Dexcom G7 Continuous Glucose Monitoring (CGM) System; Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System
K253737 · Dexcom, Inc. · QBJ · Feb 3, 2026 · Clinical Chemistry
Device Facts
Record ID
K253737
Device Name
Dexcom G7 Continuous Glucose Monitoring (CGM) System; Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System
Applicant
Dexcom, Inc.
Product Code
QBJ · Clinical Chemistry
Decision Date
Feb 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1355
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Dexcom G7 Continuous Glucose Monitoring System: The Dexcom G7 Continuous Glucose Monitoring System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older. The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes. Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System: The Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (Dexcom G7 15 Day System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older. The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G7 15 Day CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 15 Day CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.
Device Story
Wearable CGM system measures interstitial glucose via subcutaneous wire; converts glucose to electric current; transmitter calculates estimated glucose values (EGV) using onboard algorithm. Data transmitted via Bluetooth Low Energy (BLE) to mobile app or receiver. Used at home by patients; provides real-time glucose readings, trends, and alerts. Supports integration with automated insulin dosing (AID) systems. Modification introduces alternate adhesive patch (spunlace non-woven polyester acrylic).
Clinical Evidence
Clinical study conducted to validate safety and effectiveness of the G7 and G7 15 Day systems with the alternate adhesive patch. Results demonstrated that the systems meet pre-defined acceptance criteria for intended use.
Technological Characteristics
Wire-type interstitial glucose sensing; radio transmitter for wireless communication. Wearable includes sensor, transmitter, and alternate adhesive patch. Biocompatibility per ISO 10993; sterilization per AAMI TIR 28 and ANSI/AAMI/ISO 11135. Reportable range 40-400 mg/dL. Interoperable with mobile apps and receivers via secure wireless protocols.
Indications for Use
Indicated for management of diabetes in persons 2+ years (G7) or 18+ years (G7 15 Day). Contraindicated for dialysis patients, critically ill patients, and those taking hydroxyurea. Not for use during MRI or diathermy.
Regulatory Classification
Identification
An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
*e.g.,* an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
*e.g.,* abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.
Predicate Devices
Dexcom G7 Continuous Glucose Monitoring System (K240902)
Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (K243214)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K253737
B Applicant
Dexcom, Inc.
C Proprietary and Established Names
Dexcom G7 Continuous Glucose Monitoring (CGM) System,
Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QBJ | Class II | 21 CFR 862.1355 - Integrated Continuous Glucose Monitoring System | CH - Clinical Chemistry |
E Purpose for Submission:
Modifications to the adhesive patch
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Dexcom G7 Continuous Glucose Monitoring System:
The Dexcom G7 Continuous Glucose Monitoring System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.
The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.
Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System:
The Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (Dexcom G7 15 Day System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older.
The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.
The Dexcom G7 15 Day CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 15 Day CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Dexcom G7 Continuous Glucose Monitoring System:
- Users of the Dexcom G7 Continuous Glucose Monitoring System can take a standard dose of acetaminophen (up to 1 gram every 6 hours in adults) and still use the Dexcom G7 System readings. Sensor glucose readings will be falsely higher if the user is taking more than a standard acetaminophen dose.
- Sensor glucose readings will also be falsely higher if the user is taking hydroxyurea. Users should not use the Dexcom G7 System for diabetes treatment decisions if they are taking hydroxyurea.
- Don't wear any Dexcom G7 CGM System component during magnetic resonance imaging (MRI) or high-frequency electrical heat (diathermy) treatment. However, it's safe to have a CT scan if you keep the sensor out of the scanned area and cover the sensor with a lead apron during the scan.
- Don't ignore low/high symptoms: Use BG meter to make treatment decisions when the sensor readings don't match the user's low/high symptoms. If needed, seek immediate medical attention.
- Use BG meter to make treatment decisions when the Dexcom G7 CGM System doesn't show both a number and trend arrow as well as during the 30-minute sensor warmup period.
- Don't use if on dialysis or critically ill: The Dexcom G7 CGM System performance hasn't been evaluated in these populations and sensor readings may be inaccurate.
K253737 - Page 2 of 13
{3}
- Don't ignore broken or detached sensor wires. If this happens, contact 24/7 technical support.
- Insert only in the arm or buttocks: Don't wear it on other sites as it may not work as expected.
- Store sensors at room temperature or in the refrigerator, between 36° F and 86° F, but not in the freezer.
- Check settings: Make sure the smart device volume is turned up, not muted, and the speaker works. When a headphone is connected, alerts will only sound through the headphone, not on the smart device speaker.
