K253939 · RemSleep Holdings, Inc. · BZD · Jan 7, 2026 · Anesthesiology
Device Facts
Record ID
K253939
Device Name
DeltaWave Nasal Pillow System
Applicant
RemSleep Holdings, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Jan 7, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DeltaWave Nasal Pillow System channels airflow noninvasively to a patient from a noninvasive positive airway pressure device (PAP) such as CPAP, bi-level. It is intended for adult patients weighing ≥66lbs (30 Kg), and for whom positive airway pressure has been prescribed. It is intended for single patient reuse in the home and hospital/institutional/sleep center environment.
Device Story
Nasal pillow system; channels pressurized air from CPAP/bi-level devices to patient nostrils; creates sealed interface via nasal pillows; pillows gently dilate nostrils to reduce airflow restriction and pressure drop. Used in home, hospital, institutional, and sleep center environments; operated by patient or healthcare professional. Output is delivered airflow; provider monitors therapy efficacy via PAP device settings. Benefits include reduced driving pressure requirements and effective non-invasive airway pressure delivery.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included CO2 rebreathing, dead space, pressure-flow/leakage, resistance to flow, vibration, and noise. Biocompatibility and cleaning validation performed per ISO standards.
Technological Characteristics
Nasal pillow interface with 22mm tubing connection. Materials evaluated per ISO 10993 and ISO 18562 series. Conforms to ISO 17510:2015. Features exhalation vents in cannula. Non-powered, passive interface.
Indications for Use
Indicated for adult patients weighing ≥66lbs (30 Kg) prescribed noninvasive positive airway pressure (CPAP or bi-level) therapy. Intended for single-patient reuse in home, hospital, institutional, and sleep center environments.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
Respironics, Inc. DreamWear Silicone Pillows Mask (K210844)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 7, 2026
RemSleep Holdings, Inc.
% Anne Dryden Siegal
QA / RA Director
ProMedic, LLC
131 Bay Point Dr. NE
St. Petersburg, Florida 33704
Re: K253939
Trade/Device Name: DeltaWave Nasal Pillow System
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD
Dated: December 9, 2025
Received: December 9, 2025
Dear Anne Dryden Siegal:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253939 - Anne Dryden Siegal
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253939 - Anne Dryden Siegal
Page 3
Sincerely,
Binoy J. Mathews-S
Digitally signed by Binoy J. Mathews-S
Date: 2026.01.07 12:55:08
-05'00"
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253939 | |
| Device Name DeltaWave Nasal Pillow System | |
| Indications for Use (Describe) The DeltaWave Nasal Pillow System channels airflow noninvasively to a patient from a noninvasive positive airway pressure device (PAP) such as CPAP, bi-level. It is intended for adult patients weighing ≥66lbs (30 Kg), and for whom positive airway pressure has been prescribed. It is intended for single patient reuse in the home and hospital/institutional/sleep center environment. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
REMSleep Holdings, Inc.
K253939
Date Prepared: January 7, 2026
Sponsor: RemSleep Holdings, Inc.
3222 Hwy 84
Suite 101
Blackshear, GA 31516
Sponsor: Tom Wood, CEO
T: 912-590-2001
Submission Correspondent: Anne Dryden Siegal
ProMedic, LLC
Subject Device Trade Name: DeltaWave™ Nasal Pillow System
Common/Usual Name: Nasal Mask, Mask, Cannula
Classification Name: Ventilator, Non-Continuous (Respirator)
Product Code: BZD
Regulation Number: 21 CFR 868.5905
Predicate Device Trade Name: DeltaWave™ Nasal Pillow System
Common/Usual Name: Nasal Mask, Mask, Cannula
Classification Name: Ventilator, Non-Continuous (Respirator)
Product Code: BZD
Regulation Number: 21 CFR 868.5905
Clearance Number: K233415
Reference Device Trade Name: Respironics, Inc. DreamWear Silicone Pillows Mask
Common/Usual Name: Nasal Mask, Mask, Cannula
Classification Name: Ventilator, Non-Continuous (Respirator)
Product Code: BZD
Regulation Number: 21 CFR 868.5905
Clearance Number: K210844
510(k) Summary
Page 1 of 7
{5}
REMSleep Holdings, Inc.
