← Product Code [ITI](/productcode/ITI) · K253599

# Electric Wheelchair (W-A807, W-A807L) (K253599)

_Ningbo Shenyu Medical Equipment Co.,Ltd · ITI · Feb 17, 2026 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K253599

## Device Facts

- **Applicant:** Ningbo Shenyu Medical Equipment Co.,Ltd
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Feb 17, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

## Device Story

Battery-powered, four-wheeled electric wheelchair; models W-A807 and W-A807L. Input: user operation via joystick controller. Transformation: controller manages DC brushed motors on rear wheels to provide forward, backward, and turning movement; electromagnetic brake system provides stopping. Output: physical mobility for user. Used in indoor/outdoor environments by disabled or elderly patients. Healthcare provider/caregiver or patient operates device via joystick. Benefits: provides independent mobility for individuals with limited walking ability. W-A807L includes a linear actuator for backrest recline.

## Clinical Evidence

No clinical data. Bench testing only. Compliance with ISO 7176 series (stability, braking, energy consumption, dimensions, strength, climatic, obstacle-climbing, power/control systems, EMC, battery safety) and ISO 10993-1 biocompatibility testing.

## Technological Characteristics

Aluminum alloy frame; solid PU tires; 24V lithium battery; 250W DC brushed motors; electromagnetic brake system. Connectivity: none. Sterilization: N/A. Software: IEC 62304 compliant. Standards: ISO 7176 series, IEC 60601-1-2, IEC 62133-2 (batteries).

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Electric Wheelchair ([K250366](/device/K250366.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

February 17, 2026

Ningbo Shenyu Medical Equipment Co., Ltd
% Andrew Wang
Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Bldg., 1500# Century Ave.
Shanghai,
China

Re: K253599
Trade/Device Name: Electric Wheelchair (W-A807, W-A807L)
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: November 18, 2025
Received: November 18, 2025

Dear Andrew Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253599 - Andrew Wang
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K253599 - Andrew Wang
Page 3

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team

DHT5B: Division of Neuromodulation and
Physical Medicine Devices

OHT5: Office of Neurological and
Physical Medicine Devices

Office of Product Evaluation and Quality
Center for Devices and Radiological Health

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253599 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Electric Wheelchair (W-A807, W-A807L)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) #: K253599
510(k) Summary
Prepared on: 2026-01-09

# Contact Details
21 CFR 807.92(a)(1)

|  Applicant Name | Ningbo Shenyu Medical Equipment Co.,Ltd.  |
| --- | --- |
|  Applicant Address | No.1318 West of Tanjialing Road, Yuyao city, Zhejiang Province, China
Yuyao China  |
|  Applicant Contact Telephone | +86-574-6251777  |
|  Applicant Contact | Mr. Song Zhenyong  |
|  Applicant Contact Email | gm@shenyumedical.com  |
|  Correspondent Name | Shanghai SUNGO Management Consulting Co., Ltd.  |
|  Correspondent Address | 14th Floor, Dongfang Building, 1500# Century Ave. Shanghai China  |
|  Correspondent Contact Telephone | +86-17521605316  |
|  Correspondent Contact | Mr. Andrew Wang  |
|  Correspondent Contact Email | yuanju.wang@sungoglobal.com  |

# Device Name
21 CFR 807.92(a)(2)

|  Device Trade Name | Electric Wheelchair (W-A807, W-A807L)  |
| --- | --- |
|  Common Name | Powered wheelchair  |
|  Classification Name | Wheelchair, Powered  |
|  Regulation Number | 890.3860  |
|  Product Code(s) | ITI  |

# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)

|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
| --- | --- | --- |
|  K250366 | Electric Wheelchair | ITI  |

# Device Description Summary
21 CFR 807.92(a)(4)

The Electric Wheelchair is a battery powered four wheeled vehicle.

The model W-A807 consists one Lithium battery with an off-board battery charger, frame, controller, motors, seat, back support, control device (including the battery indicator, power ON/OFF, speed indicator light, decrease speed, increase speed, horn, joystick, charging port), arm supports, two rear wheels(drive wheel), two casters, foot support, anti-tip devices, handle.

