K253964 · Unimed Medical Supplies, Inc. · DXQ · Jan 9, 2026 · Cardiovascular
Device Facts
Record ID
K253964
Device Name
Disposable Neonatal NIBP Cuff (U1681S-C51N)
Applicant
Unimed Medical Supplies, Inc.
Product Code
DXQ · Cardiovascular
Decision Date
Jan 9, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Unimed disposable blood pressure cuff (Models U1681S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (3–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated.
Device Story
Disposable neonatal NIBP cuff; accessory for non-invasive blood pressure monitors using oscillometry. Input: pneumatic pressure from monitor; Output: cuff inflation/deflation to occlude/release limb artery. Used in clinical settings by healthcare professionals. Cuff wraps around neonatal limb (3-6 cm circumference for model U1681S-C51N); transmits pressure signals to monitor. Facilitates blood pressure measurement; aids clinical decision-making regarding patient hemodynamic status. Single-patient use design reduces cross-contamination risk.
Clinical Evidence
Bench testing only. Performance verified per ISO 81060-1 and ISO 80601-2-30. Biocompatibility verified per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-23 (irritation) for short-term skin contact.
Technological Characteristics
Materials: TPU film (cuff), PVC (tubing), Nylon (hook/loop). Single-tube configuration. Pressure range: 0–300 mmHg. Non-sterile. Standards: ISO 81060-1, ISO 80601-2-30, ISO 10993-1/5/10/23.
Indications for Use
Indicated for neonatal patients requiring non-invasive blood pressure monitoring via oscillometric systems. Suitable for limb circumferences 3–15 cm. For single-patient use only.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 9, 2026
Unimed Medical Supplies, Inc.
Huanyu Zeng
Regulatory Affairs Specialist
Bld#8, Nangang 3rd Industrial Park, Tangtou
Shiyan, Baoan District
Shenzhen, 518108
China
Re: K253964
Trade/Device Name: Disposable Neonatal NIBP Cuff (U1681S-C51N)
Regulation Number: 21 CFR 870.1120
Regulation Name: Blood Pressure Cuff
Regulatory Class: Class II
Product Code: DXQ
Dated: December 7, 2025
Received: December 11, 2025
Dear Huanyu Zeng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253964 - Huanyu Zeng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253964 - Huanyu Zeng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jackson Hair -S 2026.01.09
14:57:12 -05'00'
for
- LCDR Stephen Browning
- Assistant Director
- Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
- Office of Cardiovascular Devices
- Office of Product Evaluation and Quality
- Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253964 | ? |
| Please provide the device trade name(s). | | ? |
| Disposable Neonatal NIBP Cuff (U1681S-C51N) | | |
| Please provide your Indications for Use below. | | ? |
| The Unimed disposable blood pressure cuff (Models U1681S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (3–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253964
A brand of Xinwell
unimed medical supplies
# 510K Summary
## 1. Submitter
Date Prepared: Dec. 11, 2025
Submitter/Manufacturer: Unimed Medical Supplies Inc.
Bld#8, Nangang 3rd Industrial Park, Tangtou,
Shiyan, Baoan District, Shenzhen, China 518108
FDA Establishment Number: 3007307487
Contact: Zeng Huanyu
RA Specialist
Tel: +86-755 26695165
E-mail: zenghy@unimed.cn
510(k) Submission Type: This is a Special 510(k).
## 2. Proposed Device
Trade Name: Disposable Neonatal NIBP Cuff
Common Name: Blood Pressure Cuff
Model Numbers: U1681S-C51N
Classification: Medical Specialty: Cardiovascular
Regulation: 21 CFR 870.1120 – Blood Pressure Cuff
Product Code: DXQ
Class: II
## 3. Predicate Device
| Predicate Device | |
| --- | --- |
| 510(K) No. | Trade Name |
| K251045 | Disposable Neonatal NIBP Cuff (U1682S-C51N);
Disposable Neonatal NIBP Cuff (U1683S-C51N);
Disposable Neonatal NIBP Cuff (U1684S-C51N);
Disposable Neonatal NIBP Cuff (U1685S-C51N) |
## 4. Device description
The Unimed Disposable Neonatal NIBP Cuff (Model U1681S-C51N) is a single-patient-use accessory for non-invasive blood pressure (NIBP) measurement. The cuff is designed for neonatal limb circumferences ranging from 3 cm to 6 cm.
