K253807 · Convergence Medical Sciences , Ltd. · MOD · Aug 20, 2026 · Anesthesiology
Device Facts
Record ID
K253807
Device Name
Valence InVent Xtend™ (CVG-4101-VXA)
Applicant
Convergence Medical Sciences , Ltd.
Product Code
MOD · Anesthesiology
Decision Date
Aug 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Valence InVent Xtend™ is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It is a restricted medical device intended for use by qualified physicians or by a registered respiratory therapist under the direction of a qualified physician.
Device Story
Valence InVent Xtend is a single-use, passive breathing circuit accessory for differential lung ventilation using a single pressure-controlled mechanical ventilator. It consists of a manifold housing mechanical PEEP valves with 22mm connectors. In the expiratory limb, the valve adds PEEP to the ventilator baseline; in the inspiratory limb, it subtracts pressure from the ventilator's peak inspiratory pressure (PIP). An auxiliary bypass circuit with a check valve prevents CO2 rebreathing and maintains system pressure sensing. Used in hospital settings by physicians or respiratory therapists, the device allows independent titration of pressure for each lung in patients with asymmetric pulmonary pathology. By enabling differential ventilation without requiring two synchronized ventilators, it helps prevent over-distension of compliant lung tissue and improves regional ventilation distribution. All alarms and primary ventilation settings remain managed at the host ventilator.
Clinical Evidence
Bench testing confirmed ISO 5356-1:2015 connector compliance and ISO 18562:2024 biocompatibility. Cadaveric study (n=1) of asymmetric lung pathology showed statistically significant reduction in static driving pressure (p=0.0078) and improved regional ventilation distribution via Electrical Impedance Tomography. Human factors validation with 17 healthcare professionals demonstrated no uncorrected use errors across all critical tasks.
Technological Characteristics
Biocompatible thermoplastic polycarbonate (ISO 18562). Mechanical spring-loaded PEEP valves for pressure modulation (additive PEEP, subtractive PIP). 22mm connectors (ISO 5356-1:2015). Passive, non-electrical device. Single-use, disposable.
Indications for Use
Indicated for patients 10 years or older with pathological lung conditions or other medical conditions requiring endobronchial intubation, mechanical ventilation, and isolation of one lung from the other (e.g., unilateral chest trauma, asymmetric ARDS). Intended for use with a ventilator in pressure-regulated modes.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 20, 2026
Convergence Medical Sciences, Ltd.
Marvin Cervantes
Director of Quality
4121 23b St. NE
Calgary, AB T2E 7V9
Canada
Re: K253807
Trade/Device Name: Valence InVent Xtend™ (CVG-4101-VXA)
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous Ventilator
Regulatory Class: Class II
Product Code: MOD
Dated: November 26, 2025
Received: November 28, 2025
Dear Marvin Cervantes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253807 - Marvin Cervantes
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253807 - Marvin Cervantes
Page 3
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K253807
Device Name
Valence InVent Xtend
Indications for Use (Describe)
Valence InVent Xtend™ is indicated for patients with pathological lung conditions or other medical conditions that require endobronchial intubation, mechanical ventilation and isolation of one lung from the other (e.g. for unilateral chest trauma, asymmetric ARDS).
Valence InVent Xtend™ is intended to be used with a ventilator in a pressure regulated mode (e.g. Pressure Control mode or equivalent). Valence InVent Xtend™ is intended for patients 10 years or older.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
CONVERGENCE
MEDICAL SCIENCES
# Valence InVent Xtend
## 510(k) Summary
Applicant Name and Address:
Convergence Medical Sciences Ltd.
4121 23b St NE
Calgary, AB, T2E 7V9
Canada
Contact Person:
Marvin Cervantes
Director of Quality
Tel. No.: +1 (780) 909-5827
Fax No.: +1 (780) 492-7876
Email: quality@convergencemed.ca
510(k) Summary Preparation Date:
2026-08-19
Device Name:
Trade Name: Valence InVent Xtend
Common Name(s): Differential Ventilation Circuit Adapter with Adjustable Valve Enclosure
Device Identifier: CVG-4101-VXA
Device Classification:
Classification: Class II
Regulation No.: 21 CFR 868.5895 (Continuous Ventilator)
{5}
Valence InVent Xtend
510(k) Summary
[LOGO]
CONVERGENCE
MEDICAL SCIENCES
# **Intended Use:**
Valence InVent Xtend™ is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It is a restricted medical device intended for use by qualified physicians or by a registered respiratory therapist under the direction of a qualified physician.
