K253762 · Strongback Mobility AG C/O Blueaudit GmbH · IOR · Sep 22, 2026 · Physical Medicine
Device Facts
Record ID
K253762
Device Name
ErgoSteel (ES0001/696)
Applicant
Strongback Mobility AG C/O Blueaudit GmbH
Product Code
IOR · Physical Medicine
Decision Date
Sep 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The device is intended for medical purpose to provide mobility to persons restricted to a sitting position.
Device Story
ErgoSteel is a folding, manual, mechanical wheelchair for indoor/outdoor use; provides mobility to adults restricted to a sitting position. Features include four-wheel base, seat, armrests, footplates, and push handles. Key differentiator is integrated, fixed-geometry lumbar support in backrest for passive spinal support. Device is non-powered; no software/firmware; no electrical components. Operated by user or caregiver. Output is physical mobility support. Benefits include improved patient mobility and ergonomic comfort. Device is disassembled for transport.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including ISO 7176-1 (static stability), ISO 7176-3 (brakes), ISO 7176-5 (dimensions), ISO 7176-8 (fatigue strength), ISO 7176-16 (flammability), and biocompatibility assessment per ISO 10993-1.
Technological Characteristics
Mechanical wheelchair; SPCC steel frame; foldable. Components: PVC handgrips, PUR armpads, PET (Oxford fabric) seat/back with integrated lumbar support, PA footplates, PVC-coated polyester side panels. Dimensions: 24" rear wheels, 8" front casters. Manual operation; no energy source. Sterilization not applicable. Complies with ISO 7176 series and ISO 10993-1.
Indications for Use
Indicated for adults (22 years and older) restricted to a sitting position requiring mobility assistance. Suitable for indoor and outdoor use.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Zhenjiang Assure Mechanical Wheelchair, Model K1 (K231320)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG
ADMINISTRATION
September 22, 2026
Strongback Mobility AG C/O Blueaudit GmbH
Roman Hefle
Quality Manager
Schindellegistrasse 9
Wolfhagen, 34466
Germany
Re: K253762
Trade/Device Name: ErgoSteel (ES0001/696)
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: August 21, 2026
Received: November 25, 2025
Dear Roman Hefle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253762 - Roman Hefle
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K253762 - Roman Hefle
Page 3
Sincerely,
**Tushar Bansal -S**
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K253762
Device Name
ErgoSteel (ES0001/696)
Indications for Use (Describe)
The device is intended for medical purpose to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
# 510(k) SUMMARY
*Prepared in accordance with 21 CFR 807.92*
**510(k) Number K253762 — ErgoSteel Manual Wheelchair (Model ES0001 / 696)**
## 1. Submitter — 21 CFR 807.92(a)(1)
| **Submitter** | Strongback Mobility AG |
| --- | --- |
| **Address** | c/o blueAUDIT GmbH, Schindellegistrasse 73, 8808 Pfaeffikon SZ, Switzerland |
| **Telephone** | +49 178 8765829 |
| **Contact person** | Mr. Roman Hefle, Quality Manager |
| **E-mail** | roman.hefle@strongbackmobility.com |
| **Date summary prepared** | 16 September 2026 |
## 2. Device — 21 CFR 807.92(a)(2)
| **Trade / proprietary name** | ErgoSteel (ES0001/696) |
| --- | --- |
| **Common / usual name** | Mechanical wheelchair |
| **Classification name** | Wheelchair, Mechanical |
| **Regulation number** | 21 CFR 890.3850 |
| **Product code** | IOR |
| **Regulatory class** | Class I |
## 3. Legally marketed predicate device — 21 CFR 807.92(a)(3)
| **510(k) number** | K231320 |
| --- | --- |
| **Device** | Zhenjiang Assure Mechanical Wheelchair, Model K1 |
| **Product code** | IOR |
## 4. Device description — 21 CFR 807.92(a)(4)
The STRONGBACK ErgoSteel Wheelchair is a folding wheelchair with integrated ergonomic back support (lumbar support).
