GFS - Sterile Screws and Top Hat

K253858 · T.A.G. Medical Products Corporation, Ltd. · HWC · Aug 5, 2026 · Orthopedic

Device Facts

Record IDK253858
Device NameGFS - Sterile Screws and Top Hat
ApplicantT.A.G. Medical Products Corporation, Ltd.
Product CodeHWC · Orthopedic
Decision DateAug 5, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GFS Sterile Screws and Top Hat are intended to provide orthopedic surgeon a means of bone fixation and to assist in the management of fractures and reconstructive surgeries for cortical bone in the shoulder.

Device Story

Implantable metallic fixation screws and threaded bushing (Top Hat) for shoulder cortical bone; used for fixation of bone grafts or fragments. Operated by orthopedic surgeons in clinical settings. Top Hat attaches to proximal end of screw to ensure firm, safe positioning during surgery. Provided sterile for single use. Output is physical bone fixation; assists surgeon in stabilizing fractures or reconstructive sites. Benefits patient through secure bone fragment/graft stabilization.

Clinical Evidence

Bench testing only. Dimensional analysis and mechanical testing (axial pullout force and insertion torque) performed per ASTM F543 to compare subject device against predicate.

Technological Characteristics

Metallic bone fixation fasteners; includes screws (20-48 mm) and threaded bushing (Top Hat). Identical materials to predicate. Mechanical testing performed per ASTM F543. Sterile, single-use.

Indications for Use

Indicated for orthopedic surgeons for bone fixation and management of fractures and reconstructive surgeries of cortical bone in the shoulder.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} ![LOGO](U.S. Food & Drug Administration) August 5, 2026 T.A.G. Medical Products Corporation, Ltd. Shlomi Dines VP QA and RA Kibbutz Gaaton Gaaton Northern, IL 2513000 Israel Re: K253858 Trade/Device Name: GFS - Sterile Screws and Top Hat Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: July 7, 2026 Received: July 7, 2026 Dear Shlomi Dines: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253858 - Shlomi Dines Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253858 - Shlomi Dines Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253858 | ? | | Please provide the device trade name(s). | | ? | | GFS - Sterile Screws and Top Hat | | | | Please provide your Indications for Use below. | | ? | | The GFS Sterile Screws and Top Hat are intended to provide orthopedic surgeon a means of bone fixation and to assist in the management of fractures and reconstructive surgeries for cortical bone in the shoulder. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | Page 1 of 1 {4} # 510(k) Summary Pursuant to CFR 807.92, the following 510(k) Summary is provided: | Date | August 5, 2026 | | --- | --- | | Submitter Address | T.A.G. Medical Products Corporation, Ltd. Gaaton 2513000, ISRAEL www.tag-med.com | | Mfg. Phone | Tel.: 972-4-9858400 | | Contact Person | Shlomi Dines | | Trade Name | GFS Sterile Screws and Top Hat | | Device | Smooth or threaded metallic bone fixation fastener | | Classification | Class II | | Product Code | HWC | | Regulation Number | 21 CFR 888.3040 | | Predicate Device | K110763 – Latarjet Cortical Screw Set | | Reference Devices | K083096 – Latarjet Cortical Screw Set K091694 – Latarjet Cortical Screw Set K101616 – GrappLR™ and GrappLR™ Extender | | Description: | The GFS Sterile Screws and Top Hat Set consist of implantable fixation screws which are used for fixation of bone grafts or bone fragments to cortical bone in the shoulder. The screws are offered in the following size ranges 20 mm -48 mm in length. There is also a threaded bushing, “Top Hat”, included. The top hat is attached to the proximal end of the fixation screw. The purpose is to position the screw firmly and safely during the procedure. The top hat or bushing is provided in one size. The proposed screws are provided sterile single use. | | Indications for Use: | The GFS Sterile Screws and Top Hat are intended to provide orthopedic surgeon a means of bone fixation and to assist in the management of fractures and reconstructive surgeries for cortical bone in the shoulder. | | Performance Data | Dimensional analysis was compared between the subject and predicate devices. Due to differences in thread geometry, axial pullout force and insertion torque testing was completed in comparison to the predicate per ASTM F543. The performed testing demonstrates the proposed GFS Sterile Screws are substantially equivalent to the predicate Latarjet Cortical Screw Set, K110763. | | Comparison of Technological Characteristics: | In comparison to the predicate device, K110763, Latarjet Cortical Screw Set devices, the subject GFS Sterile Screws and Top Hat are identical in materials, surgical technique, and methods of manufacturing as the predicate. The proposed device, GFS Sterile Screws includes a change in dimensions, sterilization modality, shelf life, and packaging configuration. | K253858 - Page 1 of 2 {5} ## Conclusion The subject GFS Sterile Screws and Top Hat are substantially equivalent to the predicate device which the design features and intended use are identical. Differences between the subject and predicate device do not result in new or different questions of safety or effectiveness. K253858 - Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...