The model W-A807L consists one Lithium battery with an off-board battery charger, frame, controller, motors, seat, back support, control device (including the battery indicator, power ON/OFF, speed indicator light, decrease speed, increase speed, horn, lean back, backrest recline, joystick, charging port), arm supports, two rear wheels(drive wheel), two casters, foot support, anti-tip devices, handle, linear actuator.

The wheelchair can easily fold and unfold for transportation or storage.

The wheelchair uses lithium battery as its power source. The controller controls the drive left/right motor to realize the wheelchair

{5}

forward, backward and turn functions.

The frame of the device is Aluminum. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels. The rear wheels are driving wheels to control the speed and direction. The wheels are solid PU tires.

When in use, the operator drives the motor of the rear wheel by operating the controller to achieve the rear wheels movement.

The DC Brushed Motor and brake system are fixed on the rear wheels.

The max loading of the device is 136 kg. Only for one person sit.

## Intended Use/Indications for Use

21 CFR 807.92(a)(5)

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

## Indications for Use Comparison

21 CFR 807.92(a)(5)

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

## Technological Comparison

21 CFR 807.92(a)(6)

The design and technological characteristics of the Electric Wheelchair is similar to the predicate chosen. There are minor differences between the devices including Braking distance, Overall/Folded Dimension, Front/Rear wheel size, Max speed forward/backward, Max loading weight, Maximum distance of travel on the fully charged battery, Battery, Turning Radius, Maximum obstacle climbing, Motor type. All of the parameter with differences have been tested according to ISO 7176 series standards and the test records support it safety and effectiveness. There is no deleterious effect on safety and effectiveness due to the minor differences, Additionally, they do not influence the intended use of the device. And the same materials were used for parts in contact with user. Therefore, the proposed Wheelchair is substantially equivalent (SE) to the Electric Wheelchair.

{6}

|  Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark  |
| --- | --- | --- | --- |
|  Model(s) | W-A807, W-A807L | LW01301A07 | --  |
|  Classification | Class II | Class II | S.E.  |
|  Classification regulation | 21 CFR890.3860 | 21 CFR890.3860 | S.E.  |
|  Product code | ITI | ITI | S.E.  |
|  Indications for use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | S.E.  |
|  Intended user | disabled or elderly person limited to a seated position. | disabled or elderly person limited to a seated position. | S.E.  |
|  Use condition | indoor and outdoor use | indoor and outdoor use | S.E  |
|  Number of wheels | 4, including two front wheels and two rear wheels | 4, including two front wheels and two rear wheels | S.E  |
|  Function of wheels | Front wheels: driven wheels suitable for rotation, acceleration, retrograde
Rear wheels: driving wheels to control the speed and direction | Front wheels: driven wheels suitable for rotation, acceleration, retrograde
Rear wheels: driving wheels to control the speed and direction | S.E  |
|  Movement control method | By Joystick control | By Joystick control | S.E  |
|  Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | S.E  |

{7}

|  Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark  |
| --- | --- | --- | --- |
|  Brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | S.E  |
|  Braking distance | ≤1.0 m | 1.2 m | Minor difference on braking distance will not cause different performance. Shorter distance for braking will be more safety.  |
|  Maximum safe operational incline degree | 6° | 6° | S.E  |
|  Armrest | PUR | PUR | S.E  |
|  Main frame material | Aluminum Alloy | Aluminum Alloy | S.E  |
|  Back cushion | Nylon | Nylon | S.E  |
|  Seat cushion | Nylon | Nylon | S.E  |
|  Overall Dimension (length*width*height) | W-A807: 980*660*950mm
W-A807L: 990*660*950mm | 1120*640*1000mm | Minor difference on wheelchair dimension will not cause different performance. All safety and performance have been validated with the maximum rated weight dummy.  |
|  Folded Dimension (length*width*height) | W-A807: 780mm*630mm*490mm
W-A807L: 770mm*630mm*500mm | 760*640*460mm  |   |
|  Front wheel size/type | 8" PU Solid tire | 8" x 2"/PU Solid tire | Different sizes of front wheel will not affect safety and performance of the subject device as all related stability tests are performed according to standard ISO 7176 series.  |
|  Rear wheel size/type | 12.5" * 2.25" PU Solid tire | 12"x 2"/ PU Pneumatic tire | Different sizes of rear wheel will not affect safety and performance of the subject device as all related stability tests are  |

