Zeta Navigation System

K253663 · Zeta Surgical, Inc. · HAW · May 1, 2026 · Neurology

Device Facts

Record IDK253663
Device NameZeta Navigation System
ApplicantZeta Surgical, Inc.
Product CodeHAW · Neurology
Decision DateMay 1, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Zeta Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is indicated only for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites. The Zeta Stylet is intended to be used in conjunction only with the Zeta Navigation System to provide stereotaxic guidance for the placement and operation of catheters, shunts or similar products, and anatomy registration during cranial procedures. The Zeta Bolt is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures performed in conjunction only with the Zeta Navigation System using preoperative MR and/or CT imaging. These procedures include brain biopsies.

Device Story

Zeta Navigation System provides intraoperative image-guided planning and navigation for cranial procedures. Inputs: 3D preoperative CT/MRI scans via USB, CD-ROM, PACS, or Zeta Cloud. Operation: System uses optical tracking of wireless, unpowered instruments and machine vision/structured light for pinless, markerless patient registration. Software (Zeta OS) performs image segmentation, surgical planning, and real-time instrument tracking relative to patient anatomy. Output: Visual display of instrument position and trajectory on a touchscreen monitor. Used in ORs and interventional suites by neurosurgeons/staff. Benefits: Enables precise instrument positioning for biopsies, catheter, or shunt placement without requiring rigid head fixation. Healthcare providers use real-time visual feedback to guide surgical instruments, improving accuracy during neurological interventions.

Clinical Evidence

No clinical data. Bench testing only, including accuracy testing, design validation, and human factors testing. Biocompatibility verified per ISO 10993-1:2018. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Sterilization validation for single-use instruments achieved SAL 10^-6.

Technological Characteristics

Stereotaxic image guidance system; optical tracking of wireless, unpowered instruments. Materials: Titanium (Grade 23) for Stylet/Bolt. Connectivity: USB, CD-ROM, PACS, Zeta Cloud. Software: Major level of concern. Registration: Structured light and machine vision (pinless/markerless). Sterilization: Steam sterilization for single-use instruments. Standards: IEC 60601-1, IEC 62304, IEC 60601-1-2, IEC 60601-1-6.

