STRATAFIX™ Spiral PRONOVA™ Unidirectional Knotless Tissue Control Device STRATAFIX™ Spiral PRONOVA™ Bidirectional Knotless Tissue Control Device
K253572 · ETHICON, Inc. · GAW · Mar 27, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253572
Device Name
STRATAFIX™ Spiral PRONOVA™ Unidirectional Knotless Tissue Control Device STRATAFIX™ Spiral PRONOVA™ Bidirectional Knotless Tissue Control Device
Applicant
ETHICON, Inc.
Product Code
GAW · General, Plastic Surgery
Decision Date
Mar 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis. The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis.
Device Story
STRATAFIX™ Spiral PRONOVA™ Knotless Tissue Control Devices (KTCD) are monofilament, non-absorbable synthetic sutures; composed of poly(vinylidene fluoride) and poly(vinylidene fluoride-co-hexafluoropropylene) blend. Devices feature mechanical barbs cut into the suture core to anchor tissue without surgical knots. Unidirectional version includes a fixation loop; bidirectional version features a transition zone with opposing barbs. Used by trained healthcare professionals in surgical settings for soft tissue approximation. Blue pigment enhances visibility. Single-use; ethylene oxide sterilized. Eliminates need for knot-tying; provides secure tissue approximation.
Clinical Evidence
No clinical study data included. Evidence consists of non-clinical bench testing (USP <861> diameter, <871> needle attachment, <881> tensile strength), biocompatibility testing per ISO 10993-1, and a GLP-compliant pre-clinical animal study evaluating wound healing performance compared to predicate devices.
Technological Characteristics
Monofilament synthetic suture; polymer blend of poly(vinylidene fluoride) and poly(vinylidene fluoride-co-hexafluoropropylene). Mechanically barbed; blue pigment (Color Index 74160). Sterilized via ethylene oxide. Complies with USP <861>, <871>, and <881> standards. Single-use.
Indications for Use
Indicated for soft tissue approximation in patients requiring non-absorbable barbed sutures; contraindicated for surface closures through the epidermis.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 27, 2026
Ethicon, Inc.
Jenny Wang
Senior Regulatory Specialist
1000 Us-202
Raritan, New Jersey 08869
Re: K253572
Trade/Device Name: STRATAFIX™ Spiral PRONOVA™ Unidirectional Knotless Tissue Control Device
STRATAFIX™ Spiral PRONOVA™ Bidirectional Knotless Tissue Control Device
Regulation Number: 21 CFR 878.5010
Regulation Name: Nonabsorbable Polypropylene Surgical Suture
Regulatory Class: Class II
Product Code: GAW
Dated: February 27, 2026
Received: February 27, 2026
Dear Jenny Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253572 - Jenny Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253572 - Jenny Wang
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE
-S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253572 | |
| Device Name STRATAFIX™ Spiral PRONOVA™ Unidirectional Knotless Tissue Control Device STRATAFIX™ Spiral PRONOVA™ Bidirectional Knotless Tissue Control Device | |
| Indications for Use (Describe) | |
| The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis. | |
| The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 7
510(k) Summary
(K253572)
Submitter: Ethicon Inc. a Johnson & Johnson company
1000 US-202
Raritan, New Jersey 08869
Contact Person: Jenny Wang
Ethicon, Inc. a Johnson & Johnson company
e-mail: jwang347@its.jnj.com
Phone: (646)896-9786
Date Prepared: December 17, 2025
Device Trade Name: STRATAFIX™ Spiral PRONOVA™ Unidirectional Knotless Tissue Control Device
STRATAFIX™ Spiral PRONOVA™ Bidirectional Knotless Tissue Control Device
Device Common Name: Nonabsorbable polypropylene surgical suture
Class: II
Classification Name: Suture, Nonabsorbable, Synthetic, Polypropylene
(21 CFR 878.5010)
Product Code: GAW
| Legally Marketed Predicate Devices | 510(k) Number | Product Code |
| --- | --- | --- |
| Primary Predicate Device:
QUILL™ NONABSORBABLE POLYPROPYLENE BARBED SUTURE (Bidirectional)
Quill™ Polypropylene Knotless Tissue-Closure Device, Variable Loop Design (Unidirectional) | K052373
K151112 | GAW
GAW |
| Reference Predicate Devices (for Material):
PRONOVA™ NONABSORBABLE SUTURE, USP | K001625 | GAW |
* The predicate devices have not been subject to a design-related recall.
{5}
Page 2 of 7
# Device Description:
The STRATAFIX™ Spiral PRONOVA™ Knotless Tissue Control Devices (KTCD) are monofilament, synthetic non-absorbable devices composed of a polymer blend of poly (vinylidene fluoride) and poly(vinylidene fluoride-co-hexafluoropropylene). The STRATAFIX™ Spiral PRONOVA™ KTCD are pigmented blue to enhance visibility in the surgical field. The STRATAFIX™ Spiral PRONOVA™ KTCD has two variations based on the orientation of the barbs and needle attachment: STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD and STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD.
The STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD consists of barbed suture material, armed with a surgical needle on one end and a fixation loop at the opposite end. The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is designed to anchor with a closed loop at one end and a unidirectional barbed section on the other end. The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.
The STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD Design consists of barbed suture material, armed with a surgical needle on each end. The device also contains an unbarbed center transition zone that facilitates the initiation of the device use. The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device barbs allow for tissue approximation without the need to tie surgical knots.
The subject devices are for single-use and are sterilized by Ethylene Oxide. The subject devices are available in various lengths and USP diameter sizes armed with various needle sizes. STRATAFIX™ Spiral PRONOVA™ KTCD is intended for use only by Healthcare Professionals who are trained in surgical suturing techniques.
# Indications for Use:
The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis.
The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis.
The subject devices and the primary predicate devices are all indicated for soft tissue approximation.
{6}
# Comparison of Technological Characteristics with the Predicate Devices:
Substantial Equivalence Comparison Table:
| Technological Characteristic | Subject Devices: STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD | Primary Predicate Devices: QUILL™ Polypropylene Knotless Tissue Closure Device, Variable Loop Design (Unidirectional) K151112 | Reference Predicate Device for Material: PRONOVA™ NONABSORBABLE SUTURE, USP (K001625) | Conclusion |
| --- | --- | --- | --- | --- |
| | STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD | QUILL™ NONABSORBABLE POLYPROPYLENE BARBED SUTURE (Bidirectional) K052373 | | |
| Indications for Use | The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis. | Quill Unidirectional: Quill™ Knotless Tissue-Closure Device comprised of Polypropylene is indicated for soft tissue approximation, excluding closure of the epidermis. | PRONOVA™ Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and neurological procedures. | EQUIVALENT |
| | The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis. | Quill Bidirectional: Quill™ Nonabsorbable Polypropylene Barbed Sutures are indicated for use in soft tissue approximation, excluding closure of the epidermis. | | |
| Contraindications | The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is not indicated for surface closures through the epidermis as the barbs make the STRATAFIX™ Spiral PRONOVA™ Unidirectional Device removal unfeasible. | Quill Unidirectional and Bidirectional: Quill™ Barbed Suture comprised of polypropylene is not indicated for surface closures through the epidermis as the small opposing facing barbs make Quill™ Barbed Suture removal unfeasible. | None known. | EQUIVALENT |
| | The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device is not indicated for surface closures through the epidermis as the small opposing facing barbs make the | | | |
{7}
| Technological Characteristic | Subject Devices: STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD | Primary Predicate Devices: QUILL™ Polypropylene Knotless Tissue Closure Device, Variable Loop Design (Unidirectional) K151112 | Reference Predicate Device for Material: PRONOVA™ NONABSORBABLE SUTURE, USP (K001625) | Conclusion |
| --- | --- | --- | --- | --- |
| | STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD | QUILL™ NONABSORBABLE POLYPROPYLENE BARBED SUTURE (Bidirectional) K052373 | | |
| | STRATAFIX™ Spiral PRONOVA™ Bidirectional Device removal unfeasible. | | | |
| Technological Means for Achieving Intended Use | Mechanically barbed monofilament strand to approximate tissue anchored without the need to tie surgical knots. | Mechanically barbed monofilament strand to approximate tissue anchored without the need to tie surgical knots. | Monofilament strand to approximate tissue anchored with surgical Knots. | EQUIVALENT |
| Suture Design | The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device: Monofilament synthetic suture Single armed Unidirectional barbs made by cutting into the core of the suture strand Contains a welded primary loop and secondary loop | Quill Unidirectional: Monofilament synthetic suture Single armed Unidirectional barbs made by cutting into the core of the suture strand Contains a welded primary loop and secondary loop | Monofilament synthetic suture strand Single armed | EQUIVALENT |
| | The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device: Monofilament synthetic suture Double armed Opposite facing barbs from the non-barbed center transition zone | Quill Bidirectional: Monofilament synthetic suture Double armed Opposite facing barbs from the non-barbed center transition zone | | |
| Suture Material | Sterile, monofilament, synthetic non-absorbable device composed of a polymer blend of poly(vinylidene fluoride) and poly(vinylidene fluoride-co-hexafluoropropylene). | The Quill” Barbed Suture is comprised of an isotactic polypropylene polymer of high molecular weight. | Sterile, monofilament, synthetic, non-absorbable surgical suture composed of a polymer blend of poly (vinylidene fluoride) and | EQUIVALENT |
