The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adults by using the wrist cuff. The device can be used in medical facilities or at home, and only for indoor use. It is supplied for OTC use. The device is not intended for use in patients with arrhythmias or in pregnant patients.
Device Story
Fully automatic, non-invasive oscillometric blood pressure monitor; wrist-worn cuff with integrated pressure sensor/transducer. Device analyzes arterial pulsations to calculate systolic/diastolic pressure and pulse rate. Results displayed on LED screen; stored in internal memory (120 measurements/user for 2 users). Includes Bluetooth connectivity for data transmission to mobile app (non-medical function). Used in medical facilities or home environments; operated by patients or clinicians. Provides objective BP/pulse data to support clinical decision-making and patient monitoring.
Clinical Evidence
Clinical validation study conducted per ISO 81060-2:2018+A1:2020. Enrolled 85 subjects aged >12 years. Primary endpoints: mean error and standard deviation of differences in systolic and diastolic BP measurements. Results met all ISO 81060-2 limits. No adverse effects reported.
Technological Characteristics
Oscillometric measurement; wrist cuff; ABS housing/keys. Powered by 3.7V/300mAh Li-battery or 5V/1A AC adapter. Type BF applied part. Connectivity: Bluetooth (non-medical). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30. Biocompatibility: ISO 10993-1/-5/-10/-23.
Indications for Use
Indicated for adult patients requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate. Contraindicated for patients with arrhythmias or pregnant patients.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 20, 2026
Shenzhen Jamr Technology Co., Ltd.
Zhuni Chen
RA Manager
A101-301, D101-201, Jamr Science & Technology Park # 2
Guiyuan Rd. Guixiang Community Guanlan St., Longhua District
Shenzhen, Guangdong 518100
China
Re: K253570
Trade/Device Name: Blood Pressure Monitor (W06LT)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: May 3, 2026
Received: May 4, 2026
Dear Zhuni Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253570 - Zhuni Chen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253570 - Zhuni Chen
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN C. BROWNING -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253570 | ? |
| Please provide the device trade name(s). | | ? |
| Blood Pressure Monitor (W06LT) | | |
| Please provide your Indications for Use below. | | ? |
| The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adults by using the wrist cuff. The device can be used in medical facilities or at home, and only for indoor use. It is supplied for OTC use. The device is not intended for use in patients with arrhythmias or in pregnant patients. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
K253570
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92.
The assigned 510(k) number is: K253570
# 1. Submitter's information
Submitter's Names: Shenzhen Jamr Technology Co., Ltd.
Address: A101-301, D101-201, Jamr Science & Technology Park, No. 2 Guiyuan Road, Guixiang Community, Guanlan Street, Longhua District, Shenzhen 518100, PEOPLE'S REPUBLIC OF CHINA
Tel: +86-755-85292057
Applicant Contact: Zhuni Chen
Applicant Contact Email: RA.dept@jamrmed.com
# 2. Device Information
Device Trade Name: Blood Pressure Monitor
Model(s): W06LT
Common Name: Noninvasive blood pressure measurement system
Classification name: System, Measurement, Blood-Pressure, Non-Invasive
Product Code: DXN
Device Class: II
Regulation Number: 870.1130
# 3. Predicate Device Information
510(K) Number: K230409
Trade Name: Wrist Type Blood Pressure Monitor
Model(s): W05,W1101L
Classification name: System, Measurement, Blood-Pressure, Non-Invasive
Product Code: DXN
Device Class: II
Regulation Number: 870.1130
# 4. Device Description
The blood pressure monitor is a fully automatic, non-invasive measurement device using oscillometric methodology to measure systolic pressure, diastolic pressure and pulse rate. The device features an inflatable cuff that wraps around the wrist, with a built-in pressure sensor and transducer that analyze arterial pulsations to determine blood pressure values. Measurement results are clearly displayed on the LED screen.
# 5. Intended Use/Indication for use
The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adults of adults by using the wrist cuff. The device can be used in medical facilities or at home, and only for indoor use. It is supplied for OTC use. The device is not intended for use in patients with arrhythmias or in pregnant patients.
