iPredict-DR is intended for use by healthcare providers to automatically detect more than mild diabetic retinopathy in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. iPredict-DR is indicated for use with iCare's DRSPlus color fundus camera.
Device Story
iPredict-DR is a cloud-based software application for computer-aided detection of more than mild diabetic retinopathy (mtmDR). It takes macula and disc-centered color fundus images (45° field of view) as input, captured by the iCare DRSPlus retinal camera. The system uses deep-learning algorithms to analyze images for DR-suggestive features; it classifies results as 'mtmDR detected' or 'mtmDR not detected' (referable vs. non-referable). Used in primary care or endocrinology clinics by healthcare providers with minimal training. The client-side UI allows image upload; images are processed on a secure server. The system outputs a PDF report with clinical recommendations. If image quality is insufficient, the system prompts for re-acquisition or recommends referral. The device aids clinical decision-making by identifying patients requiring referral to an eye care professional, potentially benefiting patients through early detection of sight-threatening disease.
Clinical Evidence
Prospective clinical trial (N=886 enrolled, N=871 for primary analysis). Primary endpoints: sensitivity 85.58% (95% CI: 80.26%-89.65%) and specificity 91.77% (95% CI: 89.41%-93.64%). PPV 77.31%, NPV 95.10%. Imageability rate 99.08%. Ground truth established by three masked expert graders using ETDRS severity levels with majority vote/adjudication. Precision study (N=50) demonstrated repeatability (94.67% agreement) and reproducibility (91% agreement) across operator-camera pairings.
Technological Characteristics
Software-as-a-Medical-Device (SaMD) utilizing deep-learning neural network algorithms. Inputs: 45° field-of-view color fundus images. Connectivity: Cloud-based server with web-based UI client. Cybersecurity: Compliant with FDA 2023 guidance and UL 2900-1:2017. Software validation: IEC 62304 compliant. Risk management: ISO 14971:2019. Deployment: Web-based client-server architecture.
Indications for Use
Indicated for adults diagnosed with diabetes, without prior diabetic retinopathy diagnosis, to detect more than mild diabetic retinopathy using iCare's DRSPlus color fundus camera.
Regulatory Classification
Identification
A retinal diagnostic software device is a prescription software device that incorporates an adaptive algorithm to evaluate ophthalmic images for diagnostic screening to identify retinal diseases or conditions.
Special Controls
In combination with the general controls of the FD&C Act, the retinal diagnostic software device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Software verification and validation documentation, based on a comprehensive hazard analysis, must fulfill the following:
(i) Software documentation must provide a full characterization of technical parameters of the software, including algorithm(s).
(ii) Software documentation must describe the expected impact of applicable image acquisition hardware characteristics on performance and associated minimum specifications.
(iii) Software documentation must include a cybersecurity vulnerability and management process to assure software functionality.
(iv) Software documentation must include mitigation measures to manage failure of any subsystem components with respect to incorrect patient reports and operator failures.
(2) Clinical performance data supporting the indications for use must be provided, including the following:
(i) Clinical performance testing must evaluate sensitivity, specificity, positive predictive value, and negative predictive value for each endpoint reported for the indicated disease or condition across the range of available device outcomes.
(ii) Clinical performance testing must evaluate performance under anticipated conditions of use.
(iii) Statistical methods must include the following:
(A) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification.
(B) Statistical analysis must provide confidence intervals for each performance metric.
(iv) Clinical data must evaluate the variability in output performance due to both the user and the image acquisition device used.
(3) A training program with instructions on how to acquire and process quality images must be provided.
(4) Human factors validation testing that evaluates the effect of the training program on user performance must be provided.
(5) A protocol must be developed that describes the level of change in device technical specifications that could significantly affect the safety or effectiveness of the device.
(6) Labeling must include:
(i) Instructions for use, including a description of how to obtain quality images and how device performance is affected by user interaction and user training;
(ii) The type of imaging data used, what the device outputs to the user, and whether the output is qualitative or quantitative;
(iii) Warnings regarding image acquisition factors that affect image quality;
(iv) Warnings regarding interpretation of the provided outcomes, including:
(A) A warning that the device is not to be used to screen for the presence of diseases or conditions beyond its indicated uses;
(B) A warning that the device provides a screening diagnosis only and that it is critical that the patient be advised to receive followup care; and
(C) A warning that the device does not treat the screened disease;
(v) A summary of the clinical performance of the device for each output, with confidence intervals; and
(vi) A summary of the clinical performance testing conducted with the device, including a description of the patient population and clinical environment under which it was evaluated.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
2 July, 2026
iHealthScreen, Inc.