Indications for Use

Indicated for cranial surgery in patients where rigid anatomical reference is identifiable; does not require rigid patient fixation or navigated instrument guide fixation. Used for stereotaxic guidance during neurosurgical procedures, including catheter/shunt placement and brain biopsies.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION May 1, 2026 Zeta Surgical, Inc. Hieu Le Mau, COO 280 Summer St., Floor 7 Boston, Massachusetts 02210 Re: K253663 Trade/Device Name: Zeta Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: November 20, 2025 Received: November 20, 2025 Dear Hieu Le Mau: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253663 - Hieu Le Mau Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K253663 - Hieu Le Mau Page 3 See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **JAIME RABEN -S** Jaime Raben, PhD Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253663 | | | Device Name Zeta Navigation System | | | Indications for Use (Describe) Zeta Navigation System The Zeta Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is indicated only for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites. Zeta Stylet The Zeta Stylet is intended to be used in conjunction only with the Zeta Navigation System to provide stereotaxic guidance for the placement and operation of catheters, shunts or similar products, and anatomy registration during cranial procedures. Zeta Bolt The Zeta Bolt is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures performed in conjunction only with the Zeta Navigation System using preoperative MR and/or CT imaging. These procedures include brain biopsies. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Submission: K253663 In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for the Zeta Navigation System, the Zeta Stylet and the Zeta Bolt is provided below. 1. SUBMITTER Applicant: Zeta Surgical Inc. 280 Summer Street, Floor 7 Boston, MA 02210 Contact: Hieu Le Mau Chief Operating Officer Zeta Surgical Inc. 280 Summer Street, Floor 7 Boston, MA 02210 +1 (857) 225-4138 hieu.lemau@zetasurgical.com Prepared on: 4/30/2026 2. DEVICE Device Trade Name: Zeta Navigation System/Zeta OS Zeta Stylet Zeta Bolt Device Common Name: Neurological Stereotaxic Instrument Classification Name: Stereotaxic instrument, 21 CFR 882.4560 Regulatory Class: Class II Product Code: HAW 3. PREDICATE DEVICE Zeta Navigation System [K242351] Medtronic Passive Planar Blunt Probe [K050438] Cranial Reducing Tubes [K162604] 4. DEVICE DESCRIPTION The Zeta Navigation System is a stereotaxic, image guided planning and intraoperative guidance system enabling computer-assisted cranial interventional procedures. The system assists surgeons with the precise positioning of surgical instruments relative to patient anatomy by displaying the position of navigated surgical instruments relative to 3D preoperative medical scans. Zeta OS is the software application of the Zeta Navigation System. The Zeta Stylet is a guiding stylet to be used in conjunction with the Zeta Navigation System for navigated placement of catheters, shunts or similar devices during cranial procedures. The instrument is single-use and not intended for reprocessing or reuse. 510(K) SUMMARY {5} The Zeta Bolt is an instrument to be used in conjunction with the Zeta Navigation System for navigated placement of surgical instruments, including brain biopsy needle, using preoperative MR and/or CT imaging. The instrument is single-use and not intended for reprocessing or reuse. # 5. INDICATIONS FOR USE The Zeta Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is indicated only for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites. The Zeta Stylet is intended to be used in conjunction only with the Zeta Navigation System to provide stereotaxic guidance for the placement and operation of catheters, shunts or similar products, and anatomy registration during cranial procedures. The Zeta Bolt is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures performed in conjunction only with the Zeta Navigation System using preoperative MR and/or CT imaging. These procedures include brain biopsies. # 6. SUBSTANTIAL EQUIVALENCE The table below compares the key technological features of the Zeta Navigation System to the predicate device (Zeta Navigation System, K242351). | | PREDICATE DEVICE | SUBJECT DEVICE | | --- | --- | --- | | 510(K) NUMBER | K242351 | K253663 | | DEVICE | Zeta Cranial Navigation System | Zeta Navigation System | | CLASSIFICATION REGULATION | 21 CFR 882.4560 | 21 CFR 882.4560 | | PRODUCT CODE | HAW | HAW | | CLASSIFICATION | Class II | Class II | | INDICATIONS FOR USE | The Zeta Cranial Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is indicated only for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites. | The Zeta Cranial Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is indicated only for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites. | 510(K) SUMMARY {6} | | PREDICATE DEVICE | SUBJECT DEVICE | | --- | --- | --- | | INTENDED USER | Neurosurgeons, neurosurgical procedure staff | Neurosurgeons, neurosurgical procedure staff | | INTENDED USE ENVIRONMENT | Neurosurgical operating room, facilities equipped for interventional cranial procedures | Neurosurgical operating room, facilities equipped for interventional cranial procedures | | ANATOMICAL SITE | Head | Head | | PRINCIPLE OF OPERATION | Preoperative image upload, Surgical planning, Patient registration, and Instrument guidance | Preoperative image upload, Surgical planning, Patient registration, and Instrument guidance | | ACCEPTED IMAGING MODALITIES | 3D DICOM CT and MRI | 3D DICOM CT and MRI | | DATA INPUT | USB, CD-ROM, and PACS | USB, CD-ROM, PACS, and Zeta Cloud | | INSTRUMENT TRACKING TECHNOLOGY | Optical tracking of wireless, unpowered instruments | Optical tracking of wireless, unpowered instruments | | INSTRUMENT COMPATIBILITY | Brainlab Disposable Stylet, IZI Disposable Passive Blunt Probe | Brainlab Disposable Stylet, IZI Disposable Passive Blunt Probe, Zeta Stylet, Zeta Bolt | | REGISTRATION TECHNOLOGY | Structured light and machine vision | Structured light and machine vision | | GUIDANCE TECHNOLOGY | Image based, provides real-time display of instrument position relative to patient anatomy | Image based, provides real-time display of instrument position relative to patient anatomy | | MAJOR SYSTEM COMPONENTS | Cart, Sensor head, Sensor head positioning arm, Monitor, Monitor positioning arm, Tracked instruments, Software | Cart, Sensor head, Sensor head positioning arm, Monitor, Monitor positioning arm, Tracked instruments, Software | | USER INTERFACE | Non-sterile touchscreen monitor that is covered with a transparent sterile cover during the procedure | Non-sterile touchscreen monitor that is covered with a transparent sterile cover during the procedure | | WORKFLOW COMPONENTS | Upload, Segmentation, Planning, Staging/Positioning, Registration, Instrument