It is an indoor/outdoor wheelchair that has a base with four wheels with a seat. The device can be disassembled for transport, and it is foldable. The device complies with the applicable requirements of ISO 7176.
This device has direct and indirect patient-contacting components, including the seat cushion, the backrest with integrated lumbar support, the armrests with padding, the footplates, the push handles (handgrips), and the upholstery materials (seat & back fabric). This device does not contain software/firmware. This device does not require electrical safety evaluation. This device is not an in vitro diagnostic device.
{6}
## 5. Intended use and Indications for Use — 21 CFR 807.92(a)(5)
The Indications for Use of the subject device are identical to the cleared Indications for Use of the primary predicate device K231320:
"The device is intended for medical purpose to provide mobility to persons restricted to a sitting position."
The device is an Over-the-Counter (OTC) device under 21 CFR 801 Subpart C and is suitable for indoor and outdoor use.
The intended user population is adults (22 years old and greater), consistent with the cleared intended user population of the primary predicate device. The device is not intended for use by children or adolescents, i.e. persons under 22 years of age.
The intended use of the predicate device is the same as that of the subject device.
{7}
## 6. Comparison of technological characteristics with the predicate device — 21 CFR 807.92(a)(6)
The subject device and the primary predicate device K231320 are substantially equivalent in technological characteristics. The differences identified below do not raise any new questions of safety or effectiveness.
| Element of Comparison | Subject Device — ErgoSteel, 510(k) K253762 | Primary Predicate — K231320, Zhenjiang Assure Mechanical Wheelchair, model K1 | Comparison / Notes |
| --- | --- | --- | --- |
| **REGULATORY IDENTITY** | | | |
| Regulation Name | Wheelchair, Mechanical | Wheelchair, Mechanical | Identical |
| Regulation Number | 21 CFR 890.3850 | 21 CFR 890.3850 | Identical |
| Product Code | IOR | IOR | Identical |
| 510(k) Type | Traditional 510(k) | Special 510(k) | Different submission type; does not affect substantial-equivalence determination. |
| **INTENDED USE & INDICATIONS** | | | |
| Intended Use | The device is intended for medical purpose to provide mobility to persons restricted to a sitting position. | The device is intended for medical purpose to provide mobility to persons restricted to a sitting position. | Identical |
| Over-the-Counter Use | Yes | Yes | Identical |
| Use Environment | Indoor / outdoor | Indoor / outdoor | Identical |
| Target Population | Adults (22 years old and greater) | Adults (22 years old and greater) | Identical. The published K231320 510(k) Summary comparison table states the target population as 'The elderly and disabled' and does not state an age group. |
| **DESIGN & DIMENSIONS** | | | |
| Frame Material | High-quality SPCC steel pipe | High-quality SPCC steel pipe | Identical material grade and construction. ErgoSteel passed ISO 7176-8:2014 fatigue test. |
| Framework | Foldable | Foldable | Identical. |
| Casters (front) | 8' x 1.25' | 8' x 1.25' | Identical caster diameter and width. |
{8}
| Element of Comparison | Subject Device — ErgoSteel, 510(k) K253762 | Primary Predicate — K231320, Zhenjiang Assure Mechanical Wheelchair, model K1 | Comparison / Notes |
| --- | --- | --- | --- |
| Rear Wheel | 24" x 1" | 24" x 1" | Identical rear-wheel diameter and width. |
| Seat Length | 16.54" (42 cm) | 16" – 18" | Within K231320 range; no safety impact. ISO 7176 verified. |
| Seat Height | 19.69" (50 cm) | 19.5" | 0.19" (5 mm) difference; within ISO 7176-5:2008 tolerance. |
| Seat Width | 17.72" (45 cm) | 16" – 20" | Within K231320 range. |
| Max loading | 136 kg / 300 lbs | 136 kg / 300 lbs | Identical load rating. |
