Intuitive Endoluminal Gastrointestinal System (IG1000)

K253702 · Intuitive Surgical, Inc. · FDS · Aug 7, 2026 · Gastroenterology, Urology

Device Facts

Record IDK253702
Device NameIntuitive Endoluminal Gastrointestinal System (IG1000)
ApplicantIntuitive Surgical, Inc.
Product CodeFDS · Gastroenterology, Urology
Decision DateAug 7, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Intuitive Endoluminal Gastrointestinal (GI) System is intended for the robotic-assisted control of the Robotic GI Endoscope during visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach and duodenum.

Device Story

Physician-controlled electro-mechanical system for robotic-assisted navigation of GI endoscope; inputs include user commands via Motion Controller (scroll wheel, track ball, roll ring, buttons, foot pedal) and endoscopic visual data; system consists of System Cart, Motion Controller, Fluids System, and Robotic GI Endoscope (RGE); RGE features integrated vision, illumination, shape sensing, and working channels for 3rd party tools; Fluids System manages suction, irrigation, insufflation, and lens wash; physician operates system from System Cart; output is real-time endoscopic visualization on monitor; assists in navigation and control of endoscope tip to facilitate diagnostic and therapeutic procedures in upper GI tract; benefits include enhanced control and visualization for physician during complex endoluminal procedures.

Clinical Evidence

Prospective, single-center, single-arm confirmatory study (n=28) for diagnostic, therapeutic, and surveillance procedures. Subjects followed for 30 days. Primary endpoints: safety and performance. Results: 100% procedure completion rate, zero conversions to manual endoscopy, and zero device-related serious adverse events through 30-day follow-up. Study confirmed ability to safely access upper GI anatomical locations.

Technological Characteristics

Robotic-assisted electro-mechanical endoscope system. 13.2mm OD flexible endoscope (1030mm length). Features integrated vision, illumination, shape sensing, and dual working channels (3.7mm/2.8mm). Bending angles ≥200° in four directions. Connectivity: System Cart with monitor and Motion Controller. Standards: ISO 10993 (biocompatibility), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-18 (endoscopic), IEC 80601-2-77 (robotic surgical), IEC 60825-1 (laser).