Alauddin Bhuiyan
Chief Executive Officer
132-02 89th Ave. Suite 214
Richmond Hill, New York 11418
Re: K253704
Trade/Device Name: iPredict-DR
Regulation Number: 21 CFR 886.1100
Regulation Name: Retinal diagnostic software device
Regulatory Class: Class II
Product Code: PIB
Dated: November 15, 2025
Received: November 24, 2025
Dear Alauddin Bhuiyan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253704 - Alauddin Bhuiyan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253704 - Alauddin Bhuiyan
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alexander Beylin -S Date: 2026.07.02 11:09:19 -04'00'
for CAPT Bradley Cunningham, MSE, RAC
Acting Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253704 | ? |
| Please provide the device trade name(s). | | ? |
| iPredict-DR | | |
| Please provide your Indications for Use below. | | ? |
| iPredict-DR is intended for use by healthcare providers to automatically detect more than mild diabetic retinopathy in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. iPredict-DR is indicated for use with iCare's DRSPlus color fundus camera. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253704
# 510(k) Summary – iPredict-DR
K253704
| Date Prepared | June 29, 2026 |
| --- | --- |
| Official Contact (Owner) | iHealthScreen, Inc. 132-02 89th Ave, Suite 214, Richmond Hill, New York 11418, USA |
| Contact person: | Alauddin Bhuiyan, Chief Executive Officer |
| Email: | bhuiyan@ihealthscreen.org |
| Phone: | +1-718 926 9000 |
| Proprietary Name | iPredict-DR |
| --- | --- |
| Common/Usual Name | Retinal diagnostic software |
| Classification Name | Retinal diagnostic software device |
| Classification Reference | 21 CFR 886.1100 |
| Product Code | PIB |
| Regulatory Class | Class II |
Predicate Device(s):
| Trade name of the predicate device: | IDx-DR |
| --- | --- |
| 510(k) Number: | K203629 |
| Classification Reference | 21 CFR 886.1100 |
| Classification Name: | Retinal diagnostic software device |
| Product Code | PIB |
| 510(k) submitter/holder: | Digital Diagnostics Inc.; 2300 Oakdale Blvd., Coralville, IA 52241 |
Intended Use / Indications for Use
iPredict-DR is intended for use by healthcare providers to automatically detect more than mild diabetic retinopathy in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy.
iPredict-DR is indicated for use with iCare's DRSPlus color fundus camera.
Prescription Use Only: Federal law restricts this device for sale by or on the order of a physician.
{5}
K253704
## Device description
iPredict-DR software application is intended to perform computer-aided detection of more than mild diabetic retinopathy (mtmDR) in digital retinal images of diabetic patients 22 years or older, to be used by a healthcare provider (user) with basic training, in primary care or endocrinology settings. The Application hosts a software module developed using Deep-Learning algorithms to analyze digital retinal images of diagnosed diabetic patients for features suggestive of diabetic retinopathy (DR). iPredict-DR outputs a screening result of “more than mild DR detected” or “more than mild DR not detected” to the user. iPredict-DR software is designed and validated to perform at clinically acceptable sensitivity and specificity in the detection of diabetic retinopathy (DR).
iPredict-DR’s results and respective retinal images of a patient can be viewed in the form of a report made available through the application. iPredict-DR is limited to the analysis of retinal color imaging data and creating a screening result on the detection of diabetic retinopathy only.
## Supported hardware (retinal camera)
iPredict-DR software application is indicated for use with iCare’s DRSPlus retinal camera (iCare Inc. K192113). DRSPlus is available commercially and was used to validate iPredict-DR as part of this regulatory clearance application. This retinal camera is intended to be used for taking digital images of the retina of the human eye. DRSPlus camera, its hardware, and its built-in imaging functionality is out of scope for the iPredict-DR software application.