| COMPONENT-LEVEL DIMENSIONS (from K231320 Public Summary) | | | |
| Handle size | 103 × 34 mm | 103 × 34 mm | Identical. |
| Footplate size | 197 × 161 mm | 197 × 161 mm | Identical. |
| Backrest size | 16" – 20" | 16" – 20" | Identical. |
| Armrest size | 10.12" × 2.2" × 1.42" | 10.12" × 2.2" × 1.42" | Identical. |
| COMPONENTS LIST | | | |
| Handgrip | Present (PVC) | Present (material not known) | The Handgrip is made from PVC. This material has been previously used in mechanical wheelchairs cleared under the same regulation (21 CFR 890.3850) and product code IOR, with similar intended use and technology (e.g., K131529) This does not impact safety and effectiveness. |
| Footrest / footplate | Present (Polyamide footplate) | Present (Footplate) | Same component, alternative naming. |
| Backrest | Present (Oxford fabric – Polyethylene terephthalate (PET) with integrated lumbar support — see Differentiator row) | Present (Backrest) | Functional component identical. ErgoSteel adds an integrated lumbar-support contour; see dedicated row below. This PET material is included in Section B of Attachment G of FDA's guidance on the "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", and has a documented history of safe use in legally marketed medical devices in contact with intact skin (e.g., through Medical Device Reporting (MDR) analysis, literature search). |
| Armrest | Present (PUR armpad) | Present (Armrest) | Same component, alternative naming. |
{9}
| **Front caster** | Present | Present (Front caster) | Identical. |
| --- | --- | --- | --- |
| **Front fork** | Present | Present (Front fork) | Identical. |
| **Rear wheel** | Present (quick-release axle) | Present (Rear wheel) | Same component; ErgoSteel features quick-release axle for transport convenience. |
| **Skirt guard / side panel** | Present (Teslin mesh fabric) | Not separately listed in K231320 510(k) Summary; side cover function is integrated into the seat/backrest assembly per the K1 design. | Subject device adds a separate Skirt guard / side panel made of Teslin mesh fabric (PVC-coated polyester). This material has been previously used in mechanical wheelchairs cleared under the same regulation (21 CFR 890.3850) and product code IOR, with similar intended use and technology (e.g., K010094). This addition does not impact safety and effectiveness. |
| **Element of Comparison** | **Subject Device — ErgoSteel, 510(k) K253762** | **Primary Predicate — K231320, Zhenjiang Assure Mechanical Wheelchair, model K1** | **Comparison / Notes** |
| --- | --- | --- | --- |
| **Brakes** | Present (standard parking brake, lockable) | Present | Brakes are mandatory wheelchair components verified via ISO 7176-3. Predicate K231320 was tested per ISO 7176-3:2003; ErgoSteel was tested per ISO 7176-3:2012 (newer revision). Same fundamental technology and performance criteria. |
#### **STRONGBACK DIFFERENTIATOR**
{10}
| Integrated Ergonomic Lumbar Support | Present — backrest incorporates an integrated, fixed-geometry lumbar-support contour designed to passively support the lumbar spine of the seated user. Non-removable; non-mechanical; no moving parts; no electrical or pneumatic actuation. | Not present. | Additional feature of the subject device. No new questions of safety or effectiveness are raised: (i) It is a fixed, passive ergonomic shape of the existing backrest; (ii) the Seat & Back is made from Oxford fabric – Polyethylene terephthalate (PET). This PET material is included in Section B of Attachment G of FDA’s guidance on the “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process””, and has a documented history of safe use in legally marketed medical devices in contact with intact skin (e.g., through Medical Device Reporting (MDR) analysis, literature search); (iii) it does not alter the mechanical load path or the structural performance of the wheelchair, (iv) it does not affect indications for use, target population or use environment, and (v) the device passed ISO 7176-1 (static stability), ISO 7176-8 (fatigue) and ISO 7176-16 (resistance to ignition) testing with the lumbar support installed. |
