K253686 · Ge Healthcare Japan Corporation · JAK · Mar 23, 2026 · Radiology
Device Facts
Record ID
K253686
Device Name
True Definition DL
Applicant
Ge Healthcare Japan Corporation
Product Code
JAK · Radiology
Decision Date
Mar 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
High contrast spatial resolution
Deep learning-based CT reconstruction algorithm
—
—
—
—
Reader study of sample clinical data
>1 (US board-certified radiologists)
Indications for Use
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement.
Device Story
True Definition DL (TDDL) is a deep learning-based CT reconstruction algorithm; functions as an optional, user-selectable enhancement for GE Revolution CT/Apex family systems. Input: raw CT data; Output: reconstructed images with enhanced in-plane and cross-plane spatial resolution. Operated by radiology staff in clinical settings; integrated into routine reconstruction workflows. Provides three user-selectable enhancement strengths. Benefits: improved visualization of fine structural details in lung and bone imaging (e.g., inner ear, spine) without significant noise increase or algorithm-induced artifacts. Healthcare providers use enhanced images to improve diagnostic confidence in assessing high-contrast anatomical structures.
Clinical Evidence
Clinical evidence consists of a reader study using clinical data processed by TDDL. US board-certified radiologists evaluated images, confirming improved high-contrast spatial resolution in bone (inner ears, spines) and lung exams, increased diagnostic confidence, and absence of algorithm-induced artifacts or spurious structures. Bench testing on phantoms using industry-standard IQ metrics (MTF) demonstrated improved resolution without significant noise increase compared to traditional kernels.
Technological Characteristics
Deep learning-based CT reconstruction algorithm; software-only option for GE Revolution CT/Apex systems. Reconstruction matrix: 1024 x 1024. Workflow: integrated into existing console reconstruction chain with three selectable strengths (Low, Medium, High). Developed under ISO 13485 and 21 CFR 820 design controls.
Indications for Use
Indicated for high contrast spatial resolution enhancement for bone and lung imaging. May be used for patients of all ages.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 23, 2026
GE Healthcare Japan Corporation
% Yonghui Han
Senior Regulatory Affairs Leader
7-127, Asahigaoka, 4-Chome
Hino-Shi, TOKYO, 191-8503
JAPAN
Re: K253686
Trade/Device Name: True Definition DL
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed tomography x-ray system
Regulatory Class: Class II
Product Code: JAK
Dated: February 20, 2026
Received: February 20, 2026
Dear Yonghui Han:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K253686 - Yonghui Han
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253686 - Yonghui Han
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253686 | |
| Device Name True Definition DL | |
| Indications for Use (Describe) True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement for bone and lung imaging. True Definition DL may be used for patients of all ages. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
510(k) Premarket Notification Submission - True Definition DL
GE HealthCare
K253686
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
Date: February 17, 2026
Submitter: GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha, Wisconsin 53188
Primary Contact: Yonghui Han
Sr. Regulatory Affairs Leader
Phone: 1 (262) 225-1914
Email: Yonghui.han@gehealthcare.com
Secondary Contacts: Helen Peng
Sr. Regulatory Affairs Director
Phone: +1 (262) 424-8222
Email: hong.peng@gehealthcare.com.
Device Trade Name: True Definition DL
Device Classification: Class II
Regulation Number/ Product Code: 21 CFR 892.1750 Computed tomography x-ray system / JAK
# Predicate Device Information
Device Name: Deep Learning Image Reconstruction
Manufacturer: GE Medical Systems, LLC
510(k) Number: K213999
Regulation Number/ Product Code: 21 CFR 892.1750 Computed tomography x-ray system / JAK
# Device Description
Computed Tomography (CT) is an indispensable imaging modality in clinical diagnostics due to its ability to provide detailed cross-sectional images of anatomical structures. However, achieving high spatial resolution remains a persistent challenge, particularly in applications requiring the visualization of fine structural details such as in inner auditory canal imaging, vascular studies, lung imaging, and bone microarchitecture analysis. Reconstruction techniques that attempt to boost spatial resolution typically also would amplify the noise which often result in tradeoffs to be made between resolution and noise.
{6}
510(k) Premarket Notification Submission - True Definition DL
GE HealthCare
As part of the continuous innovation to solve the above challenges in CT imaging, GEHC developed a deep learning-based CT reconstruction algorithm specifically designed for high contrast lung and bone imaging, which is the subject of this premarket notification. This reconstruction algorithm, marketed under the name True Definition DL (TDDL), is an additional user-selectable recon option specifically designed for lung and bone imaging. It aims to enhance spatial resolution for both in-plane and cross-plane directions. This algorithm is incorporated into the reconstruction chain of the Revolution CT /Apex Family CT systems including Revolution CT/Revolution CT ES, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select all cleared under (K213715) and Revolution Vibe (K250941).