# **Indications for Use:**
Valence InVent Xtend™ is indicated for patients with pathological lung conditions or other medical conditions that require endobronchial intubation, mechanical ventilation and isolation of one lung from the other, e.g. for unilateral chest trauma, asymmetric ARDS.
Valence InVent Xtend™ is intended to be used with a ventilator in a pressure regulated mode (e.g. Pressure Control mode or equivalent). Valence InVent Xtend™ is intended for patients 10 years or older.
**Environment of Use:** Hospital
## Device Description & Technical Characteristics
# **Device Description:**
Valence InVent Xtend™ is a single-use, single-patient breathing circuit accessory designed to facilitate independent, differential lung ventilation using a single pressure-controlled mechanical ventilator. The device eliminates the traditional clinical requirement for two separate synchronized mechanical ventilators when managing asymmetric pulmonary pathology.
# **Technical Characteristics:**
The device consists of a custom enclosure manifold (base and cap) that houses standard, commercially available mechanical PEEP valves attached to standard 22mm connectors.
- **Expiratory Pressure Modulation (Additive PEEP):** When positioned in an expiratory circuit limb, the inline PEEP valve spring tension opposes exhaled gas flow against the baseline ventilator PEEP. Total delivered PEEP to the individual lung is additive:
$$\text{PEEP}_{\text{lung}} = \text{PEEP}_{\text{ventilator}} + \text{PEEP}_{\text{valve}}$$
- **Inspiratory Pressure Modulation (Subtractive PIP):** When positioned in an inspiratory circuit limb, the inline PEEP valve creates a controlled pressure drop during inspiration. Net Peak Inspiratory Pressure (PIP) delivered to the individual lung is subtractive:
$$\text{PIP}_{\text{lung}} = \text{PIP}_{\text{ventilator}} - \Delta\text{P}_{\text{valve}}$$
2026-08-19
page 2 of 7
{6}
Valence InVent Xtend
510(k) Summary
CONVERGENCE
MEDICAL SCIENCES
- Unidirectional Flow & Anti-Rebreathing: The integrated spring-loaded PEEP diaphragms act as unidirectional check valves. An auxiliary bypass circuit containing a dedicated check valve prevents CO₂ rebreathing and allows the host ventilator to continuously sense system pressure.
- Materials & Dimensions: Manufactured from biocompatible thermoplastic polycarbonate meeting ISO 18562 standards. Connectors conform to ISO 5356-1:2015 for standard ventilator circuits.
## Substantial Equivalence Comparison
Legally Marketed Device to which Substantial Equivalence is claimed:
| 510(k) No.: | Trade Name | Company |
| --- | --- | --- |
| K051767 | LTV 1000 | Pulmonetic Systems, Inc. |
Summary of Substantial Equivalence:
| | Proposed Device Valence InVent Xtend | Primary Predicate Device LTV 1000 (K051767) | Comparison & Equivalence Rationale |
| --- | --- | --- | --- |
| Intended Use | Valence InVent Xtend™ is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It is a restricted medical device intended for use by qualified physicians or by a registered respiratory therapist under the direction of a qualified physician. | The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. | Substantially Equivalent: The therapeutic purpose is identical, to provide continuous or intermittent ventilatory support for individuals who require mechanical ventilation, and both are restricted medical devices. The subject device states the operator qualifications, which fall within the predicate's qualified, trained personnel under the direction of a physician. |
| Indications for Use | Valence InVent Xtend™ is indicated for patients with pathological lung conditions or other medical conditions that require endobronchial intubation, mechanical ventilation and isolation of one lung from the other, e.g. for unilateral chest trauma, asymmetric ARDS. Valence InVent Xtend™ is intended to be used with a ventilator in a pressure regulated mode (e.g. Pressure Control mode or equivalent). Valence InVent Xtend™ is intended for patients 10 years or older. | Applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.) who require positive pressure ventilation delivered invasively or non-invasively; Assist/Control, SIMV or CPAP modes; and breath types including Volume, Pressure Control and Pressure Support. | Substantially Equivalent: The indications for use fall entirely within the scope of the predicate's cleared indications. The subject device addresses a specific clinical subset of that population, limited to pressure-regulated ventilation in patients 10 years and older. Narrowing an indication does not create a new intended use. |
2026-08-19
page 3 of 7
{7}
Valence InVent Xtend
510(k) Summary
[LOGO]