| --- | --- | --- | --- |
| PERFORMANCE / NON-CLINICAL TESTING | | | |
| --- | --- | --- | --- |
| ISO 7176-1 — Static stability | ISO 7176-1:2014 — PASS | ISO 7176-1:1999 — PASS | Same standard, newer revision in subject device. Acceptance criteria unchanged; ErgoSteel meets the more current revision. |
| ISO 7176-3 — Effectiveness of brakes | ISO 7176-3:2012 — PASS | ISO 7176-3:2003 — PASS | Same standard, newer revision in subject device. |
| Element of Comparison | Subject Device — ErgoSteel, 510(k) K255762 | Primary Predicate — K231320, Zhenjiang Assure Mechanical Wheelchair, model K1 | Comparison / Notes |
| --- | --- | --- | --- |
| ISO 7176-5 — Overall dimensions, mass, manoeuvring space | Verified within ISO 7176-5 ranges via dimensional inspection. | ISO 7176-5:2008 — PASS | Dimensions of ErgoSteel are within ISO 7176-5 typical ranges; explicit ISO 7176-5 retest not required for SE. |
| ISO 7176-8 — Fatigue strength | ISO 7176-8:2014 — PASS | Not separately listed in K231320 510(k) Summary | Additional testing performed by subject device sponsor at the current revision; strengthens, not weakens, the SE argument. |
{11}
| ISO 7176-11 — Test dummies | Test dummies used in fatigue and stability testing per ISO 7176-11:2012. | ISO 7176-11:1992 — PASS | Same standard, newer revision in subject device. |
| --- | --- | --- | --- |
| ISO 7176-13 — Coefficient of friction of test surfaces | Not separately reported; embedded in ISO 7176-1 and -3 testing. | ISO 7176-13:1989 — PASS | Test method standard for setting up other tests; equivalent test surfaces used. |
| ISO 7176-15 — Information disclosure and labelling | ISO 7176-15:2012 (current revision) — compliant labelling and IFU per Strongback documentation. | ISO 7176-15:1996 — PASS | Same standard, newer revision in subject device. |
| ISO 7176-16 — Resistance to ignition of upholstered parts | ISO 7176-16:2012 — PASS | ISO 7176-16:1997 — PASS | Same standard, newer revision in subject device. |
| Biocompatibility | In compliance with ISO10993-1; ISO10993-5; and ISO10993-10. | Not listed in the K231320 Special 510(k) Summary. | No new biocompatibility testing was necessary for the subject device. did not require one in its Special 510(k). No new questions raised. |
{12}
## 7. Nonclinical performance data — 21 CFR 807.92(b)(1)
The ErgoSteel Manual Wheelchair (Model ES0001 / 696) is supported by a comprehensive program of nonclinical bench testing and biocompatibility testing in accordance with FDA-recognised consensus standards and the FDA Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs.A. MECHANICAL PERFORMANCE TESTING (BENCH) — Conducted by Jiangsu Assure Medical Equipment Co., Ltd. (the OEM manufacturer) under their ISO 13485:2016 certified Quality Management System.
- ISO 7176-1:2014 — Determination of static stability (forward, rearward, lateral) — PASS
- ISO 7176-3:2012 — Determination of effectiveness of brakes — PASS
- ISO 7176-5:2008 — Determination of overall dimensions, mass and manoeuvring space — PASS
- ISO 7176-7:1998 — Measurement of seating and wheel dimensions — PASS
- ISO 7176-8:2014 — Requirements and test methods for static, impact and fatigue strengths (drop test, fatigue test on rolling drums, double-drum test) — PASS
- ISO 7176-11:2012 — Test dummies (used as reference)
- ISO 7176-13:1989 — Coefficient of friction of test surfaces (used as reference)
- ISO 7176-15:1996 — Requirements for information disclosure, documentation and labelling — PASS
- ISO 7176-16:2012 / ISO 16840-10:2021 — Resistance to ignition of postural support devices — PASS
- ISO 7176-22:2014 — Set-up procedures — PASS
# BIOCOMPATIBILITY
No new biocompatibility testing was necessary.
