Retrospective clinical datasets were used to validate the performance of the AI-based aorto-iliac outer wall segmentation algorithm, comparing automated outputs against a reference standard derived from radiologist contouring.
80 patients with at least 80% having thoracic/abdominal aortic or iliac artery diseases (e.g., aneurysm, dissection, stenosis) and at least 20% treated with stents.; Sample Size: 80 patients; Number of Sites: 11 clinical sites (7 US, 3 Europe, 1 Asia)
Reference standard (average of three US-board certified radiologists' manual contours)
Dice similarity coefficient (DSC), mean surface distance (MSD), Hausdorff distance (HD), and measurement accuracy (Dmin and Dmax).
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Aorto-iliac outer wall segmentation
AI-based algorithm
3D DSC > 0.9, 2D DSC > 0.9, MSD < 1.0 mm, HD < 3.0 mm
3D DSC 0.96, 2D DSC 0.96, MSD 0.57 mm, HD 1.68 mm
—
—
80 patients from 7 clinical sites in the US, 3 European hospitals, and one hospital in Asia.
3 (US-board certified radiologists)
Indications for Use
AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning of endovascular interventions.
Device Story
AV Vascular is a post-processing software application for CTA and MRA datasets; used by clinicians for vascular anatomy visualization, assessment, and quantification. Inputs include medical images; outputs include 2D/3D visualizations (MIP, MPR, cMPR, sMPR, VR), vessel segmentations, and quantitative measurements (diameters, areas, lengths, angles). Features include automatic bone removal, vessel centerline extraction, and ring marker placement. AI-based algorithm performs aorto-iliac outer wall segmentation; model-based algorithms handle subclavian artery extraction and ring marker placement. Used for pre-procedural planning of endovascular interventions. Healthcare providers use outputs to assess vascular pathology and plan procedures, potentially improving clinical decision-making and patient outcomes.
Clinical Evidence
Bench testing only. Performance evaluated on 80 retrospective patient datasets (independent of training data) for aorto-iliac outer wall segmentation. Primary metrics: 3D DSC 0.96 (95% CI 0.96-0.97), MSD 0.57 mm (95% CI 0.485-0.68), HD 1.68 mm (95% CI 1.23-2.08). Secondary metrics (ΔDmin/ΔDmax < 5mm) achieved >98% success. Subclavian artery extraction and ring marker placement also validated. Performance consistent across multi-vendor scanners and clinical subgroups.
Technological Characteristics
Post-processing software for CTA/MRA. Features: 2D/3D visualization, vessel segmentation, measurement tools. Algorithms: AI-based (aorto-iliac wall segmentation) and model-based (subclavian centerline, ring markers). Connectivity: Networked/PACS integration. Platform: Standalone software application.
Indications for Use
Indicated for patients with suspected or diagnosed vascular pathology (e.g., thoracic/abdominal aortic aneurysm, ectasia, dissection, stenosis) requiring visualization, assessment, and quantification of vascular anatomy on CTA/MRA for pre-procedural planning of endovascular interventions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Pie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736)
Philips Medical Systems' The Multimodality Advanced Vessel Analysis (MM AVA) application (K203216)
Reference Devices
Philips Medical Systems' Spectral CT Applications (K150665)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Philips Medical Systems Nederland B.V.
Arbel Shezaf
Regulatory Affairs Manager
Veenpluis 6
Best, 5684 PC
Netherlands
January 22, 2026
Re: K253735
Trade/Device Name: AV Vascular
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: November 24, 2025
Received: November 24, 2025
Dear Arbel Shezaf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253735 - Arbel Shezaf
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253735 - Arbel Shezaf
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Imaging Software Team
Assistant Director
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253735 | ? |
| Please provide the device trade name(s). | | ? |
| AV Vascular | | |
| Please provide your Indications for Use below. | | ? |
| AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning of endovascular interventions. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253735
# 510(k) SUMMARY
## Philips Medical Systems' AV Vascular
**Company's Name and Address**
Philips Medical Systems Nederland B.V.