Page 4 of 7
{8}
| Technological Characteristic | Subject Devices: STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD | Primary Predicate Devices: QUILL™ Polypropylene Knotless Tissue Closure Device, Variable Loop Design (Unidirectional) K151112 | Reference Predicate Device for Material: PRONOVA™ NONABSORBABLE SUTURE, USP (K001625) | Conclusion |
| --- | --- | --- | --- | --- |
| | STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD | QUILL™ NONABSORBABLE POLYPROPYLENE BARBED SUTURE (Bidirectional) K052373 | | |
| | | | poly (vinylidene fluoride-co-hexafluoropropylene). | |
| Suture Colorant | Pigmented blue with phthalocyanine blue (Color Index Number 74160) to enhance visibility | The pigment for the blue dyed suture material is Phthalocyaninato (2-) Copper | Pigmented blue with phthalocyanine blue (Color Index Number 74160) to enhance visibility | EQUIVALENT |
| USP Suture Size | The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device: 0, 2-0, and 3-0 | Quill Unidirectional: 2-0 through 2 | Sizes 10-0 through 8-0 and 6-0 through 2 | EQUIVALENT |
| | The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device: 0 and 2-0 | Quill Bidirectional: 3-0 through 2 | | |
| Suture Diameter | The suture diameters comply with diameter requirements listed in USP <861> Diameter. | The suture diameters comply with diameter requirements listed in USP <861> Diameter. | The suture diameters comply with diameter requirements listed in USP <861> Diameter. | EQUIVALENT |
| Tensile Strength | Straight Tensile meets USP <881> Knot Tensile requirements for Synthetic Non-absorbable Sutures of the same size equivalent | Strength of the Quill™ Barbed Suture can be compared to USP <881> Knot Tensile Strength of non-barbed sutures | Tensile strength complies with the tensile requirement listed in USP <881> Knot Tensile Strength. | EQUIVALENT |
| Needle Pull-Off | Meets USP <871> requirements | Meets USP <871> requirements | Meets USP <871>requirements | EQUIVALENT |
| Absorption Profile | Nonabsorbable | Nonabsorbable | Nonabsorbable | EQUIVALENT |
Page 5 of 7
{9}
| Technological Characteristic | Subject Devices: STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD | Primary Predicate Devices: QUILL™ Polypropylene Knotless Tissue Closure Device, Variable Loop Design (Unidirectional) K151112 | Reference Predicate Device for Material: PRONOVA™ NONABSORBABLE SUTURE, USP (K001625) | Conclusion |
| --- | --- | --- | --- | --- |
| | STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD | QUILL™ NONABSORBABLE POLYPROPYLENE BARBED SUTURE (Bidirectional) K052373 | | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | EQUIVALENT |
| Single Use? | Yes | Yes | Yes | EQUIVALENT |
| FDA Regulation | 21 CFR 878.5010: Nonabsorbable polypropylene surgical suture | 21 CFR 878.5010: Nonabsorbable polypropylene surgical suture | 21 CFR 878.5010: Nonabsorbable polypropylene surgical suture | EQUIVALENT |
| FDA Product Code | GAW | GAW | GAW | EQUIVALENT |
FDA’s Guidance Document “The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]” issued on July 28, 2014 was followed and it was concluded that based on the intended use, technological characteristics, and safety and performance testing, the subject devices have been shown to be appropriate for its intended use and are considered to be substantially equivalent to the legally marketed predicate devices as demonstrated in the above Substantial Equivalence Comparison table.
The slight technological differences between the subject devices and the predicate devices do not raise new questions of safety or effectiveness. The subject device met all testing criteria to demonstrate substantial equivalence to the predicate devices.
Page 6 of 7
{10}
Page 7 of 7
# Performance Data:
Mechanical bench-top testing, animal testing, and biocompatibility testing collectively demonstrated that the subject devices performed as intended and as claimed and are substantially equivalent to the predicate devices.
Non-clinical laboratory performance testing was performed demonstrating that the subject device conforms to the current USP Monograph for non-absorbable surgical sutures, including <861> suture diameter, <871> suture attachment, and <881> tensile strength. The testing was performed in accordance with FDA’s Guidance Document: “Class II Special Controls Guidance Document: Surgical Sutures” issued on June 3, 2003.
Additionally, Wound Holding related testing was conducted and the results demonstrate the substantial equivalence of STRATAFIX™ Spiral PRONOVA™ Unidirectional and Bidirectional Knotless Tissue Control Devices (KTCD) compared to the Primary Predicate Devices.
The biocompatibility evaluation for the subject devices was conducted in accordance with International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process”, as recognized by FDA.
Furthermore, GLP compliant pre-clinical animal study was performed to evaluate wound healing. The study outcomes demonstrated substantial equivalence of the subject devices, STRATAFIX™ Spiral PRONOVA™ Unidirectional and Bidirectional Knotless Tissue Control Devices, to the performances of the Primary and Reference Predicate Devices.
This submission does not include Clinical Study data.
# Conclusion:
STRATAFIX™ Spiral PRONOVA™ Unidirectional and Bidirectional Knotless Tissue Control Devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.