- Make sure Bluetooth is on. Otherwise, the user won't get readings or alerts.
- Make sure the smart device settings follow Dexcom's recommended settings. Certain phone settings such as Android's Digital Wellbeing and Apple's Screen Time may prevent notifications if enabled.
- Allow Dexcom G7 CGM System app notifications to show on Lock screen. This will ensure the user can receive Dexcom notifications and allow the user to see notifications without unlocking the phone.
- Android users must allow Location Permission, Do Not Disturb Access, and Notifications to use the app.
- Apple users must allow Critical Alerts to use the app.
- Battery: Keep the battery charged.
- Quiet Mode (Vibrate): When this setting is enabled all System alerts will vibrate. The Urgent Low Glucose and technical alerts will still escalate to sound if not acknowledged.
- Quiet Mode (Silence All): When this setting is enabled, all System alerts will be silent. The user won't receive sound or vibration for any alerts. The user will still receive visual alerts on the display device. (Exceptions: App Stopped alerts will still sound.) Check the display device frequently to avoid missing a low/high event.
- Before upgrading the smart device or its operating system, check dexcom.com/compatibility. Always update manually and verify correct device settings afterward.
- Let the date and time on the smart device automatically update when traveling across time zones or switch between standard and daylight saving times. Don't manually change the smart device time because the user may not get readings or alerts and it may make the time on the trend screen wrong.
Use electrical equipment as directed:
Portable radio frequency communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 12 inches to any part of the Dexcom G7 CGM System including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Use of this equipment adjacent to, or stacked with, other equipment should be avoided because it could result in improper operation.
Not using supplied USB charger and cable may cause the receiver battery to not charge. Don't use if the supplied USB charger or cable is damaged. Store supplied USB charger and cable safely. Misuse of the USB cable can be a strangulation risk.
In addition to the special conditions listed above for the Dexcom G7 CGM system, the following special conditions for use were included specifically for the new Direct To Watch feature in K234133:
Watch app settings: The watch app uses settings from your phone app.
K253737 - Page 3 of 13
{4}
Bluetooth wireless technology: Make sure your Bluetooth is on. If not, you won't get readings or alerts.
Make sure your smart device settings follow Dexcom's recommended settings. Certain phone settings such as Android's Digital Wellbeing and Apple's Screen Time may prevent notifications if enabled.
Apple users must allow Location Permission and Critical Alerts to use the phone app.
## Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System:
- Users of the Dexcom G7 15 Day Continuous Glucose Monitoring System can take a standard dose of acetaminophen (up to 1 gram every 6 hours in adults) and still use the Dexcom G7 15 Day CGM System readings. Sensor glucose readings will be falsely higher if the user is taking more than a standard acetaminophen dose.
- Sensor glucose readings will also be falsely higher if the user is taking hydroxyurea. Users should not use the Dexcom G7 15 Day CGM System for diabetes treatment decisions if they are taking hydroxyurea.
- Don't wear any Dexcom G7 15 Day CGM System component during magnetic resonance imaging (MRI) or high-frequency electrical heat (diathermy) treatment. However, it's safe to have a CT scan if you keep the sensor out of the scanned area and cover the sensor with a lead apron during the scan.
- Don't ignore low/high symptoms: Use BG meter to make treatment decisions when the sensor readings don't match the user's low/high symptoms. If needed, seek immediate medical attention.
- Use BG meter to make treatment decisions when the Dexcom G7 15 Day CGM System doesn't show both a number and trend arrow as well as during the sensor warmup period.
- Don't use if on dialysis or critically ill: The Dexcom G7 15 Day CGM System performance hasn't been evaluated in these populations and sensor readings may be inaccurate.
- Don't ignore broken or detached sensor wires. If this happens, contact 24/7 technical support.
- Before upgrading the smart device or its operating system, check dexcom.com/compatibility. Always update manually and verify correct device settings afterward.
- Let the date and time on the smart device automatically update when traveling across time zones or switch between standard and daylight saving times. Don't manually change the smart device time because the user may not get readings or alerts and it may make the time on the trend screen wrong.
- Not using supplied USB charger and cable may cause the receiver battery to not charge. Don't use if the supplied USB charger or cable is damaged.
## III Device Description
The Dexcom G7 Continuous Glucose Monitoring System (G7 System) and the Dexcom G7 15 Day Continuous Glucose Monitoring System (G7 15 Day System) are interoperable continuous glucose monitoring (CGM) systems intended to continuously measure the glucose in the interstitial fluid, calculate the glucose reading and make this value available to the user. The system is intended for single-patient use at home and requires a prescription. The G7 System is intended for persons with diabetes age 2 and older whereas the G7 15 Day System is intended for persons with diabetes age 18 and older.
