RealGUIDE is a stand-alone software indicated for the preparation of patient specific treatment planning by professional user and, in particular, it must be used by dental clinician, general practitioner (DDS) dentist, or specialist such as, a Periodontist or Oral Sur-geon. The software is indicated for use on patients that require dental implant placement planning, prosthesis modelling and endodontic treatments connected to the endosseous dental implant placement. RealGUIDE is, in further detail, indicated for: - Viewing in 2D/3D patient images and processing them (DICOM or STL), including volume editing and measurement - Segmenting bone, nerve canal and teeth - Virtually positioning dental implants - Designing templates for endosseous dental implant guided placement - Modelling reconstructive devices, such as bone grafts and grids - Designing dental prosthetics, such as crown, bridges and dentures - Saving dental surgery template's or prostheses' design to file RealGUIDE software allows for endosseous dental implant surgical guide and prosthetic component file creation (*.STL) for the user to further process with specific CAD systems before export to validated manufacturing centers or to the point of care. Manufacturing at the point of care requires a validated process using CAM equipment (additive manufacturing system, including software and associated tooling) and compatible material (biocompatible and sterilizable). The software is not intended to perform automatic diagnosis and to control the manufacture of devices.
Device Story
RealGUIDE is stand-alone software for dental treatment planning; inputs include 2D/3D DICOM images and STL files. Software provides tools for volume rendering, segmentation (bone, nerve, teeth), virtual implant positioning, and design of surgical guides, bone grafts, and dental prosthetics. Operated by dental professionals in clinics; available as desktop application or web-browser interface. Software transforms imaging data into STL files for downstream CAD/CAM manufacturing. AI/ML algorithms assist in segmentation and planning tasks previously performed manually. Output aids clinicians in visualizing anatomy and designing patient-specific devices; does not perform automatic diagnosis or control manufacturing. Benefits include improved surgical planning accuracy and workflow efficiency.
Clinical Evidence
Bench testing only. Evidence includes software verification and validation results and AI/ML validation. No clinical data presented.
Technological Characteristics
Stand-alone software; runs on local PC or via web browser (server-hosted). Features 2D/3D DICOM/STL processing, segmentation, and CAD modeling. Employs locked AI/ML algorithms for specific tasks. Connectivity via digital download or web access. No hardware components; relies on user-provided CAM equipment for manufacturing.
Indications for Use
Indicated for dental clinicians, general practitioners, periodontists, or oral surgeons planning dental implant placement, prosthesis modeling, and associated endodontic treatments for patients requiring such procedures.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 4, 2026
3diemme Srl
% Alesandro Avis
Quality and Regulatory Supervisor
Via Risorgimento 9
CANTÙ, 21047
ITALY
Re: K253902
Trade/Device Name: RealGUIDE
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: December 23, 2025
Received: December 23, 2025
Dear Alesandro Avis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253902 - Alesandro Avis
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253902 - Alesandro Avis
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Digitally signed by GABRIELA M. for RODAL -S
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253902 | ? |
| Please provide the device trade name(s). | | ? |
| RealGUIDE | | |
| Please provide your Indications for Use below. | | ? |