K253939
# Device Description
The DeltaWave Nasal Pillow System is the subject device for this submission. It is a non-invasive medical device that directs pressurized air from a Continuous Positive Airway Pressure (CPAP) or Bi-level medical device to the patient. The subject device is a type of medical device prescribed by a physician or authorized professional for patients who have been diagnosed with Sleep Apnea. The subject device is designed to prevent pressure drop from the CPAP machine to the patient's nasal passages. That is accomplished by maintaining the same, or greater flow space from the machine output to the patient's nostrils. To further ensure the least amount of pressure drop the nasal pillows are designed to gently dilate the patient's nostrils to avoid restrictions to incoming air entering the nostrils. This allows air to enter the patient's nostrils at a lower driving pressure.
# Principle of Operation
The device provides non-invasive means of delivering positive airway pressure to the user by creating a sealed interface using nasal pillows that fit directly into the patients' nostrils.
# Indications for Use:
The DeltaWave Nasal Pillow System channels airflow noninvasively to a patient from a noninvasive positive airway pressure device (PAP) such as CPAP, bi-level.
It is intended for adult patients weighing ≥66lbs (30 Kg), and for whom positive airway pressure has been prescribed. It is intended for single patient reuse in the home and hospital/institutional / sleep center environment.
**Patient Population:** Adult patients weighing ≥66lbs (30 Kg)
**Environments of use:** Home and hospital/institutional/sleep center environment
# Substantial Equivalence Discussion
## Indications for Use –
The indications for use for the subject and predicate device are similar.
**Discussion** – The indications for use was simplified and aligns with other CPAP masks.
## Technology and construction –
The technology of the subject and predicate device are identical.
**Discussion** – There is no difference in the technology and construction between the subject and predicate device.
## Environment of Use –
The environment of use is similar. We have expanded the patient population from single patient use in the home to single patient use in the home and hospital/institution/sleep center environment.
Therefore, we have included a reference device with a similar environments of use to the subject device.
**Discussion** – The environments of use are identical to the reference device (home and hospital/institution/sleep center).
## Patient Population –
The patient population of the subject device and predicate are identical.
**Discussion** – The patient population for both are adults weighing ≥66lbs (30 kg) who have been prescribed PAP therapy.
510(k) Summary
Page 2 of 7
{6}
REMSleep Holdings, Inc.
K253939
| | Subject Device
RemSleep Holdings, Inc.
DeltaWave Nasal Pillow System | Predicate Device
RemSleep Holdings, Inc.
DeltaWave Nasal Pillow System | Reference Device
Respironics, Inc.
DreamWear Silicone Pillows Mask | Comparison |
| --- | --- | --- | --- | --- |
| K# | K253939 | K233415 | K210844 | |
| Product Code | BZD | BZD | BZD | |
| CFR | 21 CFR 868.5905 | 21 CFR 868.5905 | 21 CFR 868.5905 | |
| Indications for Use | The DeltaWave™ Nasal Pillow System channels airflow noninvasively to a patient from a noninvasive positive airway pressure device (PAP) such as CPAP, bi-level.
It is intended for adult patients weighing ≥66lbs (30 Kg), and for whom positive airway pressure has been prescribed. It is intended for single patient reuse in the home and hospital/institutional / sleep center environment. | DELTAWAVE™ Nasal Pillow System is intended for adult patients weighing ≥66lbs (30kg), and for whom non-invasive Continuous Positive Airway Pressure (CPAP) or Bi-Level therapy has been prescribed by a physician or other authorized healthcare professional.
DELTAWAVE™ Nasal Pillow System is intended for single patient reuse in the home. | The DreamWear Silicone Pillows Mask is intended to provide an interface for application of CPAP or bi-level therapy to patients.
