The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field and multi-focal electroretinogram (ERG) recording procedures. The RM Electrode® is a single use, disposable device.
Device Story
RM Electrode (RMH 25-01) is a passive, single-use, contact lens-type corneal electrode; used in clinical settings (hospitals, clinics, research labs) by ophthalmologists, optometrists, or trained technicians. Device input: bioelectric signals from eye surface during ERG procedures. Device output: transmitted electrical signals to external, commercially available ERG recording system via 1.5mm touchproof connector. Operation: device filled with lubricating ophthalmic solution, placed on cornea; lead wire draped over ear. Features: soft silicone rubber contact lens, recessed electrode ring, and shape feature for orientation in dim lighting. Benefit: enables recording of full-field and multi-focal ERG signals to support diagnosis of retinal dysfunctions; recessed electrode minimizes direct corneal contact, reducing abrasion risk. Healthcare providers use recorded signals to assess retinal function impairment.
Clinical Evidence
Clinical testing included: 1) Full-field ERG comparison (n=5) between subject and primary predicate; amplitudes (a-wave, b-wave, flicker) and SNR were equivalent. 2) Multi-focal ERG comparison (n=10) between subject and secondary predicate (ERG Jet) using 61-hexagon stimulus; response peak amplitudes were equivalent. 3) Human ocular irritation study (n=10) with 60-minute wear time; bulbar/limbal/tarsal redness and corneal abrasion were equivalent to or less than primary predicate. No adverse events reported in any study.
Technological Characteristics
Passive corneal electrode; soft Class VI silicone rubber contact lens; gold electrode ring (recessed); 1.5mm female touchproof connector. Dimensions: standard contact lens form factor. Connectivity: wired to external ERG recording system. Sterilization: sterile, single-use. No internal power or software. Complies with IEC 60601-1 for electrical safety.
Indications for Use
Indicated for patients aged 12 years and above undergoing diagnostic full-field and multi-focal electroretinogram (ERG) recording procedures to support diagnosis of retinal dysfunctions.
Regulatory Classification
Identification
A corneal electrode is an AC-powered device, usually part of a special contact lens, intended to be applied directly to the cornea to provide data showing the changes in electrical potential in the retina after electroretinography (stimulation by light).
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 1, 2026
Retmap, Inc.
Shresta Patangay
Chief Operating Officer
832 W Superior St., Suite 302
Chicago, Illinois 60642
Re: K253586
Trade/Device Name: RM Electrode (RMH 25-01)
Regulation Number: 21 CFR 886.1220
Regulation Name: Corneal electrode
Regulatory Class: Class II
Product Code: HLZ
Dated: February 25, 2026
Received: February 25, 2026
Dear Shresta Patangay:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253586 - Shresta Patangay
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253586 - Shresta Patangay
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alexander Beylin - Date: 2026.04.01
S 12:06:28 -04'00'
for CAPT Bradley Cunningham, MSE, RAC
Acting Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253586 | ? |
| Please provide the device trade name(s). | | ? |
| RM Electrode (RMH 25-01) | | |
| Please provide your Indications for Use below. | | ? |
| The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the
measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal
dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field and
multi-focal electroretinogram (ERG) recording procedures.
The RM Electrode® is a single use, disposable device. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
RETMAP A CLEAR VISION OF EYE CARE
# 510(k) Summary
# K253586
Premarket Notification (510(k)) Number
| Submitter's Name | RetMap, Inc.
832 W Superior St., Suite 202
Chicago, IL 60642
Telephone: (312) 224-8938 |
| --- | --- |
| Contact Person | Shresta Patangay, PhD
Chief Operating Officer
RetMap, Inc.
832 W Superior St., Suite 202
Chicago, IL 60642 |
| Date Prepared | February 25, 2026 |
| Device Trade Name | RM Electrode
Model Number: RMH 25-01 |
| Device Common Name | Corneal or ERG Electrode |
| Classification Name | Corneal Electrode |
| Product Code | HLZ |
| Device Classification | Class II |
| Panel | Ophthalmology |
| Regulatory Classification | 21 CFR 886.1220 |
| Type of 510(k) Submission | Traditional |
| Legally Marketed Predicate Devices | RM Electrode (K232273) (primary predicate device)
ERG Jet Electrode (K813399) (secondary predicate device) |
{6}
RETMAP A CLEAR VISION OF EYE CARE
# Device Description
The RM Electrode® is a contact lens-type corneal electrode for use in electroretinography (ERG) procedures. Like other corneal electrodes classified under 21 CFR 886.1220 and cleared under product code HLZ, the RM Electrode is used for the measurement and recording of ERG signals, to support the diagnosis of retinal dysfunctions. Like the original version of the device cleared under K232273, the subject device (RM Electrode (RMH 25-01)) has four components, including (1) the contact lens (substrate), which supports the conductive element (electrode ring) and allows light from the stimulus source to enter the eye; (2) the electrode ring, which is the conductive element of the device; (3) the lead wire, which is connected to the electrode ring extension and terminates in an industry standard 1.5mm female "touchproof" connector that connects to a commercially available ERG recording system (not part of the subject device); and (4) heat shrink tubing which covers the electrode ring extension and junction with the lead wire.
