Knee+ is a stereotaxic system including an intraoperative software as a medical device and surgical instruments. Knee+ is intended for primary Total Knee Replacement, to assist the surgeon in determining reference alignment axes in relation to anatomical landmarks, in order to position the cutting guide regarding computed mechanical axis. The Knee+ includes smart glasses as a Head Mounted Device (HMD) for displaying information to the user intraoperatively. The smart glasses should not be relied upon solely and should always be used in conjunction with traditional methods.
Device Story
Knee+ is a stereotaxic surgical navigation system for primary Total Knee Replacement (TKR). It consists of surgical instruments with markers and software running on a wearable Head Mounted Device (HMD) (smart glasses with integrated camera and external battery). During surgery, the system tracks instrument markers in a 3D reference frame to collect anatomical landmarks. Software algorithms process these coordinates to calculate the mechanical axis and determine the orientation of femoral and tibial cutting planes. The HMD displays this guidance information to the surgeon via a near-eye display. The surgeon uses this real-time feedback to adjust cutting guide orientation and resection levels. The device is used in an operating room environment by surgeons; it is intended to supplement, not replace, traditional surgical methods. The system provides intraoperative guidance to improve the accuracy of bone resections, potentially benefiting the patient through better implant alignment.
Clinical Evidence
Bench testing (ASTM F2554) confirmed repeatability and accuracy. Human factors validation (n=15 surgeons) in a simulated OR environment confirmed usability of the new platform. Software verification/validation followed IEC 62304. Biocompatibility assessed per ISO 10993. A quantitative cadaver study (n=3 surgeons, 9 knees) compared the subject device to the predicate in a simulated setting, demonstrating equivalent performance.
Technological Characteristics
Stereotaxic navigation system; includes surgical instruments and HMD (smart glasses). Tracking via optical markers. Software-based 3D coordinate processing. Complies with ASTM F2554 (accuracy), IEC 62366-1 (usability), IEC 62304 (software), IEC 60601-1/1-2 (electrical/EMC), and ISO 10993 (biocompatibility). Platform consists of smart glasses with integrated camera and external battery.
Indications for Use
Indicated for adults (22+ years) undergoing primary Total Knee Replacement to assist surgeons in determining reference alignment axes and positioning cutting guides relative to the mechanical axis.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 24, 2026
Pixee Medical S.A.S.
François Beaumont
Regulatory Affairs Engineer
18 Rue Alain Savary
Besançon, 25000
France
Re: K253805
Trade/Device Name: Knee+
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: SBF
Dated: March 26, 2026
Received: March 26, 2026
Dear François Beaumont:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253805 - François Beaumont
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{3}
K253805 - François Beaumont
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
Page 1 of 1.
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253805 | ? |
| Please provide the device trade name(s). | | ? |
| Knee+ | | |
| Please provide your Indications for Use below. | | ? |
| Knee+ is a stereotaxic system including an intraoperative software as a medical device and surgical
instruments.
Knee+ is intended for primary Total Knee Replacement, to assist the surgeon in determining reference
alignment axes in relation to anatomical landmarks, in order to position the cutting guide regarding
computed mechanical axis. The Knee+ includes smart glasses as a Head Mounted Device (HMD) for
displaying information to the user intraoperatively. The smart glasses should not be relied upon solely and
should always be used in conjunction with traditional methods. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is
to be used. | ☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☐ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater) | ? |
{5}
pixee
K253805 510(k) SUMMARY
Pixee Medical - KNEE+
# 510(k) Submitter:
Name: Pixee Medical
Address: 18 rue Alain Savary
25000 Besançon
France
Phone: (+33) 4 58 10 13 65
Fax: (+33) 4 58 10 14 51
Date Prepared: March 25th, 2026
# Device:
K253805
Trade name: Knee+
Common name: Surgical Navigation Software and Instruments
Classification name: Orthopedic Augmented Reality (21 CFR §882.4560)
Product code: SBF
Regulatory class: II
Classification Panel: Orthopedic
# Predicate Device:
Knee+ is substantially equivalent to the previous version of Knee+, legally marketed:
| Applicant Name | Device Name | Product code | 510(k) Number |
| --- | --- | --- | --- |
| Pixee Medical | Knee+ | SBF | K243975 |
No reference devices were used in this submission.
# Device Description:
The main purpose of Knee+ is to assist the surgeon during the primary Total Knee Replacement (TKR) intervention. Knee+ includes software and surgical instruments.