| RealGUIDE is a stand-alone software indicated for the preparation of patient specific treatment planning by professional user and, in particular, it must be used by dental clinician, general practitioner (DDS) dentist, or specialist such as, a Periodontist or Oral Sur-geon. The software is indicated for use on patients that require dental implant placement planning, prosthesis modelling and endodontic treatments connected to the endosseous dental implant placement. RealGUIDE is, in further detail, indicated for: - Viewing in 2D/3D patient images and processing them (DICOM or STL), including volume editing and measurement - Segmenting bone, nerve canal and teeth - Virtually positioning dental implants - Designing templates for endosseous dental implant guided placement - Modelling reconstructive devices, such as bone grafts and grids - Designing dental prosthetics, such as crown, bridges and dentures - Saving dental surgery template's or prostheses' design to file RealGUIDE software allows for endosseous dental implant surgical guide and prosthetic component file creation (*.STL) for the user to further process with specific CAD systems before export to validated manufacturing centers or to the point of care. Manufacturing at the point of care requires a validated process using CAM equipment (additive manufacturing system, including software and associated tooling) and compatible material (biocompatible and sterilizable). The software is not intended to perform automatic diagnosis and to control the manufacture of devices. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| | RealGUIDE® Software Traditional 510(k) - Device Modification 510(k) Summary | Medical Device Class II K253902 |
| --- | --- | --- |
| | Latest revision date: 30/07/2026 | Pag. 1 of 5 |
# 1 Submitter
3DIEMME Srl
Via Risorgimento 9 22063 Cantù (CO)
Italy
Contact person: Alessandro Avis
Role: Quality & Regulatory Supervisor
Phone: +39.031.7073353
Fax: +39.031.710284
Email: avis@3diemme.it
US agent: Ruth Lopez
BIOMET 31
Palm Beach Gardens
4555 Riverside Drive
Palm Beach Gardens, FL 33410, United States
Phone: 561 7766700
Email: 3ipbg-Regulatoryaffairs@Zimvie.Com
# 2 Subject device
Proprietary Name: RealGUIDE
Classification Name: Medical Image Management And Processing System
Common Name: Automated Radiological Image Processing Software
Regulation: 892.2050
Primary Product Code: QIH
Additional Product Code(s): LLZ
# 3 Predicate device
Name of Device: 3Diemme RealGUIDE
Manufacturer: 3Diemme
K Number: K173041
Product Code: LLZ
Reference Device #1:
Name of Device: Relu Creator
Manufacturer: Relu BV
K Number: K233925
Product Code: QIH
3DIEMME S.r.l.
P.IVA 02938110133
www.3diemme.it
support@3diemme.it
Registered Office:
Via Risorgimento,9
22063 Cantù – (CO)
Italia
Operative Headquarters:
Via Delle Baraggette,10
22060 Figino Serenza – (CO)
Italia
Phone:
+39 031 7073353
Page 1 of 5
{5}
| | RealGUIDE® Software Traditional 510(k) - Device Modification 510(k) Summary | Medical Device Class II K253902 |
| --- | --- | --- |
| | Latest revision date: 30/07/2026 | Pag. 2 of 5 |
Reference Device #2:
Name of Device: Atomica Planner
Manufacturer: Atomica Technology, Inc.
K Number: K230012
Product Code: LLZ
#### 4 Device description
RealGUIDE is a stand-alone software that supports all the common 3D medical imaging functionalities used by professional doctors to support their diagnosis. It includes various Volume and IsoSurface rendering, segmentation tools, masking and sculpting, MPR, 2D and 3D measurement and analysis tools. Since 2D imaging is still an important feature, it is possible to switch with a single click to a 2D view, use an even more sophisticated MPR view or switch back to the 3D view.
RealGUIDE is characterized by its intuitive user interface, 2D, MPR and 3D imaging, prime image quality and extensive visualization options, fast image rendering, measurement and analysis tools, and easy integrated reporting. The software is integrating all the surface and volume modelling tools necessary to integrate the diagnostic and virtual planning functions to any CAD/CAM and rapid prototyping system for further processing and manufacturing.
The output format of the software is a STL file, mainly focused on dental surgery. A list of the possible devices that can be modelled with the software is reported below:
- Surgical guides for endosseous dental implants
- Bone graft models for mandible/maxilla regenerative procedures
- Dental and maxillofacial prosthesis
RealGUIDE Software is available for digital download or via web browser. When used on a local computer, the software is installed on the user's machine and run on its hardware. When used via browser, the user accesses the version available for all users, installed on the server.
Each desktop user has to install a version of RealGUIDE while browser users can access the version available on the server through their preferred internet application. A brief example of user access is given in the following diagram.