The mask is for single patient use in the home or multi-patient use in the hospital/institutional environment.
The mask is to be used on patients (>66lbs/30kg) for whom CPAP or bi-level therapy has been prescribed. | We have updated the Indications for Use statement to be clearer.
The only difference between the Subject and Predicate devices are the expanded environment of use for the Subject device.
Unlike the reference device, the Subject device is not intended for multi-patient use. |
| Patient Population | Adults weigh ≥66 lbs. (30kg) | Adults weigh ≥66 lbs. (30kg) | Patients weigh >66lbs (30kg) | Identical |
| Environment of Use | Home, Hospital/Institution / Sleep Centers | Home | Home and Hospital/Institutional environmental | Similar
We have expanded the environment use, which is similar to the reference device |
| Availability to Patient | Rx | Rx | Rx | Identical |
| Mask Type | Nasal Pillow with 2 ports fitting into the entrance of the patient’s nostrils | Nasal Pillow with 2 ports fitting into the entrance of the patient’s nostrils | Silicone nasal pillows cushion with tips that seal at the entrance to the nares. | Identical |
| Exhalation Vent | Many tiny air vents in the Cannula of the nasal pillow | Many tiny air vents in the Cannula of the nasal pillow | Exhalation ports incorporated into the front of the cushion and elbow | Identical |
| Tubing to connect to CPAP Device | 22mm | 22mm | 22mm | Identical |
510(k) Summary
Page 3 of 7
{7}
REMSleep Holdings, Inc.
K253939
| | Subject Device
RemSleep Holdings, Inc.
DeltaWave Nasal Pillow System | Predicate Device
RemSleep Holdings, Inc.
DeltaWave Nasal Pillow System | Reference Device
Respironics, Inc.
DreamWear Silicone Pillows Mask | Comparison |
| --- | --- | --- | --- | --- |
| Dead Space volume
for nasal pillows | 1.19 cu inch | 1.19 cu inch | S Frame
S Cushion – 72.8 ml
M Cushion – 74.0 ml
MW Cushion – 75.3 ml
L Cushion – 77.8 ml
Medium Frame
S Cushion – 75.8 ml
M Cushion – 77.0 ml
MW Cushion – 78.3 ml
L Cushion – 80.8 ml
Large Frame
S Cushion – 78.7 ml
M Cushion – 79.9 ml
MW Cushion – 81.2 ml
L Cushion – 83.7 ml | Identical |
| Pressure Range | 4 to 20 cmH₂O | 4 to 20 cmH₂O | 4 cm H₂O to 30 cm H₂O | Identical |
| Exhaust Flow Rates | 4 cm H₂O – 24.5 LPM
8 cm H₂O – 36.1 LPM
12cm H₂O – 45.8 LPM
16cm H₂O – 53.7 LPM
20 cm H₂O – 61.2 LPM | 4 cm H₂O – 24.5 LPM
8 cm H₂O – 36.1 LPM
12cm H₂O – 45.8 LPM
16cm H₂O – 53.7 LPM
20 cm H₂O – 61.2 LPM | Unknown | Identical |
| Resistance –
Pressure Drop | 0.3cm H₂O at 50 L/min
1.2 cm H₂O at 100 L/min | 0.3cm H₂O at 50 L/min
1.2 cm H₂O at 100 L/min | 50 SPLM
S Cushion – 1.3 cm H₂O
M Cushion – 1.19 cm H₂O
MW Cushion – 1.02 cm H₂O
L Cushion – 1.11 cm H₂O
100 SPLM
S Cushion – 4.71 cm H₂O
M Cushion – 4.29 cm H₂O
MW Cushion – 3.7 cm H₂O
L Cushion – 4.13 cm H₂O
50 SPLM (one tube occluded)
S Cushion – 2.2 cm H₂O
M Cushion – 2.1 cm H₂O
MW Cushion – 1.94 cm H₂O
L Cushion – 2.1 cm H₂O | Identical |
510(k) Summary
Page 4 of 7
{8}
REMSleep Holdings, Inc.
