K253786 · Siemens Medical Solutions USA, Inc. · QIH · Jun 25, 2026 · Radiology
Device Facts
Record ID
K253786
Device Name
syngo.CT Coronary Cockpit
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
QIH · Radiology
Decision Date
Jun 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, PCCP, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K253786 · Jun 25, 2026
syngo.CT Coronary Cockpit
Siemens Medical Solutions USA, Inc.
Retrospective clinical CT angiography datasets
The sponsor used a retrospective cohort of 425 clinical CT scans to validate the performance of coronary vessel tracing and lumen/wall segmentation algorithms.
Patients undergoing cardiac CT angiography (CCTA) for evaluation of coronary artery disease; Sample Size: 425 CT data sets; Number of Sites: Data from 5 US states and 3 European countries
syngo.CT Coronary Cockpit is an image analysis software package for evaluating cardiac CT angiography volume data sets. Combining automated and semi-automated image processing, visualization and reporting tools, the software package is designed to assist the physician in confirming the presence or absence of coronary lesions and their evaluation, documentation, follow-up as well as treatment planning.
Device Story
Software application for cardiac CT angiography (CCTA) analysis; functions as a plug-in to MM Reading application on syngo.via. Inputs: CCTA volume datasets. Processing: Automated/semi-automated algorithms for vessel centerline tracing, lumen segmentation, and vessel wall segmentation. Outputs: 2D/3D visualizations, profile curves, coronary unfolded vessel views, plaque quantification/classification, stenosis grading, and reporting. Used in clinical settings by physicians (radiologists/cardiologists) to confirm coronary lesions, assess stenosis/plaques, and plan treatment. Affects clinical decision-making by providing quantitative metrics (lumen diameter/area, plaque volume, remodeling index) to support diagnostic and therapeutic planning.
Software-only medical image management and processing system. Operates as a plug-in to syngo.via. Implements automated/semi-automated algorithms for vessel tracing, lumen/wall segmentation, and plaque quantification. Complies with IEC 62304 (software lifecycle), NEMA PS 3.1-3.20 (DICOM), ISO 14971 (risk management), and ISO 20417/15223-1 (labeling).
Indications for Use
Indicated for physicians to evaluate cardiac CT angiography volume data sets for the presence, absence, and evaluation of coronary lesions, including documentation, follow-up, and treatment planning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 25, 2026
Siemens Medical Solutions USA, Inc.
% Kenny M Bello
Regulatory Affairs Professional
810 Innovation Dr.
KNOXVILLE, TN 37932
Re: K253786
Trade/Device Name: syngo.CT Coronary Cockpit
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: May 22, 2026
Received: May 26, 2026
Dear Kenny M Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253786 - Kenny M Bello
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{3}
K253786 - Kenny M Bello
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253786
Device Name
syngo.CT Coronary Cockpit
Indications for Use (Describe)
syngo.CT Coronary Cockpit is an image analysis software package for evaluating cardiac CT angiography volume data sets. Combining automated and semi-automated image processing, visualization and reporting tools, the software package is designed to assist the physician in confirming the presence or absence of coronary lesions and their evaluation, documentation, follow-up as well as treatment planning.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
K253786

## 510(k) Summary
syngo.CT Coronary Cockpit (VC10)
### 1. Identification of the Submitter
| **Submitter / Primary Contact Person** | Kenny M Bello Regulatory Affairs monsuru.bello@siemens-healthineers.com +1(202) 856-6099 |
| --- | --- |
| **Secondary Contact Person** | Alaine Medio Regulatory Affairs Manager alaine.medio@siemens-healthineers.com +1 (865)206-0337 |
| **Submitter Address** | SIEMENS MEDICAL SOLUTIONS USA, INC. 810 INNOVATION DR. KNOXVILLE, TN 37932 Registration Number: 1034973 |
| **Location of Manufacturing Site** | Siemens Healthineers AG SIEMENSSTRASSE 1 -OR- Rittigfeld 1 FORCHHEIM Bavaria, DE 91301 Registration Number: 3004977335 |
| **Importer/Distributor** | SIEMENS MEDICAL SOLUTIONS USA, INC. 40 Liberty Boulevard Malvern, PA 19355 Registration Number: 2240869 |
### 2. Device Name and Classification
Product Name: *syngo.CT* Coronary Cockpit
Propriety Trade Name: *syngo.CT* Coronary Cockpit
Classification Name: Medical image management and processing system
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Product Code: QIH
### 3. Predicate Devices
#### Predicate Device
Trade Name: Cleerly Labs v2.0
Classification Name: Medical image management and processing system
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Product Code: LLZ
K-Number: K202280
Siemens Medical Solutions USA, Inc.