Veenpluis 6
5684 PC Best
The Netherlands
**Contact Person**
Arbel Shezaf
Regulatory Affairs Manager
MATAM building 34, Haifa, 3100202, Israel
+972544487597
arbel.shezaf@philips.com
**Date**
January 22, 2026
**Device Trade name**
AV Vascular
**Classification Name**
Automated radiological image processing software
**Primary Product Code**
QIH
**Secondary Product Code**
LLZ
**Classification**
21 CFR 892.2050
**Primary Predicate Device**
Pie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736)
**Secondary Predicate Device**
Philips Medical Systems' The Multimodality Advanced Vessel Analysis (MM AVA) application (K203216)
**Reference Device**
Philips Medical Systems' Spectral CT Applications (K150665)
**Device Description**
AV Vascular is a post-processing software application intended for visualization, assessment, and quantification of vessels in computed tomography angiography (CTA) and magnetic resonance angiography (MRA) data with a unified workflow for both modalities.
AV Vascular includes the following functions:
- Advanced visualization: the application provides all relevant views and interactions for CTA and MRA image review: 2D slides, MIP, MPR, curved MPR (cMPR), stretched MPR (sMPR), path-aligned views (cross-sectional and longitudinal MPRs), 3D volume rendering (VR).
- Vessel segmentation: automatic bone removal and vessel segmentation for head/neck and body CTA data, automatic vessel centerline, lumen and outer wall extraction and labeling for the main branches of the vascular anatomy in head/neck and body CTA data, semi-automatic and manual creation of vessel centerline and lumen for CTA and MRA data, interactive two-point vessel centerline extraction and single-point centerline extension.
- Vessel inspection: enable inspection of an entire vessel using the cMPR or sMPR views as well as inspection of a vessel locally using vessel-aligned views (cross-sectional and longitudinal MPRs) by selecting a position along a vessel of interest.
- Measurements: ability to create and save measurements of vessel and lumen inner and outer diameters and area, as well as vessel length and angle measurements.
- Measurements and tools that specifically support pre-procedural planning: manual and automatic ring marker placement for specific anatomical locations, length measurements of the longest and shortest curve along the aortic lumen contour, angle measurements of aortic branches in clock position style, saving viewing angles in C-arm notation, and configurable templated
{5}
measurements.
- Saving and export: saving and export of batch series and customizable reports.
## Indications for Use
AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning of endovascular interventions.
## Verification and Validation
Software verification and validation activities were performed to verify that the software meets the product requirements.
### Verification and Validation
Verification was performed according to the verification plan.
Product requirement specifications were tested and found to meet the requirements.
### Validation
Validation was performed according to the validation plan. User requirement specifications were tested and found to meet the requirements. The validation results provide evidence that the product meets its intended use and user requirements.
## Performance Data
Performance testing was conducted for subclavian artery extraction on body cases (model-based algorithm), aorto-iliac wall segmentation (AI-based algorithm) and ring marker placement (model-based algorithm). The datasets represent the targeted US patient population in terms of key demographics, clinical and technical characteristics.
### Subclavian Artery Centerline Extraction
CT angiography scans of the body were collected for testing. The automatically extracted subclavian artery centerlines were compared with a reference standard.
### Ring Marker Placement
CT angiography scans of the body were collected for testing, with the majority coming from patients with aortic or iliac aneurysms. Automatically placed ring markers at the ostia of the renal and superior mesenteric arteries were evaluated by expert vascular interventionalists for their clinical acceptability without any adjustment.
### Aorto-iliac Outer Wall Segmentation
To validate the performance of the aorto-iliac wall segmentation, body CT angiography scans were collected from 80 patients.
The aorto-iliac outer wall contour automatically generated by the AI-based algorithm in the subject device was compared with a reference standard in terms of Dice similarity coefficient (DSC), mean surface distance (MSD), and Hausdorff distance (HD). The measurement accuracy (minimum and maximum diameter, $D_{\min}$ and $D_{\max}$) was also evaluated.
