Zimmer Biomet Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The Zimmer Biomet Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The titanium Glenosphere components are indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium Glenospheres is not recommended for patients who lack suspected material sensitivity to cobalt alloy. The humeral components are indicated for either cemented or uncemented (biological fixation). Compatible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
Device Story
Modular reverse shoulder arthroplasty system components; includes humeral bearings, humeral trays, glenospheres, and adapters. Designed for intraoperative flexibility; allows restoration of glenohumeral joint kinematics; enables conversion from anatomical to reverse shoulder arthroplasty while retaining well-fixed humeral stems. Used by orthopedic surgeons in clinical settings. Components interface with existing Identity and Comprehensive Reverse shoulder systems. Provides modularity for individual anatomical adjustment; restores joint stability and motion. Benefits patients through pain relief and functional restoration in cases of severe rotator cuff deficiency.
Clinical Evidence
No clinical or animal testing was required. Evidence is based on non-clinical bench testing and engineering evaluations, including Finite Element Analysis, fatigue strength, accelerated corrosion fatigue, micromotion/moment analysis, lever-out resistance, static axial/torsional strength, wear assessment, range of motion analysis, and MRI compatibility testing.
Indicated for patients with grossly deficient rotator cuff and severe arthropathy, or failed shoulder replacement with grossly deficient rotator cuff. Requires functional deltoid muscle and suitable anatomy. Indicated for primary, fracture, or revision total shoulder replacement for pain relief and disability due to gross rotator cuff deficiency. Titanium Glenospheres indicated for patients with suspected cobalt alloy sensitivity.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 13, 2026
Zimmer, Inc.
Alexandria Irwin
Regulatory Affairs Senior Specialist
1800 W. Center St.
Warsaw, Indiana 46850
Re: K253592
Trade/Device Name: Zimmer Biomet Reverse Shoulder Prosethesis(Alliance Humeral Bearings, Identity Humeral Trays, Comprehensive Reverse Humeral Trays, Comprehensive Reverse 32mm Glenospheres and Adapter)
Regulation Number: 21 CFR 888.3670
Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis
Regulatory Class: Class II
Product Code: MBF, KWT, KWS, HSD, PHX
Dated: November 17, 2025
Received: March 12, 2026
Dear Alexandria Irwin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253592 - Alexandria Irwin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253592 - Alexandria Irwin
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FARZANA SHARMIN -S
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253592 | ? |
| Please provide the device trade name(s). | | ? |
| Zimmer Biomet Reverse Shoulder Prosethesis(Alliance Humeral Bearings, Identity Humeral Trays,
Comprehensive Reverse Humeral Trays, Comprehensive Reverse 32mm Glenospheres and Adapter) | | |
| Please provide your Indications for Use below. | | ? |
| Zimmer Biomet Reverse Shoulder products are indicated for use in patients whose shoulder joint has a
grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with
a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the
implants and a functional deltoid muscle is necessary.
The Zimmer Biomet Reverse Shoulder is indicated for primary, fracture, or revision total shoulder
replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
The titanium Glenosphere components are indicated for patients with suspected cobalt alloy sensitivity. The
wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium Glenospheres
is not recommended for patients who lack suspected material sensitivity to cobalt alloy.
The humeral components are indicated for either cemented or uncemented (biological fixation).