The RM Electrode® is not a powered device. As with the primary predicate device (RM Electrode®, K232273), the secondary predicate device (ERG Jet Electrode, K813399), and other commonly used ERG electrodes, the subject RM Electrode® (RMH 25-01) is a passive sensor for recording bioelectric signals from the surface of the eye and transmitting them to an ERG recording system. From an electrical perspective, the device is a single conductor that connects the eye surface to a data recorder.
# Indications for Use
The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field and multi-focal electroretinogram (ERG) recording procedures.
The RM Electrode® is a single use, disposable device.
# Primary Predicate Device
The RM Electrode® (RMH 25-01) is substantially equivalent to the RM Electrode® cleared under K232273 (primary predicate device). Both devices are electrodes intended to be applied on the cornea for the measurement and recording of ERG signals, to support the diagnosis of retinal dysfunctions. Both the subject and primary predicate devices have the same intended use and nearly identical technological characteristics, but the subject RM Electrode® (RMH 25-01) has optical characteristics that are suitable for use in multi-focal ERG testing (as well as the previously cleared use of the predicate device in full-field ERG testing), and a shape feature
{7}
RETMAP A CLEAR VISION OF EYE CARE
intended to improve user workflow in distinguishing the eye-contact side from the distal (non-eye contact) side under the dim lighting conditions often present when preparing ERG electrodes prior to installation.
## Description of Operation
As part of an ERG procedure, the device is filled with a lubricating ophthalmic solution and placed on the eye. The lead wire is gently draped over the ipsilateral ear, secured with tape if needed, and the connector is plugged into a compatible ERG recording system. The device can be used for any full-field or multi-focal ERG protocol to record bioelectric signals from the surface of the eye and transmit them to the ERG data recording system. After the ERG test procedure, the device is removed from the patient's eye.
## Technological Characteristics Summary
The subject RM Electrode (RMH 25-01) is substantially equivalent to the currently marketed RM Electrode (K232273) and the secondary predicate device (ERG Jet Electrode, K813399) with respect to intended use, design, principle of operation, technological characteristics and performance. Other than the shape feature included in the subject device described above), the overall shape and design are identical to the primary predicate device. Both the subject and primary predicate RM Electrodes include a soft contact lens (eye contact) portion, electrode, and lead wire with connector. Like the secondary predicate device, the subject device (RM Electrode (RMH 25-01)) can be used for both multi-focal and full-field ERG procedures. These differences do not raise new questions of safety or effectiveness.
{8}
RETMAP A CLEAR VISION OF EYE CARE
# Substantial Equivalence Summary Comparison Table
| Features | RM Electrode (RMH 25-01) (Subject Device) | RM Electrode Primary Predicate Device (K232273) | ERG Jet Electrode Secondary Predicate Device (K813399) | Comparison |
| --- | --- | --- | --- | --- |
| Intended Use | Corneal (contact lens) electrode to transmit the retinal electrical signal in diagnostic ERG procedures | Corneal (contact lens) electrode to transmit the retinal electrical signal in diagnostic ERG procedures | Corneal (contact lens) electrode to transmit the retinal electrical signal in diagnostic ERG procedures | Identical |
| Indications for Use | The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field and multi-focal electroretinogram (ERG) recording procedures. | The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field electroretinogram (ERG) recording procedures. | The ERG-Jet supports the diagnosis of retinal dysfunctions, such as the assessment of retinal function impairment that can occur with retinal disease and/or neurological conditions.
The ERG-Jet is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It can be | Equivalent.
Both the subject RM Electrode (RMH 25-01) and the primary predicate device are indicated for full-field ERG testing in patients aged 12 years and above.
Both the subject device and the secondary predicate device are intended for use in full-field and multi-focal ERG procedures. |
{9}
RETMAP A CLEAR VISION OF EYE CARE
| Features | RM Electrode (RMH 25-01) (Subject Device) | RM Electrode Primary Predicate Device (K232273) | ERG Jet Electrode Secondary Predicate Device (K813399) | Comparison |
| --- | --- | --- | --- | --- |
| | The RM Electrode® is a single use, disposable device. | The RM Electrode® is a single use, disposable device. | used to perform either full-field or multifocal ERG assessment. | The subject and both predicate devices are single-use, disposable devices. |
| Product Code | HLZ | HLZ | HLZ | Identical |
| Target Population | Patients aged 12 years and above undergoing ERG recording procedures | Patients aged 12 years and above undergoing ERG recording procedures. | The target patients are adults and adolescents that require an ERG testing. | Identical.
