UltraPrint-Dental Hard Splint UV

K253798 · Guangzhou Heygears IMC., Inc. · MQC · Mar 2, 2026 · DE

Device Facts

Record IDK253798
Device NameUltraPrint-Dental Hard Splint UV
ApplicantGuangzhou Heygears IMC., Inc.
Product CodeMQC · DE
Decision DateMar 2, 2026
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The UltraPrint-Dental Hard Splint UV device is indicated for the fabrication of orthodontic and dental appliances such as mouthguards, nightguards, splints, repositioners and retainers.

Device Story

UltraPrint-Dental Hard Splint UV is a light-curing, biocompatible acrylate-based resin for 3D printing dental appliances. Used with DLP printers (385 nm UV-LED), the resin undergoes photo-initiated polymerization and cross-linking to form solid, customized orthodontic/dental devices (mouthguards, nightguards, splints, retainers). Operated by dental professionals in clinical or laboratory settings, the material is processed via additive manufacturing. The healthcare provider uses the final printed appliance to provide therapeutic support or protection for the patient's dentition. The device benefits patients by enabling the rapid, precise fabrication of patient-specific dental appliances that meet mechanical and biocompatibility standards.

Clinical Evidence

No clinical data. Evidence consists of bench testing including physical/mechanical performance (flexural strength, modulus, water sorption, solubility, fracture work per ISO 20795-2 and ASTM D790/D2240), biocompatibility (ISO 10993/ISO 7405), and additive manufacturing process validation. Accelerated aging (ASTM F1980) and unsealing lifetime verification confirmed stability and performance over a 2-year shelf life.

Technological Characteristics

Light-curing acrylate-based resin; 3D printing (DLP) via 385 nm UV-LED. Mechanical properties: flexural strength ≥50MPa, flexural modulus ≥1500MPa. Compliant with ISO 20795-2 (mechanical/physical), ISO 10993-1 (biocompatibility), and AAMI ST98:2022 (cleaning).

