The surgical transmission instrument is indicated for: Drilling and screwing of osteosynthesis screws, implants and plate systems into organic hard tissue in oral/maxillofacial surgery.
Device Story
Angled Screwdriver MED (M-SD090) is a surgical transmission instrument for drilling and screwing osteosynthesis screws, implants, and plates in maxillofacial surgery. Device operates manually via hand drive or mechanically via motor drive with ISO 3964 coupling. Integrated automatic gearbox increases motor torque by 6.5x during drilling. When used with W&H AMADEO (K213221) control unit, torque limiter ensures precise screw insertion. Designed for anatomically restricted areas; provides improved access. Used by clinicians in surgical settings. Output is mechanical rotation for bone fixation. Benefits include safe, precise screw placement in restricted surgical sites.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via mechanical/thermal safety testing per ISO 14457, biocompatibility evaluation per ISO 10993 series (parts 1, 5, 10, 18, 23) and ISO 7405, and reprocessing validation per FDA guidance.
Technological Characteristics
Handpiece with 90° head angle; ISO 3964 coupling. Features integrated automatic planetary gearbox (4:1) and head reduction (1.6:1) for total 6.5:1 reduction. Max input speed 20,000 rpm; max drill speed 3,056 rpm. Manual or motor-driven operation. Compliant with ISO 14457. Sterilizable.
Indications for Use
Indicated for drilling and screwing of osteosynthesis screws, implants, and plate systems into organic hard tissue in oral/maxillofacial surgery.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 31, 2026
W&H Dentalwerk Bürmoos GmbH
Gerhard Weidler
Regulatory Affairs Manager
Werner-Bader-Straße 1
Bürmoos, Salzburg 5111
AUSTRIA
Re: K253655
Trade/Device Name: Angled Screwdriver MED (M-SD090)
Regulation Number: 21 CFR 872.4120
Regulation Name: Bone Cutting Instrument And Accessories
Regulatory Class: Class II
Product Code: DZI, DZJ
Dated: June 23, 2026
Received: June 23, 2026
Dear Gerhard Weidler:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253655 - Gerhard Weidler
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253655 - Gerhard Weidler
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253655 | ? |
| Please provide the device trade name(s). | | ? |
| Angled Screwdriver MED (Type: M-SD090) | | |
| Please provide your Indications for Use below. | | ? |
| The surgical transmission instrument is indicated for: Drilling and screwing of osteosynthesis screws, implants and plate systems into organic hard tissue in oral/maxillofacial surgery. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Angled Screwdriver MED
Page 11 of 50
{5}
510(k) #: K253655
# 510(k) Summary
Prepared on: 2026-07-29
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | W&H Dentalwerk Bürmoos GmbH |
| --- | --- |
| Applicant Address | Werner-Bader-Straße 1 Bürmoos Salzburg 5111 Austria |
| Applicant Contact Telephone | +4362746236102 |
| Applicant Contact | Dr. Gerhard Weidler |
| Applicant Contact Email | gerhard.weidler@wh.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Angled Screwdriver MED (M-SD090) |
| --- | --- |
| Common Name | Bone cutting instrument and accessories |
| Classification Name | Drill, Bone, Powered |
| Regulation Number | 872.4120 |
| Product Code(s) | DZI, DZJ |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K082649 | Synthes Screwdriver 90° | DZI |
| K213221 | WS-75 handpiece of "AMADEO, M-UK1015 (incl. attachments and accessories)" submission | ERL |
Device Description Summary
21 CFR 807.92(a)(4)
W&H's Angled Screwdriver (M-SD090) is suitable for drilling holes and screwing of osteosynthesis screws, implants and plate systems in organic hard tissue in the field of maxillofacial surgery. Screwing can be performed manually with a hand drive or mechanically with a motor drive.
The angled screwdriver enables safe drilling with any drive motor with an ISO 3964 coupling, thanks to an integrated automatic gearbox that increases the torque of the motor (which the respective motor delivers at a speed of 20,000 rpm) by a factor of 6.5. The gearbox is only active during drilling when a motor with ISO 3964 coupling (or shorter couplings) is connected to the angled screwdriver.
