K253910 · GM Dos Reis Industria e Comercio Ltda. · GAT · Aug 22, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253910
Device Name
Stitch - Suture Wires and Tapes
Applicant
GM Dos Reis Industria e Comercio Ltda.
Product Code
GAT · General, Plastic Surgery
Decision Date
Aug 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Stitch - Suture Wires and Tapes are indicated for use in general soft tissue approximation and/or ligation.
Device Story
Stitch - Suture Wires and Tapes are nonabsorbable surgical sutures and tapes used for soft tissue approximation and ligation during surgery. The device consists of UHMWPE (ultra-high molecular weight polyethylene) sutures and AISI300 series stainless steel needles. It is provided in various diameters and sizes (USP 3-0 to 5; tapes 1.3mm to 2.0mm) to accommodate different patient anatomies. The device is intended for single use and is supplied sterile via ethylene oxide. Surgeons use the sutures to hold soft tissues together until healing occurs. The device provides mechanical support for tissue union, benefiting patients by facilitating surgical repair and wound closure.
Clinical Evidence
No clinical data were included in this submission. Performance was demonstrated through bench testing according to USP standards (Diameter <861>, Needle Attachment <871>, Tensile Strength <881>) and biological evaluation per ISO 10993-1:2018.
Technological Characteristics
Materials: UHMWPE suture, AISI300 stainless steel needles (ASTM F899). Design: Braided nonabsorbable suture. Sizes: USP 3-0, 2-0, 0, 1, 2, 5; tapes 1.3mm, 1.6mm, 2.0mm. Sterilization: Ethylene oxide (ISO 11135, SAL 10⁻⁶). Colorants: Chromium-cobalt-aluminum oxide (<2%) and D&C Black (<1%).
Indications for Use
Indicated for general soft tissue approximation and/or ligation in adult patients.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 22, 2026
GM Dos Reis Industria e Comercio Ltda.
Guilherme Esteves Pontes
Senior Regulatory Affairs Analyst
Avenida Pierre Simon De Laplace, 600
Campinas, 13069320
Brazil
Re: K253910
Trade/Device Name: Stitch - Suture Wires and Tapes
Regulation Number: 21 CFR 878.5000
Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture
Regulatory Class: Class II
Product Code: GAT
Dated: July 16, 2026
Received: July 16, 2026
Dear Guilherme Esteves Pontes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K253910 - Guilherme Esteves Pontes
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253910 - Guilherme Esteves Pontes
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N.
LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253910 | ? |
| Please provide the device trade name(s). | | ? |
| Stitch - Suture Wires and Tapes | | |
| Please provide your Indications for Use below. | | ? |
| Stitch - Suture Wires and Tapes are indicated for use in general soft tissue approximation and/or ligation. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
K253910
Page 1 of 3
# 510(k) Summary
(K253910)
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
# I. Submitter:
GM Dos Reis Industria e Comercio Ltda
Avenida Pierre Simon de La Place 600
Campinas, São Paulo, Brazil 13069-320
Guilherme Esteves Pontes, Regulatory Affairs Specialist
Telephone: +55 (19) 3765-9900
Email: qualidade4@gmreis.com.br
Date prepared: August 12, 2026
# II. Device Name:
Stitch - Suture Wires and Tapes
Common Name:
Suture, Nonabsorbable, Synthetic, Polyethylene
Classification Name:
Nonabsorbable poly(ethylene terephthalate) surgical suture
Device Class:
II
Product Codes:
GAT
Regulation Number:
21 CFR 878.5000
# III. Predicate Devices:
Legally marketed devices to which we are claiming “Substantial Equivalence” is the following:
HS Fiber Sutures - Riverpoint Medical (K231163) (Primary)
Arthrex Suture - Arthrex Inc. (K122374) (Additional)
Stitch Cerclage - Suture Tapes - GMReis (K243905) (Reference Device)
EXPERT - Joint Fixation System - GMReis (K200332) (Reference Device)
# IV. Device Description:
The purpose of this submission is to obtain marketing clearance for Stitch - Suture Wires and Tapes. Stitch - Suture Wires and Tapes are used in surgeries for approximations and for union of soft tissues until healing occurs. The surgical sutures and tapes are made of UHMWPE (ultra-high molecular weight polyethylene), and the needles are made in AISI300 steel series, according to the standard ASTM F899. They are found in a variety of diameters to meet the range of anatomies of the patients; it is provided in sterile condition sterilized by Ethylene Oxide. The sutures are classified as non-absorbable and meet the USP standards. This product is for single use only. This device comes in sizes USP 3-0, 2-0, 0, 1, 2 and 5 and suture tape sizes 1.3mm, 1.6mm and 2.0mm.
{6}
K253910
Page 2 of 3
The color additives used are accepted by FDA, according to the guidance “Summary of Color Additives for Use in the United States in Foods, Drugs, Cosmetics, and Medical Devices”; Color Additives Approved for Use in Medical Devices, as follows:
- Chromium-cobalt-aluminum oxide (CFR §73.1015) for the blue color in concentration lower than 2% by weight of the suture material.
- D&C Black (CFR §74.3054) for the black color in concentration lower than 1% by weight of the suture material.
# V. Indications for Use of the Device:
Stitch - Suture Wires and Tapes are indicated for use in general soft tissue approximation and/or ligation.
