K253919 · Xiamen Weiyou Intelligent Technology Co., Ltd. · IRP · Jun 10, 2026 · Physical Medicine
Device Facts
Record ID
K253919
Device Name
Cryo And Thermo Therapy System
Applicant
Xiamen Weiyou Intelligent Technology Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Jun 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cryo And Thermo Therapy System ( Model:VU-COT02) : The device provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings and home settings. Cryo And Thermo Therapy System (Model: VU-COT01-V1, VU-COT01-V2, VU-COT03) : The device provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings.
Device Story
AC-powered, software-controlled multi-modality system; provides localized heat exchange and intermittent pneumatic compression. Inputs: user-selected therapy programs (heat, cold, contrast, compression). Operation: circulates water through cuffs; refrigeration/heating system (or external water source for VU-COT02) regulates temperature; pneumatic pump applies cyclic compression. Output: thermal and mechanical compression therapy to patient limbs/body. Used in rehabilitation, clinics, athletic settings (and home for VU-COT02) by healthcare professionals. Output allows clinicians to manage post-surgical/trauma edema, swelling, and pain. Benefits: non-invasive symptom management.
Clinical Evidence
No clinical data. Bench testing only: temperature accuracy, pressure accuracy, pressure leak performance, cuff seam strength, and durability testing (6 months/182.5 hours). Electrical safety and EMC testing per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Technological Characteristics
AC-powered (100-240V, 50/60Hz). Refrigeration/hydraulic system for thermal exchange; pneumatic pump for compression. Cuffs: 3-layer fabric, dual-chamber (compression/thermal). Interface: segment or touchscreen. Connectivity: standalone. Sterilization: non-sterile. Software: electronic pressure/time/temperature monitoring.
Indications for Use
Indicated for patients with post-surgical or trauma injuries requiring reduction of edema, swelling, and pain via cold and compression, or post-traumatic/post-surgical conditions requiring localized thermal therapy (hot, cold, or contrast).
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 10, 2026
Xiamen Weiyou Intelligent Technology Co., Ltd.
% Eva Li
Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
Rm. 1401, Dongfang Bldg.
#1500 Century Ave.
Shanghai, 200122
China
Re: K253919
Trade/Device Name: Cryo And Thermo Therapy System
Regulation Number: 21 CFR 890.5650
Regulation Name: Powered Inflatable Tube Massager
Regulatory Class: Class II
Product Code: IRP, ILO
Dated: May 12, 2026
Received: May 13, 2026
Dear Eva Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253919 - Eva Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253919 - Eva Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**Tushar Bansal -S**
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253919 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Cryo And Thermo Therapy System (VU-COT01-V1; VU-COT01-V2; VU-COT02; VU-COT03) Cuff Models: Arm Cuff (CC-EA); Waist Cuff (CC-WA); Knee Cuff (CC-KA); Shoulder Cuff (CC-SA); Foot/Calf Cuff (CC-FA); Ankle Cuff (CC-AA); Leg Cuff (CC-LA); Knee Cuff (CC-K); Arm Cuff (CC-E). | | |
| Please provide your Indications for Use below. | | ? |
| Cryo And Thermo Therapy System ( Model:VU-COT02) : The device provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings and home settings. | | |
| Cryo And Thermo Therapy System (Model: VU-COT01-V1, VU-COT01-V2, VU-COT03) : The device provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Xiamen Weiyou Intelligent Technology Co., Ltd.---
# 510(K) Summary
**(K253919)**
## 1. General Information
| Applicant name | Xiamen Weiyou Intelligent Technology Co., Ltd. |
| --- | --- |
| Address | Unit 301 No. 6 Xianghong Road, Torch Hi-Tech Zone Industrial Park, Xiang'an District, Xiamen, P.R.China |
| Contact | Kelly Guo |
Date prepared: June 08, 2026
### Device name
Device Trade Name: Cryo And Thermo Therapy System
Common Name: Powered inflatable tube massager
Classification name: Massager, Powered Inflatable Tube
Regulation Number: 890.5650
Product Code: IRP, ILO
Legally Marketed Predicate Device
K242754
Z-ONE, ZT Cube (ZC3), ZT Clinic
ZAMAR Medical d.o.o.
## 2. Device Description
AC-powered, software-controlled multi-modality devices, designed to impart localized heat exchange and compression therapies. The devices have been developed with the possibility of choosing between different programs: cold therapy and heat therapy. Moreover, they also possess an intermittent pneumatic compression activity, which is the effect of an alternating compression and decompression.