## iPredict-DR architecture and modes of operation
iPredict-DR software system is intended to be used with color fundus images obtained using DRSPlus retinal camera with 45 degrees field of view. The camera is connected to a computer where iPredict-DR client software is available for use. iPredict-DR client software is a web-based user interface (UI) module/component that is the user facing side of the application for data and image collection, which connects to the AI based processing modules in the server to process retinal images and evaluate the patient’s mtmDR disease status. The UI allows uploading of images by healthcare providers as drag/drop or using “upload image” option. Other information, including patient name and date of birth can be added as well. A computer with internet connection is required to access iPredict-DR. The components of iPredict-DR are described in Figure 1.
For each patient, iPredict-DR analyzes left and right eye’s retinal images and determines whether more than mild DR (mtmDR) in the individual eye is detected. If either of the eyes has mtmDR, the patient is considered at “referable DR” stage and recommended to be referred to an eye care professional. The mtmDR output is determined according to the Early Treatment Diabetic Retinopathy Study (ETDRS) grading protocol for no-DR, mild, moderate, and severe DR, from which iPredict-DR classifies the output of “mtmDR detected or referable DR (includes moderate or severe DR)” or “mtmDR not detected or non-referable DR (includes no or mild DR)”. A report is generated with the “mtmDR detected” or “mtmDR not detected” output and with the respective recommendation of “referable DR” or “non-referable DR”.
{6}
K253704

Figure 1. iPredict-DR system
iPredict-DR Client: It is a cloud-based software application's web/user interface (UI) running on a computer with internet connection, and connected to the fundus camera, located at the customer site. Using this UI, the patient's color retinal images and patient data are transferred to the DR diagnostic module or the AI server for DR diagnosis, and the UI receives the results back as a pdf report with a recommendation of referable or non-referable DR. The images and data are saved to the database server after processing. If images cannot be analyzed due to inadequate image quality, quality feedback is provided to help the operator acquire high quality images and successfully obtain a result after resubmission of images. If the image quality is poor and automatic evaluation is not feasible, the device provides an output that a referral to ophthalmologist is recommended "Refer to eye care professional due to insufficient quality image".
- iPredict-DR-server: iPredict-DR-server contains a webserver front-end that securely handles incoming requests, a database that stores customer information, and a logging system that records information about each transaction through iPredict-DR-service. iPredict-DR webserver is also responsible for device cybersecurity. The report generation module in the web server provides a pdf report of the patient's status of the disease and returns the report to user-interface with a recommendation of referable or non-referable DR status of the patient.
- iPredict-DR AI module: The AI module (deep learning algorithms and decision modules) processes the images in server and returns the results as "mtmDR detected or mtmDR not detected" and "referable or non-referable DR" to the webserver and to the report generation module.
### Target Environment
The target environment for iPredict-DR is a primary care or endocrinology clinical setting with standard room, lighting, and other clinical environmental conditions. The targeted users are healthcare providers – clinician or healthcare assistant with self-training (maximum half an hour of training).
{7}
K253704
# Predicate Device Description
# Predicate Device(s): IDx-DR
IDx-DR (K203629) is an artificial intelligence-supported retinal image analysis system indicated for use by healthcare providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. IDx-DR is indicated for use with the Topcon NW400 retinal camera.
iPredict-DR has the same intended use and same indications for use (IFU) as the predicate IDx-DR device. A comprehensive summary of Substantial Equivalence in Table 1 below, details the similarities and differences between the subject device, iPredict-DR, and the predicate device, IDX-DR; and demonstrates the substantial equivalence between the subject device and predicate device in terms of the intended use and indications for use, technological characteristics, safety and effectiveness, clinical performance, non-clinical testing, and intended patient population.
An overview of the technological and functional characteristics between iPredict-DR and IDx-DR (K203629) is presented below. The subject device and predicate device have the same technological principle as AI-based technology utilized to analyze specific pathologic features from retinal fundus images. Specifically, both iPredict-DR and IDX-DR screen for DR in diabetic patients, and both devices are used for the screening of retinal diseases using a proprietary algorithm that was trained on extensive datasets and further validated based on prospective data. The proposed new device was evaluated using the same methods utilized for the cleared predicate (and consistent with the special controls for the device type), and the results demonstrate substantial equivalence to the predicate.