{8}

|  Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark  |
| --- | --- | --- | --- |
|   |  |  | performed according to standard ISO 7176 series.  |
|  Max speed forward | Up to 5.04 km/h (1.4 m/s), adjustable | Up to 6.84 km/h (1.9 m/s), adjustable | Different max speed forward will not affect safety and performance of the subject device as all related stability tests are performed according to standard ISO 7176 series.  |
|  Max Speed backward | Less than 2.52 km/h (0.7 m/s) | Less than 3.96 km/h (1.1 m/s) | Different max speed backward will not affect safety and performance of the subject device as all related stability tests are performed according to standard ISO 7176 series.  |
|  Max loading weight | 136 kg | 120 kg | Difference on loading weight will not cause different performance. All safety and performance have been validated with the maximum loading weight.  |
|  Maximum distance of travel on the fully charged battery | 16.4 km | 17.6 km | The subject device complies with ISO 7176-4: 2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range, these differences do not affect safety and effectiveness.  |
|  Battery | li-ion battery pack; rechargeable, 24V, 20Ah, | li-ion battery pack; rechargeable, 24 VDC 13Ah | Minor difference on battery capacity will not cause different performance. Larger  |

{9}

|  Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark  |
| --- | --- | --- | --- |
|   |  |  | battery capacity will contribute to the maximum distance.  |
|  Motor | Brushed Motor, 24VDC, 250W, 2pcs | Brushless DC motor; 24VDC; 250W; 2pcs | Although the Subject Device uses brushed motors while the Predicate uses brushless motors, both have identical power ratings (250W) and operating voltages (24VDC). The difference in motor type primarily relates to the commutation method and does not affect the intended use or basic driving performance. The safety and performance of the motors are further validated through compliance with ISO 7176-14 and IEC 60601-1. This difference does not raise new questions of safety and effectiveness.  |
|  Electronic controller | Brushless dual-drive rocker controller | Brushless motor controller | S.E  |
|  Turning Radius | 900 mm | 958 mm | Minor difference on Turning Radius will not cause different performance.  |
|  Maximum obstacle climbing | 15 mm | 30 mm | The predicate device and subject device have different dimensions. Both comply with ISO 7176-10:2008 Wheelchairs – Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs.  |

{10}

|  Item | Proposed Device | Predicate Device | Results  |
| --- | --- | --- | --- |
|  Biocompatibility | Comply with ISO 10993-1, FDA Guidance | Comply with ISO 10993-1, FDA Guidance | S.E.  |
|  EMC | ISO 7176-21, IEC 60601-1-2, IEC 60601-4-2 | ISO 7176-21, IEC 60601-1-2, IEC 60601-4-2 | S.E.  |
|  Performance | ISO 7176 series | ISO 7176 series | S.E.  |
|  Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E.  |
|  Item | Proposed Device | Predicate Device | Results  |
| --- | --- | --- | --- |
|  ISO 7176-1 | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | S.E.  |
|  ISO 7176-2 | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | S.E.  |
|  ISO 7176-3 | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | S.E.  |
|  ISO 7176-4 | The theoretical distance range has been determined after the testing according to the ISO 7176-4, and | The theoretical distance range has been determined after the testing according to the ISO 7176-4, and | S.E.  |
|   | test results meet its design specification. | test results meet its design specification. |   |

