Retrospective clinical imaging data was used to validate the performance of the deep learning algorithm for prostate segmentation and volumetric quantification, comparing device output against expert radiologist consensus.
Retrospective clinical data; Multi-center validation; AI algorithm validation; Prostate MRI
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective Performance Validation; Retrospective observational study
Adult males aged 22 or over undergoing prostate MRI; Sample Size: 170 patients (70 internal, 100 external); Number of Sites: Multi-site (US and UK)
Prostate Core™ is a software medical device intended for automatic segmentation and volumetric quantification of the prostate. Prostate Core™ is intended to be used by trained healthcare professionals as an aid to reading/assessing prostate MRI images of adult males aged 22 or over. Prostate Core™ is not intended as a replacement for clinical review or of the judgement of healthcare professionals. The software does not modify the original medical images and does not provide a direct diagnosis. Prostate Core™ is not intended for patients: • who received a prostate biopsy within 6 weeks before the MRI examination • with prior prostate or urethral treatment or surgery that substantially affects the prostate gland • whose MRI scan quality is severely impaired, for example by patient motion, breathing, peristaltic motion, rectal gas, or metallic implants such as hip prostheses • where a radiologist considers that the MRI examination is not of diagnostic quality • with MRI examinations that substantially deviate from PI-RADS®v2.1 scanning protocols • whose MRI images have been captured using an endo-rectal coil.
Device Story
Software-only device for automated prostate segmentation and volumetric quantification from DICOM-formatted MRI (1.5T/3T). Uses locked deep learning (CNN) to process T2-weighted images; generates segmentation masks and volume measurements (ml). Deployed on-premises on hospital servers; integrates with PACS via DICOM networking. No GUI; outputs exported as DICOM-SEG and RTSTRUCT files. Used by radiologists/trained professionals as adjunctive tool; does not perform CADe/CADx, score lesions, or modify original images. Benefits include standardized, reproducible volumetric data to assist clinical assessment.
Clinical Evidence
No prospective clinical study. Retrospective observational study using 170 independent cases (70 internal, 100 external). Ground truth: consensus annotations from three expert radiologists. Metrics: Dice Similarity Coefficient (DSC), Mean Surface Distance (MSD), Relative Absolute Volume Error (RAVE). External validation results: Median DSC 0.93 (95% CI: 0.92–0.93), Median MSD 0.27 mm (95% CI: 0.25–0.28), Median RAVE 3.99% (95% CI: 3.17–4.80). All predefined acceptance criteria met.
Technological Characteristics
Software-only; Python 3/PyTorch-based pipeline. Deep learning (CNN) algorithm for voxel-level segmentation. Inputs: DICOM MRI (T2-weighted, PI-RADS v2.1). Outputs: DICOM-SEG, RTSTRUCT, structured reports. On-premises deployment; networked via DICOM/TLS. Locked algorithm; no adaptive learning. Compliant with IEC 62304 and ISO 14971.
Indications for Use
Indicated for adult males aged 22+ undergoing prostate MRI. Contraindicated for patients with recent biopsy (<6 weeks), prior prostate/urethral surgery, severe MRI artifacts (motion, implants), non-diagnostic quality scans, non-PI-RADS v2.1 compliant protocols, or endo-rectal coil usage.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 16, 2026
Lucida Medical, Ltd.
Toby Davis
VP of Engineering
Future Business Centre
King's Hedges Rd.