The subject and both predicate devices are used in adolescent and adult patients undergoing ERG procedures. |
| Types of ERG Procedures | Full-field and multi-focal ERG protocols | Full-field ERG protocols | Full-field and multi-focal ERG protocols. | The currently marketed RM Electrode (primary predicate device) is intended for full-field ERG protocols only.
The subject RM Electrode (RMH 25-01) and secondary predicate are both intended for use in multi-focal and full-field ERG procedures. |
5
{10}
RETMAP A CLEAR VISION OF EYE CARE
| Features | RM Electrode (RMH 25-01) (Subject Device) | RM Electrode Primary Predicate Device (K232273) | ERG Jet Electrode Secondary Predicate Device (K813399) | Comparison |
| --- | --- | --- | --- | --- |
| Wear Duration | Typically 20 minutes; range 5-60 minutes.
The RM Electrode® (RMH 25-01) is not indicated to be used for more than one hour. | Typically 20 minutes; range 5-60 minutes.
The RM Electrode® is not indicated to be used for more than one hour. | Typically 20 minutes; range 5-60 minutes
The ERG-Jet is not indicated to be used more than one hour. | Identical |
| Intended Users | Ophthalmologists, optometrists, trained medical technicians and professionals and vision science researchers | Ophthalmologists, optometrists, trained medical technicians and professionals and vision science researchers | The user(s) profile are ophthalmologists and optometrists (under the supervision of an ophthalmologist) practicing in ophthalmic hospitals, clinics, and medical settings. | Equivalent. |
| Environment of Use | Hospitals, clinics, physician offices, research labs | Hospitals, clinics, physician offices, research labs | Ophthalmic hospitals, clinics, and medical settings. | Equivalent. |
| Types of Device Components | Contact lens, electrode ring, lead wire with 1.5mm touchproof connector that connects to ERG data recorder. | Contact lens, electrode ring, lead wire with 1.5mm touchproof connector that connects to ERG data recorder. | Contact lens that is highly transparent in the visible spectrum.
Gold film electrode ring | The subject and primary predicate device have identical types of components.
The subject device additionally includes a shape feature to |
6
{11}
RETMAP A CLEAR VISION OF EYE CARE
| Features | RM Electrode (RMH 25-01) (Subject Device) | RM Electrode Primary Predicate Device (K232273) | ERG Jet Electrode Secondary Predicate Device (K813399) | Comparison |
| --- | --- | --- | --- | --- |
| | Subject device additionally has a shape feature intended to improve user workflow in distinguishing the eye-contact side from the distal (non-eye contact) side under the dim lighting conditions often present when preparing ERG electrodes prior to installation | | Lead wire approximately 1 meter in length with 1.5mm touchproof connector that connects to ERG data recorder. | improve user workflow by making the eye-contact and distal sides of the device easier to distinguish during device preparation prior to installation.
The subject and secondary predicate devices have equivalent types of components. |
| Materials | Biocompatible materials | Biocompatible materials | Biocompatible materials | Equivalent materials
The material of the subject and primary predicate RM Electrodes that contacts the eye is a soft Class VI silicone rubber, which minimizes the risk of corneal abrasion.
The materials of the secondary predicate |
7
{12}
RETMAP A CLEAR VISION OF EYE CARE
| Features | RM Electrode (RMH 25-01) (Subject Device) | RM Electrode Primary Predicate Device (K232273) | ERG Jet Electrode Secondary Predicate Device (K813399) | Comparison |
| --- | --- | --- | --- | --- |
| | | | | device that contact the eye are hard acrylic and the metal electrode ring.
In the subject and primary predicate RM Electrodes, the electrode ring is recessed within the contact lens portion so that it makes no direct contact with the eye. |
| Compatible ERG Recorders | ERG recording system accepting 1.5mm touchproof connector that complies with all applicable device safety standards, such as IEC 60601-1 | ERG recording system accepting 1.5mm touchproof connector that complies with all applicable device safety standards, such as IEC 60601-1 | All ERG recording systems accepting 1.5mm touchproof connectors that comply with all applicable device safety standards, such as IEC 60601-1 | Identical |
8
{13}
RETMAP A CLEAR VISION OF EYE CARE
# Performance Testing
As described below, a comprehensive battery of ocular irritation, bench, and clinical studies was submitted to confirm product conformance with device requirements and substantial equivalence to the predicate. Performance data demonstrated the absence of ocular irritation in humans and rabbits and substantially equivalent performance to the predicate device(s) (as applicable) in terms of ERG signal quality, lead wire integrity, electrical impedance, and optical characteristics.