True Definition DL offers three strengths that the user can choose depending on enhancement preferences. The benefits provided by this enhancement include improved spatial resolution measured by MTF.
## Intended Use
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement.
## Indications for Use
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement for bone and lung imaging. True Definition DL may be used for patients of all ages.
## Technological Characteristics
Since the True Definition DL is a recon option for the Revolution CT/Apex system, it utilizes the same hardware and software platform for reconstruction. The table below summarizes the substantive feature/technological differences between the predicate device, and the proposed device. The changes described below do not change the fundamental technology of the predicate device or host system and do not raise any new questions of safety and effectiveness.
{7}
510(k) Premarket Notification Submission - True Definition DL
GE HealthCare
| Specification | Predicate Device
Deep Learning Image Reconstruction
(K213999) | Proposed Device
True Definition DL |
| --- | --- | --- |
| Patient Population | Patients of all ages | Same |
| Reconstruction Matrix | 512 x 512
1024 x 1024 | 1024 x 1024 |
| Anatomies | Head, Body (including bone and lung), Cardiac | Bone and lung |
| Hardware | DLIR is a recon option implemented on the operator console for Revolution Apex systems. | Same |
| Clinical workflow | Built into routine clinical workflow with 3 strengths to select from (Low, Medium, High) | Same |
## Determination of Substantial Equivalence
True Definition DL has successfully completed the design control testing per GE HealthCare’s quality system. No additional hazards were identified, and no unexpected test results were observed. True Definition DL was designed under GE HealthCare’s QMS per the Quality System Regulations of 21CFR 820 and ISO 13485. GEHC believes that the successfully completed design verification & validation, as well as the extensive bench testing is sufficient for FDA’s substantial equivalence determination.
The following quality assurance measures have been applied to the development of the system:
- Requirement Definition
- Risk Analysis and Control
- Technical Design Reviews
- Formal Design Reviews
- Software Development Lifecycle
- Code Review
- Software Unit Implementation
- Software Integrations and Integration Testing
- System Testing
- Safety Testing (Verification)
- Image Performance Testing (Verification)
- Simulating Use Testing (Validation)
- Software Release
The testing and results did not raise different questions of safety and effectiveness than associated with predicate device.
## Summary of Non-Clinical Testing
A range of bench testing on phantoms was performed to measure the performance of True Definition DL. The tests include well established test methods using industry-standard IQ metrics. The test data demonstrated acceptable overall image quality and substantially equivalent performance of True Definition DL to the predicate measured on the Revolution CT/Apex scanner’s traditional metrics.
{8}
510(k) Premarket Notification Submission - True Definition DL
GE HealthCare
The result of the bench testing showed that True Definition DL improves the in-plane and cross-plane resolution for body (lung and bone) scans as well as bony structures in the head acquired from Revolution CT/Apex without significantly increasing the image noise, in comparison with a sharper resolution kernel.
To evaluate the algorithm robustness of the True Definition DL algorithm, GEHC conducted qualitative and quantitative test utilizing the entire image reconstruction pipeline. The testing also included assessment of hallucination risks and their mitigations with the results indicating that True Definition DL consistently enhanced anatomical edges without introducing spurious structures, preserved noise texture under realistic dose conditions, and maintained strong correlation with input and target images across all frequency bands.
## Summary of Clinical Testing
The clinical testing was carried out in the form of a reader study of sample clinical data processed using True Definition DL software. The images were evaluated by US board-certified radiologists.
The result of this reader study validated that True Definition DL provides diagnostic value by improving high contrast spatial resolution in routine bone (including inner ears and spines) and lung exams and improves confidence in the assessment of high contrast structures in these types of images.
In addition, the readers confirmed that the images enhanced by the subject device do not present algorithm-induced artifacts or enhancements not supported by underlying anatomies.
## Substantial Equivalence
True Definition DL was developed under GEHC's quality system. Design verification, along with bench testing and the clinical reader study provided in this submission demonstrates that True Definition DL is substantially equivalent to the predicate device. Verification & Validation testing and risk management processes did not raise any new questions of safety or effectiveness, or unexpected results, or adverse effects stemming from the changes. GE HealthCare believes that True Definition DL is substantially equivalent to the predicate, hence it is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.