The G7 System and G7 15 Day System consist of the following primary components: a wearable, consisting of a G7 sensor and transmitter worn on the body and a display device,
K253737 - Page 4 of 13
{5}
which can be a G7 Mobile Application1 (Mobile App) on an iOS or Android OS smart device or a G7 Receiver (Receiver). The G7 System and G7 15 Day System use Bluetooth Low Energy (BLE) for wireless communication between these components.
To achieve the intended functions and performance of both the G7 System and the G7 15 Day System, one sensor and at least one display device (app or receiver) must be used together. The receiving device acts as the primary display device and the user may choose to use only the app (with optional watch app) or the receiver as the sole display device; they may also choose to use multiple display devices simultaneously. The user must pair the display device(s) with each unique sensor to enable communication and start a sensor session. During an active session, the sensor reports new glucose data to a receiving device every 5 minutes. The receiving device then displays glucose data and provides alerts and information signals to the user. The reportable glucose range for both the G7 System and the G7 15 Day System is 40 g/dL to 400 mg/dL. Glucose values below this range are reported as 'LOW' and glucose values above this range are reported as 'HIGH'. The sensor has an expected wear period of up to 10 days for the G7 System and up to 15 days for the G7 15 Day System, both with an extended 12-hour grace period after the sensor session. The grace period allows additional time for the user to change the sensor at a convenient time.
The proposed G7 System and G7 15 Day System are based on the same principle of operation and fundamental design as their respective predicate devices (K240902 and K243214) with identical user functionality, intended use, indications for use, display devices, sensor, and alerts/alarms.
The G7 System and G7 15 Day System comprise the following subsystems, which are described in comparison to the respective predicate below:
Glucose Sensing Subsystem (GSS): Wearable (Sensor, Transmitter, and Patch) and Applicator The Sensor is a small and flexible wire inserted by the applicator into subcutaneous tissue where it converts glucose into electric current. The Transmitter is pre-connected to the sensor and is worn on the body by the adhesive patch. The Transmitter measures the electric current produced by the sensor and converts these measurements into estimated glucose values (EGV) using an onboard algorithm. The Transmitter is capable of sending glucose data to a display device (mobile app, receiver, or watch app).
Each GSS box also includes an overpatch, which is a general adhesive tape that helps with the adhesion of the wearable to the user's body.
Comparison to Predicate Device (K240902 and K243214)
The proposed G7 System and the proposed G7 15 Day System include an alternate adhesive patch (MT-1000356). There are no changes to the Transmitter, Sensor or Applicator as a result of this premarket notification.
K253737 - Page 5 of 13
{6}
K253737 - Page 6 of 13
# Alternate Adhesive Patch
Comparison of G7 System and G7 15 Day System Predicate Adhesive Patch and Proposed Adhesive Patch
| Item | Subject Device (G7 and G7 15 Day) | Predicate (K240902 and k243214) |
| --- | --- | --- |
| Adhesive Patch supplier and material | Solventum, Spunlace non-woven polyester acrylic adhesive with tackifier | Adhesives Research, NPU Acrylic adhesive |
Mobile Applications: iOS and Android G7 Mobile Application (iOS App and Android App)
The G7 Mobile App is available on both iOS and Android platforms for users with a compatible Bluetooth Low Energy (BLE)-enabled smart device. The Mobile App can be used as a primary display to receive and display glucose readings and trend graphs via BLE communication. The Mobile App provides in-app guidance for the user on how to apply and set up the wearable and how to create a user account. Once the set-up has been completed, the Mobile App receives information from the transmitter and acts as a user interface by indicating the system state (e.g., warm-up period, signal loss, etc.), displaying glucose readings and trend graphs, and alerting the user when glucose levels are outside of a target zone and when specific system states occur. The iOS App (Apple users) can also be used with the Watch Application. When the user downloads the iOS Mobile App from the Apple App Store, the Dexcom Watch App is, by default, automatically downloaded.
# Comparison to Predicate Device (K240902 and K243214)
The proposed G7 System and the proposed G7 15 Day System utilize the same G7 Mobile App, which is the same app as their respective predicate devices (K240902 and K243214). There are no changes to the G7 Mobile App as a result of this pre-market notification.
