Medical records/Chart review; Post-market registry/observational study
The RUBY study, a retrospective chart review of 800 patients, was used to evaluate the safety and effectiveness of the Jada System in patients with gestational age ≥ 28 weeks, supporting a labeling update.
RUBY study; Multicenter, observational, chart review study; Follow-up/Duration: Duration of patient's admission for delivery; Study Period: October 2020 - March 2022
Patients treated with the Jada System for abnormal postpartum uterine bleeding or PPH; Sample Size: 800 patients; Number of Sites: 16
Not applicable for this study
Treatment success (control of abnormal bleeding > 500 mL or PPH ≥ 1000 mL without treatment escalation or recurrence)
Indications for Use
The Jada System (Jada-2002) is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted.
Device Story
Jada System is a single-use, silicone intrauterine device for managing postpartum hemorrhage. Device consists of an intrauterine loop with 21 vacuum pores and a donut-shaped cervical seal. Physician inserts compressed loop transvaginally into uterus; inflates cervical seal with 60-120 mL sterile fluid at external cervical os. External vacuum source is connected to the device and applied at a maximum of 90 mmHg. Vacuum induces uterine wall collapse, creating a tamponade effect on bleeding vessels. Device is secured to the patient's thigh. Output is the mechanical tamponade of the uterus. Clinical benefit is rapid, effective control of postpartum bleeding, reducing the need for treatment escalation.
Clinical Evidence
Evidence includes a multicenter, observational, post-market chart review (RUBY study) of 800 patients (786 analyzed) treated for postpartum hemorrhage. Primary endpoint: treatment success (control of bleeding without escalation). Success rate was 89.4% (95% CI: 87.0%-91.5%). Subgroup analysis for preterm births (28 to <34 weeks) showed 88.9% success. Results are consistent with the pivotal PEARLE trial (94.3% success). Adverse event profile is consistent with previous clinical data.
Technological Characteristics
Materials: Silicone, polycarbonate, ABS. Sensing/Actuation: Vacuum-induced mechanical tamponade. Energy: External vacuum source (max 90 mmHg). Dimensions: 41 cm length. Connectivity: Manual connection to vacuum tubing. Sterilization: SAL 10^-6. Single-use, disposable. No software or electronic components.
Indications for Use
Indicated for control and treatment of abnormal postpartum uterine bleeding or hemorrhage in patients where conservative management is warranted, including patients with gestational age ≥ 28 weeks.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 15, 2026
Organon, LLC
Riddhish Patel
Associate Principle Scientist, Regulatory Affairs
30 Hudson St.
Floor 33
Jersey City, New Jersey 07302
Re: K253642
Trade/Device Name: Jada System (Jada-2002)
Regulation Number: 21 CFR 884.4530
Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II
Product Code: OQY
Dated: April 14, 2026
Received: April 15, 2026
Dear Riddhish Patel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253642 - Riddhish Patel
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253642 - Riddhish Patel
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253642 | |
| Device Name Jada System (Jada-2002) | |
| Indications for Use (Describe) The Jada System (Jada-2002) is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Jada® System 510(k)
K253642
Page 1 of 4
510(k) Summary – K253642
I. SUBMITTER
Organon, LLC.
Address: 30 Hudson St.
Floor 33, Jersey City, NJ
07302 USA
Contact: Riddhish Patel,
Associate Principle Scientist, Regulatory Affairs
Phone: 718-801-0666
Email: riddhish.patel@organon.com
Date Prepared: May 14, 2026
II. DEVICE
Name of Device: Jada System (Jada-2002)
Common or Usual Name: Vacuum-induced Hemorrhage Control System
Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument
Regulation Number: 21 CFR § 884.4530
Regulatory Class: II
Product Code: OQY (Intrauterine Tamponade Balloon)
III. PREDICATE DEVICE
The predicate device is the Jada System, K212757. This predicate has not been subject to a design-related recall.
IV. DEVICE DESCRIPTION
The subject of this submission is the Jada System kit, which includes 1) the Jada System, 2) a commercially available pre-sterilized vacuum tubing, and 3) a commercially available pre-sterilized luer lock syringe.
The Jada System is a 41 cm long intrauterine device made of silicone. The device consists of an intrauterine loop on the distal end of a translucent tube. The proximal end of the tube has a vacuum connector for connection to the vacuum tubing. Proximal to the connection of the intrauterine loop is a donut-shaped cervical seal. The cervical seal is filled with and emptied of 60-120 mL of sterile fluid by attaching a luer lock syringe to the seal valve. The intrauterine loop consists of a loop tube with 21 vacuum pores oriented toward the inside diameter of the intrauterine loop. On the outer surface of the intrauterine loop is a shield which overhangs the vacuum pores to protect tissue from vacuum and to prevent the vacuum pores from plugging with tissue and blood clots.