The new angled screwdriver can be exclusively used for mechanical screwing-in in combination with W&H's AMADEO (K213221) device. Here, a torque limiter on the control unit ensures precise and torque-accurate screwing in of the bone screws.
The WS-75 handpiece (K213221) of the AMADEO device is used as a reference device within this submission.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The surgical transmission instrument is indicated for:
Drilling and screwing of osteosynthesis screws, implants and plate systems into organic hard tissue in oral/maxillofacial surgery.
Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use statements of the new M-SD090 and the predicate device are identical.
{6}
Both devices can be used manually (main application) or powered via a surgical unit and are indicated for the insertion (which includes pre-drilling, drilling and screwing) of osteosynthesis (bone fixation) screws in maxillofacial surgery.
The reference device, WS-75, is also intended to be used for the insertion (which includes pre-drilling, drilling and screwing) of osteosynthesis (bone fixation) screws in maxillofacial surgery.
## Technological Comparison
21 CFR 807.92(a)(6)
Both devices, the M-SD090 and the predicate device, share the same technological characteristics, like the basic functions (manual or mechanical use), the design, the ISO 3964 coupling, as well as the reprocessing instructions.
The reference device, the WS-75 handpiece, has a similar basic functionality, it supports drilling and the placement of implant screws in dental and maxillofacial surgery, can be connected via an ISO 3964 coupling, has the same reprocessing instructions and shares a similar design concept.
The new device and the predicate device share a similar maximum input speed (M-SD090 - 20.000 rpm vs. predicate device - 15.000 rpm) as well as the standard gear ratio (M-SD090 - 1.6:1 vs. predicate device 2:1).
The differing maximum drill speed of the new device with 3,056 rpm (predicate 7,500 rpm) is based on the drill manufacturer's specifications. The W&H defined drive speed is 20,000 rpm and results in a drill speed of 3,056 rpm due to the reduction ratio of 6.5:1 (head reduction ratio 1.6:1 and automatic planetary gear reduction ratio 4:1) which allows the new device also to be used with less powered drive units.
The differences of the WS-75 handpiece compared to the M-SD090 device are in the design (shorter neck, no manual use, no 90° head angle) and in technological parameters (gear ratio, max. input speed).
The new screwdriver is specifically designed to enable drilling and screw insertion in anatomically restricted treatment areas, improving access while maintaining established safety and performance characteristics. The basic technological characteristics of the WS-75 are retained in the M-SD090.
The WS-75 handpiece, cleared under K213221, provides appropriate technical context to support the evaluation of the subject device's technological characteristics. The similarities in intended use, functional principles, materials, and reprocessing methods demonstrate that the identified differences do not affect safety or effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Nonclinical testing to support submission:
To verify the performance requirements of the proposed device, the following tests were performed in accordance with standards and FDA guidance documents. All tests were completed successfully.
Mechanical and thermal Safety testing according to ISO 14457 "Dentistry — Handpieces and motors" was performed and shows that the new device is in compliance with this standard.
A biocompatibility evaluation of patient contacting components of the proposed device according to the requirements ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", ISO 10993-5 "Biological evaluation of medical devices Part 5: Test for cytotoxicity", ISO 10993-10 "Biological evaluation of medical devices - Part 10: Tests for skin sensitization", ISO 10993-18 "Biological evaluation of medical devices – Part 18: Chemical characterization of materials", ISO 10993-23 "Biological evaluation of medical devices – Part 23: Tests for irritation" and ISO 7405 "Dentistry - Evaluation of biocompatibility of medical devices used in dentistry" was conducted.
Reprocessing validation (Cleaning, Sterilization) according to FDA Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" was performed.
Clinical Testing:
No clinical testing was required to support this submission.
Conclusions drawn from conducted performance tests:
Results of performance testing demonstrate that the functionality, integrity, as well as safety and effectiveness of the new device, W&H's Angled Screwdriver MED (M-SD090), is sufficient for its intended use. Verification test results demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.