K253972 · Karl Storz SE & CO. KG · OWN · Feb 12, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K253972
Device Name
IMAGE1 (TC400US); X-to-4U Adapter (TC040)
Applicant
Karl Storz SE & CO. KG
Product Code
OWN · Gastroenterology, Urology
Decision Date
Feb 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IMAGE1, a camera control unit (CCU), when used with compatible camera heads or videoendoscopes and ancillary equipment, is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescent imaging and documentation of endoscopic and microscopic procedures in adults and pediatrics. The X-to-4U Adapter, as ancillary equipment, is intended to connect compatible camera heads or videoendoscopes to the IMAGE1 CCU.
Device Story
IMAGE1 CCU is an endoscopic video imaging system for OR use; processes visual and NIR fluorescent signals from compatible camera heads/videoendoscopes. Performs color processing, filtration, and image enhancement (S-Clara, S-Chroma, SPECTRA A/B). X-to-4U Adapter provides electromechanical interface between X-Line flexible videoendoscopes and IMAGE1 4U-receptacle. System supports software application hosting, 4K-UHD resolution, and digital outputs (12G/3G-SDI, DisplayPort, DVI-D). Operated by medical professionals; provides real-time visualization on monitors to assist in surgical/microscopic procedures. Benefits include enhanced visualization of anatomy and fluorescence for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including optical performance (color accuracy, resolution, SNR, ICG sensitivity/penetration depth), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and software verification/validation per IEC 62304.
Technological Characteristics
CCU with 4K-UHD resolution; supports VIS and NIR imaging. Features include automatic light/shutter control, digital zoom, and image enhancement (S-Clara, S-Chroma, SPECTRA A/B). Connectivity via 12G/3G-SDI, DisplayPort, DVI-D, and Ethernet (KS HIVE). Software-defined architecture supports downloadable applications. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and ISO 17664-2.
Indications for Use
Indicated for real-time visible (VIS) and near-infrared (NIR) fluorescent imaging and documentation of endoscopic and microscopic procedures in adult and pediatric patients.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
KARL STORZ ICG Imaging System, Image1S Camera Control Unit (K233333)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 12, 2026
Karl Storz SE & CO. KG
Winkie Wong
Director Regulatory Affairs
Dr.-Karl-Storz-Straße 34
Tuttlingen, Baden-Wurttemberg 78532
Germany
Re: K253972
Trade/Device Name: KARL STROZ IMAGE1(TC400US); X-to-4U Adapter (TC040)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FET, OWN
Dated: December 11, 2025
Received: December 11, 2025
Dear Winkie Wong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253972 - Winkie Wong
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K253972 - Winkie Wong
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.02.12
15:06:17 -05'00'
Tanisha Hithe, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253972 | ? |
| Please provide the device trade name(s). | | ? |
| KARL STORZ IMAGE1 (TC400US);
X-to-4U Adapter (TC040) | | |
| Please provide your Indications for Use below. | | ? |
| The IMAGE1, a camera control unit (CCU), when used with compatible camera heads or videoendoscopes
and ancillary equipment, is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescent
imaging and documentation of endoscopic and microscopic procedures in adults and pediatrics.
The X-to-4U Adapter, as ancillary equipment, is intended to connect compatible camera heads or
videoendoscopes to the IMAGE1 CCU. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
STORZ
KARL STORZ-ENDOSKOPE
KARL STORZ Premarket Notification
IMAGE1 TC400US & X-to-4U Adapter
510(k) Summary
# 510(k) Summary: K253972
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92 and the FDA guidance document titled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued on July 28, 2014. All data included in this document is accurate and complete to the best of KARL STORZ SE & Co. KG knowledge.
| Submitter: | KARL STORZ SE & Co. KG
Dr.-Karl-Storz-Straße 34
78532 Tuttlingen, Germany |
| --- | --- |
| Contact: | Winkie Wong
Director of Regulatory Affairs
Phone: (424) 218-8618
Email: winkie.wong@karlstorz.com |
| Date of Preparation: | December 11, 2025 |
| Type of 510(k) Submission: | Traditional |
| Device Identification: | KARL STORZ IMAGE1 (TC400US);
X-to-4U Adapter (TC040) |
| Regulatory Class: | II |
| Product Code: | FET: Endoscopic Video Imaging System/Component
OWN: Confocal Optical Imaging |
| Classification Name: | 21 CFR 876.1500 (Endoscope and Accessories) |
| Device Panel: | Gastroenterology & Urology |
| Predicate Device(s): | KARL STORZ ICG Imaging System, Image1S Camera Control Unit (K233333)
The predicate has not been subjected to a design related recall. |
| Device Description: | The IMAGE1 CCU is the next generation KARL STORZ CCU designed for use by medical professionals in an OR setting to provide real-time visible (VIS) and near-infrared (NIR) fluorescent imaging and documentation of endoscopic and microscopic procedures. The IMAGE1 CCU performs image processing functions including color processing, color filtration, and image enhancement and is compatible with many KARL STORZ C-Line and 4U connector type camera heads. Software extensions and applications can be purchased by the user and downloaded onto the CCU to provide additional features beyond those of the default set.