The Cryo And Thermo Therapy System have 4 models :
VU-COT01-V1;
VU-COT01-V2;
VU-COT02;
VU-COT03
And the Cuff has different option models:
Arm Cuff (CC-EA);
Waist Cuff (CC-WA);
Knee Cuff (CC-KA);
Shoulder Cuff (CC-SA);
Foot/Calf Cuff (CC-FA);
Ankle Cuff (CC-AA);
1 / 10
{5}
Xiamen Weiyou Intelligent Technology Co., Ltd.
Leq Cuff (CC-LA);
Knee Cuff (CC-K);
Arm Cuff (CC-E).
2.1 VU-COT01-V1/VU-COT01-V2
These models are designed to deliver both cooling and heating therapy. The temperature range for heat is 38°C-40°C, while that for cold is 3°C-15°C. The treatment pressure range is from 30 to 120 mmHg.
It is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings.
2.2 VU-COT02
This model is designed to circulate water only. The device cannot provide cooling or heating. The water temperature is determined by the temperature of the water that the user adds to the reservoir.
It is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings, and home setting.
2.3 VU-COT03
This model is designed to deliver both cooling and heating therapy. The temperature for heat is one option 40°C, while that for cold is range from 5°C to 15°C.
It is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings.
3. Indication for use
Cryo And Thermo Therapy System (Model:VU-COT02) :
The device provides a combination of cold, heat, contrast, and compression therapies.
It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated.
It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated.
The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings and home settings.
Cryo And Thermo Therapy System (Model: VU-COT01-V1, VU-COT01-V2, VU-COT03) :
The device provides a combination of cold, heat, contrast, and compression therapies.
It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated.
2 / 10
{6}
Xiamen Weiyou Intelligent Technology Co., Ltd.---
It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings.
#### **4. Substantial equivalence comparison**
In the table below a detailed comparison of the subject devices Cryo And Thermo Therapy System (VU-COT01-V1, VU-COT01-V2, VU-COT02, VU-COT03) and predicate device Z-ONE, ZT Cube, ZT Clinic
3 / 10
{7}
Xiamen Weiyou Intelligent Technology Co., Ltd.
Comparison table
| Characteristic | Predicate device(K242754) Z-ONE, ZT Cube, ZT Clinic | Subject device(K253919) Cryo And Thermo Therapy System | Comparison |
| --- | --- | --- | --- |
| Manufacturer | ZAMAR Medical D.o.o | Xiamen Weiyou Intelligent Technology Co., Ltd. | |
| Classification Name | Powered inflatable tube massager | Powered inflatable tube massager | same |
| Classification Product Code | IRP | IRP | Same |
| Subsequent Product Code | ILO | ILO | same |
| Indications for use | The Heat Exchange Equipment for Cryotherapy and Thermotherapy provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings | VU-COT01-V1, VU-COT01-V2, VU-COT03: The device provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, and athletic training settings VU-COT02: The device provides a combination of cold, heat, contrast, and compression therapies. It is intended to treat post-surgical and trauma injuries to reduce edema, swelling and pain, for which cold and compression are | Different*1 |
4 / 10
{8}
Xiamen Weiyou Intelligent Technology Co., Ltd.
| | | indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions, for which localized thermal therapy (hot or cold or contrast) are indicated. The equipment is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home setting | |
| --- | --- | --- | --- |
| Intended Users | Health Care Professionals only (Prescription use) | Health Care Professionals only (Prescription use) | same |
| Where used (hospital, home, ambulance, etc) | Intended for indoor use only such as rehabilitation clinics, outpatient clinic, athletic training settings. | VU-COT01-V1, VU-COT01-V2, VU-COT03: Intended for indoor use only such as rehabilitation clinics, outpatient clinic, athletic training settings VU-COT02: Intended for indoor use only such as rehabilitation clinics, outpatient clinic, athletic training settings, and home setting. | Different *1 |
| Number of Patients that can be treated | Two patients with single locations or two anatomical locations on one patient. 4 possible connections with ZT Clinic | 2 options: Two patients with single locations or two anatomical locations on one patient. 1 patient | similar |
| **Physical unit** | | | |
| Energy used and/or delivered | AC powered | AC powered | same |
| User Interface | Display touchscreen Z-ONE: 5' ZT Cube: 7' ZT Clinic: 10,4' | VU-COT01-V1, VU-COT02: 5''Segment screen VU-COT01-V2: 5'' touch scree VU-COT03: 4.3''touch scree | similar |
| Design | Z-ONE: portable unit, equipped with a handle for moving. ZT Cube: portable unit, equipped with a handle for moving. | VU-COT01-V1/ VU-COT01-V2: portable unit, equipped with a handle and Back strap for moving | similar |
5 / 10
{9}
Xiamen Weiyou Intelligent Technology Co., Ltd.