#### 5 Indications for use
RealGUIDE is a stand-alone software indicated for the preparation of patient specific treatment planning by professional user and, in particular, it must be used by dental clinician, general practitioner (DDS) dentist, or specialist such as, a Periodontist or Oral Sur-geon. The software is indicated for use on patients that require dental implant placement planning, prosthesis modelling and endodontic treatments connected to the endosseous dental implant placement. RealGUIDE is, in further detail, indicated for:
[LOGO]
3DIEMME S.r.l.
P.IVA 02938110133
www.3diemme.it
support@3diemme.it
Registered Office:
Via Risorgimento,9
22063 Cantù – (CO)
Italia
Operative Headquarters:
Via Delle Baraggette,10
22060 Figino Serenza – (CO)
Italia
Phone:
+39 031 7073353
Page 2 of 5
{6}
| | RealGUIDE® Software Traditional 510(k) - Device Modification 510(k) Summary | Medical Device Class II K253902 |
| --- | --- | --- |
| | Latest revision date: 30/07/2026 | Pag. 3 of 5 |
- Viewing in 2D/3D patient images and processing them (DICOM or STL), including volume editing and measurement
- Segmenting bone, nerve canal and teeth
- Virtually positioning dental implants
- Designing templates for endosseous dental implant guided placement
- Modelling reconstructive devices, such as bone grafts and grids
- Designing dental prosthetics, such as crown, bridges and dentures
- Saving dental surgery template's or prostheses' design to file
RealGUIDE software allows for endosseous dental implant surgical guide and prosthetic component file creation (*.STL) for the user to further process with specific CAD systems before export to validated manufacturing centers or to the point of care. Manufacturing at the point of care requires a validated process using CAM equipment (additive manufacturing system, including software and associated tooling) and compatible material (biocompatible and sterilizable).
The software is not intended to perform automatic diagnosis and to control the manufacture of devices.
## 6 Comparison of technological characteristics with the predicate device
The software functions available in RealGUIDE are designed to give the user tools
- to visualize DICOM images for diagnostic purpose. Based on the image's characteristics, the user will base his clinical decisions. In this circumstance, RealGUIDE is only capable of visualizing the images, giving the user the ability to set filers and remove unwanted parts of it from the field.
- To virtually place implants within the DICOM images, to plan the dental guided surgery to be performed in a later stage. Given the fact that RealGUIDE is intended to drive clinical management only, thus aid in diagnosis and treatment, the information it provides is necessary for the next treatment intervention. Implant position and surgery are always the responsibility of the qualified professional carrying out the surgery. RealGUIDE does not provide diagnosis.
- To virtually design the surgical guide for dental guided surgery. The characteristics of the guide are always based on the previous implant planning and on the anatomical features detected with the diagnostic imaging systems.
- To virtually model the prothesis. Crown or dental prosthesis modelling is always based on previous implant planning.
The device under evaluation is an update of the predicate device, with two major characteristics:
1. Creation of a specific version of RealGUIDE WEB to be installed on a server, functionally equivalent to the existing desktop version, and to be accessed by the user from a web browser.
2. Introduction of AI/ML based algorithms to limited functions, currently only available in manual mode.
The AI/ML algorithms introduced are locked and the device cannot learn from the data it is provided with, once it has been released to the market.
[LOGO]
3DIEMME S.r.l.
P.IVA 02938110133
www.3diemme.it
support@3diemme.it
Registered Office:
Via Risorgimento,9
22063 Cantù – (CO)
Italia
Operative Headquarters:
Via Delle Baraggette,10
22060 Figino Serenza – (CO)
Italia
Phone:
+39 031 7073353
Page 3 of 5
{7}
| | RealGUIDE® Software Traditional 510(k) - Device Modification 510(k) Summary | Medical Device Class II K253902 |
| --- | --- | --- |
| | Latest revision date: 30/07/2026 | Pag. 4 of 5 |
A comparison with the predicate device and reference devices shows that the technological characteristics of the devices are similar.