K253939
| | Subject Device
RemSleep Holdings, Inc.
DeltaWave Nasal Pillow System | Predicate Device
RemSleep Holdings, Inc.
DeltaWave Nasal Pillow System | Reference Device
Respironics, Inc.
DreamWear Silicone Pillows Mask | Comparison |
| --- | --- | --- | --- | --- |
| | | | 100 SPLM (one tube occluded)
S Cushion – 7.7 cm H2O
M Cushion – 7.41 cm H2O
MW Cushion – 6.66 cm H2O
L Cushion – 7.21 cm H2O | |
| **Sound Pressure & Power Level** | A-weighted Sound Pressure is 29.7 dBA (at 10 cm H2O)
A-weighted Sound Power Level 37.7 (at 10 cm H2O) | A-weighted Sound Pressure is 29.7 dBA (at 10 cm H2O)
A-weighted Sound Power Level 37.7 (at 10 cm H2O) | Measured Sound Pressure Level: 27dBA
Measured Sound Power Level: 19dBA | Identical |
| **Reprocessing methods** | Mild soap and water | Mild soap and water | Air path and non-air path components – Cleaning with liquid dish detergent
Air path components – High level chemical and thermal disinfection
Non-air path components thermal disinfection | Identical |
| **CO2 Rebreathing** | Results:
4 cm H2O -2.0%
5 cm H2O -2.0%
10 cm H2O -4.0% | Results:
4 cm H2O -2.0%
5 cm H2O -2.0%
10 cm H2O -4.0% | Results:
MW Cushion with Large Frame
4 cm H2O – 5.0%
5 cm H2O – 5.0%
10 cm H2O – 5.0% | The CO2 Rebreathing is consistent with ISO 17510:2015 requirements. |
510(k) Summary
Page 5 of 7
{9}
REMSleep Holdings, Inc.
K253939
# Non-Clinical Testing Summary
## Bench testing
Performance testing demonstrated that the subject device met its acceptance criteria, though no new testing was performed on the subject device. Testing on the predicate device included biocompatibility, performance and cleaning validation testing. The following is a summary of the tests performed and standards conformance:
## Bench Testing
- Cleaning Validation Testing
- Shelf-life, Storage and Transportation
- CO2 Rebreathing
- Dead Space
- Pressure – Flow Curve/Leakage
- Resistance to Flow/Pressure Drop
- Vibration and Noise
## Standards/Guidance Conformation
- ISO 17510:2015 Medical devices — Sleep apnoea breathing therapy — Masks and application accessories
- ISO 10993-1: 2025 Biological evaluation of medical devices Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process
- ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2017 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity -Acute Systemic Toxicity
- ISO 10993-17:2023 Biological evaluation of medical devices Part 17: Toxicological risk assessment of medical device constituents
- ISO 10993-18:2020 Biological evaluation of medical devices Part 18: Chemical characterization of medical device materials within a risk management process
- ISO 18562-1: 2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds (VOCs)
- FDA Biocompatibility Guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process": Guidance for Industry and Food and Drug Administration Staff (09/08/2023)
- FDA Reprocessing Guidance - Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling : Guidance for Industry and Food and Drug Administration Staff (03/17/2015)
510(k) Summary
{10}
REMSleep Holdings, Inc.
K253939
## Discussion of Differences
The predicate device – DeltaWave™ Nasal Pillow System - K233415 was cleared for home use only. In this new submission, the environment of use is being expanded to Home, Hospital/Institution, and Sleep Centers. There are no other changes to the subject device.
We have included a reference device – Respironics, Inc. DreamWear Silicone Pillows Mask (K210844), which has clearance for Home and Hospital/Institution environments
## Substantial Equivalence Conclusion
The subject device is substantially equivalent to the predicate device. No new questions of safety or efficacy have been raised compared to the predicate.
510(k) Summary
Page 7 of 7
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.