| | ZT Clinic: wheeled model, with brakes. | VU-COT02/ VU-COT03: portable unit, equipped with a handle for moving | |
| --- | --- | --- | --- |
| Dimensions | Z-ONE: 11' x 10' x 12'H ZT Cube: 15' x 12' x 13'H ZT Clinic: 15' x 22' x 37'H | VU-COT01-V1/ VU-COT01-V2: 20' x 11' x 11'H VU-COT02: 13' x 8' x 10'H VU-COT03: 11' x 7' x 13'H | similar |
| Weight empty | Z-ONE: 20 lbs (9kg) ZT Cube: 31 lbs (14kg) ZT Clinic: 95 lbs (43kg) | VU-COT01-V1/ VU-COT01-V2: 22 lbs (10kg) VU-COT02: 5.8 lbs (2.6 kg) VU-COT03: 11.6lbs (5.3kg) | similar |
| Weight fully loaded | Z-ONE: 26 lbs (12kg) ZT Cube: 37 lbs (17kg) ZT Clinic: 106 lbs (48kg) | VU-COT01-V1/ VU-COT01-V2: 44lbs (20kg) VU-COT02: 22lbs (10 kg) VU-COT03: 28.6 lbs (13kg) | similar |
| Chilling Mechanism | Combination of a refrigeration system with a hydraulic system | VU-COT01-V1/ VU-COT01-V2/ VU-COT03: Combination of a refrigeration system with a hydraulic system VU-COT02: External cold water circulation | similar |
| Heating Mechanism | Combination of a refrigeration system (with the aid of 4-way valve to force the reverse cycle) with a hydraulic system. | VU-COT01-V1/ VU-COT01-V2/ VU-COT03: Combination of a refrigeration system with a hydraulic system VU-COT02: External hot water circulation | similar |
| Reservoir Fluid Capacity | Single reservoir of about one liter | VU-COT01-V1, VU-COT01-V2: hot water tank 1.7L and cold water tank 0.7L VU-COT02: single water tank 1.7L VU-COT03: single water tank 4.5L | similar |
| **Functions** | | | |
| Heat Therapy | Available within the specific temperature ranges of the models, either without compression or with low compression. | VU-COT01-V1, VU-COT01-V2, VU-COT03: Available within the specific temperature ranges of the models, either without compression or with low compression. | similar |
6 / 10
{10}
Xiamen Weiyou Intelligent Technology Co., Ltd.
| | Z-ONE: only cold therapy | VU-COT02: The device cannot be heated and it depends on the external water temperature. | |
| --- | --- | --- | --- |
| Heat Therapy Treatment time | 5 to 30 minutes. | 5 to 99minutes adjustable Default: 30 minutes | similar |
| Cold Therapy | Available without and with compression in 4 levels (Low, Medium-low, Medium, High) | Available without and with compression | similar |
| Cold Therapy Treatment Time | 5 to 90 minutes. | VU-COT01-V1, VU-COT02: 5 to 99 minutes. VU-COT01-V2, VU-COT03: 5 to 90 minutes. | similar |
| Contrast Therapy | Available for ZT Cube and ZT Clinic. Treatment cycles of Cryotherapy, Thermotherapy, Pause and Repeat can be individually set. Treatment up to 90 minutes. Not available for Z-ONE. | Available for VU-COT01-V1, VU-COT01-V2. Not available for VU-COT02, VU-COT03. | similar |
| Compression range | Available in four levels Low: 15 mmHg Medium-low: 40 mmHg Medium: 60 mmHg High: 80 mmHg Possibility of selecting compression time applied to the treated part, intermittent compression. | VU-COT01-V1,VU-COT01-V2,VU-COT02:10 level adjustable ranging 30-120mmHg VU-COT03: low-40mmHg, medium-70mmHg, high-100mmHg | similar |
| Temperature range | Z-ONE: 41°F - 59°F ZT Cube: 34°F - 104°F ZT Clinic: 34°F – 113°F | **Cold:** VU-COT01-V1,VU-COT01-V2:37°F-59°F VU-COT03: 41°F - 59°F **Heat:** VU-COT01-V1, VU-COT01-V2:100°F-104°F VU-COT03:104°F | similar |
| Software features | Electronic pressure control and therapy time and temperature monitoring. | Electronic pressure control and therapy time and temperature monitoring. | same |
| **Electrical characteristic** | | | |
| Line Voltage | 115V | 100-240V | similar |
| Line Frequency | 60 Hz | 50/60Hz | similar |
7 / 10
{11}
Xiamen Weiyou Intelligent Technology Co., Ltd.
