K253636 · Termosalud S.L. · NGX · Mar 12, 2026 · Physical Medicine
Device Facts
Record ID
K253636
Device Name
Zionic Pro Max (EMS)
Applicant
Termosalud S.L.
Product Code
NGX · Physical Medicine
Decision Date
Mar 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Zionic Pro Max EMS function of the device is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the muscles by the Zionic Pro Max programs is not suitable for rehabilitation or physiotherapy.
Device Story
Zionic Pro Max is a floor-standing electrical muscle stimulator (EMS) for professional use; generates electrical impulses delivered to subcutaneous tissue via handpieces in contact with intact skin. Device features console with user interface and 4 output channels; delivers symmetrical bi-phasic square wave pulses. Operates via AC line power (120V, 60Hz). Includes safety features: automatic overload/no-load trip, automatic shut-off, and patient remote safety stop switch. Healthcare professionals operate the device to stimulate healthy muscles; output parameters (frequency 3-40 Hz, pulse width 610uS) are adjusted to facilitate muscle performance. Benefits include improved muscle performance in healthy populations. Device is not for rehabilitation or physiotherapy.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance requirements for nerve/muscle stimulators (IEC 60601-2-10), software lifecycle (IEC 62304), risk management (ISO 14971), and biocompatibility (ISO 10993-1: cytotoxicity, sensitization, irritation).
Indicated for healthy individuals requiring muscle stimulation to improve or facilitate muscle performance. Not indicated for rehabilitation or physiotherapy.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 12, 2026
Termosalud S.L.
% Connie Hoy
Regulatory Consultant
Hoy & Associates Regulatory Consulting, LLC
1830 Bonnie Way
Sacramento, California 95825
Re: K253636
Trade/Device Name: Zionic Pro Max (EMS)
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: NGX
Dated: August 31, 2025
Received: November 19, 2025
Dear Connie Hoy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253636 - Connie Hoy
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253636 - Connie Hoy
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253636 | ? |
| Please provide the device trade name(s). | | ? |
| Zionic Pro Max (EMS) | | |
| Please provide your Indications for Use below. | | ? |
| The Zionic Pro Max EMS function of the device is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the muscles by the Zionic Pro Max programs is not suitable for rehabilitation or physiotherapy. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
Applicant TermoSalud S.L.
Address Ataulfo Friera Tarfe, 8 -33211 Gijón, Spain
Contact Person Connie Hoy Consultant
Contact Information conniehoy@hoyregulatory.com
Preparation Date March 9, 2026
Device Trade Name ZIONIC Pro Max EMS
Common Name Electromagnetic Muscle Stimulator
Regulation Number 21 CFR 890.5850
Product Code NGX
Regulatory Class II
Legally Marketed Predicate Device Compex Sport Elite 3.0 K201653
## Device Description:
The ZIONIC Pro Max is an innovative electrical impulse generator that generates electrical muscle stimulation (EMS) energy to subcutaneous tissue by way of contact with intact skin. The device includes and console and user interface that power a selection of handpieces that deliver EMS to intact skin. The device is intended for use by medical professionals.
## Indications for use:
The Zionic Pro Max EMS function of the device is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the muscles by the Zionic Pro Max programs is not suitable for rehabilitation or physiotherapy.
Page 1 of 7
{5}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
| Device Name, Model | Zionic Pro Max (Subject) | Compex Sport Elite 3.0 (primary predicate) | Comparison |
| --- | --- | --- | --- |
| 510K # | K253636 | K201653 | N/A |
| Indication | Intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the muscles by the Zionic Pro Max programs is not suitable for rehabilitation or physiotherapy. | Intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the muscles by the Compex Sport programs are not suitable for rehabilitation or physiotherapy. | Same |
| Prescription Use/Over-the-Counter Use | Prescription Use | Over-the-Counter Use | The subject device is restricted to prescription use; however, it has the same intended use and comparable technological characteristics as the predicate. This difference does not impact safety or effectiveness and does not affect the determination of substantial equivalence. |
| Power Source
*Line Voltage
*Frequency
*Isolation Method | 120V, 60 Hz, 200VA
*Galvanic isolation | * Li-ion rechargeable 3.7V battery, *10 W | Different, Zionic Pro Max is AC Line powered whereas other predicates are battery powered. The power source does not impact safety or effectiveness in that it's output waveform etc. are substantially equivalent to the predicate device. |
Page 2 of 7
{6}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
| Device Name, Model | Zionic Pro Max (Subject) | Compex Sport Elite 3.0 (primary predicate) | Comparison |
| --- | --- | --- | --- |
| Number of Output Modes | One (NMES) | Two (NMES/TENS) | Different, Zionic provides only a single waveform type in both preset & free "modes". This waveform is used for NMES and the subject does not provide for a TENS type waveform. This difference does not impact safety or effectiveness. |
| Number of Output Channels
*Synchronous or Alternating *Method of Channel Isolation | 4 channels
*synchronous but never 2 channels at same time
*isolated by independent high Z H-bridges | 4 channels
*synchronous but never 2 channels at same time
*isolated by independent high Z H-bridges | Same |
| Regulated Current or Voltage? | Regulated Current | Regulated Current | Same |