Cambridge, CB4 2HY
United Kingdom
Re: K253794
Trade/Device Name: Prostate Core™
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: June 15, 2026
Received: June 15, 2026
Dear Toby Davis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253794 - Toby Davis
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253794 - Toby Davis
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NINGZHI LI -S Digitally signed by NINGZHI LI -S
for
Daniel M. Krainak, Ph.D
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253794 | ? |
| Please provide the device trade name(s). | | ? |
| Prostate Core™ | | |
| Please provide your Indications for Use below. | | ? |
| Prostate Core™ is a software medical device intended for automatic segmentation and volumetric quantification of the prostate. Prostate Core™ is intended to be used by trained healthcare professionals as an aid to reading/assessing prostate MRI images of adult males aged 22 or over. Prostate Core™ is not intended as a replacement for clinical review or of the judgement of healthcare professionals. The software does not modify the original medical images and does not provide a direct diagnosis. | | |
| Prostate Core™ is not intended for patients: • who received a prostate biopsy within 6 weeks before the MRI examination • with prior prostate or urethral treatment or surgery that substantially affects the prostate gland • whose MRI scan quality is severely impaired, for example by patient motion, breathing, peristaltic motion, rectal gas, or metallic implants such as hip prostheses • where a radiologist considers that the MRI examination is not of diagnostic quality • with MRI examinations that substantially deviate from PI-RADS®v2.1 scanning protocols • whose MRI images have been captured using an endo-rectal coil. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
LUCIDA Medical
# 510(K) Summary
K253794
Date of summary: 5 \( ^{th} \) August 2026
Submitter's name: Lucida Medical Ltd
Submitter's address: Lucida Medical Ltd, Future Business Centre, King's Hedges Road, Cambridge. CB4 2HY. United Kingdom
Submitter's contact: Toby Davis
Telephone number: +44-1223-921904
Device Proprietary Name: Prostate Core™
Device Common Name(s): Medical image management and processing system
Classification Name: Class II
Product Code: QIH
Regulation No: 21 C.F.R. §892.2050
Classification Panel: Radiology Devices
Prostate Core is Substantially Equivalent to the following Legally Marketed device:
Legally Marketed Predicate Devices
| 510(k) Number | Trade Name | Manufacturer | Product Code |
| --- | --- | --- | --- |
| K203582 | qp-Prostate | Quibim S.L | LLZ |
## Device Description
Prostate Core™ is a software-only medical device intended for the automated segmentation and volumetric measurement of the prostate gland from prostate magnetic resonance imaging (MRI) data.
Prostate Core™ uses a deep learning-based algorithm to automatically segment the entire prostate gland from DICOM-formatted MRI studies and to calculate the corresponding gland volume. The algorithm is based on a convolutional neural network architecture that operates on image data to generate segmentation outputs. The model is a locked algorithm that does not adapt or learn from new data during clinical use. Model development included training and validation on independent, multi-center datasets with expert radiologist annotations, and the final model is fixed at the time of release.
Prostate Core™ supports prostate MRI studies acquired on 1.5T and 3T scanners from major
1
{5}
LUCIDA Medical
manufacturers (e.g., Siemens, GE Healthcare, Philips) using acquisition protocols consistent with PI-RADS v2.1. Prostate Core™ is not patient-contacting and does not include any hardware components or accessories. The device operates as an image processing tool and does not perform computer-aided detection (CADe) or computer-aided diagnosis (CADx). It does not analyze, score, or characterize lesions, nor does it generate heatmaps or lesion-specific outputs.
Prostate Core™ is designed to be integrated into the clinical imaging workflow and does not include a graphical user interface. Outputs are viewed via the healthcare facility's Picture Archiving and Communication System (PACS) or other image management platforms. Following processing, the segmentation mask and volumetric data are exported in standardized DICOM formats (including DICOM-SEG and RTSTRUCT) to the PACS or other designated systems for storage, review, and subsequent clinical interpretation by appropriately trained healthcare professionals.
Prostate Core™ uses a modular, pipeline-based architecture developed in Python 3, employing industry-standard libraries for medical imaging and AI inference, including PyTorch.
Prostate Core™ is deployed on-premises, on a hospital server connected to the local imaging network.
Prostate Core™ includes the following key components:
1. Data Loading Module: Receives and validates DICOM-formatted MRI input data and allows specification of relevant MRI sequences (T2 axial, ADC axial, high B-value axial) as required by varying site protocols.
2. Segmentation Module: Applies a validated deep learning algorithm to generate a whole-gland prostate segmentation.
3. Volume Calculation Module: Computes the prostate gland volume (in milliliters) derived from the segmentation mask
4. Export Module: Converts the segmentation and volumetric data into standardized DICOM outputs for transfer to the PACS or compatible imaging systems.
Inputs: any DICOM-compliant prostate MRI data that meets PI-RADS v2.1 imaging standards.
Output: DICOM-SEG and RTSTRUCT files containing segmentation masks and volumetric measurements, fully compatible with common PACS and radiology workstations.