# Watch Application: G7 Watch App (Watch App)
Users with compatible smartwatches can use the G7 Watch App to view glucose information and notifications. The Watch App is bundled as part of the iOS Mobile App software code and is automatically downloaded at the same time the user downloads their Mobile App from the Apple App Store. The Watch App operates in two modes:
- Standard Smartwatch mode – the Watch App connects to the G7 Mobile App and mirrors the G7 Mobile App UI. The user’s smart phone is required to be within close proximity of the watch to utilize the Watch App, as the Watch App receives all information from the Mobile App.
- Direct to Watch (DTW) mode – the Watch App connects directly with the G7 Transmitter after the sensor session has started and the user has enabled the DTW feature on their iOS App. This means that users are not required to keep their smart phones in close proximity to their Watch App in order to receive EGVs and alerts and can instead receive glucose data on their smartwatch directly from the Transmitter.
# Comparison to Predicate Device (K240902 and K243214)
The proposed G7 System and the proposed G7 15 Day System utilize the same G7 Watch App, which is the same app as their respective predicate devices (K240902 and K243214). There are no changes to the G7 Watch App as a result of this pre-market notification.
{7}
Receiver Subsystem (RVS): G7 Receiver (Receiver)
The Receiver is a small handheld device that can also be used as a primary display to receive and display glucose readings and trend graphs via BLE communication with the Transmitter. Similar to the Mobile App, the Receiver can alert the user when glucose levels are outside of a target zone and when specific system states occur.
Comparison to Predicate Device (K240902 and K243214)
The proposed G7 System and the proposed G7 15 day System use the same Receiver Subsystem (RVS) as their respective predicate devices (K240902 and K243214). There are no changes to the RVS as a result of this pre-market notification.
Interoperability
The GSS is designed to communicate to all display devices simultaneously. The proposed G7 System and the proposed G7 15 Day System, same as their respective predicate devices (K240902 and K243214), are also designed to communicate estimated glucose values, trend, and system information to other compatible electronic interfaces via the following secure wireless connections:
- Wireless communication from the transmitter directly to an interoperable device communicating through the same protocol
- The G7 Mobile App communicates to another app on a single mobile platform
- The G7 Mobile App communicates through the cloud to another software device.
- Dexcom Partner Web APIs: The Dexcom Partner Web APIs enable secure and reliable communication of CGM data to authorized client software intended to receive the data through the cloud. The Partner Web APIs is not intended to be used by automated insulin delivery systems (AID).
IV Substantial Equivalence Information:
A Predicate Device Name(s):
Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System
B Predicate 510(k) Number(s):
K240902, K243214
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K253737 | K240902 | K243214 |
| --- | --- | --- | --- |
| Device Trade Name | Dexcom G7 Continuous Glucose Monitoring (CGM) System
Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System | Dexcom G7 CGM System | Dexcom G7 15 Day CGM System |
K253737 - Page 7 of 13
{8}
K253737 - Page 8 of 13
| General Device Characteristic Similarities | | | |
| --- | --- | --- | --- |
| Intended Use/Indications For Use | Dexcom G7 Continuous Glucose Monitoring (CGM) System:
The Dexcom G7 Continuous Glucose Monitoring (CGM) System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older.
The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.
The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.
Dexcom G7 15 Day Continuous Glucose | The Dexcom G7 Continuous Glucose Monitoring (CGM) System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older.
The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.
The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical | The Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (Dexcom G7 15 Day CGM System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older.
The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments |
{9}
K253737 - Page 9 of 13
| | **Monitoring (CGM) System:**
The Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (Dexcom G7 15 Day CGM System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older.
The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. | devices for the purpose of managing diabetes. | |
| --- | --- | --- | --- |
| Principle of Operation | Amperometric measurement of current proportional to glucose concentration in interstitial fluid via glucose oxidase chemical reaction. | Same | Same |
| Compatibility with connected devices | Compatible with digitally connected devices, including automated insulin dosing (AID) systems | Same | Same |
| Measurement Range | 40 - 400 mg/dL | Same | Same |
| **General Device Characteristic Differences** | | | |
| Adhesive Patch | Solventum, Spunlace non-woven polyester acrylic | Adhesives Research, NPU Acrylic adhesive | Adhesives Research, NPU Acrylic adhesive |
{10}
K253737 - Page 10 of 13
| | adhesive with tackifier | | |
| --- | --- | --- | --- |
| Sensor Shelf-Life | 18-months | 18-months | 12-months |
V Standards/Guidance Documents Referenced:
- ISO 14971:2019, Medical devices - Application of Risk Management to Medical Devices
- ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-12:2021, Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- AAMI TIR 28:2016, Product adoption and process equivalence for ethylene oxide sterilization
- ANSI AAMI ISO 11135:2014, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices - Amendment 1: Revision of Annex E, Single batch release
- ANSI AAMI ISO 11737-2:2019, Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ANSI AAMI ISO 11737-1:2018, Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
- ISO 11138-1:2017, Sterilization of health care products - Biological indicators - Part 1: General requirements
- ISO 10993-2:2022, Biological Evaluation of medical devices - Part 2: Animal welfare requirements
VI Performance Characteristics:
A. Analytical Performance
1. Precision/Reproducibility:
Previously established in K240902 and K243214.