Before placing the Jada System device inside the uterus, the intrauterine loop is compressed. The compressed loop is inserted into the uterus transvaginally. The cervical seal is placed within the vagina, at the external cervical os, and inflated and filled with 60-120 mL of sterile fluid. The vacuum tubing is attached to the vacuum connector on the Jada System and vacuum is then applied to a maximum value of 90 mmHg until bleeding is controlled. The Jada System should be fixed to the thigh along the tube.
The Jada System convenience kit is packaged in a secondary cardboard carton. The packaging
{5}
Jada® System 510(k)
K253642
Page 2 of 4
configuration consists of a) the Jada System, intrauterine device packaged in a sterile blister tray with Tyvek lid, b) a 60ml plastic luer lock syringe packaged in a sterile peelable pouch c) vacuum tubing packaged in a sterile peelable pouch and d) accompanying documentation.
The secondary carton and individual primary packs contain the required labelling information and symbols for communication to the end user based on international consensus standards.
The Jada System convenience kit is intended for single use (i.e., disposable) and has a three year shelf-life.
# V. INDICATIONS FOR USE
The Jada System (Jada-2002) is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted.
# VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Attribute | K253642 Subject Device: Jada System (Jada-2002) | K212757 Subject Device: Jada System | Comparison |
| --- | --- | --- | --- |
| Manufacturer | Organon | Alydia Health | N/A |
| Product Code | OQY | OQY | Same |
| Indications for Use | The Jada System (Jada-2002) is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted. | The Jada System is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted. | Same |
| Mechanism of Action | Inserted into the uterus and establishes a vacuum to cause the uterine walls to press against one another, producing a tamponade of the bleeding vessels. | Inserted into the uterus and establishes a vacuum to cause the uterine walls to press against one another, producing a tamponade of the bleeding vessels. | Same |
| Design | Inflatable cervical seal and intrauterine loop with 21 vacuum pores. The seal valve features luer lock syringe attachment. | Inflatable cervical seal and intrauterine loop with 21 vacuum pores. The seal valve features luer lock syringe attachment. | Different* |
| Rx/OTC | Rx | Rx | Same |
| Materials | Silicone, polycarbonate, Acrylonitrile- Butadiene-Styrene (ABS) | Silicone, polycarbonate, Acrylonitrile- Butadiene-Styrene (ABS) | Same |
| Sterile | SAL 10^{-6} | SAL 10^{-6} | Same |
| Single Use | Yes | Yes | Same |
The subject and predicate device have identical indications for use statements and have the same intended use – the treatment of abnormal uterine bleeding when conservative management is warranted.
*Since the clearance of JADA system convenience kit under K212757, Organon has made several changes which individually or cumulatively did not require a submission of a new 510(k). These changes were
{6}
K253642 Page 3 of 4
Jada® System 510(k)
assessed using the FDA guidance “Deciding When to Submit a 510(k) for a Change to an Existing Device – Guidance for Industry and Food and Drug Administration Staff” issued October 25, 2017 and these changes are captured within the write up below as technological differences.
The following minor technological differences exist between the subject and predicate devices:
- The subject device may utilize silicones procured from an alternate supplier for the following device components: Drain Shield (Tube Shield), Bilumen Shaft (Bilumen Tube), Tubing Connector, Y-Connector.
- The subject device has kitted accessory components from alternate suppliers. The 60mL luer lock syringe and 12' suction vacuum tubing are off the shelf devices available on US market and are considered like for like alternate accessory devices for use with Jada System. The convenience kit performance specifications remain unchanged.
- Manufacturing related changes including qualification of alternate supplier of SLS mold release agent (production aid for ejecting silicon parts from the mold tooling during molding), and qualification of new replacement molds with additional cavities for increased production efficiency.
The differences in technological characteristics also do not raise different questions of safety and effectiveness.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Non-clinical performance was leveraged from K212757.
## Clinical Test Summary and Conclusions
Study data from the following study: RUBY – Treating abnormal postpartum uterine bleeding or postpartum hemorrhage with the Jada System was utilized to update device labeling to support the use of Jada System in pregnant patients who give birth at a gestational age of ≥ 28 weeks. This was a multicenter, observational, chart review study. This study was a post-market study of the Jada System in commercial distribution. Data was collected through the duration of the patient’s admission for their delivery (i.e., starting with admission and continuing through discharge). Regular chart review occurred from the date the site began commercial use of the device (October 2020-March 2022). Sixteen sites participated in this study. Patients treated with the Jada System for abnormal postpartum uterine bleeding or PPH at these sites were included in the registry.