## Technological Characteristics
Prostate Core™ is a software-only medical device that performs automated segmentation and volumetric quantification of the prostate from T2-weighted MRI images. Prostate Core™ utilizes a supervised deep learning algorithm based on a convolutional neural network architecture to generate voxel-level probability maps that are converted into a binary prostate segmentation. The processing pipeline includes image pre-processing and post-
2
{6}
LUCIDA Medical
processing steps to improve segmentation consistency and reduce false positives.
Prostate Core™ receives DICOM-format MRI data via standard DICOM networking interfaces and outputs DICOM-compliant segmentation objects (e.g., DICOM SEG, RTSTRUCT) and structured reports, including prostate volume measurements. The software is deployed within a healthcare provider network environment and interfaces with PACS or other designated DICOM systems for data input and output. The device does not modify original images and does not retain patient data after processing. The algorithm has been trained and validated using multi-center datasets with expert radiologist consensus annotations, and performance has been evaluated using established quantitative metrics for segmentation accuracy and volumetric agreement, consistent with similar automated image analysis devices.
## Indications for Use
Prostate Core™ is a software medical device intended for automatic segmentation and volumetric quantification of the prostate. Prostate Core™ is intended to be used by trained healthcare professionals as an aid to reading/assessing prostate MRI images of adult males aged 22 or over. Prostate Core™ is not intended as a replacement for clinical review or of the judgement of healthcare professionals. The software does not modify the original medical images and does not provide a direct diagnosis.
Prostate Core™ is not intended for patients:
- who received a prostate biopsy within 6 weeks before the MRI examination
- with prior prostate or urethral treatment or surgery that substantially affects the prostate gland
- whose MRI scan quality is severely impaired, for example by patient motion, breathing, peristaltic motion, rectal gas, or metallic implants such as hip prostheses
- where a radiologist considers that the MRI examination is not of diagnostic quality
- with MRI examinations that substantially deviate from PI-RADS®v2.1 scanning protocols
- whose MRI images have been captured using an endo-rectal coil.
3
{7}
LUCIDA
Medical
## Substantial Equivalence Discussion
The subject device and the predicate device, qp-Prostate (K203582), are both Class II, software-only PACS tools intended for the processing of prostate MR images and adjunctive information for review by trained healthcare professionals. Neither device is intended for diagnostic use or as a stand-alone clinical decision support system. Both use DICOM MRI inputs, integrate with PACS, and output quantitative data for radiologist interpretation.
There are important differences between the subject device and the predicate device.
The subject device is limited to automated prostate segmentation and volumetric measurement, whereas the predicate also provides perfusion/diffusion analysis and structured reporting. This difference does not introduce any new or different type of risk as the subject device offers a subset of the predicate functionality, limiting the complexity and the risks associated with the additional functionality.
The predicate device (qp-Prostate (K203582)) includes a user interface and features to enable visualization, editing and review of segmentation results, while the subject device exports segmentation outputs directly to PACS or or other designated DICOM system, without a dedicated GUI. The absence of a GUI is a workflow choice that simplifies integration into existing PACS environments and is accompanied by suitable labeling to ensure that output segmentations are reviewed prior to onward use. This design reflects a workflow configuration difference rather than a difference in intended use or risk profile. Appropriate labeling ensures that segmentation outputs are reviewed by qualified healthcare professionals prior to clinical use.
Both subject and predicate devices are standalone software devices which process DICOM compliant MR images and output prostate segmentation and quantitative volumetric results.
A technical difference: The subject device uses a deep learning-based algorithm for automatic whole-gland segmentation while the predicate device uses automated prostate segmentation techniques. Despite this technological difference, both devices perform the same fundamental clinical task and produce equivalent types of outputs.
The performance of the subject device has been validated against expert reference standards using established quantitative metrics, demonstrating accuracy within clinically acceptable limits for adjunctive use. This technical difference does not raise new questions of safety or effectiveness. Automated segmentation of anatomical structures from MRI data is a well-established methodology, and the performance of the subject device has been validated against expert manual segmentation reference standards using accepted quantitative metrics, including Dice similarity coefficient and volumetric error analysis.
