The STEDI Scan Device is a bioimpedance spectroscopy device indicated to measure impedance ratios to aid in the assessment of cerebral fluid volume asymmetry between the cerebral hemispheres in adult patients undergoing neurological assessment. The STEDI Scan Device is an adjunct to standard methodologies for clinical evaluation.
Device Story
Device uses RF coil-based sensors to non-invasively monitor bioimpedance on left and right sides of patient head; operates via multi-frequency bioimpedance spectroscopy. Software processes signal data to calculate Bioimpedance Asymmetry Score (BAS). Used in clinical settings by healthcare providers as adjunct to standard neurological evaluation. Output provides clinician with quantitative BAS reflecting cerebral fluid-volume status; assists in clinical decision-making regarding fluid-volume changes. Device is mains-powered; consists of sensor and system controller.
Clinical Evidence
No formal clinical performance testing required. Bench testing included a healthy volunteer study (n=79) to establish normative bioimpedance measurements and variability. Additional bench testing utilized a head phantom with variable saline/tissue mixtures to validate the device's ability to detect and quantify bioimpedance asymmetry, showing high correlation between BAS and measured conductivity.
Technological Characteristics
Multi-frequency bioimpedance spectroscopy; 0.5-30MHz frequency range; RF coil-based sensor; mains-powered with medical-grade power supply. Complies with IEC 60601-1, IEC 60601-1-2, and IEC TS 60601-4-2. Software-based signal processing.
Indications for Use
Indicated for adult patients undergoing neurological assessment to aid in the assessment of cerebral fluid volume asymmetry between cerebral hemispheres.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
July 22, 2026
Stroke DX
Richard Lilly
Consultant
300 Corporate Pointe
Suite 310
Culver City, California 90230
Re: K253840
Trade/Device Name: STEDI Scan Device
Regulation Number: 21 CFR 870.2770
Regulation Name: Impedance Plethysmograph
Regulatory Class: Class II
Product Code: QAF
Dated: June 22, 2026
Received: June 22, 2026
Dear Richard Lilly:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K253840 - Richard Lilly
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance)
{3}
K253840 - Richard Lilly
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Patrick Antkowiak -S
for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253840 | ? |
| Please provide the device trade name(s). | | ? |
| STEDI Scan Device | | |
| Please provide your Indications for Use below. | | ? |
| The STEDI Scan Device is a bioimpedance spectroscopy device indicated to measure impedance ratios to aid in the assessment of cerebral fluid volume asymmetry between the cerebral hemispheres in adult patients undergoing neurological assessment | | |
| The STEDI Scan Device is an adjunct to standard methodologies for clinical evaluation | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
STEDI Scan Device
Page 8 of 43
{5}
NeuraNova
Traditional 510(k) Notification
STEDI Scan Device
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.87 and 21 CFR § 807.92.
510 (k) number: K253840
# I. SUBMITTER
Name: NeuraNova Inc.
Address: 30 Corporate Pointe
Unit 310
Culver City, CA 90230
Phone: 408-307-3759
Email: rick@Neura-Nova.com
Contact Person: Richard Lilly
Date Prepared: 19 JUN 2026
# II. DEVICE INFORMATION
| Name of Device: | STEDI Scan Device |
| --- | --- |
| Common or Usual Name: | N/A |
| Regulation Name: | Impedance Plethysmography |
| Regulation Number: | 21CFR § 870.2770 |
| Classification Panel: | Neurology |
| Regulatory Class: | Class II |
| Product Code: | QAF |
{6}
NeuraNova
Traditional 510(k) Notification
STEDI Scan Device
### III. PREDICATE DEVICE
| Name | Manufacturer | 510(k)# |
| --- | --- | --- |
| Visor System | Cerebrotech Medical Systems | K182967 |
This Predicate has not been subject to a design-related recall. No reference devices were used in this submission.
### IV. DEVICE DESCRIPTION
The STEDI Scan Device is a bioimpedance spectroscopy device used to monitor patients who are predisposed to cerebral fluid-volume changes. The Device utilizes sensor technology to non-invasively assess bioimpedance on the left and right sides of a patient's head by monitoring bioimpedance changes from biological tissue in response to an RF coil. The Device's software processes this signal data and provides a single value to the treating clinician that reflects the patient's cerebral fluid-volume status in the form of a Bioimpedance Asymmetry Score (BAS).