Compatible Glenoid Baseplate components are intended for cementless applications with the addition of
screw fixation. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K253592 | 510(k) Summary | Prepared on: 2026-04-13 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Zimmer, Inc. | |
| Applicant Address | 1800 W. Center Street Warsaw IN 46850 United States | |
| Applicant Contact Telephone | 574-373-0167 | |
| Applicant Contact | Mrs. Alexandria Irwin | |
| Applicant Contact Email | alexandria.irwin@zimmerbiomet.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Zimmer Biomet Reverse Shoulder Prosethesis(Alliance Humeral Bearings, Identity Humeral Trays, Comprehensive Reverse Humeral Trays, Comprehensive Reverse 32mm Glenospheres and Adapter) | |
| Common Name | Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis | |
| Classification Name | Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented | |
| Regulation Number | 21 CFR 888.3670 | |
| Product Code(s) | MBF, KWT, KWS, HSD, PHX | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K240876 | Identity Shoulder System | MBF |
| K193373 | Comprehensive Reverse Shoulder | PHX |
| K181611 | Comprehensive Reverse(Mini) Shoulder System | PHX |
| K121543 | Zimmer Trabecular Metal Reverse Shoulder System | KWT |
| K233481 | Altivate Reverse Glenoid | PHX |
| K200209 | Persona Personalized Knee System | JWH |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| This submission is seeking approval of Alliance Humeral Bearings, Identity Humeral Trays, Comprehensive Reverse Humeral Trays, Comprehensive Reverse 32mm Glenospheres and Adapter to provide compatibility with the Identity Shoulder System and/or the Comprehensive Reverse System. This submission is a collection of components designed with the intention of providing the modularity and adaptability necessary to facilitate individual anatomical adjustment and restoration of the glenohumeral joint during reverse shoulder arthroplasty. Modular joints provide intraoperative flexibility for accurate and reproducible reconstruction of the glenohumeral joint to restore limb kinematics. The modularity of the humeral components also allows convertibility from an anatomical total shoulder arthroplasty to a reverse shoulder arthroplasty with retention of the well-fixed humeral stem in case of failure of the prosthesis or the rotator cuff. | | |
{5}
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
Zimmer Biomet Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
The Zimmer Biomet Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
The titanium Glenosphere components are indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium Glenospheres is not recommended for patients who lack suspected material sensitivity to cobalt alloy.
The humeral components are indicated for either cemented or uncemented (biological fixation).
Compatible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use are similar to the primary predicate device.
Technological Comparison
21 CFR 807.92(a)(6)
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: Identical to predicate devices
- Indications for Use: Similar to predicate devices
- Materials: Identical to predicate devices
- Alliance Humeral Bearings: Identical to primary predicate
- Identity Trays: Identical to primary predicate
- Comprehensive Reverse Trays: Identical to predicate device.
- Comprehensive Glenospheres: Co-Cr-Mo Glenospheres: Identical to predicate device: No surface treatment processing applies to the cobalt chrome variant. Ti-6Al-4V Glenospheres: Similar to predicate device. The Ti-6Al-4V glenospheres include a surface hardening treatment process identical to reference device K200209.
- Comprehensive Taper Adapter: Identical to predicate device
- Manufacturing Process: Similar to predicate devices
- Alliance Humeral Bearings: Machined from Vitamin E-stabilized highly crosslinked polyethylene; marked, cleaned, and packaged; sterilized via ethylene oxide. Similar to primary predicate polyethylene bearings.
- Identity Trays: Machined from wrought titanium alloy; marked, cleaned and packaged; sterilized via gamma irradiation. Similar to primary predicate humeral trays.
- Comprehensive Reverse Trays: Machined from wrought titanium alloy; marked, cleaned and packaged; sterilized via gamma irradiation. Similar to predicate device humeral trays.
- Comprehensive Glenospheres: Machined from wrought titanium or cobalt chrome alloy with polished surfaces; titanium variants include surface hardening comparable to predicate devices; cleaned, packaged, and sterilized via gamma irradiation. Similar to predicate device glenospheres.
- Comprehensive Taper Adapter: Machined from wrought titanium alloy; marked, cleaned and packaged; sterilized via gamma irradiation. Similar to predicate device taper adapters.
- Design Features: Similar to predicate devices
- Alliance Humeral Bearings: Polyethylene components that articulate with glenospheres and engage with humeral trays through a locking mechanism. Offered in multiple configurations to support joint stability and motion. Design and function are similar to primary predicate humeral bearing components.
- Identity Trays: Metallic components that interface with the humeral stem and support modular bearing attachment through a locking mechanism. Offered in multiple configurations to accommodate patient anatomy. Design and function are similar to primary predicate humeral trays.