Validation studies demonstrate that the subject device achieves clinically acceptable accuracy for prostate segmentation and volumetric measurement. Performance testing was conducted using multi-vendor MRI datasets and compared against independently generated expert manual segmentations. Results demonstrate that the device performs reliably within clinically
4
{8}
LUCIDA Medical
acceptable limits for adjunctive use.
Therefore, although technological differences exist, these differences do not introduce new questions of safety or effectiveness. The subject device is as safe and effective as the predicate device qp-Prostate (K203582) and is substantially equivalent
| Feature | Subject Device (Prostate CoreTM) | Predicate Device (qp-Prostate, K203582, LLZ) |
| --- | --- | --- |
| Regulation / Product Code | 21 CFR 892.2050, PACS, Class II, QIH | 21 CFR 892.2050, PACS, Class II, LLZ |
| Intended Use / Indications | Automated segmentation and volumetric measurement of the prostate gland from MRI; adjunctive, not diagnostic; reviewed by qualified professionals. | Image viewing, processing, and analysis of prostate MRI; includes ADC and perfusion maps, segmentation, structured reporting; adjunctive, not diagnostic. |
| Inputs | DICOM-formatted prostate MRI (1.5T/3T, PI-RADS v2.1 compliant). | DICOM prostate MRI (T2 + DWI and/or DCE). |
| Outputs | DICOM-SEG and RTSTRUCT prostate masks and gland volume (ml), DICOM reports including Structured Reports | ADC and perfusion maps, segmentation masks, structured reports. |
| Segmentation Method | Deep learning AI for whole-gland segmentation. | Automated prostate segmentation algorithm. |
| Automation | Fully automatic; no manual interaction. | Automated but allows user verification/editing. |
| User Interface | None; outputs only to PACS or other designated DICOM systems. | Web-based viewer, structured reporting GUI. |
| Workflow Integration | On-premises server encrypted DICOM transfer; outputs routed to PACS or other designated DICOM systems. | On-premises client-server model; PACS query/retrieve; browser access |
5
{9}
LUCIDA Medical
| Performance Data | Validated on multi-vendor MRI datasets; metrics include Dice coefficient and volumetric error. | Validated on prostate MR datasets; compared against Olea Sphere. |
| --- | --- | --- |
## Performance Data
The following performance data were provided in support of this submission.
### Software Verification and Validation Testing
Non-clinical software verification and validation testing was conducted in accordance with a defined software development lifecycle consistent with IEC 62304 and ISO 14971. Testing demonstrated that the device meets its software system requirements, product requirements, and user needs.
Verification activities included unit, integration, and system-level testing. System-level testing was performed in a representative healthcare IT environment and included end-to-end validation of device operation from data input through to output generation.
Testing included:
- DICOM data handling and interoperability testing, including receipt of input MRI studies from PACS systems and transmission of output segmentation results (e.g., DICOM-SEG, RTSTRUCT) to external systems.
- Data validation and error handling, including rejection of incomplete, invalid, or non-DICOM inputs and verification of safe system behavior in these scenarios.
- Configuration and workflow testing, including verification of correct routing of outputs to configured destinations.
- Algorithm integration testing, confirming correct execution of the processing pipeline and generation of segmentation and volumetric outputs.
### Cybersecurity and Network Communication Testing
Cybersecurity-related verification activities were conducted to ensure secure handling of patient data and protection of device functionality within a networked environment. Testing included:
- Verification of secure data transmission, including Transport Layer Security (TLS) implementation for DICOM communication
- Validation of system behaviour under secure and non-secure configurations, including appropriate handling of configuration settings
- Assessment of data integrity protections, ensuring that input data is validated prior to processing and that outputs are not generated from invalid or incomplete data
- Verification of system behaviour under fault conditions, including network interruption and communication failure scenarios
6
{10}
LUCIDA Medical
## Bench (non-Clinical) and System Testing
For this software-only device, bench testing consists of non-clinical system and interoperability testing performed using representative datasets and simulated clinical workflows.
Testing includes end-to-end processing of prostate MRI studies from input through segmentation and output generation, verification of compatibility with multi-vendor MRI data (e.g., Siemens, GE Healthcare, Philips), validation of output formats (DICOM-SEG, RTSTRUCT, structured reports) for compatibility with standard PACS systems and verification of system performance across a range of imaging conditions and acquisition parameters
These testing activities demonstrate that the device performs as intended within its operating environment and that risks associated with interoperability, data handling, and network communication are appropriately controlled.