CONVERGENCE
MEDICAL SCIENCES
| | Proposed Device Valence InVent Xtend | Primary Predicate Device LTV 1000 (K051767) | Comparison & Equivalence Rationale |
| --- | --- | --- | --- |
| Environment of Use | Hospital | Hospital, Home, or Transport | Substantially Equivalent: The subject device claims a subset of the predicate's cleared environments. |
| Ventilation Modes | Used with a ventilator in a pressure regulated mode (e.g. Pressure Control mode or equivalent). The device is passive and does not select, generate or alter the ventilator's mode. | Per the predicate labeling: Control, Assist/Control, SIMV, CPAP, NPPV and Apnea Backup modes, with Volume Control, Pressure Control, Pressure Support and Spontaneous breath types. | Substantially Equivalent: The subject device is used with Pressure Control mode or equivalent, which the predicate already provides and which is a subset of those the predicate is cleared for. |
| Peak Inspiratory Pressure (PIP) adjustment | Up to 20 cmH₂O per lung, set by reading the graduated cap of the in-circuit mechanical PEEP valve. The minimum decrement is determined by the lowest setting of the valve in use. The adjustment is subtractive, reducing the peak inspiratory pressure reaching that lung by the amount set. | Peak inspiratory pressure is set at the ventilator and titrated in Pressure Control. | Substantially Equivalent: Peak inspiratory pressure is titrated to the patient in both the subject device and the predicate device. In the subject device the titration is made in the breathing circuit, using the same mechanical spring PEEP valve technology the predicate uses to adjust pressure in the circuit. The adjustment is subtractive only and cannot present a lung with a pressure higher than the ventilator already delivers. Bench testing at representative settings showed a predictable relationship between the setting and the pressure delivered. |
| Positive End-Expiratory Pressure (PEEP) adjustment | Up to 20 cmH₂O per lung, set by reading the graduated cap of the in-circuit mechanical PEEP valve. The minimum increment is determined by the lowest setting of the valve in use. The adjustment is additive, raising the end expiratory pressure of that lung by the amount set. | In-circuit mechanical PEEP valve, 0 to 20 cmH₂O. | Identical: The adjustment is made by the same means, an in-circuit mechanical spring PEEP valve, to the same 20 cmH₂O maximum. The subject device places one valve in each limb of each lung circuit. |
| Compatible ventilation settings and parameters | Used only in pressure-regulated modes. Peak inspiratory pressure and positive end expiratory pressure are the only parameters the device adjusts, and they are adjusted for each lung. Tidal volume is not a set parameter; the volume each lung receives results from the pressure set for that lung. All other ventilation controls remain on the ventilator. The compatible ventilator settings are specified in the labeling. | All ventilation settings are made at the ventilator. In Pressure Control, the delivered volume follows the set pressure. | Substantially Equivalent: Ventilation is pressure controlled in both, and the delivered volume follows the pressure that is set. Pressure is the only parameter the subject device adjusts, and it adjusts it for each lung. |
2026-08-19
page 4 of 7
{8}
Valence InVent Xtend
510(k) Summary
CONVERGENCE
MEDICAL SCIENCES
| | Proposed DeviceValence InVent Xtend | Primary Predicate DeviceLTV 1000(K051767) | Comparison & Equivalence Rationale |
| --- | --- | --- | --- |
| Alarm requirements | The device is mechanical and requires no electrical power. All alarms are set and managed on the ventilator. The labeling addresses the ventilator high pressure, low pressure and volume alarms, and provides guidance on setting the high pressure and low tidal volume limits. | All alarms are set and managed on the ventilator. The ventilator provides high pressure, low pressure and volume alarms. | Substantially Equivalent: Alarms are set and managed on the ventilator in both, and the subject device adds no alarm. The alarms its labeling addresses are the same ventilator alarms the predicate provides. |
| Conformance to Connector Consensus Standards | Yes (ISO 5356-1:2015) | Yes (ISO 5356-1:2015) | Identical: Standard circuit interface. |
| Gas Path Materials | Biocompatible Thermoplastic (Polycarbonate) | Biocompatible Thermoplastic (Polycarbonate) | Identical: Common medical grade ventilator circuit materials. |
| Cleaning of Circuit/PEEP valve | Disposable | Disposable | Identical: Disposable. |
### Basis for the Predicate:
The LTV 1000 (K051767) is the primary predicate device for the Valence InVent Xtend.