Table 1: Comparison with the predicate device
| Features and Characteristics | IDx-DR (Predicate device) 510(k) K203629 | iPredict-DR (Subject device) | Discussion of Differences and Comments |
| --- | --- | --- | --- |
| Regulation Number/Name | 21 CFR 886.1100 / Retinal diagnostic software device | 21 CFR 886.1100 / Retinal diagnostic software device | Equivalent |
| Product code | PIB, Class II | PIB, Class II | Equivalent |
{8}
K253704
| Intended Use / Indications for use | IDx-DR is intended for use by healthcare providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. IDx-DR is indicated for use with the Topcon NW400. | iPredict-DR is intended for use by healthcare providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. iPredict-DR is indicated for use with iCare's DRSPlus retinal camera. | Both devices have the same intended use per 21 CFR 886.1100 / Retinal diagnostic software device. They are substantially equivalent. |
| --- | --- | --- | --- |
| Prescription/ over-the-counter use | Prescription Use | Prescription Use | Equivalent |
| Technological principle | Artificial Intelligence software as a medical device that performs automated analysis of retinal images using Deep Learning neural network algorithms | Artificial Intelligence software as a medical device that performs automated analysis of retinal images using Deep Learning neural network algorithms | Equivalent |
| Mode of Action | Image processing software device intended to detect more than mild diabetic retinopathy in retinal images of patients diagnosed with diabetes | Image processing software device intended to detect more than mild diabetic retinopathy in retinal images of patients diagnosed with diabetes | Both devices are intended to detect DR-specific findings on acquired retinal images per 21 CFR 886.1100 / Retinal diagnostic software device. They are substantially equivalent. |
| Inputs | Macula and disc centered color fundus images with 45° field of view, 2 per eye | Macula and disc centered color fundus images with 45° field of view, 2 per eye | Equivalent |
| Outputs | Detection of diabetic retinopathy | Detection of diabetic retinopathy | |
{9}
K253704
| | More than mild diabetic retinopathy (mtmDR): mtmDR not detected, mtmDR detected, or insufficient quality | More than mild diabetic retinopathy (mtmDR): mtmDR not detected, mtmDR detected, or insufficient quality | They are substantially equivalent. |
| --- | --- | --- | --- |
| End Users | Healthcare providers | Healthcare providers | Equivalent |
| Location of use (primary) | Primary care settings, healthcare clinics, hospitals | Primary care settings, healthcare clinics, hospitals | Equivalent |
| Deployment | Computer-based Client software Application | Computer-based Client software Application | Equivalent |
| Architecture | Client software (user facing) transfers images to and receives results from Analysis through Web Server. | Client software (user facing) transfers images to and receives results from Analysis through Web Server. | Equivalent |
| Software Level of Concern | Major level of concern | Enhanced level of documentation to be submitted to the FDA | iPredict-DR software's level of concern is described as the "Enhanced level documentation" as per the latest FDA regulations and the Risk analysis performed. This does not present any additional risk and only indicates compliance with the updated regulatory language used by the FDA to translate the "level of concern" into "level of documentation". They are substantially equivalent. |
{10}
K253704
| Indicated Cameras | Topcon NW400 | iCare's DRSPlus | The legally marketed cameras specified for use are being used to capture macula and disc centered with 45° field of view; for both subject device (iPredict-DR) and predicate device (IDX-DR). The clinical performance data supports the use of iPredict-DR with the indicated cameras. Both cameras produce retinal images. They are substantially equivalent. |
| --- | --- | --- | --- |
| Workflow | Labeling (the Quick Reference Guide) guides the user through the image acquisition workflow and submission of the exam. The graphical user interface includes onscreen prompts to guide the user through the image acquisition workflow one image at a time and submission of the exam. | Labeling (the Quick Reference Guide) guides the user through the image acquisition workflow and submission of the exam. The graphical user interface includes onscreen prompts to guide the user through the image acquisition workflow one image at a time and submission of the exam. | Equivalent |
### Conclusion on substantial equivalence:
Based upon the technical comparison information presented, it is concluded that iPredict-DR is substantially equivalent to the previously cleared predicate device, IDx-DR. Both devices have the same intended use with regard to retinal disease diagnosis based on the analysis of digital retinal images. The indications for use are the same in that both devices import images from ophthalmic devices for analysis, detect retinal features suggestive of retinal disease, and display disease diagnosis results.