## Stand-alone Performance Testing
No prospective clinical study was conducted to support this submission. Clinical performance was established through non-clinical evaluation using retrospectively collected clinical imaging data. The level of evidence is consistent with an observational study using retrospectively collected clinical imaging data and a reference standard derived from expert radiologist consensus.
The device was evaluated using independent, multi-center datasets of T2-weighted prostate MRI studies representative of the intended use population, including variation in scanner types, imaging protocols, and patient characteristics. Ground truth segmentations were established using consensus annotations from multiple expert radiologists with relevant subspecialty expertise.
Performance was assessed using established quantitative metrics for medical image segmentation, including Dice similarity coefficient (DSC), mean surface distance (MSD), and relative absolute volume error (RAVE). Results from both internal testing and independent external validation datasets demonstrated that predefined acceptance criteria were met for segmentation accuracy and volumetric agreement, supporting the device's performance as an aid to the assessment of prostate MRI.
## AI/ML Algorithm Validation and Performance
The device incorporates a locked, supervised deep learning algorithm for automated prostate segmentation and volumetric quantification. The algorithm was developed using annotated prostate MRI datasets. The final model is fixed at release and does not perform continuous or adaptive learning in the field.
## Training Data:
The algorithm was trained using a multi-source dataset comprising public and proprietary data, including PROSTATEx, PAIR-1, PRIME, and U.S. clinical datasets. Training included 775 cases, all with expert-annotated prostate segmentations.
7
{11}
LUCIDA Medical
## Validation Data and Study Design:
Performance validation was conducted using independent datasets not used in model development, ensuring separation between training and test data. Evaluation included:
- Internal testing dataset: 70 patients
- External validation dataset: 100 patients
- Total validation population: 170 patients
The study design is consistent with a retrospective observational study using independently collected clinical imaging data.
Ground truth was established using consensus annotations from three expert radiologists, with voxel-wise agreement used to define the reference standard.
## Dataset Characteristics:
The validation datasets were designed to be representative of the intended use population and clinical imaging conditions, including:
- Age: Mean ~66–67 years (range ≥42 years)
- Sex: Male patients (consistent with prostate imaging population)
- Ethnicity/Race: Includes White/Caucasian, Black/African American, Hispanic/Latino, and cases with unavailable ethnicity data
- Geographic distribution: Multi-site data from the United States and United Kingdom
- Clinical subgroups: PIRADS categories (1–5)
- Scanner manufacturers: Siemens, GE Healthcare, Philips
- Magnetic field strength: 1.5T and 3T
- Other variables: BMI, site type (academic and community), and imaging protocol variability
All validation datasets were independent of the development and training datasets and were not accessible to the algorithm development team during model development or training. External validation data were held out specifically for regulatory performance evaluation.
## Imaging Protocols:
All cases consisted of T2-weighted prostate MRI acquired using standard clinical protocols consistent with PI-RADS v2.1 recommendations. Imaging parameters included:
- In-plane resolution ≤ 0.8 mm
- Slice thickness ≤ 4 mm
- Whole prostate coverage
- Multi-vendor and multi-site acquisition
8
{12}
LUCIDA Medical
### Performance Metrics and Acceptance Criteria:
Algorithm performance was evaluated using established quantitative segmentation metrics:
• Dice Similarity Coefficient (DSC) – overlap accuracy
• Mean Surface Distance (MSD) – boundary accuracy
- Relative Absolute Volume Error (RAVE) – volumetric accuracy
Predefined acceptance criteria across the whole dataset were:
DSC ≥ 0.90
- MSD ≤ 0.7 mm
- RAVE ≤ 7.09%
### Performance Results
Performance on independent validation datasets demonstrates that all predefined acceptance criteria were met for the overall validation dataset. Subgroup analyses demonstrated generally consistent performance across clinically relevant demographic, imaging and clinical categories.