| Software/Firmware/μProc Control? | Yes | Yes | Same |
| Automatic Overload Trip? | Yes, | Yes | Same |
| Automatic No-Load Trip? | Yes | Yes | Same |
| Automatic Shut Off? | Yes | Yes | Same |
Page 3 of 7
{7}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
| Device Name, Model | Zionic Pro Max (Subject) | Compex Sport Elite 3.0 (primary predicate) | Comparison |
| --- | --- | --- | --- |
| Patient Override Control? | Yes, patient remote stop/ safety stop switch | Yes, On/Off switch pauses output | Same |
| Indicator Display *On/Off Status? | Yes, LCD display | Yes | Same |
| Indicator Display *Low Battery? | N/A | Yes | N/A |
| Indicator Display *Voltage/ Current Level? | Yes | Yes | Same |
| Weight | 32 kg | .21 kg | Different, Zionic Pro Max is a floor standing unit whereas predicate is handheld. This difference does not impact safety and effectiveness. |
| Dimensions (inches) [W x H x D] | W: 68.4cm H: 142cm L: 64.6cm | W: 80mm H: 24.6mm L: 140mm | Different, Zionic Pro Max is a floor standing unit whereas predicates are handheld. This difference does not impact safety and effectiveness. |
| Housing Materials & Construction | Main case: polyurethane + metalic back panel. Handpieces: Polyamide 12. | Plastic ABS with PMMA windows, silicone | Different. This difference does not impact safety and effectiveness. |
| Waveform / Shape | Symmetrical bi-phasic square wave | Symmetrical bi-phasic square wave | Same |
| Maximum Output Voltage (± 10%) | 48 V @ 500 Ω
162 V @ 2 kΩ
178 V @ 10 kΩ | 60 V @ 500 Ω
165 V @ 2 kΩ
165 V @ 10 kΩ | Different; Considering the electrode impedance, this voltage acts to provide a max current density substantially equivalent to |
Page 4 of 7
{8}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
| Device Name, Model | Zionic Pro Max (Subject) | Compex Sport Elite 3.0 (primary predicate) | Comparison |
| --- | --- | --- | --- |
| | | | other class II PMS devices cleared under 21 CFR 890.5850 operate very close to this range (e.g., K940301 at 100mA). This difference does not impact safety or effectiveness. |
| Maximum Output Current (± 10%) | 96 mA @ 500 Ω
81 mA @ 2 kΩ
17 mA @ 10 kΩ | 120 mA @ 500 Ω
82 mA @ 2 kΩ
16 mA @ 10 kΩ | Different; other class II PMS devices cleared under 21 CFR 890.5850 operate substantially equivalent to this range (e.g., K940301 at 100mA). This difference does not impact safety or effectiveness. |
| Pulse Width | 610uS | 200-400uS | Different; other class II PMS devices cleared under 21 CFR 890.5850 operate substantially equivalent to this range (e.g., K940301 at 15-800uS). This difference does not impact safety or effectiveness. |
| Pulse Frequency | 3-40 Hz | 1-120 Hz | Different. The pulse frequency is within the range of the cited predicate device. This difference does not impact safety and effectiveness. |
| Interferential Modes only
*Beat Frequency | N/A | N/A | N/A |
Page 5 of 7
{9}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
| Device Name, Model | Zionic Pro Max (Subject) | Compex Sport Elite 3.0 (primary predicate) | Comparison |
| --- | --- | --- | --- |
| Net Charge (μC/pulse) if zero state method | 0μC @500Ω excitation pulse fully compensated | 0μC @500Ω excitation pulse fully compensated | Same |
| Maximum Phase Charge (μC) | 58 uC | 48 uC | Different; 2 other class II PMS devices cleared under 21 CFR 890.5850 operate above and below this range (e.g., K940301, K011880). This difference does not impact safety or effectiveness. |
| Maximum Current Density | 1.19mA/cm² @500Ω | 1.49mA/cm² @500Ω | Different; Subject max current = 96mA vs. 120mA for cited predicate resulting in max current density ~25% below cited predicate although other class II PMS devices cleared under 21 CFR 890.5850 operate in this range (e.g., K940301). This difference does not impact safety or effectiveness. |
| Maximum Power Density | 8.98mW/cm² @500Ωmw | 27.6mW/cm² @500Ω | Different; Zionic max frequency is 40Hz Vs. 120Hz for cited predicate resulting in a lower max power density. At 40Hz, predicate = 9.2mW/cm². This difference does not impact safety or effectiveness. |
Page 6 of 7
{10}
510(K) Summary
ZIONIC PRO MAX – EMS
K253636
Performance Testing
Verification and validation activities were successfully completed and establish that the ZIONIC Pro Max control unit performs as intended. Testing included the following:
- IEC 60601-1 Edition 3.2 2020-08; Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- IEC 60601-1-2:2014 + A1:2020; Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests;
- IEC 60601-2-10:2012+AMD2:2023 specifies particular safety and performance requirements for nerve and muscle stimulators, including TENS and EMS devices used in physical medicine
- IEC 62304:2006+A1:2015; Medical Device – Software Life Cycle Processes;
- EN ISO 14971:2019+A11:2021; Medical Devices – Application Of Risk Management To Medical Devices
The biocompatibility evaluation for the Zionic Pro Max EMS device was conducted in accordance with the FDA guidance on “Use of International Standard ISO-10993, ‘Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,’”, and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
Software verification and validation testing was conducted, and documentation provided in accordance with FDA’s Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices.
Clinical Testing – no clinical testing was conducted as part of this submission
Comparison to the predicate device and conclusion
The subject device shares similar indications for use, general design, and technical parameters as the cited predicate device. The technological differences of the subject device do not impact the safety or effectiveness. It can be concluded, therefore, that the subject device is substantially equivalent to the predicate device.
Page 7 of 7
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.