Calibration, Navigation | Upload, Segmentation, Planning, Staging/Positioning, Registration, Instrument Calibration, Navigation | | PLANNING FEATURES | Multiple target point selection, multiple instrument selection | Multiple target point selection, multiple instrument selection, cloud pre-planning app | | NAVIGATION FEATURES | Target point projection, instrument rendering, instrument extended trajectory, physical distance measurement, multiple perspectives | Target point projection, instrument rendering, instrument extended trajectory, physical distance measurement, multiple perspectives | | REGISTRATION METHOD | Automatic, pinless, and markerless | Automatic, pinless, and markerless | 510(K) SUMMARY {7} The table below compares the key technological features of the Zeta Stylet to the predicate device (Medtronic Passive Planar Blunt Probe, K050438). | | PREDICATE DEVICE | SUBJECT DEVICE | | --- | --- | --- | | 510(K) NUMBER | K050438 | K253663 | | DEVICE | Medtronic Passive Planar Blunt Probe | Zeta Stylet | | DEVICE TYPE | Slender, rod-like, hand-held manual surgical instrument (metal probe/stylet) | Slender, rod-like, hand-held surgical stylet (metal instru-ment) | | INTENDED USE | Aid for precisely locating anatomical structures in open or percutaneous neurosurgical procedures, for any medi-cal condition where stereotactic surgery is appropriate | The Zeta Stylet is intended to be used in conjunction only with the Zeta Navigation System to provide stereotaxic guidance for the placement and operation of catheters, shunts or similar products, and anatomy registration during cranial procedures | | FDA CLASSIFICATION | Class II | Class II | | REGULATION | 21 CFR 882.4560 — Neurological Stereotaxic Instrument | 21 CFR 882.4560 — Neurological Stereotaxic Instrument | | PRODUCT CODE | HAW | HAW | | PRIMARY MATERIAL | Stainless steel | Titanium (grade 23) | | NAVIGATION TECHNOLOGY | Passive optical (reflective sphere markers) — compatible with Medtronic StealthStation electromagnetic/optical navigation systems | Passive optical — compatible with Zeta Navigation Sys-tem | | SINGLE-USE OR REUSABLE | Reusable | Single use | | STERILITY | Non-sterile | Non-sterile | 510(K) SUMMARY {8} The table below compares the key technological features of the Zeta Bolt to the predicate device (Cranial Reducing Tubes, K162604). | | PREDICATE DEVICE | SUBJECT DEVICE | | --- | --- | --- | | 510(K) NUMBER | K162604 | K253663 | | DEVICE | Cranial Reducing Tubes | Zeta Bolt | | INTENDED USE | The Cranial Reducing Tubes are intended to maintain the position of instruments or devices placed using stereotactic guidance during planning and operation of neurological procedures performed in conjunction with the use of the Medtronic StealthStation System Image Guided Workstation. The devices were not tested for MR compatibility and are not intended for use in an MRI environment. | The Zeta Bolt is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures performed in conjunction only with the Zeta Navigation System using preoperative MR and/or CT imaging. These procedures include brain biopsies. | | FDA CLASSIFICATION | Class II | Class II | | REGULATION | 21 CFR 882.4560 — Neurological Stereotaxic Instrument | 21 CFR 882.4560 — Neurological Stereotaxic Instrument | | PRODUCT CODE | HAW | HAW | | PRIMARY MATERIAL | Stainless steel | Titanium (grade 23) | | STERILITY | Non-sterile | Non-sterile | | REUSABLE/SINGLE USE | Reusable | Single Use | | GUIDE FOR INTRU-MENTS AND DEVICES | Yes | Yes | | TRAJECTORY ALIGNED BY IMAGE GUIDED WORK-STATIONS | Yes | Yes | | ABLE TO INTERFACE STE-REOTACTIC NAVIGATION TOOL FOR GUIDANCE | Yes, optical tracking | Yes, optical tracking | | NAVIGATIONAL PERFOR-MANCE | Mean Error ≤2.00 mm and ≤2.00° | Mean Error ≤2.00 mm and ≤2.00° | | NAVIGATION SYSTEM IN-TEGRATION | Medtronic StealthStation System Image Guided Work-station | ZETA Navigation System | # 7. PERFORMANCE DATA # 7.1. Biocompatibility Testing The biocompatibility of all patient contact materials of the Zeta Stylet and Zeta Bolt was verified according to ISO 10993-1:2018 and FDA guidance on the use of ISO 10993-1, September 2023. No new issues of safety or effectiveness were raised. 510(K) SUMMARY {9} 510(K) SUMMARY PAGE 6 OF 7 ## 7.2. Electrical safety and electromagnetic compatibility (EMC) The Zeta Navigation System was tested in accordance with the following standards: - IEC 60601-1:2005 Ed. 3 + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical electrical equipment: Part 1: General requirements for basic safety and essential performance including US deviations, except Clause 11.7 regarding biocompatibility. The device passed all tests. - IEC 62304:2006+ Amd 1:2015, Medical device software - Software life cycle processes. The device passed all tests. - IEC 60601-1-2:2014+A1:2021, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests. The device passed all tests. - IEC 60601-1-6, Ed. 3.2 2020-07, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. The device passed all tests. ## 7.3. Software Verification and Validation Testing Software verification and validation testing was conducted, and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device was considered a Major level of concern. Cybersecurity documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices and Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions. ## 7.4. Sterilization, Cleaning, and Shelf Life The Zeta Navigation System is reusable, provided non-sterile, and is not sterile when used. Cleaning instructions are provided in the labeling. Shelf-life is not applicable due to the low likelihood of time-dependent product degradation. The Zeta Stylet and Zeta Bolt are single use, provided clean and non-sterile, and are required to be sterilized before use. Sterilization and cleaning procedures are provided in the labeling. The subject devices underwent a steam sterilization validation to demonstrate that they can be expected to be sterile and have a sterility assurance level (SAL) of $10^{-6}$ or greater after processing. All requirements were met, and no new issues of safety or effectiveness were raised. Shelf-life is not applicable since sterilization occurs at point-of-use. ## 7.5. Bench Testing The following bench testing was performed to demonstrate substantial equivalence: - Accuracy testing - Design validation testing - Human factors testing, following the FDA Guidance Document, Applying Human Factors and Usability Engineering to Medical Devices. ## 7.6. Clinical Testing No clinical testing was required for the subject devices. {10} 510(K) SUMMARY # 8. CONCLUSION The results of testing described above demonstrate that the Zeta Navigation System, Zeta OS, Zeta Stylet, and Zeta Bolt are as safe and effective as the predicate devices and support a determination of substantial equivalence. PAGE 7 OF 7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...