Indications for Use

Indicated for patients requiring visualization of the upper digestive tract for observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION August 7, 2026 Intuitive Surgical, Inc. Einav Yemini Regulatory Technical Lead 1266 Kifer Rd. Sunnyvale, California 94086 Re: K253702 Trade/Device Name: Intuitive Endoluminal Gastrointestinal System (IG1000) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDS, FET Dated: July 6, 2026 Received: July 7, 2026 Dear Einav Yemini: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K253702 - Einav Yemini Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K253702 - Einav Yemini Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **SIVAKAMI VENKATACHALAM** **-S** *for* Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253702 | ? | | Please provide the device trade name(s). | | ? | | Intuitive Endoluminal Gastrointestinal System (IG1000) | | | | Please provide your Indications for Use below. | | ? | | The Intuitive Endoluminal Gastrointestinal (GI) System is intended for the robotic-assisted control of the Robotic GI Endoscope during visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach and duodenum. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} K253702 Page 1 of 4 Intuitive Endoluminal Gastrointestinal System Traditional 510(k) 1 ### 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) 1990 and 21 CFR 807.92. ### Submitter | 510(k) Owner: | Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 | | --- | --- | | Contact: | Einav Yemini Regulatory Technical Lead Intuitive Surgical, Inc. 1266 Kifer Road, Sunnyvale, CA 94086 Ph: 206-300-0910, Email: einav.yemini2@intusurg.com | | Date Prepared: | June 29, 2026 | ### Subject Device Information | Manufacturer Name: | Intuitive Surgical, Inc. | | --- | --- | | Trade Name: | Intuitive Endoluminal Gastrointestinal System (IG1000) | | Common Name: | Endoscope and Accessories | | Classification: | Class II, 21 CFR 876.1500 | | Product Code: | FDS, FET | | Review Panel: | Gastroenterology/Urology | ### Predicate Device FUJIFILM Medical Systems U.S.A., Inc. Endoscope Model EI-740D/S (K210162) ### Reference Device 1. Intuitive Surgical, Inc. Ion Endoluminal System, Model IF1000 (K182188) 2. Olympus Medical Systems Gastrointestinal Videoscope GIF-1100 (K222584) ### Device Description The Intuitive Endoluminal Gastrointestinal (GI) System (IG1000) is a physician-controlled, electro-mechanical system intended to assist qualified physicians to navigate a GI endoscope and 3rd party endoscopic tools (placed or used through the working channels of the endoscope) in the upper GI tract using endoscopic visualization of the esophagus, stomach, and duodenum for observation, diagnosis, and therapeutic procedures. It consists of a System Cart, Motion Controller, Fluids System, Robotic GI Endoscope (RGE), and accessories. The System Cart contains the Flexible Instrument Manipulator, electronics for the follower portion of the servomechanism, and a monitor. It allows the user to navigate the Robotic GI endoscope using the Motion INTUITIVE © 2026 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders. Intuitive Surgical Confidential and Proprietary. {6} Intuitive Endoluminal Gastrointestinal System Traditional 510(k) K253702 Page 2 of 4 2 Controller and a leader/follower control system. For optimal viewing, the physician can position and orient the System Cart monitor in both the vertical and horizontal axes. The Fluids System resides on top of the System Cart and controls and directs the flow of gases and fluids for suction, irrigation, insufflation, and lens wash from their sources and through the endoscope in response to commands from the user through the Motion Controller. The Motion Controller is the user input device for the system. It provides controls to command insertion, retraction, and articulation of the RGE via a scroll wheel, track ball, and roll ring. It also has buttons and a foot pedal to control fluidics functions (lens wash, insufflation, suction, and irrigation). Lastly, it includes a passive button to enter passive mode and an emergency stop button to enter a Safe Fault Reaction Logic (FRL) state. The Robotic GI Endoscope is a 13.2mm OD flexible endoscope with a working length of 1030 mm. It contains integrated vision, illumination light fibers, shape sensing, 3.7 and 2.8 mm working channels compatible with third party endoscopic tools, and channels for irrigation, lens wash, insufflation, and suction. ### Intended Use/Indications for Use The Intuitive Endoluminal Gastrointestinal (GI) System is intended for the robotic-assisted control of the Robotic GI Endoscope during visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach and duodenum. ### Comparison of Technological Characteristics with the Predicate Device The subject device (IG1000 system) and predicate device (FUJIFILM EI-740D/S) are both intended for visualization during gastrointestinal procedures. They share overall the same functions of providing access, navigation, visualization, allowing control of the Endoscope tip, providing a working channel for endoscopic instruments, and enabling use of insufflation, irrigation, lens wash and suction. The differences between the subject and predicate devices are in their mode of operation and endoscope controls, software-enabled functionality and slight differences in dimensions, articulation and visualization. A summary of the technological characteristics of the subject device compared to the predicate device is provided below Table 0-1: Comparison of the Subject and Predicate Devices | Description | Subject