{11}
K253704
## Labeling
The User Manual or Instructions for Use (IFU) and the Installation Instructions for iPredict-DR are provided in the document intended to be used by healthcare providers at primary care or endocrinology clinics; for adult patients diagnosed with diabetes. A soft copy of the user manual is included with iPredict-DR web application. A user is eligible to request a paper/printed copy of the user manual/IFU at no additional cost, by contacting iHealthScreen. The user manual includes a description of the intended patient population, retinal imaging process, iPredict-DR functioning and processing steps, the factors that may impact iPredict-DR screening results, complete set of warnings, precautions, and contraindications. These warnings were found to be appropriate by other similar legally marketed devices, including the predicate device.
## Software
The iPredict-DR software version 4.0 was determined to require an Enhanced Level Documentation; as defined in the FDA guidance document "Content of Premarket Submissions for Device Software Functions, JUNE 2023". Software validation testing was performed as per the FDA guidelines for software validation and IEC 62304 standard requirements which showed compliance. The software verification and validation were performed for unit, integration, and system level testing. iPredict-DR passed the tests as intended (i.e. met all specifications).
The software hazard analysis was performed as part of system risk and safety analysis. The hazards of the software and the incorrect operations of the system by the user that could affect the software's appropriate functioning, were handled as part of the system hazard analysis. iPredict-DR risk control measures have been designed and implemented adequately to mitigate all identified hazards to acceptable levels. iPredict-DR analyzed and implemented the cybersecurity requirements to be compliant according to the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
## Clinical testing
### Study design and methods
In a pivotal prospective clinical trial, 922 participants were enrolled, of whom 905 were eligible for recruitment, and 886 subjects completed the study procedures. For the primary sensitivity/specificity/PPV/NPV analysis, the number of subjects was 871.
After enrollment of subjects meeting the inclusion criteria and obtaining informed consent, retinal photos of subjects' non-dilated eyes (both eyes) were taken with the DRSPlus Digital Retinography System (CenterVue, Inc), at each of the investigational sites. For ground truth imaging, dilated 4 W-D stereoscopic and OCT imaging fundus photos were taken with the Topcon Maestro2.
All images were graded by three certified, masked expert graders at the Wisconsin Reading Center (WRC). Each eye was independently evaluated (i.e., graders were blinded to each other's gradings) using ETDRS severity levels. The final ground truth was determined by a majority vote
{12}
K253704
(2 of 3 graders). If all three grades differed, adjudication was performed.
The WRC grading was used to determine the clinical reference standard for subject eyes as follows:
Clinical reference standard for more than mild DR not detected or non-referable DR
- ETDRS level 10 and 20
- Center-involving diabetic macular edema (CI-DME) absent
Clinical reference standard for more than mild DR (mtmDR) or referable DR
- ETDRS Level 20 or above (excluding subjects with ETDRS level 60 with prior panretinal photocoagulation)
- CI-DME present
If a subjects clinical reference standard images were graded ETDRS level 90 (ungradable) there were three possible outcomes in the primary outcome analysis:
1. ETDRS 90 (considered positive for analysis) and device output positive: Agreement
2. ETDRS 90 (considered positive for analysis) and device output negative: Disagreement
3. ETDRS 90 and device output ungradable: Exclude image from full analysis set/ungradable
Study demographics
Table 2. Demographic and baseline characteristics of the study population, N = 886 (Note that binary performance endpoints use N=871)
| SL. No. | Characteristic | Category / statistic | All subjects (N = 886) |
| --- | --- | --- | --- |