Clinical performance testing comparing performance of the subject device in multi-focal ERG (mfERG) procedures to the secondary predicate device (ERG Jet Electrode cleared under K813399), which is another legally marketed ERG electrode indicated for use in mfERG procedures, demonstrated equivalence, as did clinical testing comparing performance of the subject and predicate RM Electrodes® in full-field ERG procedures.
The ocular irritation, bench, shelf life, and clinical studies summarized below concluded that the device is as safe, as effective, and performs as well or better than the legally marketed predicate device.
# Ocular Irritation Testing
Ocular irritation in human participants was evaluated following 60 minutes of wear time (approximately 3 times the typical 20-minute wear duration) in 10 subjects with both the subject and primary predicate devices. Bulbar redness, limbal redness, tarsal redness, and corneal abrasion associated with wear of the subject device were found to be equivalent to, or less than, that associated with wear of the predicate device.
Additionally, ocular irritation testing was successfully completed in rabbits per ISO 10993-23 Biological evaluation of medical devices – Part 23: Tests for irritation. The test articles were modified RM Electrodes (subject device, model RMH 25-01) in their final, finished, sterilized form. Testing was performed by an independent third-party testing laboratory. All measures of irritation of the cornea, iris and conjunctiva, in all six rabbits, and at all four time points (1, 24, 48 and 72 hours after exposure) were identical to the control eyes.
# Bench Testing
The subject and primary predicate devices were tested under tensile load to determine the failure load for the lead wire-electrode ring junction. The subject device failure load was substantially equivalent to the predicate device. In addition, all samples of the subject device met the prospectively defined acceptance criteria for the minimum failure load.
{14}
RETMAP A CLEAR VISION OF EYE CARE
The subject and primary predicate devices were also evaluated for electrical impedance measured at three frequencies within the passband of the ERG signal. At all frequencies, the electrical impedance of the subject device was substantially equivalent to the predicate device.
Additionally, relevant device optical properties (local blur and distortion) were compared between the subject, primary, and secondary predicate devices to demonstrate equivalence and appropriateness for the subject device intended use.
## Shelf-Life Testing
The subject device and its packaging successfully completed evaluations after accelerated aging to support the labeled shelf life. Integrity of the sterile packaging was confirmed before and after accelerated aging through seal strength and bubble leak testing. Device properties were confirmed before and after accelerated aging through 1) evaluation of material hardness of the eye contact portion of the device using a calibrated durometer, 2) evaluation of the electrical impedance of the device measured at three frequencies within the passband of the ERG signal and 3) evaluation of the failure load of the lead wire-electrode junction. All samples met the acceptance criteria demonstrating no significant changes over the labeled shelf life. All samples tested met the prospectively defined acceptance criteria for impedance, material hardness and failure load.
## Clinical Testing
Clinical performance testing was compared using the subject and primary predicate devices in 5 subjects who sequentially wore both devices in the same eye in a randomized order. Subjects were healthy with no history of retinal disease or corneal surgery. Full-field flash (ISCEV LA 3.0 stimulus) and 30 Hz flicker ERG protocols were followed. Amplitudes (a-wave, b-wave and flicker) and signal to noise ratios were substantially equivalent for both devices. There were no adverse events or complications related to the subject device.
Additional clinical performance testing in 10 subjects compared the subject device and the secondary predicate device (ERG Jet Electrode – K813399) for performance in multi-focal ERG (mfERG) testing. A 61-hexagon stimulus conforming to ISCEV standards for clinical electroretinography was used. Subjects wore the subject and secondary predicate devices in the same eye in a randomized order. Amplitudes of response peaks were substantially equivalent for both devices, thus demonstrating suitability of the proposed use of the device in mfERG procedures. As noted above, the secondary predicate (ERG Jet Electrode) was used for this comparative testing since it is indicated for use in mfERG testing while the primary predicate RM Electrode® is indicated for use in only full-field ERG procedures. There were no adverse events or complications related to the subject device.
10
{15}
RETMAP A CLEAR VISION OF EYE CARE
In addition, as noted in the Ocular Irritation Testing section above, human ocular irritation was completed following 60 minutes of wear time (approximately 3 times the typical wear duration) in 10 subjects with both the subject and primary predicate devices. Subjects were normally sighted aged 18 years and above. Bulbar redness, limbal redness, tarsal redness, and corneal abrasion associated with wear of the subject device were found to be equivalent to, or less than, that associated with wear of the primary predicate device. There were no adverse events or complications related to the study device.
## Summary
In summary, the RM Electrode (RMH 25-01) is substantially equivalent to the primary predicate RM Electrode (currently marketed, K232273) and the secondary predicate (ERG Jet Electrode, K813399) with respect to intended use, technological characteristics and performance.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.