- Comprehensive Reverse Trays: Metallic components that interface with the humeral stem and support modular bearing attachment through a locking mechanism. Offered in multiple configurations to accommodate patient anatomy. Design and function are similar to predicate device humeral trays.
{6}
o Comprehensive Glenospheres: Metallic spherical components that articulate with humeral bearings and connect to the baseplate via a modular taper adapter. Offered in multiple configurations to support joint stability and motion. Design and function are similar to predicate device glenosphere components.
o Comprehensive Taper Adapter: Metallic component that provides a modular connection between the glenosphere and baseplate through dual taper interfaces. Design and function are similar to predicate device taper adapter components.
- Packaging: Similar to predicate devices
o Alliance Humeral Bearings: Similar to primary predicate, both packaged in double sterile tray with protective packaging.
o Identity Trays: Identical to primary predicate, however primary predicate includes a physical IFU, while subject device has an eIFU.
o Comprehensive Reverse Trays: Similar to predicate device, both packaged in double sterile configuration with protective packaging.
o Comprehensive Glenospheres: Similar to predicate device, both packaged in double sterile tray with protective packaging.
o Comprehensive Taper Adapter: Similar to predicate device, both packaged in double sterile configuration with protective packaging.
- Sterilization Method: Identical to predicate device
o Alliance Humeral Bearings: Sterilized via ethylene oxide. Method is identical to primary predicate crosslinked polyethylene bearing component
o Identity Trays: Sterilized via gamma irradiation. Method is identical to the primary predicate humeral tray components.
o Comprehensive Reverse Trays: Sterilized via gamma irradiation. Method is identical to predicate device humeral tray components.
o Comprehensive Glenospheres: Sterilized via gamma irradiation. Method is identical to predicate device glenosphere components.
o Comprehensive Taper Adapter: Sterilized via gamma irradiation. Method is identical to predicate device taper adapter components.
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-Clinical Tests/Justifications:
Non-clinical testing and engineering evaluations were conducted to assess mechanical performance and functional integrity of the subject device under worst case conditions relative to predicate devices.
- Finite Element Analysis (FEA): Used to identify worst case configurations and loading conditions.
- Fatigue Strength Evaluation: Conducted to assess durability of components and constructs under cyclic loading
- Accelerated Corrosion Fatigue (ACF) Testing: Performed to evaluate fretting and corrosion of modular junctions as well as construct strength.
- Micromotion & Moment Analysis: Conducted to evaluate construct strength, modular junction demand and initial fixation behavior under physiological loading conditions.
- Lever Out Resistance (ASM F1820): evaluated locking mechanism resistance to disengagement.
- Static Axial and Torsional Strength (ASTM F2009, ASTM F1820): Evaluated strength of modular connections under axial and torsional loading.
- Wear Assessment (ASTM F1378): Evaluated articulating geometries, materials, and surface characteristics to confirm no new worst case condition.
- Range of Motion Analysis (ASTM F1378): Evaluated system-level range of motion and potential impingement.
- Magnetic Resonance Imaging (MRI, ASTM F2052, F2213, F2119): Conducted to establish MR-conditional use.
- rTSA Humeral Construct Performance Requirement: rTSA performance requirements were defined using predicate device strength and supported by biomechanical literature.
The subject devices have the same intended use and similar indications for use as the predicate device. The subject devices have similar technological characteristics to the predicates.
Non-clinical testing demonstrates that the subject device meets design requirements and performs as intended under worst case conditions. Results confirm that the subject device is equivalent to or bounded by predicate device performance, including fatigue strength, modular connection integrity, resistance to fretting and corrosion and construct level initial fixation under physiological loading conditions. No new risk or failure modes were identified.
No animal or clinical testing was required.
The subject device demonstrates safety and effectiveness equivalent to the legally marketed predicate device.
and the information provided herein demonstrates that:
- any differences do not raise different questions of safety and effectiveness; and
- the subject devices are as safe and effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.