For compatibility with the X-link connector type camera head and videoendoscopes, IMAGE1 requires the X-to-4U adapter. The X-to-4U Adapter provides an electromechanical interface between the X-Line family of Flexible Videoendoscopes and the 4U-Receptacle of the IMAGE1 CCU. |
| Indications for Use: | IMAGE1, a camera control unit (CCU), when used with compatible camera heads or videoendoscopes and ancillary equipment, is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescent imaging and documentation of |
{5}
| | endoscopic and microscopic procedures in adults and pediatrics.
The X-to-4U Adapter, as ancillary equipment, is intended to connect compatible camera heads or videoendoscopes to the IMAGE1 CCU. | | |
| --- | --- | --- | --- |
| Physical and Technological Characteristics: | | Subject Device
KARL STORZ IMAGE1 & Adapter | Predicate Device K233333
KARL STORZ Image1S CCU |
| | Resolution | Up to 4K-UHD | Same as the subject |
| | Brightness control | Yes | Same as the subject |
| | Enhancement control | Yes (S-Clara, S-Chroma, SPECTRA A/B) | Yes (S-Clara, S-Chroma) |
| | Light Source Control | Automatic | Same as the subject |
| | Shutter Control | Automatic | Same as the subject |
| | Image/Video Capture | Yes | Same as the subject |
| | Zoom | 1x, 1.2x, 1.5x, 1.75x, 2x, 2.25x, 2.5x | Same as the subject |
| | Adaptive Zoom | Yes | Same as the subject |
| | Digital Outputs | 12G/3G-SDI, DisplayPort, DVI-D | HDMIx2, 3/12G-SDI x1, SDVoE via 10Gig Ethernet |
| | Communication Interface | KS HIVE (Ethernet) | KS HIVE (Ethernet), SCB |
| | Peripheral | Mouse, keyboard, monitor | Same as the subject |
| | Remote Service | Yes | Same as the subject |
| | SW application hosting | Yes | No |
| | Software Revision | 1.0 | 4.5 |
| Non-Clinical Performance Data: | The following non-clinical performance data were provided in support of the determination for substantial equivalence.
**Electrical Safety and Electromagnetic Compatibility (EMC)**
The system complies with the following standards:
• IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION / ANSI/AAMI ES60601-1:2005/A2:2021
• IEC 60601-1-2:2014 [Including AMD 1:2021]
• IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION
• IEC 60601-2-18: Edition 3.0 2009-08
• ANSI/AAMI IEC 62366-1:2015 + AMD1:2020 (Consolidated Text)
**Reprocessing Validation**
The reprocessing data submitted complies with the following standards:
• ISO 17664-2: Processing of health care products — Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices
**Bench Performance Testing**
The bench performance data was conducted to establish substantial equivalence between the subject device and the predicate device for comparison and characterization of white light and ICG optical performance characteristics. | | |
{6}
The following standard was used to establish test methodology in field of view bench tests:
- ISO 8600-3:2019 Endoscopes — Medical endoscopes and endotherapy devices - Part 3: Determination of field of view and direction of view of endoscopes with optics
The bench testing conducted on the subject devices includes the following:
Optical performance testing (White Light)
Color Accuracy and CCER
Distortion
- Field of View
- Instantaneous Dynamic Range and Detection Linearity
- Spatial Frequency Response
- White Light Linearity & Dynamic Range
- 2D Autolevel Drift and Noise
- 2D Autolevel Settling Time
- 3D System Zero Parallax
- On-Axis and Off-Axis Resolution
- Illumination Detection Uniformity
- Depth of Field
- Signal to Noise Ratio & Sensitivity
Latency
- SPECTRA A and B Color Accuracy and Color Contrast Enhancement Ratio
- SPECTRA A and B SNR and Sensitivity
- SPECTRA A and B Instantaneous Dynamic Range
- SPECTRA A and B Dynamic Range
- SPECTRA A and B Illumination Detection Uniformity
Optical performance testing (ICG Imaging)
ICG Coregistration
Minimum Detectable Concentration of ICG
- ICG Fluorescence Minimum Sensitivity and ICG SNR and Noise
ICG Penetration Depth
ICG Spatial Resolution
ICG Color Reproducibility
ICG Linearity and Dynamic Range
ICG Depth of Field
- ICG Illumination Detection Uniformity
CPU Resource Testing – SW application hosting
Software Life Cycle Processes
The system complies with the following standard:
- IEC 62304:2006 + A1:2016 – Medical device software – Software life cycle processes
Software Verification & Validation
The system complies with the following guidance:
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
{7}
| **Clinical Performance Data:** | Clinical testing was not required to demonstrate the substantial equivalence to the predicate device. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence to the predicate device. |
| --- | --- |
| **Conclusion:** | The conclusions drawn from the non-clinical performance data demonstrate that the subject device is as safe as and as effective as the predicate device. As such, we conclude that the substantial equivalence of the subject and the predicate device has been met, and the differences between the subject and the predicate device do not raise new questions of safety and effectiveness. |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.