1
{6}
K253786

# **Reference Device #1:**
Trade Name: syngo.CT Coronary Analysis
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
CFR Section: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
K-Number: K173637
# **Reference Device #2:**
Trade Name: Philips Comprehensive Cardiac Analysis (CCA) – Plaque Assessment Tool
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
CFR Section: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
K-Number: K092747
# **4. Device Description**
The medical device, syngo.CT Coronary Cockpit, is a software application, which is intended to be used as an interactive tool for viewing and analyzing coronary computed tomography angiography (CCTA) datasets. It is applied on the images acquired during the regular work-up and provides postprocessing and visualization functionalities for the evaluation of coronary artery disease. syngo.CT Coronary Cockpit VC10 is available as an extension for the MM Reading application on syngo.via. The syngo.CT Coronary Cockpit is an extension to One-Viewer based applications, i.e., it is not a standalone application but a plug-in to the hosting environment.
# **5. Indications for Use**
syngo.CT Coronary Cockpit is an image analysis software package for evaluating cardiac CT angiography volume data sets. Combining automated and semi-automated image processing, visualization and reporting tools, the software package is designed to assist the physician in confirming the presence or absence of coronary lesions and their evaluation, documentation, follow-up as well as treatment planning.
Siemens Medical Solutions USA, Inc.
2
{7}
K253786
SIEMENS^{}[] Healthineers
## 6. Indications for Use Comparison to the Predicate Device
| Subject Device | Predicate Device |
| --- | --- |
| syngo.CT Coronary Cockpit (SOMARIS/8 VC10) | Cleerly Labs v2 |
| syngo.CT Coronary Cockpit is an image analysis software package for evaluating cardiac CT angiography volume data sets. Combining automated and semi-automated image processing, visualization and reporting tools, the software package is designed to assist the physician in confirming the presence or absence of coronary lesions and their evaluation, documentation, follow-up as well as treatment planning. | Cleerly Labs is a web-based software application that is intended to be used by trained medical professionals as an interactive tool for viewing and analyzing cardiac computed tomography (CT) data for determining the presence and extent of coronary plaques (i.e. atherosclerosis) and stenosis in patients who underwent Coronary Computed Tomography Angiography (CCTA) for evaluation of CAD or suspected CAD. This software post processes CT images obtained using any Computed Tomography (CT) scanner. The software provides tools for the measurement and visualization of coronary arteries. The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present. |
The syngo.CT Coronary Cockpit and Cleerly Labs v2 software applications are intended to assist physicians in the evaluation of cardiac computed tomography (CT) angiography datasets. These applications provide functionality for image processing, visualization, quantitative assessment, and reporting to support physicians in confirming the presence or absence of coronary artery lesions. Both devices provide an assessment of stenosis and coronary plaques, which in the case of syngo.CT Coronary Cockpit is reflected by the term 'coronary lesions and their evaluation'. The indication for treatment planning in syngo.CT Coronary Cockpit is not present in the predicate device.
Siemens Medical Solutions USA, Inc.
3
{8}
K253786
## 7. Comparison of Technological Characteristics with the Predicate Device
The differences between the above referenced predicate device are listed at a high-level in the following table:
| Functionality | Subject Device | Predicate Device | Comparison Result |
| --- | --- | --- | --- |
| | syngo.CT Coronary Cockpit (SOMARIS/8 VC10) | Cleerly Labs v2.0 | |
| **Basic Visualization** | | | |
| 2D image display | yes | yes | Comparable |
| 3D image display | yes | yes | Comparable |
| Vessel display: CPR, cross-sections | yes | yes | Comparable |
| Profile curve | yes | no | Comparable to reference device #1 *syngo.CT Coronary Analysis* |
| **Vessel Definition and Visualization** | | | |
| Automatic Centerline Tracing | yes | yes | Comparable |
| Centerline Editing | Yes | yes | Comparable |
| Coronary Unfolded Vessel View visualization | yes | no | Comparable to Globe View feature of reference device #2 *Philips CT Comprehensive Cardiac Analysis* |
| **Anatomical Evaluation** | | | |
| Lumen Segmentation Algorithm | yes | yes | Comparable |
| Lumen Segmentation Display and Editing | yes | yes | Comparable |
| Definition and Evaluation of lesions | yes | yes | Comparable |
| **Plaque Evaluation** | | | |
| Vessel Wall Segmentation Algorithm | yes | yes | Comparable |
| Vessel Wall Segmentation Display and Editing | yes | yes | Comparable |
Siemens Medical Solutions USA, Inc.