| 1 | Age (years) | Mean ± SD | 62.9 ± 12.18 |
| | | Median | 64 |
| | | Range | 22 – 92 |
| | | 22-44 years, n (%) | 70 (7.9%) |
| | | 45-60 years, n (%) | 269 (30.4%) |
| | | 61-75 years, n (%) | 424 (47.9%) |
| | | 76-92 years, n (%) | 123 (13.9%) |
| 2 | Sex / Gender | Female, n (%) | 421 (47.5%) |
| | | Male, n (%) | 465 (52.5%) |
| | | Asian | 72 (8.1%) |
| | | Black or African American | 192 (21.7%) |
| | | White | 610 (68.8%) |
{13}
K253704
| 3 | Race, n (%) | Other (incl. American Indian / Alaska Native and Native Hawaiian / Pacific Islander) | 12 (1.4%) |
| --- | --- | --- | --- |
| 4 | Ethnicity, n (%) | Hispanic or Latino | 272 (30.7%) |
| | | Not Hispanic / Latino | 614 (69.3%) |
| 5 | HbA1c (%) | Mean ± SD | 7.81 ± 1.62 |
| | | Median | 7.2 |
| | | Range | 6.5 – 15.5 |
| | | < 10%, n (%) | 794 (89.6%) |
| | | ≥ 10%, n (%) | 92 (10.4%) |
| 6 | Duration of diabetes (years) | Mean ± SD | 13.1 ± 10.20 |
| | | Median | 10 |
| | | Range | 0 – 55 |
| | | < 10 years, n (%) | 369 (41.6%) |
| | | ≥ 10 years, n (%) | 517 (58.4%) |
| 7 | Diabetes type | Type 2 – 843 subjects Type 1 – 42 subjects Type 3c - 1 subject | 886 (100%) |
Summary of clinical study results: Table 3 below describes the validation study results for primary endpoints.
Table 3: Primary-analysis performance (subject-level, N = 871)
| Metric | Estimate (x / n) | 95% Wilson CI |
| --- | --- | --- |
| Sensitivity | 85.58% (184 / 215) | [80.26%, 89.65%] |
| Specificity | 91.77% (602 / 656) | [89.41%, 93.64%] |
| Positive predictive value (PPV) ¹ | 77.31% (184 / 238) | [71.58%, 82.17%] |
| Negative predictive value (NPV) | 95.10% (602 / 633) | [93.13%, 96.53%] |
| Imageability (1 – IIQ rate) | 99.08% (863 / 871) | [98.20%, 99.53%] |
| Prevalence (mtmDR+ in ITS) | 24.68% (215 / 871) | [21.94%, 27.65%] |
IIQ - Insufficient Image Quality.
{14}
K253704
# Clinical validation results summary:
The iPredict-DR algorithm performed with sensitivity of 85.58% and specificity of 91.77%. Considering that CRS ungradable images were considered CRS positive in the iPredict-DR primary analysis, the results of this clinical study support a determination of substantial equivalence between iPredict-DR and IDx-DR.
# Precision Study (Repeatability and Reproducibility)
iHealthScreen conducted a precision study (repeatability and reproducibility) using data from a cohort comprised of pivotal study subjects. A total of 50 subjects were recruited, with 24 subjects with "mtmDR detected" and 26 subjects with "mtmDR not detected." The "mtmDR not detected" category included normal and mild DR cases, while the "mtmDR detected" category encompassed moderate, severe, and proliferative DR cases, as well as DME. The precision study evaluated the iPredict-DR results when retinal photography was repeated using the DRSPlus camera with different operator-camera pairs. There were 3 operator-camera pairings consisting of 3 different operators using 2 different camera units of the same model (each operator operated a given camera unit). The study objective was to demonstrate that the results are repeatable and reliable. Study success required at least 80% overall agreement for both inter- and intra-operator use.
- Repeatability (Intra-operator variability) analysis: Repeatability was assessed to determine whether the same operator produced consistent iPredict-DR results when imaging the same eye twice using the same operator-camera pairing. Each operator-camera pairing consisted of one operator assigned to one of two identical DRSplus camera units. For every eye, each operator-camera pairing acquired two repeat image sets (Attempt 1 and Attempt 2). Each image set consisted of a macula-centered image and a disc-centered image, which together form a single iPredict-DR result for that eye. Across the study, this produced a total of 600 repeatability pairs (3 operators × 50 subjects × 2 eyes × 2 cameras). The observed intra-operator overall agreement was 94.67%, therefore meeting the 80% criterion for success. Additionally, average agreement for positive results (mtmDR detected) was 94.55%, and average agreement for negative results (mtmDR not detected) was 95.48%.