The Valence InVent Xtend is a passive breathing circuit accessory rather than a full-function ventilator. The predicate is appropriate because the subject device delivers its effect entirely through the same physiological parameters the predicate is cleared to titrate, peak inspiratory pressure and positive end expiratory pressure, and does so using the same technology the predicate uses to adjust pressure in the breathing circuit: a mechanical spring PEEP valve. The predicate provides that adjustment by the same means and to the same 20 cmH \( _{2} \) O maximum.
The subject device has the same intended use as the predicate. The differences in technological characteristics do not raise different questions of safety and effectiveness. The subject device adjusts only the two pressures the predicate is cleared to titrate, and it adjusts them within the range of the mechanical valve, up to 20 cmH \( _{2} \) O per lung, subtractively on inspiration and additively at end expiration. Every ventilator control and every alarm remains at the ventilator and is set there by the user. The subject device is indicated for a narrower patient population, in a narrower environment, and for pressure-regulated ventilation only, and each of these is a subset of the predicate's cleared use.
2026-08-19
page 5 of 7
{9}
Valence InVent Xtend
510(k) Summary
[LOGO]
CONVERGENCE
MEDICAL SCIENCES
# Summary of Supporting Performance Evidence
Non-clinical bench testing, pre-clinical cadaveric testing, and human factors evaluations were conducted to evaluate unique safety risks associated with differential circuit branching.
## A. Non-Clinical Bench Performance Testing
- Conical Connector & Mechanical Integrity (ISO 5356-1:2015): Passed the dimensional requirements for standard conical connectors.
- Biocompatibility Evaluation (ISO 18562:2024): Confirmed compliance for breathing gas pathways regarding particulate matter and volatile organic substances.
- Ventilation Independence & Pressure Modulation Validation: Bench testing verified that the device allows individualization of inspiratory and expiratory pressures settings for each lung independently.
## B. Non-Clinical Cadaveric Performance Testing
- A human cadaveric model of asymmetric lung pathology evaluated ventilatory mechanics using single-lumen ventilation versus dual-lumen ventilation with Valence InVent Xtend.
- Driving Pressure Reduction: Dual-lung ventilation using Valence InVent Xtend™ achieved a statistically significant reduction in static driving pressure compared to single-lumen ventilation (p = 0.0078).
- Ventilation Balance: Electrical Impedance Tomography (EIT) demonstrated improved regional right-to-left ventilation distribution, preventing over-distension of compliant lung tissue.
## C. Human Factors & Usability Validation
- Study Cohort: Tested across seventeen (17) participants from the intended user group, healthcare professionals experienced in ventilating patients and managing ventilated patients.
- Results: No uncorrected use error occurred in any participant. All 17 participants completed the critical tasks evaluated and all three scenarios.
# Conclusion
The Valence InVent Xtend™ has the same intended use as the cleared predicate device, and is indicated for a subset of the predicate's cleared patient population, environment of use and ventilation modes. It has the same breathing gas pathway materials and the same standard circuit interface. It acts on the same two parameters the predicate is cleared to titrate, peak inspiratory pressure and positive end expiratory pressure, using the same in-circuit mechanical spring PEEP valve technology the predicate uses to adjust pressure in the breathing circuit.
2026-08-19
page 6 of 7
{10}
Valence InVent Xtend
510(k) Summary
CONVERGENCE
MEDICAL SCIENCES
The differences in technological characteristics do not raise different questions of safety and effectiveness, and the subject device is as safe and as effective as the predicate device. Performance bench testing, pre-clinical cadaveric models, and human factors validation demonstrate that the device performs as intended.
**Therefore, Valence InVent Xtend™ is substantially equivalent to the predicate device.**
2026-08-19
page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.