### V. INDICATIONS FOR USE
The STEDI Scan Device is a bioimpedance spectroscopy device indicated to measure impedance ratios to aid in the assessment of cerebral fluid volume asymmetry between the cerebral hemispheres in adult patients undergoing neurological assessment.
The STEDI Scan Device is an adjunct to standard methodologies for clinical evaluation.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
#### A. INTENDED USE
| | Predicate Visor System | Subject Device STEDI Scan Device | Comment |
| --- | --- | --- | --- |
| Indication for Use | The Visor System is a bioimpedance spectroscopy device indicated to measure impedance ratios to aid in the assessment of cerebral fluid volume asymmetry between the cerebral hemispheres in adult patients undergoing neurologic assessment. | The STEDI Scan Device is a bioimpedance spectroscopy device indicated to measure impedance ratios to aid in the assessment of cerebral fluid volume asymmetry between the cerebral hemispheres in adult patients undergoing neurological assessment. | Same as Predicate |
{7}
NeuraNova
Traditional 510(k) Notification
STEDI Scan Device
| | Predicate Visor System | Subject Device STEDI Scan Device | Comment |
| --- | --- | --- | --- |
| | The Visor System is an adjunct to standard methodologies for clinical evaluation. | The STEDI Scan Device is an adjunct to standard methodologies for clinical evaluation. | |
| 21 CFR Section | 21CFR § 870.2770 | 21CFR § 870.2770 | Same as Predicate |
| Product Code | QAF | QAF | Same as Predicate |
| Target Population | Adult | Adult | Same as Predicate |
# B. TECHNOLOGICAL CHARACTERISTICS
| Attribute | Predicate Visor System | Subject Device STEDI Scan | Comment |
| --- | --- | --- | --- |
| Monitoring Mode | Periodic. Baseline, then repeated as required | Periodic. Baseline, then repeated as required | Same as Predicate |
| Operating Principle | Multi-Frequency Bioimpedance Spectroscopy | Multi-Frequency Bioimpedance Spectroscopy | Same as Predicate |
| Software | Yes | Yes | Same as Predicate |
| Type of Energy | Low-Power Radio Frequency Current | Low-Power Radio Frequency Current | Same as Predicate |
| Frequency Range | 30-310 MHz | 0.5-30MHz | Modified See Note 1. |
| Displayed Parameter | Bioimpedance ratio | Bioimpedance ratio | Same as Predicate |
| Device Configuration | Non-invasive, antenna modules, and system controller | Non-invasive, RF coil-based sensor and system controller | Substantially Equivalent |
{8}
NeuraNova
Traditional 510(k) Notification
STEDI Scan Device
| Attribute | Predicate Visor System | Subject Device STEDI Scan | Comment |
| --- | --- | --- | --- |
| Physical Dimensions | Weight: 1.1 lbs Length: 9.0" Width: 9.0" Height: 2.8" | Weight: 18.6 lbs Length: 19.8" Width: 16.8" Height: 15.1" | Substantially Equivalent. Changes in size and weight do not raise new safety or efficacy concerns. |
| Power Source | Rechargeable Battery Removable Li-ion batteries | Mains powered with a medical-grade power supply | Substantially Equivalent Changes in the power delivered to the device do not raise new safety or effectiveness concerns. |
| Performance | Device detects and monitors bioimpedance ratios | Device detects and monitors bioimpedance ratios | Same as Predicate |
| Voltage | 100 – 240 | 100 – 240 | Same as Predicate |
| Frequency | 50 – 60 Hz | 50 – 60 Hz | Same as Predicate |
| Current | 2.1 amp Power Adapter (for charging the Tablet) 1 amp for the Scanner Battery Charger | 2.71 amp for Power Adapter 1.08 amp for Device | Substantially Equivalent Both systems use medical-grade power supplies to provide power to their components. |
NOTE 1: Frequency Range: The STEDI Scan Device operates at the same or lower frequency range than the predicate device. Lower frequencies generally carry less energy and are considered appropriate for use in biological tissue. Performance data indicate that this change does not adversely affect sensor performance. No new safety or effectiveness questions are raised by this difference.