The data was collected from 800 patients. Among these 800 patients, 50 (6.3%) gave birth at < 34 weeks of gestation, including 36 at ≥ 28 and < 34 weeks of gestation (23 patients with Cesarean delivery, 13 patients with vaginal delivery). The primary effectiveness outcome was treatment success, defined as control of abnormal bleeding (> 500 mL) or postpartum hemorrhage (≥ 1000 mL) after device insertion with no treatment escalation or bleeding recurrence. Treatment success rates were 85.7% for preterm birth at less than 28 wGA (81.8% vaginal [9/11], 100% cesarean [3/3]). Treatment success rates were 88.9% for preterm births 28 to less than 34 wGA (100% vaginal [13/13], 82.6% cesarean [19/23]). Results were consistent with births greater than or equal to 34 weeks of gestation. Overall, the effectiveness was similar among patients who delivered between ≥ 28 and < 34 weeks of gestation, and those who delivered at ≥ 34 weeks of gestation as referenced in the warning section of current Jada System label, in both cesarean delivery and vaginal delivery. There was only 1 (4.3%) adverse event reported as severe and serious among the patients who delivered at between ≥ 28 and < 34 weeks of gestation. This unique adverse event was not reported as possibly or definitely related to the JADA System. Among the patients who delivered at ≥ 34 weeks of gestation, there were 37 (4.9%) adverse events (AE) reported (11 mild, 15 moderate and 11 severe), among which 32 (4.3%) were serious AE. Eight AE were reported as adverse device effects, and 3 AE were reported as serious adverse device effects.
{7}
Jada® System 510(k)
K253642
Page 4 of 4
For both vaginal and Cesarean births in real-world settings, the Jada System provided rapid, effective, and safe control of bleeding. The adverse event profile and device related adverse events was consistent with that observed in the registrational PEARLE trial.
## Conclusion
The overall effectiveness and safety of the Jada system among 28 < 34 weeks of gestation group is comparable to patients who delivered ≥ 34 weeks of gestation. This conclusion is based on a multicenter, observational, chart review, post-market study including 800 patients with abnormal postpartum bleeding treated with the Jada system, of which 786 were considered for this analysis (36 patients at 28 < 34 weeks of gestation, and 750 patients ≥ 34 weeks of gestation). Additionally, the overall effectiveness (treatment success rate defined as control of bleeding without escalation of treatment was 89.4% (95% CI: 87.0% to 91.5%)) and safety results of this real-world evidence study are similar with the results of the pivotal PEARLE study (overall effectiveness for the same endpoint was 94.3% (95% CI 88.1%-97.9%)) submitted for initial clearance of the Jada System.
## References
Kara M. Rood, Angela Bianco, Joseph R. Biggio, Marcela C. Smid, Hyagriv N. Simhan, James Li, Candice Yong, Patricia I. Carney, Damien J. Croft & Dena Goffman (2025) Real-world use of a vacuum-induced hemorrhage-control device in births <34 weeks gestational age, The Journal of Maternal-Fetal & Neonatal Medicine, 38:1, 2451658, DOI: 10.1080/14767058.2025.2451658
D'Alton, Mary E. MD; Rood, Kara M. MD; Smid, Marcela C. MD; Simhan, Hyagriv N. MD, MS; Skupski, Daniel W. MD; Subramaniam, Akila MD; Gibson, Kelly S. MD; Rosen, Todd MD; Clark, Shannon M. MD; Dudley, Donald MD; Iqbal, Sara N. MD; Paglia, Michael J. MD, PhD; Duzyj, Christina M. MD, MPH; Chien, Edward K. MD; Gibbins, Karen J. MD; Wine, Kathryn D. MPH; Bentum, Nana Ama A. MD; Kominiarek, Michelle A. MD; Tuuli, Methodius G. MD; Goffman, Dena MD. Intrauterine Vacuum-Induced Hemorrhage-Control Device for Rapid Treatment of Postpartum Hemorrhage. Obstetrics & Gynecology 136(5):p 882-891, November 2020. | DOI: 10.1097/AOG.0000000000004138
## VIII. CONCLUSIONS
The clinical performance data described above demonstrate that the Jada System (Jada-2002) is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.