- Testing Data, Demographics, & Independence from Training Data The demographic information of the 80 patients included for performance testing is:
| Demographics | Number of patients (percentage) |
| --- | --- |
| Geographics | North America: 58 (72.5%) |
| | Europe: 3 (3.75%) |
| | Asia: 19 (23.75%) |
| | None of these testing data were from the same clinical sites which provided the training data. |
| Sex | Male: 59 (73.75%) |
| | Female: 21 (26.25%) |
| Age (years) | 21-50: 2 (2.50%) |
| | 51-70: 31 (38.75%) |
{6}
The performance testing data were collected retrospectively from 7 clinical sites in the US, 3 European hospitals, and one hospital in Asia. All performance testing datasets were acquired from clinical sites distinct from those which provided the algorithm training data, and the algorithm developers had no access to the testing data, ensuring complete independence between training and testing data. At least $80\%$ of the included patients had thoracic and/or abdominal aortic diseases and/or iliac artery diseases (e.g., thoracic/abdominal aortic aneurysm, ectasia, dissection, and stenosis). At least $20\%$ of patients had been treated with stents. The composed testing data were considered representative of the intended patient population in the US.
Reference Standard (i.e., Truthing Process)
For each of the CT angiographic images, three US-board certified radiologists (i.e., truthers) independently performed manual contouring of the outer wall along the aorta and iliac arteries on cross-sectional planes. The order of the cases was randomized for each truther to reduce bias. After quality control, these three aortic and iliac arterial outer wall contours were averaged as the reference standard contour.
Performance & Acceptance Criteria
For primary performance metrics (2D and 3D DSC, MSD, and HD) the performance statistics and acceptance criteria are:
| Metrics | | Acceptance criteria | Mean (98.75% confidence intervals) |
| --- | --- | --- | --- |
| DSC | 3D | DSC > 0.9 | 0.96 (0.96, 0.97) |
| | 2D | DSC > 0.9 | 0.96 (0.95, 0.96) |
| MSD [mm] | | MSD < 1.0 mm | 0.57 (0.485, 0.68) |
| HD [mm] | | HD < 3.0 mm | 1.68 (1.23, 2.08) |
For secondary performance metrics $(\mathsf{D}_{\min}$ and $\mathsf{D}_{\max})$ , the performance statistics and acceptance criteria are:
| Metrics | Acceptance criteria | Proportion (95% confidence intervals) |
| --- | --- | --- |
| ΔDmin | >95% |ΔDmin|<5 mm | 98.8% (98.3-99.2%) |
| ΔDmax | >95% |ΔDmax|<5 mm | 98.5% (97.9-98.9%) |
Clinical Subgroups & Confounders
The performance of automatic aorto-iliac outer wall segmentation was evaluated internally using independent testing CT angiographic data acquired with multi-vendor scanners (GE Healthcare, Siemens, Philips, Toshiba/Canon) from patients of which at least $80\%$ had aortic and/or iliac artery aneurysm, dissection, stents, calcification, stenosis, and/or plaque. Primary performance metrics (DSC, MSD, and HD) were analyzed across subgroups (defined by sex, age group, data source, vessel, pathology, reconstruction method, slice thickness, and manufacturer). The algorithm's performance was consistent across multiple subgroups and comparable to the inter-truther variability. The outer wall segmentation accuracy, however, might be reduced in cases of giant aneurysms (diameter $>10~\mathrm{cm}$ ), which is rare in the intended patient population.
{7}
Therefore, the performance of all three algorithms is state of the art and on par with the expectation.
# Substantial Equivalence
AV Vascular is as safe and effective as its predicate device. Both devices have similar intended use and indications for use, technological characteristics, and principles of operation. The minor technological differences between AV Vascular and its predicate device raise no new issues of safety or effectiveness.
Performance data demonstrate that Philips Medical Systems' AV Vascular is safe and effective. Thus, AV Vascular is substantially equivalent to Pie Medical Imaging's 3mensio. Table 1 below summarizes the substantive feature/technological similarities and differences between the subject and predicate devices.