# VII. PERFORMANCE DATA:
ELECTRICAL, THERMAL, AND MECHANICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY TESTING (EMC):
{9}
**NeuraNova**
Traditional 510(k) Notification^{}[] STEDI Scan Device
Electrical, thermal, mechanical safety, and EMC testing were conducted on the STEDI Scan Device, consisting of the STEDI Device, STEDI Controller, and a medical-grade power supply. The system complies with IEC 60601-1:2005 + AMD1:2012 + AMD2:2020, IEC 60601-1-2:2014 + AMD1:2020, and IEC TS 60601-4-2:2024.
### **SOFTWARE VERIFICATION AND VALIDATION TESTING:**
Software verification and validation testing were conducted, and documentation was provided as recommended by the FDA guidance document, “Content of Premarket Submissions for Device Software Functions,” issued on June 14, 2023. The software documentation level for this Device was considered Basic because a failure or flaw of any device software function would not present a hazardous situation with a probable risk of death or serious injury.
### **SUMMARY OF PERFORMANCE TESTING:**
| Study | Description |
| --- | --- |
| Normative Human Bioimpedance Measurements | 79 healthy adult volunteers established normative human bioimpedance and person-to-person variability. |
| Detection of bioimpedance asymmetry | A human head phantom with a septum dividing the right and left halves was used to measure bioimpedance asymmetry between the two hemispheres. |
| Mathematical model | A mathematical model validated the accuracy of the bioimpedance asymmetry readings. |
| Measurement Stability (Human) | Demonstrated day-to-day measurement stability in human subjects. |
| Conductivity Detection Test | Demonstrated the ability of the STEDI Device to detect different levels of electrical conductivity, and thus bioimpedance, in a test target with correlation to calibrated laboratory measurement tools. |
### **CLINICAL TESTING:**
Formal clinical performance testing was not required to support this medical Device, as the indication for use and technology are equivalent to those of the predicate device. A healthy volunteer study was conducted in a clinical setting using the STEDI Scan
{10}
NeuraNova
Traditional 510(k) Notification
STEDI Scan Device
Device to generate normative human bioimpedance measurements for the non-clinical performance testing section. The results from this study are presented below.
# Results – all Scans
| Parameter | Value |
| --- | --- |
| Subject | 79 |
| Trials | 267 |
| Mean ± Standard Deviation | 1.025 ± 0.047 |
| 90% Confidence interval of the Mean | 1.021 – 1.030 |
| 95% confidence interval of the Mean | 1.020 – 1.031 |
# Results – First Scan Only
| Parameter | Value |
| --- | --- |
| Subjects | 79 |
| Mean ± Standard Deviation | 1.024 ± 0.045 |
| 90% Confidence interval of the Mean | 1.016 – 1.033 |
| 95% confidence interval of the Mean | 1.014 – 1.035 |
# Bench Testing
To demonstrate that the STEDI System can detect and quantify bioimpedance asymmetry, a series of bench tests were performed on a head phantom. The head phantom is a plastic cylinder with a dome-shaped top and a diameter derived from the median dimensions of a human head. The phantom contains two separate chambers on the left and right sides and is mounted to the STEDI Device in place of the head cradle.
The chambers are filled with different mixtures of saline and tissue, representative of different amounts of fluid asymmetry in the brain. The right-side chamber remained sealed during the experiment, while the left-side chamber had varying ratios of saline to tissue. For each trial, the electrical conductivity and impedance of the variable mixture were measured, and the STEDI Device performed a scan and reported the BAS, as shown below.
{11}
NeuraNova
Traditional 510(k) Notification
STEDI Scan Device

The BAS was highly correlated with the measured conductivity and bioimpedance of the variable mixture. This demonstrates that the STEDI System can detect and quantify bioimpedance asymmetry.
### VIII. CONCLUSIONS:
The NeuraNova STEDI Scan Device is substantially equivalent to the predicate device with respect to indications for use, technological characteristics, and performance. This 510(k) submission provides supporting details, data, and validation activities demonstrating that the differences between the subject device and the predicate device do not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.