Indications for Use

Indicated for the fabrication of orthodontic and dental appliances including mouthguards, nightguards, splints, repositioners, and retainers for patients requiring such dental devices.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD &amp; DRUG ADMINISTRATION March 2, 2026 Guangzhou Heygears IMC., Inc. Meifang Liu Regulatory Affairs Block B2, 501,601, Enterprise Accelerator, Kaifa District Guangzhou, Guangdong 510700 CHINA Re: K253798 Trade/Device Name: UltraPrint-Dental Hard Splint UV Regulatory Class: Unclassified Product Code: MQC, KMY Dated: January 30, 2025 Received: January 30, 2025 Dear Meifang Liu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K253798 - Meifang Liu Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K253798 - Meifang Liu Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253798 | | | Device Name UltraPrint-Dental Hard Splint UV | | | Indications for Use (Describe) The UltraPrint-Dental Hard Splint UV device is indicated for the fabrication of orthodontic and dental appliances such as mouthguards, nightguards, splints, repositioners and retainers. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {5} 510(k) summary Keygears UltraPrint-Dental Hard Splint UV K253798 Page 1 of 4 # 510K summary ## I. SUBMITTER GUANGZHOU HEYGEARS IMC.INC Block B2, 501, 601, Enterprise Accelerator, Kaifa District, Guangzhou, Guangdong, 510700, China Phone: +86-020-28186347 Contact Person: Meifang Liu Date Prepared: February 28, 2026 ## II. DEVICE Name of Device: UltraPrint-Dental Hard Splint UV Trade name of Device: UltraPrint-Dental Hard Splint UV; UltraPrint Hard Splint UV; Hard Splint UV; Common or Usual Name: Dental Resin for 3D printing, Mouthguard Classification Name: Mouthguard, Unclassified; Positioner, 872.5525 Regulatory Class: Unclassified Product Code: MQC ## III. PREDICATE DEVICE KeyPrint@ KeySplint Hard K203000 No reference devices were used in this submission ## IV. Device Description UltraPrint-Dental Hard Splint UV is a light-curing resin based on acrylate, biocompatible, for the generative fabrication of splint for DLP printers with UV-LED 385 nm. The material is a polymer undergoes polymerization and cross-linking reaction under the action of photo-initiation when exposed to UV light, which makes the resin material polymerize to a solid state at high speed. Key property as shown below. | Ultimate flexural strength | ≥50MPa | | --- | --- | | Flexural elastic modulus | ≥1500MPa | ## V. Indications for Use The UltraPrint-Dental Hard Splint UV device is indicated for the fabrication of orthodontic and dental appliances such as mouthguards, nightguards, splints, repositioners and retainers. ## VI. Comparison of technological characteristics with predicate device UltraPrint-Dental Hard Splint UV is similar to predicate device in raw material and processing. Both resins are manufactured using very similar components and are {6} 510(k) summary Heygears UltraPrint-Dental Hard Splint UV processed into the final device using 3D printing and light curable polymerization. They share same indication for use, same material composition(UV curable resin,photo Initiator and pigments ),same DLP printing technology, same compliance with ISO 20795-2 regarding chemical,physical and mechanical performance, and both compliant with ISO10993-1 in terms of biological safety. A difference between UltraPrint-Dental Hard Splint UV and predicate device KeyPrint® KeySplint Hard lies in the specific formula/chemical compositions. The formulation of UltraPrint-Dental Hard Splint UV was developed based on FDA-marketed dental resin/monomer and the ratio of those composition is tailored to achieve both ISO20795 compliance and clinical needs.The biological concern for potential difference in formula/chemical composition is addressed and further biological evaluation(chemical characterization, toxicological evaluation, and biological test). The result shows that UltraPrint-Dental Hard Splint UV is biocompatible for its intended use. # VII. Non-Clinical DATA # Physical /Mechanical performance The physical properties of the subject device were determined and tested according to ISO 20795-2:2013, ASTM standard and internal inspection standard, and the test results demonstrated the qualification and substantial equivalence when compared to the predicate device. | Physical / Mechanical Property | Standard | Acceptance Criteria | If the test met the acceptance criteria? | | --- | --- | --- | --- | | Ultimate Flexural Strength | ISO 20795-2 | >50MPa | Yes | | Flexural Modulus | | >1500MPa | Yes | | Water Sorption | | <32ug/mm3 | Yes | | Solubility | | <5ug/mm3 | Yes | | K factor | | >1.1 MPa m1/2 | Yes | | Fracture work | | >250 J/m2 | Yes | | Free Monomer Extraction | | <2.2% | Yes | | Flexural Strength | ASTM D790 | >50MPa | Yes | | Flexural Modulus | | >1500MPa | Yes | | Shore D Hardness | ASTM D2240 | >80D | Yes | | IZOD Impact(Notched) | ASTM D256 | N/A | N/A | | Color | N/A | N/A | N/A | | Density | N/A | N/A | N/A | # Additive Manufacturing Process validation According to FDA's guidance Technical Considerations for Additive Manufactured Medical Devices(issued on December 5, 2017), studies were performed to demonstrate repeatability, physical properties, build orientation, reuse&amp;recycling, and validation of compatible system. Results show that under various conditions as challenged, the final device as manufactured meet the performance requirements as specified in ISO 20795-2:2013. {7} 510(k) summary Heygears UltraPrint-Dental Hard Splint UV Page 3 of 4 # Biocompatibility The compositions of the subject device do not contain any non-conventional chemicals compared to the legally marketed predicate and reference devices. Biocompatibility tests were performed fully following the ISO 7405 and ISO 10993 standards, including Cytotoxicity, Irritation, Sensitization, Chemical Characterization and Toxicological assessment and results are sufficient to prove the subject device is substantially equivalent to the predicate device. The device met the acceptance criteria of the ISO 7405 and ISO 10993 standards. No clinical testing were performed for UltraPrint-Dental Hard Splint UV. # Reprocessing The cleaning method is validated according to AAMI ST98:2022 (Cleaning validation of health care products-Requirements for development and validation of a cleaning process for medical devices). # Shelf life The shelf life of UltraPrint-Dental Hard Splint UV is 2 years. # Accelerated /Real aging Following ASTM F1980 - 21 (Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices) and ISO 20795-2:2013(Dentistry — Base polymers Part 2: Orthodontic base polymers), accelerated aging test were conducted, which involves inspection in the homogeneity, surface characteristics, polishability, colour, freedom from porosity, ultimate flexural strength, flexural modulus, maximum stress intensity factor, total fracture work, residual methyl methacrylate monomer, water absorption, water solubility and packaging integrity. The result shows that after 2 years and 3 months accelerated aging, the device meet ISO20795-2:2013 and the packaging integrity remain its conformance with packaging specification. Real-time aging is in process, so far no nonconformity was detected. # Unsealing lifetime verification Heygears has conducted 45-day unseal aging testing and confirmed its conformity with ISO 20795-2 requirement. It is verified that the material stability of UltraPrint-Dental Hard Splint UV in unsealed bottles can be maintained for at least 30 days. # VIII. CONCLUSIONS The intended use, technical characteristic, are substantially equivalent to the predicate device. The stability of resin/final finished device, and the additive manufacturing method/cleaning method as recommended in IFU are verified according to FDA recognized standards/guidance. While the formula of subjective resin is different from the Predicate, both are light curable acrylate resins used in additive manufacturing and are from the same material category, commonly used for fabricating customized dental splints. The additive manufacturing processes both use the light curable resins, validated printers and curing {8} 510(k) summary Heygears UltraPrint-Dental Hard Splint UV units. Biocompatibility evaluation according to FDA recognized standard/guidance demonstrate positive result. The differences, in comparison to the Predicate device, raise no new questions of safety and effectiveness. Based on the physical performance and biocompatibility test results of the UltraPrint-Dental Hard Splint UV material, it has been confirmed that this device meets all the necessary standards and prove the subjective device is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...