| Feature | RealGUIDE | 3Diemme RealGUIDE | Relu Creator | Atomica Planner |
| --- | --- | --- | --- | --- |
| 510(k) number | - | K173041 | K233925 | K230012 |
| DICOM 2D/3D reconstruction | Yes | Yes | Yes | Yes |
| Panoramic image reconstruction | Yes | Yes | No | |
| Segmentation of anatomy and dentures (nerve canals, bone, crown, teeth) | Yes | Yes | Yes | Yes |
| Import and matching STL files | Yes | Yes | Yes | Yes |
| Implant planning from library | Yes | Yes | No | Yes |
| Teeth setup | Yes | Yes | No | Yes |
| Surgical guides design | Yes | Yes | No | Yes |
| Model design | Yes | Yes | Yes | Yes |
| Prosthesis design | Yes | Yes | No | No |
| Freeform modelling & design | Yes | Yes | No | Yes |
| Virtual articulator | Yes | Yes | No | No |
| AI/ML enabled functionality | Yes | No | Yes | Yes |
| PC version | Yes | Yes | No | Yes |
| Mac version | Yes | Yes | No | No |
| Mobile version | Yes | Yes | No | No |
| WEB version | Yes | No | Yes | No |
Overall, the technological characteristics of the subject device are the same of the predicate and reference devices. The new features of the subject device are equivalent to the ones already available on reference devices such as Relu (WEB version) and Atomica (AI/ML functions).
## 7 Performance data
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- Software verification & validation results
- AI/ML validation
[LOGO]
3DIEMME S.r.l.
P.IVA 02938110133
www.3diemme.it
support@3diemme.it
Registered Office:
Via Risorgimento,9
22063 Cantù – (CO)
Italia
Operative Headquarters:
Via Delle Baraggette,10
22060 Figino Serenza – (CO)
Italia
Phone:
+39 031 7073353
Page 4 of 5
{8}
| | RealGUIDE® Software Traditional 510(k) - Device Modification 510(k) Summary | Medical Device Class II K253902 |
| --- | --- | --- |
| | Latest revision date: 30/07/2026 | Pag. 5 of 5 |
The software under evaluation is the subsequent update of the previously marketed version of RealGUIDE (K152078), being considered as predicate device.
Testing demonstrates the implemented functions perform as intended, and differences between the Device and the predicates do not raise additional concerns with the Device's safety and effectiveness.
## 8 Conclusion
For the purpose of demonstrating the substantial equivalence of RealGUIDE (subject device) to already cleared and marketed devices, the following aspects have been considered.
- Indications for use: no substantial difference in intended user, medical application and patient population.
- Principle of operation: to perform image processing, virtual dental implant positioning, surgical guide design and prosthetic modelling the software works on a proprietary algorithm using commonly available libraries and self-developed software code. With no changes to the software architecture to implement the new function, the subject device now has two different ways (manual and automatic) of performing a function that was already available. Thus, no substantial change to the operating principle has been made
- Software features: the main functions of the subject device, with respect to the predicate devices have remained unchanged. The addition of the web based application to access the server installation of RealGUIDE does not represent a change affecting safety and effectiveness of the device. All AI/ML based functions were already available in RealGUIDE, to be performed only manually.
- Primary product Code (QIH): the change is administrative, considering the Product Code is belonging to the same regulations. No change in risk classification. The difference in product code reflect the presence of AI/ML enabled functions that have been introduced to the Radiological Image Processing System (LLZ) previously cleared, and considered as Predicate Device.
Based on the comparison of intended use, indications, principle of operations, features, technical data and the test results, the RealGUIDE software is thought to be substantially equivalent in safety and effectiveness to the predicate and reference devices listed.
[LOGO]
3DIEMME S.r.l.
P.IVA 02938110133
www.3diemme.it
support@3diemme.it
Registered Office:
Via Risorgimento,9
22063 Cantù – (CO)
Italia
Operative Headquarters:
Via Delle Baraggette,10
22060 Figino Serenza – (CO)
Italia
Phone:
+39 031 7073353
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.