# VI. Comparison of Technological Characteristics with The Predicate Devices:
The subject device is substantially equivalent in indications and design principles to the following predicate devices:
K231163 - HS Fiber Sutures - Riverpoint Medical
K122374 - Arthrex Suture - Arthrex Inc.
The subject and predicate devices have equivalent intended use and equivalent technological characteristics. They are all manufactured from identical materials and share equivalent design characteristics like physical dimensions. Any difference in the technological characteristics do not raise new issues of safety or efficacy. The performance of the subject device was compared with the specific FDA guidance and was demonstrated through mechanical testing according to USP standards and comparison side-by-side with the predicate. No clinical data were included in this submission. The subject device is provided sterile by Ethylene Oxide method. The sterilization process, parameters, validation and shelf life is the same as the reference device cleared under the 510(k) K243905 and K200332.
| Specification | Subject Device | Primary Predicate Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- |
| | GM Dos Reis Stitch - Suture Wires and Tapes | Riverpoint (K231163) HS Fiber Sutures | Arthrex Inc. (K122374) Arthrex Suture | GM Dos Reis (K243905) Stitch Cerclage - Suture Tapes | GM Dos Reis (K200332) EXPERT - Joint Fixation System - GMReis |
| Indications for Use | Stitch - Suture Wires and Tapes are indicated for use in general soft tissue approximation and/or ligation. | HS Fiber sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular surgery, and the use of allograft tissues for orthopedic procedures. | The Arthrex Suture is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft or autograft tissue are used for repair. | Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries as bone fixation cerclage, specifically: • Sternotomy indications including the “rewiring” of osteomized sternums; • Repair of long bone fractures due to trauma or reconstruction. | The EXPERT – Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting. The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following: Mini EXPERT When used for fixation of bone-to-bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as: - Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions; - Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and - Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the |
{7}
K253910
Page 3 of 3
| | | | | | healing process of the reconstruction of the ligament at the base of the thumb metacarpal by providing stabilization between the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis. EXPERT Knotless and EXPERT Knotless Dual Syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures. |
| --- | --- | --- | --- | --- | --- |
| Materials | UHMWPE and Stainless Steel | UHMWPE and Stainless Steel | UHMWPE and Stainless Steel | UHMWPE and Stainless Steel | UHMWPE and Stainless Steel |
| Design | Braided Sutures | Braided Sutures | Braided Sutures | Braided Sutures | Braided Sutures |
| Performance Testing | USP Standards | USP Standards | USP Standards | USP Standards | USP Standards |
| Application Procedure | Soft tissue fixation | Soft tissue fixation | Soft tissue fixation | Soft tissue fixation | Soft tissue fixation |
| Presentation | Sterile | Sterile | Sterile | Sterile | Sterile |
| Regulation Name | Nonabsorbable poly(ethylene terephthalate) surgical suture | Nonabsorbable poly(ethylene terephthalate) surgical suture | Nonabsorbable poly(ethylene terephthalate) surgical suture | Bone Fixation Cerclage | Single/multiple component metallic bone fixation appliances and accessories. |
| Regulation Number | 21 CFR 878.5000 | 21 CFR 878.5000 | 21 CFR 878.5000 | 21 CFR 888.3010 | 21 CFR 888.3030 |
| Product Code | GAT | GAT | GAT | JDQ | HTN |
| Class | II | II | II | II | II |
| Sizes | Suture Wire: 3-0 / 2-0 / 0 / 1 / 2 and 5 Suture tape: 1.3 / 1.6 and 2.0mm | Suture Wire: 3-0 / 2-0 / 0 / 1 / 2 and 5 Suture tape: 1.3 / 1.6 and 2.0mm | Suture Wire: 3-0 / 2-0 / 0 / 1 / 2 and 5 Suture tape: 1.3 / 1.7 and 2.0mm | Suture tape: 2.5mm | Sutures from 2-0 to 5 Plates from Ø2.6 to Ø5.5 |
### VII. Performance Data:
The non-clinical tests performed and the Recognized Consensus Standards to which the SStitch - Suture Wires and Tapes have been tested are specifically the following:
- Sutures - Diameter <861>
- Sutures - Needle Attachment <871>
- Tensile Strength <881>
Materials used were evaluated per ISO 10993-1:2018 – Biological Evaluation of Medical Devices. The subject device components are provided sterile. The sterilization method is ethylene oxide exposure. The EO sterilization has been validated to a sterility assurance level (SAL) of 10⁻⁶, according to ISO 11135 Sterilization of health care products - Ethylene oxide - Requirements for the development, validation, and routine control of a sterilization process for medical devices. The shelf life is the same as the reference device cleared, and they share equivalent characteristics ensuring that no new worst-case scenario for shelf-life and sterilization is identified. The medical device in its final finished form is identical to Stitch Cerclage - Suture Tapes (K243905) (previously marketed device) in formulation, processing, design, sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, additives, additives, additives, cleaning agents, mold release agents).
### VIII. Conclusions:
As was established in this submission, the subject Stitch - Suture Wires and Tapes are substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics, intended use, indications for use, material composition, anatomical region, multiple sizes, and basic design features compared to its predicate devices. Any differences between the subject and the predicate devices are considered minor and do not raise different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.