{11}

|  Item | Proposed Device | Predicate Device | Results  |
| --- | --- | --- | --- |
|   | test results meet its design specification. | test results meet its design specification. |   |
|  ISO 7176-5 | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | S.E.  |
|  ISO 7176-6 | The dimensions, mass has been determined after the testing according to the ISO 7176-6. | The dimensions, mass has been determined after the testing according to the ISO 7176-6. | S.E.  |
|  ISO 7176-7 | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7. | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7. | S.E.  |
|  ISO 7176-8 | All test results meet the requirements in Clause 4 of ISO 7176-8. | All test results meet the requirements in Clause 4 of ISO 7176-8. | S.E.  |
|  ISO 7176-9 | The test results shown that the device under tests could continue to function according to manufacturer’s specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9. | The test results shown that the device under tests could continue to function according to manufacturer’s specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9. | S.E.  |
|  ISO 7176-10 | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | S.E.  |
|  ISO 7176-11 | The test dummies used in the testing of ISO 7176 series are | The test dummies used in the testing of ISO 7176 series are | S.E.  |

{12}

|  Item | Proposed Device | Predicate Device | Results  |
| --- | --- | --- | --- |
|   | meet the requirements of ISO 7176-11. | meet the requirements of ISO 7176-11. |   |
|  ISO 7176-13 | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved. | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved. | S.E.  |
|  ISO 7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14. | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14. | S.E.  |
|  ISO 7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15. | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15. | S.E.  |
|  ISO 7176-16/ ISO 16840-10 | The performance of resistance to ignition meet the requirements of ISO 16840-10:2021. | The performance of resistance to ignition meet the requirements of ISO 16840-10:2021. | S.E.  |
|  ISO 7176-21 | The EMC performance results meet the requirements of ISO 7176-21. | The EMC performance results meet the requirements of ISO 7176-21. | S.E.  |
|  ISO 7176-25 | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25. | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25. | S.E.  |

## 2 Substantial Equivalence Discussion

The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO

{13}

7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-21, ISO 7176-25, and FDA guidance Submission for Power Wheelchair.

The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/ control criteria following above mentioned standards. And the test results show that the subject product is substantially equivalent to the predicate device in performance.

The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability (tipping angle), The Dynamic stability (Safe Gradient Maximum Gradient), Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to ignition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers.

The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.

## 3 Substantially Equivalency Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices, K250366 Electric Wheelchair from Anhui Longway Medical Technology Co., LTD.

{14}

Non-Clinical and/or Clinical Tests Summary &amp; Conclusions 21 CFR 807.92(b)

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ISO 7176-1: 2014, Wheelchairs - Part 1: Determination of static stability
ISO 7176-2: 2017, Wheelchairs - Part 2: Determination of dynamic stability of Powered Wheelchairs
ISO 7176-3: 2012, Wheelchairs - Part 3: Determination of effectiveness of brakes
ISO 7176-4, Third edition 2008-10-01, Wheelchairs - Part 4: Energy consumption of electric wheelchairs and wheelchairs for determination of theoretical distance range
ISO 7176-5, Second edition 2008-06-01, Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
ISO 7176-6: 2018, Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of Powered Wheelchairs
ISO 7176-7, Wheelchairs - Part 7: Measurement of seating and wheel dimensions
ISO 7176-8: 2014, Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
ISO 7176-9: 2009, Wheelchairs - Part 9: Climatic tests for Powered Wheelchairs
ISO 7176-10: 2008, Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
ISO 7176-11: 2012, Wheelchairs - Part 11: Test dummies.
ISO 7176-13, First edition 1989-08-01, Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
ISO 7176-14: 2022, Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and wheelchairs - Requirements and test methods
ISO 7176-15: 1996, Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
ISO 16840-10: 2021, Wheelchair seating Part 10: Resistance to ignition of postural support devices - Requirements and test method.
ISO 7176-21: 2009, Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and wheelchairs, and battery chargers
ISO 7176-25: 2022, Wheelchairs - Part 25: Lead-acid batteries and chargers for powered wheelchairs - Requirements and test methods
IEC 62133-2: 2017/AMD1: 2021, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary lithium cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
IEC 62304: 2006/AMD1: 2015, Medical device software - Software life cycle processes
IEC 60601-1-2: 2014/AMD1: 2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC TS 60601-4-2: 2024, Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
ISO 10993-1: 2018, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process

Not Applicable.

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, Electric Wheelchair, model(s): W-A807, W-A807L, is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K250366.

---

**Source:** [https://fda.innolitics.com/device/K253599](https://fda.innolitics.com/device/K253599)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