#### External Validation Dataset (n = 100):
• DSC: Median 0.93 (95% CI: 0.92–0.93)
• MSD: Median 0.27 mm (95% CI: 0.25–0.28)
• RAVE: Median 3.99% (95% CI: 3.17–4.80)
#### Internal Testing Dataset (n = 70):
DSC: Median 0.92
• MSD: Median 0.28 mm
• RAVE: Median 4.60%
### Subgroup Analysis
Performance was evaluated across clinically relevant subgroups, including PIRADS categories, scanner manufacturer and field strength, geographic region and patient demographics. Results demonstrated consistent performance across racial, ethnic and PI-RADS scoring subgroups, and across scanner types. One geographic subgroup (US Midwest, n = 4) had a median RAVE value exceeding the predefined whole-dataset acceptance threshold. Given the very small subgroup size, this isolated finding was not considered indicative of a systematic reduction in device performance. Not all geographical subgroups are of sufficient size to make meaningful evaluations, and performance targets are defined across the whole dataset.
9
{13}
LUCIDA Medical
| Subgroup | Category | n | Median DSC | Median MSD (mm) | Median RAVE (%) |
| --- | --- | --- | --- | --- | --- |
| Overall | - | 100 | 0.93 | 0.27 | 3.99 |
| Race | White/Caucasian | 35 | 0.93 | 0.27 | 2.54 |
| | Black/African American | 24 | 0.93 | 0.30 | 3.65 |
| | Hispanic/Latino | 10 | 0.92 | 0.33 | 3.78 |
| Ethnicity | Not Hispanic/Latino | 60 | 0.93 | 0.28 | 3.18 |
| | Hispanic/Latino | 10 | 0.92 | 0.33 | 3.78 |
| Geographic Region | US Southeast | 49 | 0.93 | 0.30 | 2.81 |
| | UK East of England | 15 | 0.92 | 0.27 | 5.85 |
| | UK South West England | 15 | 0.94 | 0.22 | 5.51 |
| | US Midwest | 9 | 0.93 | 0.22 | 1.75 |
| | US Southern | 8 | 0.91 | 0.25 | 5.11 |
| | US Midwest | 4 | 0.92 | 0.32 | 10.15 |
| PI-RADS Score | PI-RADS 1 | 13 | 0.93 | 0.28 | 3.88 |
| | PI-RADS 2 | 33 | 0.93 | 0.26 | 4.09 |
| | PI-RADS 3 | 19 | 0.93 | 0.27 | 5.91 |
| | PI-RADS 4 | 20 | 0.93 | 0.26 | 3.21 |
| | PI-RADS 5 | 15 | 0.93 | 0.28 | 3.62 |
| Scanner Manufacturer | Siemens | 56 | 0.93 | 0.29 | 3.50 |
| | GE Healthcare | 24 | 0.92 | 0.26 | 5.07 |
| | Philips | 20 | 0.94 | 0.23 | 5.29 |
| Field Strength | 1.5T | 48 | 0.93 | 0.26 | 4.12 |
| | 3T | 52 | 0.93 | 0.27 | 3.68 |
Performance was generally consistent across clinically relevant subgroups, including scanner types, field strengths, PI-RADS categories, and demographic groups, with all median Dice scores \( \geq0.91 \) and within predefined acceptance criteria.
The validation results demonstrate that the algorithm provides accurate and reliable prostate segmentation and volumetric measurements across a representative clinical population and imaging conditions. These results support that the device performs as intended.
## Conclusion
The information provided in this submission demonstrates that Prostate Core™ is substantially equivalent to the predicate device qp-Prostate (K203582) with respect to intended use, technological characteristics, and performance.
Non-clinical verification and validation testing confirmed that the device meets its software system requirements and performs reliably within its intended operating environment. Performance evaluation using independent, multi-centre clinical imaging datasets demonstrated that the device achieves predefined acceptance criteria for segmentation accuracy and volumetric measurement, with generally consistent performance across clinically relevant subgroups, imaging conditions, and scanner types.
10
{14}
LUCIDA Medical
Although the subject device utilises a deep learning-based segmentation algorithm, this technological difference does not raise new questions of safety or effectiveness. The algorithm is a locked model, and its performance has been validated against expert radiologist reference standards using established quantitative metrics.
Based on the totality of the evidence, Prostate Core™ is as safe and effective as the predicate device and does not raise new questions of safety and effectiveness.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.