4
{9}
K253786
**SIEMENS**^{}[] **Healthineers**
| Functionality | Subject Device | Predicate Device | Comparison Result |
| --- | --- | --- | --- |
| | syngo.CT Coronary Cockpit (SOMARIS/8 VC10) | Cleerly Labs v2.0 | |
| Plaque Classification and Display | yes | yes | Comparable |
| Plaque Quantification | yes | yes | Comparable |
| Remodeling Index | yes | yes | Comparable |
| **Stent Landing Zone Evaluation** | | | |
| Definition and Evaluation of Stent Landing Zone Markers | yes | No | Comparable to definition of stenosis markers in reference device #1 *syngo.CT Coronary Analysis* |
| Export to Angio System as DICOM SSO | yes | No | Comparable to SSO Export in reference device #1 *syngo.CT Coronary Analysis* |
| **General** | | | |
| Results creation and Reporting | Yes | Yes | Comparable |
| Integration with Spectral Imaging | Yes | No | Comparable to spectral imaging support in reference device #1 *syngo.CT Coronary Analysis* |
## 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Software Verification and Validation
Software Documentation for Enhanced documentation Level per FDA’s Guidance Document “Guidance for the Content of Premarket Submissions for Device Software Functions” issued on June 14, 2023, is also included as part of this submission. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
### Summary of Performance Testing
The coronary vessel tracing algorithm has been validated on 425 CT data sets from individual subjects. Data originated from 5 different US states and 3 different European countries. Data covers a mix of slice thicknesses from 0.2 to 1.0 mm and tube voltage from 70 KVP to 140 KVP. The coronary lumen and wall segmentation algorithms have been evaluated using a random subset of 150 cases of the above data. Additional details about the data distribution are shown below.
Siemens Medical Solutions USA, Inc.
5
{10}
K253786
**SIEMENS**^{}[] **Healthineers**
| Category | Property | Count Vessel Tracing | Count Lumen and Wall segmentation |
| --- | --- | --- | --- |
| Manufacturer | Canon/Toshiba | 72 | 23 |
| | GE Med Sys | 67 | 25 |
| | Philips | 61 | 27 |
| | Siemens | 225 | 75 |
| Patient Sex | F | 159 | 49 |
| | M | 264 | 101 |
| | other / unknown | 2 | |
| Patient Age | 22-49 | 65 | 13 |
| | 50-59 | 103 | 30 |
| | 60-69 | 144 | 60 |
| | 70-79 | 90 | 36 |
| | ≥80 | 22 | 10 |
| | unknown | 1 | 1 |
| CAD-RADS score according to clinical report | 0 | 90 | 8 |
| | 1 | 58 | 24 |
| | 2 | 64 | 26 |
| | 3 | 61 | 30 |
| | 4A | 62 | 27 |
| | 4B | 47 | 20 |
| | 5 | 43 | 15 |
For coronary vessel tracing, data was annotated by an ARRT-certified technologist with >15 years of experience and reviewed by a US-board certified radiologist (SCCT-Level 3 equivalent experience) or a US-board certified cardiologist (CBCC-Level 3 certification), both having >20 years of experience.
For coronary lumen and wall segmentation, two independent annotations were generated by a team of seven US-board certified radiologists or cardiologists with SCCT Level III or equivalent certification (years of experience ≥ 5, average 8.0 years). Adjudication was performed by a team of 2 US-board certified cardiologists with SCCT Level III or equivalent certification (7 and 25 years of experience, respectively).
For vessel tracing, the algorithm failed in 0% of the cases. Pointwise sensitivity and precision to identify vessels ≥ 1.5 mm were found to be 95.8% and 96.1%, respectively, with an average DICE of 95.7%.