| Electrical Safety Standards | - IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION - ANSI AAMI ES60601 1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] - IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Type BF | VU-COT01-V1,VU-COT01-V2,VU-COT03: - IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION - ANSI AAMI ES60601 1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] - IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Type BF VU-COT02: - IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION - ANSI AAMI ES60601 1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] - IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Type BF - IEC 60601-1-11:2020 | Similar |
| --- | --- | --- | --- |
| **Environmental characteristics** | | | |
| Operating Temperature | 50°F - 86°F (10°C – 30°C) | 10°C~40°C | similar |
| Storage Temperature | 50°F - 104°F (10°C – 40°C) | -40°C~70°C | similar |
| Operating Humidity | Relative humidity 10% - 50% | Relative humidity 30%-85%RH | similar |
| Storage Humidity | Relative humidity 30%-60% | Relative humidity 10%-100%RH | similar |
| Operating Atmospheric Pressure | 0 – 6,500 Ft (0 – 2,000 m) | 50kPa-106kPa | similar |
| Expected lifetime | 5 years | 5 years | same |
8 / 10
{12}
Xiamen Weiyou Intelligent Technology Co., Ltd.
| Cuff | | | |
| --- | --- | --- | --- |
| Types of cuff | The cuffs are available for different parts of the body: full leg, shoulder, thigh, knee calf, lumbar, elbow wrist, ankle, knee, hip. | The cuffs are available for different parts of the body: arm cuff-all in one, waist cuff-all in one, knee cuff-all in one, shoulder cuff-all in one, foot+calf cuff-all in one, ankle cuff-all in one, leg, arm, knee | similar |
| Cuff technology | Flexible cuffs, formed by two separated chambers: one for the compression activity and one with a coil allowing the passage of the NON TOX liquid. | Flexible cuffs, The cuff is formed by pressing three layers of fabric into two chambers. One chamber is used for applying pressure, and the other for cold or hot water compresses. | similar |
| Patient Contacting Material | Polyether Polyurethane | Nylon | Different*2 |
| Expected lifetime of accessories | 1 years or 200 applications | 6 months | Different*2 |
| Sterility | Provided non-sterile only | Provided non-sterile only | same |
# Discussion:
Different*1, Model VU-COT02 pass the testing of IEC 60601-1-11, so home setting is safety and effective for this model.
Different*2, The material and expected lifetime of the wraps are different. These differences are mitigated by limiting skin contact with the wraps and with labeling addressing appropriate integrity conditions of wraps. Therefore, these differences do not raise questions of safety or effectiveness.
9 / 10
{13}
Xiamen Weiyou Intelligent Technology Co., Ltd.
## 5. Non- Clinical testing
Performance of the subject device have been evaluated and tested by performing
- Temperature accuracy
- Pressure accuracy
- Pressure leak performance
- Pressure resistance of cuff (seam strength)
Results of the tests indicated that the devices conform to theirs predetermined specifications and operate within safety limits.
The durability of the thermal wraps has been tested under simulated real-use conditions, demonstrating the device's ability to maintain its integrity and performance over 6 month (Equal to 182.5 hours) repeated application cycles and maintain safety limits.
The device also pass the following testing:
(a) ANSI AAMI ES60601-1 "Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1)".
(b) IEC 60601-1-2 "Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
And the device(VU-COT02) passed IEC 60601-1-11 General requirements for basic safety and essential performance Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
## 6. Clinical testing
N/A
## 7. Conclusion.
In conclusion, the testing data validates that the subject devices perform as intended under their proposed use conditions. The Cryo And Thermo Therapy System meets all safety and performance requirements, confirming substantial equivalence to the predicate device (K242754) in terms of indications for use and technological characteristics.
10 / 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.