Device | Predicate Device (K210162) | | --- | --- | --- | | Manufacturer | Intuitive Surgical Inc. | FUJIFILM Medical Systems U.S.A., Inc. | | Trade Name | Intuitive Endoluminal Gastrointestinal System (IG1000) | Endoscope Model EI-740D/S | | Common Name | Endoscope and Accessories | Identical to the subject device | | Regulation Number | 876.1500 | Identical to the subject device | INTUITIVE © 2026 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders. Intuitive Surgical Confidential and Proprietary. {7} K253702 Page 3 of 4 Intuitive Endoluminal Gastrointestinal System Traditional 510(k) 3 | Description | Subject Device | Predicate Device (K210162) | | --- | --- | --- | | Product Code | FDS, FET | FDS, FAM | | Intended Use / Indications for Use | The Intuitive Endoluminal Gastrointestinal (GI) System is intended for the robotic-assisted control of the Robotic GI Endoscope during visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach and duodenum. | FUJIFILM Endoscope Model EI-740D/S is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This device is also intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and sigmoid colon. | | Endoscope Working Length | 1030 mm | Identical to the subject device | | Distal End Diameter | 15 mm | 12.8 mm | | Flexible Portion Diameter | 13.2 mm | 12.8 mm | | Working Channels | Two- 3.7 & 2.8 mm | Two- 3.7 & 3.2 mm | | Bending Angles | Minimum of 200° in all four directions | Up 210° Down 90° Left 100° Right 100° | | Viewing Direction | 0° | Identical to the subject device | | Field of View | 120° | 140° | | Observation Range | 3mm - 100mm | Identical to the subject device | ### Performance Data The following performance testing was conducted to demonstrate substantial equivalence to the predicate device. Biocompatibility of the IG1000 system was evaluated in accordance with ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2023, ISO 10993-11:2018, ISO 10993-12:2021, ISO 10993-17:2023, and ISO 10993-18:2020, along with FDA Guidance “Use of International Standard ISO 10993-1, ‘Biological INTUITIVE © 2026 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders. Intuitive Surgical Confidential and Proprietary. {8} K253702 Page 4 of 4 Intuitive Endoluminal Gastrointestinal System Traditional 510(k) 4 evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”, issued on September 8, 2023. **Software verification and validation** included unit, subsystem integration, and system level testing. The software testing results demonstrate that the IG1000 System Software meets design specifications and user needs. Software documentation has been provided per FDA Guidance “Content of Premarket Submissions for Device Software Functions,” issued on June 14, 2023. **Cybersecurity testing** included cybersecurity verification and validation, evaluated per the FDA’s final guidance, “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”, issued on September 27, 2023. The cybersecurity verification and validation test results demonstrate the adequacy of the implemented cybersecurity controls. **Electrical safety and electromagnetic compatibility** testing was conducted using third-party labs in accordance with the current versions of IEC 60601-1 (basic safety and essential performance), IEC 60601-1-2 (electromagnetic disturbances), IEC 60601-2-18 (endoscopic equipment), IEC 80601-2-77 (robotically assisted surgical equipment) and IEC 60825-1 (laser safety). **Bench testing** demonstrates that the IG1000 System’s design output meets the design input requirements and that the system performs as intended. The testing conducted consisted of dimensional measurements, mechanical and functional verification, summative usability testing, and reprocessing, sterility, and shelf-life testing. **Animal and cadaver testing** was performed for the evaluation of the safety and performance of the IG1000 system and software. Testing consisted of Design Validation and performance and safety equivalence testing, compared to the predicate device. The Design Validation studies evaluated the IG1000 System’s intended use, safety, reachability, and performance. The performance and safety equivalence study evaluated the performance and safety of the IG1000 System compared to the predicate device, during its intended use of visualization of the upper digestive tract to support observation, diagnosis, and therapeutic procedures. **Clinical data** was collected via a prospective, single center, single arm confirmatory clinical study. The study enrolled 28 subjects for indicated diagnostic, therapeutic, and surveillance procedures. Subjects were followed for 30 days. All study procedures were completed as planned with no conversions to manual endoscopy and no device-related serious adverse events through 30 days post-procedure. Results from the study demonstrate that the IG1000 System can safely access key anatomical locations throughout the Upper GI tract to enable diagnostic, therapeutic and/or surveillance procedures and confirm the safety and performance profile established through pre-clinical testing. ## Conclusion Based on the intended use, indications for use, technological characteristics, and performance data, the Intuitive Endoluminal Gastrointestinal (GI) System (Model IG1000) is substantially equivalent to its predicate device, the FUJIFILM Endoscope Model EI-740D/S. INTUITIVE © 2026 Intuitive Surgical, Inc. All rights reserved. Product names are trademarks or registered trademarks of their respective holders. Intuitive Surgical Confidential and Proprietary.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...