Charging Time | 4 hours | Not publicly available | Same battery charging time. |
| Device Dimensions | 203 × 64 × 70 mm | Not publicly available | Similar device form factor. |
| Device Weight | 250 g | Not publicly available | Similar handheld form factor. |
| Safety Classification | Class I laser | Not publicly available | Same laser safety classification. |
# 510(k) Summary Table — Technical Characteristics Comparison (Part 3)
# Table Footnote:
Output values for the MR5 ACTIV PRO are reported using the Lateral Epicondylitis treatment protocol (ACTIV PRO platform, 250 Hz SPL frequency, 60-second treatment duration per site). This is the only configuration clinically evaluated for the adjunctive use of temporary pain relief associated with Lateral Epicondylitis. The MR5 ACTIV PRO platform supports SPL frequencies from 5 Hz to 5000 Hz; however, only the 250 Hz configuration is in the scope of this 510(k). Other device platforms are 510(k)-exempt or outside the scope of this submission.
# *Pulse Timing Footnote:
For electronically pulsed LED emitters, pulse timing is determined by repetition frequency and duty-cycle operation rather than output power. At $50\%$ duty cycle, the 660 nm red LED channel operating at 2 Hz produces a 0.5 second pulse period with 250 ms ON-time and 250 ms OFF-time in both devices. For the 850 nm infrared LED channel, the MR5 ACTIV PRO operating at 250 Hz produces a 4 ms pulse period with 2 ms ON-time and 2 ms OFF-time, while the predicate operating at 1000 Hz produces a 1 ms pulse period with 0.5 ms ON-time and 0.5 ms OFF-time. These timing differences reflect native operating configuration and pulse architecture, while optical output power remains an independent operating characteristic.
# **Fluence / Irradiance Calculation Basis Footnote:
The 905 nm SPL output is calculated using the SPL optical spot area because the SPL beam does not fill the full treatment aperture. For the MR5 ACTIV PRO, the SPL spot area is $0.44\mathrm{cm}^2$ . Red and infrared LED outputs are divergent and overfill the treatment aperture; therefore, red LED, infrared LED, and combined aperture-based irradiance/fluence values are calculated using the effective treatment aperture area: $4\mathrm{cm}^2$ for MR5 ACTIV PRO.
{9}
6
# 510(k) Summary – Clinical Testing
## Study Objective
The purpose of the supportive clinical study was to determine the effectiveness of the MR5 ACTIV PRO device for the temporary relief of pain associated with lateral epicondylitis.
## Study Design
The study was a prospective, 1:1 randomized, triple-blind, placebo-controlled multi-center design.
## Study Subject Population
A total of fifty (50) subjects completed the study across three (3) clinical study sites and randomized in a 1:1 ratio: 25 subjects randomized to receive the active device treatment and 25 subjects randomized to receive the placebo (sham) treatment. Subjects were adults diagnosed with unilateral lateral epicondylitis, evenly divided between genders, with 52% males (13/25) in the active treatment group and 48% (12/25) males in the placebo treatment group. The average subject age was 34.64 years for active group subjects and 35.12 years for placebo group subjects. At study entry, subject three-day average pain rating on the 0-100 Visual Analog Scale (VAS) for the affected lateral epicondyle was 79.36 for the active treatment group and 75.60 for the placebo treatment group.
## Study Treatments
Subjects received six (6) treatment administrations with the MR5 ACTIV PRO device (active or sham) according to the defined protocol to the affected lateral epicondyle across a consecutive three-week period: two (2) treatments per week, each treatment three (3) to four (4) days apart. Subjects agreed to use only the medication(s), treatment(s) and therapy(ies) determined during the individualized medication stabilization phase of the study to relieve any lateral epicondyle pain, as needed, throughout the study.
## Study Results
### Primary Effectiveness Endpoint
The primary effectiveness endpoint was the proportion of subjects achieving at least a 30% reduction in pain, as measured using the 0 to 100 Visual Analog Scale (VAS), at study endpoint relative to baseline. Primary efficacy success was predefined as the difference in the proportion of subjects between active and placebo groups who attained the primary effectiveness endpoint. Overall study success was predefined as at least a 30% difference in the proportion of individual subject successes between the active and placebo treatment groups.
The clinical study met its pre-defined primary effectiveness endpoint. The proportion of subjects meeting the primary effectiveness endpoint was larger in the treatment group than in the placebo group. As such, the study results support that the subject device is as effective as the predicate device for the temporary relief of pain associated with lateral epicondylitis.
### Safety Outcomes
Two (2) of the 25 active treatment group subjects reported adverse events during the study: one (1) being a heat sensation and one (1) being a biting sensation. Both events were reportedly mild in intensity, fully resolved upon cessation of the treatment session, and were determined by the study investigator to be possibly related to application of the study device.
{10}
7
# Conclusion
The clinical study findings demonstrate that the MR5 ACTIV PRO device is safe and effective as the predicate device for the temporary relief of pain associated with lateral epicondylitis.
# Performance Testing
Verification and validation activities were successfully completed. Testing included evaluation of bench thermal testing, electrical safety, electromagnetic compatibility, software lifecycle processes, risk management, laser safety classification, and thermal performance.
Testing was conducted in accordance with the following recognized standards:
- IEC 60601-1:2005 + A2:2020, Medical Electrical Equipment — Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2:2014 + A1:2020, Medical Electrical Equipment — Part 1-2: Electromagnetic Disturbances — Requirements and Tests
- IEC 62304:2006 + A1:2015, Medical Device Software — Software Life Cycle Processes
- ISO 14971:2019, Medical Devices — Application of Risk Management to Medical Devices
- IEC TR 60601-4-2:2016, Electromagnetic Immunity Guidance and Interpretation
- IEC 60825-1:2014, Safety of Laser Products — Equipment Classification and Requirements
Biocompatibility assessment in accordance with "ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process", confirmed that all patient-contacting materials are biocompatible for their intended use.
# Conclusions
The non-clinical data provided in this submission demonstrate that the MR5 ACTIV Pro complies with recognized standards for electrical safety, electromagnetic compatibility, software lifecycle processes, and risk management. Thermal characterization testing conducted under representative clinical ambient conditions and physiologic baseline skin temperature conditions demonstrated controlled skin surface temperature responses consistent with non-heating operation, with no burns, erythema, skin sensitivity, pigmentary alteration, blistering, or thermal tissue injury observed across Fitzpatrick Skin Types I through VI.
Clinical performance data further supports the use of the device for adjunctive use in the temporary relief of pain associated with Lateral Epicondylitis under labeled treatment conditions.
Based on the technological comparison, performance testing, thermal characterization, and clinical evidence provided in this submission, the MR5 ACTIV Pro is as safe and effective as the legally marketed predicate device K212189.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.