2. Linearity:
Previously established in K240902 and K243214.
3. Analytical Specificity/Interference:
Previously established in K240902 and K243214.
4. Assay Reportable Range:
Previously established in K240902 and K243214.
{11}
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The Dexcom G7 sensor and transmitter have a storage shelf-life of up to 18 months. Shelf life was evaluated at 140°F. Sensors should be stored at 36°-86°F and 10-90% relative humidity. Previously established in K243214 for the Dexcom G7 15 Day CGM System.
6. Detection Limit:
Previously established in K240902 and K243214.
7. Assay Cut-Off:
Not applicable
8. Accuracy (Instrument):
Not applicable
9. Carry-Over:
Not applicable
B. Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable. Accuracy is determined by comparing device values to an FDA cleared laboratory grade glucose analyzer.
2. Matrix Comparison:
Not applicable. Interstitial fluid is the only indicated matrix.
C. Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Sensor Life
A total of 210 sensors were evaluated to determine the percentage of sensors that lasted through the 15-day sensor life. Of the 210 Sensors, 81.9% of sensors lasted the full 15
K253737 - Page 11 of 13
{12}
days. No sensors were excluded from the sensor life evaluation. Three sensors were removed by the sponsor at the last scheduled clinic visit for the subject (censored from Kaplan Meier analysis). 32 sensors (15.2%) had "early sensor shut-off" (ESS), which is when a sensor automatically ends a sensor session as a result of self-diagnostics. One participant withdrew from the study and their sensor was censored on Day 7. 163 sensors survived past 15 days. Survival rates were calculated using the Kaplan Meier method.
An additional analysis of the survival study data was conducted to evaluate the survival rate for only the adhesive (i.e., considering only adhesive-related failures, such as when a sensor falls off). There were a total of 5 failures attributed to the adhesive. The 15-day survival rate for the adhesive (adhesive patch + overpatch) was 97.5%.
Sensor Survival Rate and Adhesive Survival Rate Over Wear Duration
| Day of wear* | Number of sensors (n = 210) | Sensor survival rate(%) | Adhesive survival rate(%) |
| --- | --- | --- | --- |
| 1 | 208 | 99.0 | 99.5 |
| 2 | 208 | 99.0 | 99.5 |
| 3 | 208 | 99.0 | 99.5 |
| 4 | 207 | 99.0 | 99.5 |
| 5 | 207 | 99.0 | 99.5 |
| 6 | 205 | 98.1 | 99.0 |
| 7 | 202 | 97.6 | 98.6 |
| 8 | 200 | 97.6 | 98.6 |
| 9 | 199 | 97.1 | 98.6 |
| 10 | 196 | 95.7 | 98.1 |
| 11 | 194 | 94.7 | 98.1 |
| 12 | 189 | 92.2 | 98.1 |
| 13 | 181 | 88.8 | 97.5 |
| 14 | 172 | 84.9 | 97.5 |
| 15 | 163 | 81.9 | 97.5 |
* Refers to the number of sensors that survived the full wear day
This same study was also used to evaluate the adhesive survival rate of the 10-day version of the G7 sensor in adults and pediatrics using the results observed through day 10. The adhesive survival rate through day 10 was 98.1% in adults and 100% in pediatrics (n=32 subjects, 2 years to 17 years old).
## D. Clinical Cut-Off:
Not applicable.
## E. Expected Values/Reference Range:
Not applicable.
## F. Other Supportive Instrument Performance Characteristics Data
The following supportive performance characteristics were established through nonclinical
K253737 - Page 12 of 13
{13}
testing and are applicable to the G7 and the G7 15 Day Continuous Glucose Monitoring System in this 510(k):
- Biocompatibility
- Sterilization Validation
- Packaging Validation
The following performance characteristics were verified or validated through studies conducted on the subject device, G7 and G7 15 Day CGM Systems:
**Shelf-Life:**
Shelf-Life testing was performed to evaluate the stability of the G7 and G7 15 Day CGM System under real time anticipated storage conditions and supported its useful life to be up to 18 months. The test results for the G7 and G7 15 Day Continuous Glucose Monitoring System met specifications.
**VII** Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
**VIII** Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K253737 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.