For lumen segmentation, the algorithm failed in 0% of the cases. For coronary lumen segmentation, the following results were obtained:
| Metric | Pearson's r |
| --- | --- |
| Lumen volume (normalized by vessel length) | 0.906 |
| Lumen diameters | 0.861 |
| Lumen area | 0.859 |
| Stenosis grading (diameter-based) | 0.728 |
| Stenosis grading (area-based) | 0.755 |
Siemens Medical Solutions USA, Inc.
6
{11}
K253786
SIEMENS^{}[] Healthineers
For plaque localization, 94.8% of the ground truth plaque volume was part of a plaque cluster detected by the algorithm. Vice versa, 96.4% of the plaque volume calculated by the algorithm was part of a ground truth plaque cluster.
For plaque quantification the following results were obtained:
| Metric | Pearson's r |
| --- | --- |
| Plaque volume (cluster-based) | 0.883 |
| Plaque volume (vessel-based) | 0.883 |
| Calcified | 0.951 |
| Non-calcified | 0.727 |
| High density | 0.730 |
| Low density | 0.582 |
| Remodeling index (lesion-based) | 0.671 |
For all algorithms, consistent results were observed for all relevant subgroups. In summary, all performance targets were met.
### Risk Analysis
The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Siemens hereby certifies that syngo.CT Coronary Cockpit meets the following FDA Recognized Consensus standards listed below:
| Standard | Version | Content | FDA Recognition Number (if applicable) |
| --- | --- | --- | --- |
| IEC 62304:2006 +AMD1:2015 (Ed.1.1) | :2015 | 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes | 13-79 |
| NEMA PS 3.1 - 3.20 2024e | :2024 | Digital Imaging and Communications in Medicine (DICOM) Set | 12-363 |
| ISO 14971 | :2019 | Application of Risk Management to Medical Devices | 5-125 |
| IEC 62366-1 | Edition 1.1 2020-06 | Medical devices - Part 1: Application of usability engineering to medical devices | 5-129 |
Siemens Medical Solutions USA, Inc.
7
{12}
K253786
SIEMENS Healthineers
| Standard | Version | Content | FDA Recognition Number (if applicable) |
| --- | --- | --- | --- |
| | CONSOLIDATED VERSION | | |
| ISO 15223-1 | Fourth edition 2021-07 | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements | 5-134 |
| ISO 20417:2021 | First edition 2021-04 Corrected version 2021-12 | Medical devices - Information to be supplied by the manufacturer | 5-135 |
## 9. Predetermined Change Control Plan (PCCP)
Syngo.CT Coronary Cockpit has a Predetermined Change Control Plan (PCCP) authorized by FDA. The plan allows for the following modification of the lumen and wall segmentation algorithm:
- including additional training data,
- modification of hyper-parameters for algorithm training, and
- modification of pre- and post-processing parameters of the algorithm.
Any modified algorithm will undergo performance testing and software verification and validation described in Section 8. Only if both, performance testing and verification and validation, have been successfully passed, the modification will be implemented into the next suitable software version.
Users will be informed about updates, including modifications implemented under the authorized Predetermined Change Control Plan (PCCP), through multiple established communication mechanisms to ensure transparency and traceability. Specifically, users are informed through:
- Release notes provided with each software version, describing the implemented modifications, including PCCP-authorized changes and their impact on device performance where applicable;
- Updates to the Instructions for Use (IFU) and user documentation, reflecting relevant modifications and revised performance information;
- Direct customer communication by Siemens Healthineers Customer Service (CS) in connection with deployment activities, including information on the update content and version installed;
- Standard marketing and customer communication channels, as defined in the Siemens Healthineers Quality Management System (QMS), ensuring consistent dissemination of update-related information.
Through these combined measures, users are systematically informed of PCCP-authorized modifications and the corresponding software version in which they are implemented.
Siemens Medical Solutions USA, Inc.
8
{13}
K253786
SIEMENS Healthineers
## 10. Conclusion
syngo.CT Coronary Analysis has the same intended use and similar indication for use as the predicate device. The subject device syngo.CT Coronary Analysis does not have changes in fundamental scientific technology compared to the predicate device. The technological characteristics are the same as the predicate device. For the subject device, syngo.CT Coronary Analysis, Siemens used the same testing with the same workflows as used to clear the predicate device. Siemens considers syngo.CT Coronary Cockpit to be as safe, as effective, and with performance substantially equivalent to the commercially available predicate device.
Siemens Medical Solutions USA, Inc.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.