- Reproducibility (Inter-operator variability) analysis: Reproducibility was assessed to determine whether different operators, using different but identical DRSplus camera units, produced consistent iPredict-DR results for the same eye under routine imaging conditions. Each operator-camera pairing acquired one image set (Attempt 1) for every eye. Each image set consisted of a macula-centered and a disc-centered image, which together form a single combined iPredict-DR result per eye. Inter-operator reproducibility was evaluated by comparing the three iPredict-DR outputs (one from each operator-camera pairing) for the same eye using only the first attempt from each operator. Agreement was quantified using pairwise 3×3 tables (two-operator reproducibility) and using the full three-rater 3×3×3 agreement table (three-operator reproducibility). Overall agreement was 91%, therefore meeting the 80% criterion for success. In addition, average agreement for positive results (mtmDR detected) was 94.22% %, and average agreement for negative results (mtmDR not detected) was 94.65%.
{15}
K253704
# Non-clinical (Bench) testing
1. Verification and Validation: The verification and validation testing reports summarize the evaluation of the iPredict-DR Diabetic Retinopathy Screening Software. The verification and validation testing concluded that the iPredict-DR Software meets the requirements, is suitable for its intended use, and can be released. The testing report demonstrates that the algorithm developed to determine the quality of retinal images and to detect the presence or absence of diabetic retinopathy (DR) in retinal images, meets or exceeds the sensitivity and specificity requirements.
All testing demonstrated that the software met the predetermined pass/fail criteria. All reported issues were minor with no adverse impact to the safety and efficacy of this application. The test failures reported do not impact safety or efficacy based on the rationale provided for each test failure. All protocol deviations were evaluated in accordance with the iPredict-DR Software Development Plan, and it was determined that the intent of the modified test steps was not affected by the changes. The protocol deviations did not exclude any aspect of the requirements being verified and the functional requirements traced to each protocol are still valid.
2. Risk analysis and management: We reviewed the risk management outputs and have concluded that the risk management effort complies with the iPredict-DR Risk Management Plan, and in doing so, has complied with the requirements of ISO 14971:2019. Based on review of the relevant documentation, we have concluded that the overall residual risk posed by iPredict-DR is acceptable. The likelihood of harm is considered generally low, and the controls for usability are believed to support safe, proper, and effective use. Adequate risk control measures were designed and implemented to mitigate all identified hazards to acceptable levels.
3. Cybersecurity assessment report: This document summarizes the results of the cybersecurity assessment and mitigation efforts undertaken for the iPredict-DR system and aligns with the FDA's 2023 Guidance titled "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, Guidance for Industry and Food and Drug Administration Staff" and UL 2900-1:2017 (Updated in year 2025). This document addresses cybersecurity within the iPredict-DR application and the cloud server to which it communicates, considering data security, data confidentiality, data integrity, denial of service attacks, and malware. It provides a complete assessment of the system design and threat model, requirements verification, malformed input testing, structured penetration testing, vulnerability testing, data/attack targets, system interfaces/attack vectors, system weaknesses, risk mitigations, and post-market management.
This cybersecurity analysis provides sufficient evidence of safe and effective software performance and ensures that the data security and privacy is maintained according to the industry standards. The "Software Maintenance Plan" lists the procedures involved with maintaining the software and managing cybersecurity vulnerabilities following the release, as part of the lifecycle management.
4. Human factors testing: The human factors data support the safety and effectiveness of the iPredict-DR. The usability of iPredict-DR was analyzed through a human factor validation study, for the understanding and implementation of the instructions in the user manual, as well as the usability of iPredict-DR's user interface in submitting the retinal images for processing and receiving the diagnostic results and reports. The study was
{16}
K253704
conducted in the setting in which the device is intended to be used, to validate that the iPredict-DR application is safe and effective for the intended users and uses, in the intended use environment.
**Conclusion on non-clinical testing:** The performance data for the non-clinical or bench tests, including software validation, risk analysis, cybersecurity, and human factor validation (summative testing), demonstrate that the iPredict-DR device is safe and effective for use, and substantially equivalent to the predicate device.
**Conclusion:** iPredict-DR is substantially equivalent to the predicate device, IDx-DR. iPredict-DR has the same intended use and equivalent indications for use. The technological characteristics are similar. The technological differences between iPredict-DR and its predicate device raise no new issues of safety or effectiveness. Performance data support the substantial equivalence of iPredict-DR to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.