# 6. Comparisons of technological characteristics with the predicate device
The substantial equivalence chart is provided as follows:
Page 1 of 5
{6}
Page 2 of 5
K253570
| Substantial Equivalence Comparison | | | |
| --- | --- | --- | --- |
| Elements of Comparison | Subject Device | Predicate Device (K230409) | Judgment |
| Models | W06LT | W05,W1101L | / |
| Company | Shenzhen Jamr Technology Co., Ltd. | Shenzhen Jamr Technology Co., Ltd. | Same |
| Device Name | Blood Pressure Monitor | Blood Pressure Monitor | Same |
| Product code | DXN | DXN | Same |
| Regulation # | 21CFR 870.1130 | 21CFR 870.1130 | Same |
| Intended use | The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adults by using the wrist cuff. The device can be used in medical facilities or at home, and only for indoor use. It is supplied for OTC use. The device is not intended for use in patients with arrhythmias or in pregnant patients. | This device is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult by using the wrist circumference 12.5-21.5cm, it can be used in medical facilities or at home. It is supplied for OTC use. | Same |
| Anatomical site | Wrist | Wrist | Same |
| Patient population | Adults | Adults | Same |
| Measurement Item | SYS, DYS, Pulse rate | SYS, DYS, Pulse rate | Same |
| Principle | Oscillometric | Oscillometric | Same |
| BP measurement range | Cuff pressure display range:0~295mmHg; Systolic Blood Pressure: 60~230mmHg; Diastolic Blood Pressure: 40~130mmHg | Cuff pressure display range: W05:0~295mmHg; W1101L:0~300mmHg; Systolic Blood Pressure: 60~230mmHg; Diastolic Blood Pressure: 40~130mmHg | Same as W05 |
| BP accuracy | ±3mmHg | ±3mmHg | Same |
| PR measurement range | 40~199 Beats/Min | W05:40~170 Beats/Min W1101L:40~170 Beats/Min | Similar,refer to Note 1 |
| PR measurement accuracy | ±5% of reading | ±5% of reading | Same |
{7}
| Power supply | Powered by a) d.c. 3.7V/ 300mAh lithium battery; b) AC adapter INPUT: a.c. 100-240V 50/60HZ, OUTPUT: d.c. 5V 1A | For W05: Powered by d.c. 3.0V, 2 x1.5V AAA batteries; For W1101L: Powered by a) d.c. 3.7V/ 300mAh lithium battery; b) AC adapter INPUT: a.c. 100-240V 50/60HZ, OUTPUT: d.c. 5V 1A | Same as W1101L |
| --- | --- | --- | --- |
| Degree of protection against electric shock | Type BF applied part | Type BF applied part | Same |
| Model of operation | Continuous operation | Continuous operation | Same |
| Cuff size suitable for arm size | 12.5-21.5cm | 12.5-21.5cm | Same |
| Sets of memory | Automatically stores the last 120 measurements for 2 users (total 240) | Automatically stores the last 120 measurements for 2 users (total 240) | Same |
| Irregular heartbeat detector | Yes | Yes | Same |
| Voice | Yes | Yes | Same |
| Operation environment | Temperature: 5°C~40°C; Humidity: 15~90%RH; Atmospheric Pressure:70 kPa~106 kPa | Temperature: 5°C~40°C; Humidity: 15~93%RH; Atmospheric Pressure:70 kPa~106 kPa | Similar,refer to Note 1 |
| Transport/Storage environment | Temperature: -20°C ~ +60°C, Humidity: 10%RH ~ 93%RH; Atmospheric Pressure:70 kPa~106 kPa | Temperature: -20°C ~ +70°C, Humidity: ≤ 93%RH; Atmospheric Pressure:70 kPa~106 kPa | Similar,refer to Note 1 |
| Performance | Compliance with Compliance with IEC 80601-2-30 | Compliance with Compliance with IEC 80601-2-30 | Same |
| Transmit data via Bluetooth to the APP | Yes | No | Different,refer to Note 2 |
| Clinical | Compliance with Compliance with ISO 81060-2 | Compliance with Compliance with ISO 81060-2 | Same |
| Material | ABS housing and ABS keys | ABS housing and ABS keys | Same |
| Biocompatibility | All the patient contacting materials are compliance with ISO 10993-1/-5/-10/-23 | All the patient contacting materials are compliance with ISO 10993-1/-5/-10/-23 | Same |
| Electrical Safety | Compliance with IEC 60601-1 and IEC 60601-1-11 | Compliance with IEC 60601-1 and IEC 60601-1-11 | Same |
| EMC | Compliance with IEC 60601-1-2 | Compliance with IEC 60601-1-2 | Same |
Page 3 of 5
K253570
{8}
Note 1:
Although there are slight differences in the "Operation Environment", "Transport/Storage Environment" and "PR measurement range" between the subject device and the predicate device, both the subject devices and predicate devices meet the requirements of the standards IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, and IEC 80601-2-30. Relevant test reports demonstrate that these differences do not impact the safety or effectiveness of the devices.
Note 2:
There is wireless communication with Bluetooth function low energy on the subject device, this wireless function is not for medical purposes, but an additional function, it can transmit the measurement results to the Bluetooth app for storage. The measurement results are also stored in the Blood Pressure Monitor itself. Therefore, the monitor can achieve medical purposes even if it does not use Bluetooth. Therefore, this function has nothing to do with the clinical performance and safety of the product. Besides, the software validation and performance test and FCC test demonstrated the difference of Bluetooth function does not raise any new performance and safety questions.
Conclusion:
Based on the comparative analysis provided in this submission, it is concluded that the subject device is substantially equivalent to the predicate device in terms of safety and effectiveness.
7. Brief discussions of the non-clinical tests
Performance testing:
The subject device conforms to the following guidances and standards:
- IEC 60601-1: Medical electrical equipment–Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirement for basic safety and essential performance–Collateral standard: Electromagnetic compatibility – Requirements and tests
- IEC 80601-2-30: Medical electrical equipment–Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 60601-1-11: Medical electrical equipment–Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Biocompatibility testing:
The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
The biocompatibility testing includes the following tests:
- Cytotoxicity
- Sensitization
- Irritation
8. Brief discussions of clinical tests
The device was tested in accordance with ISO 81060-2:2018+A1:2020 (Non-invasive sphygmomanometers – Part 2: Clinical validation of automated measurement type).
Clinical study was conducted using appropriately sized cuff, enrolling 85 eligible subjects. The study included participants aged >12 years.
Page 4 of 5
K253570
{9}
The mean error and standard deviation of differences in systolic and diastolic blood pressure measurements were all within the limits specified by ISO 81060-2:2018+A1:2020. No adverse effects or complications were reported.
The test results confirm that the Blood Pressure Monitor fully complies with the requirements of IEC 80601-2-30:2018 and ISO 81060-2:2018+A1:2020.
## Final Conclusion:
The subject device maintains the same core technology and indications for use as the predicate device (K230409). The minor differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device (K230409).
K253570
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.