Knee+ provides information to help locate and orientate the main femoral and tibial cutting planes as required in knee replacement surgery. Knee+ allows the surgeon to adjust the cutting plane orientation and the resection level. This includes means for the surgeon to collect anatomical references during the TKR intervention using the surgical instruments. The software locates in a 3D reference frame the instruments which include markers. All collected coordinates are treated by
{6}
pixee
K253805 510(k) SUMMARY
Pixee Medical - KNEE+
software algorithms to provide the surgeon with relevant orientation of the tracked cutting guide. Knee+ software is installed on a wearable Head Mounted Device (HMD) which includes a camera and displays intraoperative information to the user. A near-eye display allows the surgeon to look at the HMD screen or the field of view when needed.
## Intended Use / Indications for Use:
Knee+ is a stereotaxic system including an intraoperative software as a medical device and surgical instruments. Knee+ is intended for primary Total Knee Replacement, to assist the surgeon in determining reference alignment axes in relation to anatomical landmarks, in order to position the cutting guide regarding computed mechanical axis. The Knee+ includes smart glasses as a Head Mounted Device (HMD) for displaying information to the user intraoperatively. The smart glasses should not be relied upon solely and should always be used in conjunction with traditional methods.
## Summary of Technological characteristics
The device subject of this Premarket Notification is a modification from the device legally authorized under K243975, also manufactured by Pixee Medical.
There are no differences between the subject device and the predicate with respect to indications and intended use.
The main change from the previous version (K243975) is the addition of a qualified execution platform compatible with the KneePlus software.
The previous qualified execution platform had 4 components while the new one has only two (smart glasses with integrated camera and external battery).
The surgical protocol and the technics to determine and display the pose of the instruments are identical to the predicate device.
The tracking system technology, the use of a Head Mounted Device to provide information and the claimed accuracy for cut orientation and resection level also remain the same as the most recent cleared version (K243975).
The main change as well as other minor changes were assessed through risk management activities, including relevant verification and validation information, produced under design controls procedures. The results of the design controls activities were provided as a summary in the Premarket Notification.
Substantial equivalence was therefore supported by performance data which demonstrated that the modified Knee+ device is still safe and effective for its intended use. The differences from the previous version do not raise any concern regarding the safety and effectiveness of the device.
## Nonclinical Performance Data:
Performance data were necessary to demonstrate that the same performance is still achieved with the use of the new qualified execution platform. The same methods, protocols and acceptance criteria
Page 2 / 3
{7}
k253805
pixee
K253805 510(k) SUMMARY
Pixee Medical - KNEE+
used to support the previously cleared K243975 were applied to evaluate the change, as well as FDA-recognized standards methods:
- Bench testing was conducted in order to demonstrate that Knee+ performs according to its requirements and specifications and use conditions. In particular, repeatability and accuracy were tested according to ASTM F2554.
- User needs validation – The system was validated in accordance with IEC 62366-1 and FDA guidance "Applying Human factors and usability engineering to medical devices". A human factor validation study was conducted with 15 surgeons with different types of experiences performing the complete Knee+ workflow up to completion of the guided bone resections in a representative operating room environment (sawbones, surgeon positioning, wearable platform worn under surgical helmet or without) insuring the coverage of all surgeon's device usages through the entire surgery. The purpose of this evaluation was to determine whether the new execution platform affects human factor validation.
- Software verification and validation testing were conducted as required by IEC 62304 and documentation was provided as recommended by FDA Guidance "Content of Premarket Submissions for Software Contained in Medical Devices".
- EMC and electrical safety testing have been performed on the qualified execution platform in accordance with IEC 60601-1, IEC 60601-1-2, EN 62368-1, the ETSI EN 301 489 series, and other applicable standards.
- Biocompatibility assessment has been performed according to ISO 10993 standard following instrument modification and confirmed that the patient-contacting materials do not introduce new risks and that the Knee Tools instruments were safe and biocompatible.
- Quantitative Cadaver Validation - A study was conducted using three surgeons and involving 9 knees conducted in a clinically simulated setting using cadaver specimens. The objectives of this study was to:
- Observe device utilization in a clinically simulated setting using cadaver specimens,
- Compare performance between the device under evaluation and its predicate device,
- Collect complementary data: observations or recommendations.
In parallel to the procedure performed by the surgeon, an operator wore the predicate device. The data collected of the new augmented reality device was then compared to the predicate.
All performance testing demonstrates that Knee+ performs according to its specifications and functions as intended.
## Conclusion:
Knee+ has the same intended use, indications for use and technological characteristics as its predicate device (i.e., the most recent cleared version K243975). The addition of a qualified execution platform and other minor resulting modifications do not alter the intended surgical use of the device and do not raise new questions of safety and effectiveness. Performance data demonstrated that Knee+ is as safe and effective as its previous version, also manufactured by Pixee Medical. Thus, the Knee+ is substantially equivalent to the legally marketed predicate device.
Page 3 / 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.