The Armpad is made from Polyurethane (PUR), the Seat & Back is made from Oxford fabric – Polyethylene terephthalate (PET), and the Footplate is made from Polyamide (PA). These materials (PUR, PET and PA) are included in Section B of Attachment G of FDA's guidance on the "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"", and have a documented history of safe use in legally marketed medical devices in contact with intact skin (e.g., through Medical Device Reporting (MDR) analysis, literature search).
The Handgrip is made from PVC and the Side Panels are made from (Teslin mesh / PVC-coated polyester). These materials (PVC, and PVC-coated polyester) with similar nature of intact skin contact have been previously used in mechanical wheelchairs cleared under the same regulation (21 CFR 890.3850) and product code IOR, with similar intended use and technology (e.g., K131529 and K010094). Since FDA has previously cleared devices under the same regulation and product code with comparable intended use, technology, and similar materials (PVC and PVC-coated polyester) without identifying safety or effectiveness concerns, these material differences do not impact safety and effectiveness of the subject device.
{13}
## 8. Clinical performance data — 21 CFR 807.92(b)(2)
Not Applicable. No clinical investigation, clinical trial, or other prospective clinical study was conducted in support of this 510(k) submission.
## 9. Conclusions — 21 CFR 807.92(b)(3)
The nonclinical bench testing and biocompatibility information described above demonstrates that the ErgoSteel Manual Wheelchair (Model ES0001 / 696) is as safe, as effective, and performs as well as the legally marketed primary predicate device K231320 (Zhenjiang Assure Mechanical Wheelchair, model K1), and meets the applicable FDA-recognised consensus standards for mechanical wheelchairs (ISO 7176 series), biological evaluation (ISO 10993 series), risk management (ISO 14971:2019) and labelling (ISO 7176-15:1996 / ISO 20417:2021). Specifically:
- • Mechanical safety and performance — All required ISO 7176 series tests (static stability, brake effectiveness, dimensions, mass, manoeuvring space, static / impact / fatigue strengths, flammability, labelling, set-up) demonstrated PASS results equivalent to or exceeding the requirements applicable to Class I mechanical wheelchairs under 21 CFR 890.3850 (Product Code IOR).
- • Biocompatibility — No new biocompatibility testing was necessary. All intact skin contacting components (Armpad (PUR), Seat & Back Oxford fabric (PET), Footplate (PA), Handgrip (PVC), and Side Panels (Teslin mesh/PVC-coated polyester)), are either included in Section B of Attachment G of FDA's guidance on the "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"", or have been previously used in FDA-cleared mechanical wheelchairs under the same regulation and product code, with identical intended use and intact skin contact classification. No new biocompatibility risks are introduced by these materials. Therefore, these material differences do not impact safety and effectiveness.
- • Substantial equivalence — The technological characteristics of the ErgoSteel and the primary predicate K231320 (manufactured by the same OEM group, Zhenjiang Assure / Jiangsu Assure Medical Equipment Co., Ltd.) are substantially equivalent in indications for use, intended user population, regulation, classification, dimensions, materials, brake system, frame design, performance characteristics and biocompatibility profile. The differences identified in the substantial equivalence Comparison Table do not raise any new questions of safety or effectiveness.
- • Risk management — A complete ISO 14971:2019-compliant risk management file is maintained in the Strongback Mobility AG Device Master Record. No residual risks were identified that would change the safety or effectiveness conclusion.
Therefore, the ErgoSteel Manual Wheelchair is substantially equivalent to the legally marketed predicate device K231320.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.