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
{5}
K253964
A brand of Xinwell
unimed medical supplies
The cuff is intended to be used with compatible blood pressure monitors employing oscillometry. It is non-sterile, and is disposed of after single-patient use to reduce cross-contamination risks.
# 5. Intended use/Indications for use
The Unimed disposable blood pressure cuff (Models U1681S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (3–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated.
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
{6}
K253964
A brand of Xinwell
unimed
medical supplies
# 6. Comparison to predicate devices
| Features | Subject Device (U1681S-C51N) | Predicate Device (K251045) | Comment / Notes |
| --- | --- | --- | --- |
| Intended Use & Indications for Use | The Unimed disposable blood pressure cuff (Models U1681S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (3–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated. | The Unimed disposable blood pressure cuff (Models U1682S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (4–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated. | Same basic intended use; add U1681S-C51N to the product family |
| Prescription / Over-the-Counter Use | Prescription | Prescription | Same |
| Target Population | Neonatal | Neonatal | Same |
| Application Site | Arm (for neonatal use) | Arm (for neonatal use) | Same general application site. |
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
Page 3 of 6
{7}
K253964
A brand of xmwell
unimed
medical supplies
| Limb Circumference (cm) | - U1681S-C51N: Neonate 1 (3–6 cm) | - U1682S-C51N: Neonate 2 (4–8 cm)
- U1683S-C51N: Neonate 3 (6–11 cm)
- U1684S-C51N: Neonate 4 (7–13 cm)
- U1685S-C51N: Neonate 5 (8–15 cm) | Difference #1: Different circumference due to difference cuff size |
| --- | --- | --- | --- |
| Material | - Cuff: TPU film | - Cuff: TPU film | Same |
| | - Tubing: PVC | - Tubing: PVC | |
| | - Hook: Molded Nylon | - Hook: Molded Nylon | |
| | - Loop: Nylon | - Loop: Nylon | |
| Tube Configuration | One (single-tube) | One (single-tube) | Same |
| Pressure Limits | 0–300 mmHg | 0–300 mmHg | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Standards Met | ISO 81060-1, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 | ISO 81060-1, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 | Same |
Notes on Key Differences
- Difference #1 (Limb Circumference Range)
The subject device specifically adds U1681S-C51N to provide 3-6 cm circumference range. The safety and effectiveness of the product is not affected.
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
Page 4 of 6
{8}
K253964
A brand of Xinwell
unimed medical supplies
# 7. Verification and validation testing
## Non-clinical test data
The only potential risk-initiating difference between the subject device and the predicate device is the difference in cuff printing color. Non-clinical tests were conducted to verify that the proposed blood pressure cuff does not produce new risk in safety. The testing was performed in accordance with the following recognized consensus standards:
### Performance testing standards:
- ISO 81060-1:2007 – Non-invasive sphygmomanometers – Part 1: Requirements and test methods for non-automated measurement type
- ISO 80601-2-30:2018 – Medical electrical equipment — Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers
### Biocompatibility testing standards:
- ISO 10993-5 – Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 – Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23 – Biological evaluation of medical devices – Part 23: Tests for irritation
### Performance testing
The subject device is demonstrated to have the same performance as the predicate device through performance testing conducted in accordance with ISO 81060-1:2007 and ISO 80601-2-30:2018. The results demonstrated that the subject device meets the applicable requirements for its intended use and supports substantial equivalence to the predicate device.
### Biocompatibility testing
The subject device consists of patient-contacting materials, such as the cuff bladder and outer sleeve, which are in direct contact with intact skin. The contact duration is classified as short-term (i.e., less than 24 hours cumulative use) per ISO 10993-1:2018. Accordingly, the biocompatibility assessment was designed to address the relevant endpoints for short-term skin contact, and included the following tests:
- Cytotoxicity (ISO 10993-5)
- Skin Sensitization (ISO 10993-10)
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
{9}
K253964
A brand of Xinwell
unimed
medical supplies
- Skin Irritation (ISO 10993-23)
The results from these evaluations demonstrated that the materials used in the proposed cuff are biocompatible and suitable for its intended use.
## 8. Substantial Equivalence Statement
Based on the comparison, analysis, and the submitted verification data, Unimed believes that the subject device is as safe and effective and are substantially equivalent to the predicate device.
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.