Table 1. Substantial Equivalence
| Comparison Feature | Subject Device Philips Medical Systems' AV Vascular | Predicate Device Pie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) | Secondary Predicate Device Philips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Reference Device Philips Medical Systems' Spectral CT Applications (K150665) | Comparison between the subject and predicate devices (identical/ different) |
| --- | --- | --- | --- | --- | --- |
| Device Class | Class II | Class II | Class II | Class II | Identical |
| Classification Panel | Radiology | Radiology | Radiology | Radiology | Identical |
| Product Code | QIH, LLZ | LLZ | JAK | JAK, LLZ | Similar to primary predicate device |
| Regulation Description | Medical image management and processing system | Medical image management and processing system | Computed tomography x-ray system | Computed tomography x-ray system Medical image management and processing system | Identical to primary predicate device |
| Regulation Number | 892.2050 | 892.2050 | 892.1750 | 892.1750 892.2050 | Identical to primary predicate device |
| Indications for Use | AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy | 3mensio Workstation is a standalone software for medical image analysis intended for advanced | The Multimodality Advanced Vessel Analysis (MM AVA) application is intended for visualization, assessment and | The Philips Spectral CT Applications support viewing and analysis of images at energies selected | Different |
| | | | spectral CT | | |
{8}
| Comparison Feature | Subject Device | Predicate Device | Secondary Predicate Device | Reference Device | Comparison between the subject and predicate devices (identical/different) |
| --- | --- | --- | --- | --- | --- |
| | on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning for endovascular interventions. | visualization and quantitative analysis for diagnostic and/or for assistance during treatment in the field of cardiology or radiology by means of enabling visualization and measurement of structures of the heart and vessels for:
• Pre-operational planning and sizing for cardiovascular interventions and surgery
• Postoperative evaluation
• Support of clinical diagnosis by quantifying dimensions in coronary arteries
• Support of clinical diagnosis by quantifying calcifications (calcium scoring) in the coronary arteries
To facilitate the above, the 3mensio Workstation provides general functionality such as: | quantification of vascular datasets. | from the available spectrum in order to provide information about the chemical composition of the body materials and/or contrast agents. The Spectral CT Applications provide for the quantification and graphical display of attenuation, material density, and effective atomic number. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures. The Spectral enhanced Advanced Vessel Analysis (sAVA) application is intended to assist clinicians in viewing and evaluating CT images, for the inspection of contrast-enhanced vessels. | |
5
{9}
| Comparison Feature | Subject Device | Predicate Device | Secondary Predicate Device | Reference Device | Comparison between the subject and predicate devices (identical/different) |
| --- | --- | --- | --- | --- | --- |
| | Philips Medical Systems' AV Vascular | Pie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) | Philips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Philips Medical Systems' Spectral CT Applications (K150665) | |
| | | • Segmentation of cardiovascular structures
• Automatic and manual centerline detection
• Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched CMRP, Slabbing, MIP, AIP, MinIP
• Measurement and annotation tools Reporting tools | | The Spectral enhanced Comprehensive Cardiac Analysis (sCCA) application is intended to assist clinicians in viewing and evaluating cardiovascular CT images.
The Spectral enhanced Tumor Tracking (sTT) application is intended to assist clinicians in viewing and evaluating CT images, for the inspection of tumors. | |
| Clinical Characteristics | | | | | |
| Intended body part | Head and neck, body, peripherals | Heart and vessels | Head and neck, body, peripherals | Head and neck, body, peripherals | Identical to secondary predicate device |
| Type of scans | CTA and MRA | CT Angiography | CTA and MRA | CT Angiography | Identical to secondary predicate device |
| Technological features | | | | | |
| Spectral capabilities | Yes | No | No | Yes | Identical to reference device |
| Subclavian Artery | Automatic vessel centerline extraction of the head, neck and | Automatic vessel centerline extraction | Automatic vessel centerline | Automatic vessel centerline | Different |
{10}
| Comparison Feature | Subject Device | Predicate Device | Secondary Predicate Device | Reference Device | Comparison between the subject and predicate devices (identical/different) |
| --- | --- | --- | --- | --- | --- |
| Centerline Extraction | body including shoulder region | | Philips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Philips Medical Systems' Spectral CT Applications (K150665) | |
| Aorto-iliac Outer Wall Segmentation | Automatic non-AI based vessel contouring of major vessels.
Automatic AI-based vessel contouring specifically for the aorta and common and external iliac arteries. | No | Automatic non-AI based vessel contouring of major vessels, including the aorta and common and external iliac arteries. | Automatic non-AI based vessel contouring of major vessels, including the aorta and common and external iliac arteries. | Different |
| Review Marker tool | Review Marker tool including manual ring markers, and non-AI based automatic ring markers for the ostia of the SMA, LRA and RRA | Review Marker tool including manual ring markers | Review Marker tool | Review Marker tool | Different |
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.