The Luminoah Flow™ Enteral Nutrition System is intended for the delivery of supplemental nutrition and fluids to the gastrointestinal system including nutritional formula, fluids, and/or water. The system is intended for use by healthcare professionals (HCP) to laypersons. The Luminoah Flow™ Enteral Nutrition System is indicated for use in individuals who are unable to obtain sufficient nutrition through eating and or swallowing. The system is indicated for use with the Luminoah Flow™ Administration Set and Luminoah branded accessories. The Luminoah Flow™ System can be used by an adult layperson, caretaker, or healthcare professional, for both adult and pediatric patients, ages 2 years and older. This system is not intended for use with neonates. The Luminoah Flow™ Enteral Nutrition System is intended for hospital and acute care settings, as well as for home health use. It is intended to be used in both stationary and ambulatory conditions.
Device Story
Enteral nutrition pump system delivering formula/fluids via rotary peristaltic pumping; utilizes disposable administration sets with ENFit connectors. Operated by HCPs, laypersons, or caretakers in hospital/home settings. Input: user-defined feed rates/volumes via menu-controlled interface. Transformation: motor-driven peristaltic action; optical sensor for air-in-line detection; motor current/pulse width modulation monitoring for occlusion detection. Output: controlled fluid delivery; visual/audible alarms/notifications (e.g., occlusion, air-in-line, battery status). Benefits: provides supplemental nutrition for patients unable to ingest food orally; supports ambulatory use.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including software verification/validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), flow rate accuracy (AAMI TIR101:2021), alarm detection, human factors engineering, and biocompatibility (ISO 10993).
Indicated for individuals unable to obtain sufficient nutrition via eating/swallowing; pediatric (ages 2+) and adult patients; hospital, acute care, or home settings; stationary or ambulatory use. Contraindicated for neonates.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets and Kangaroo™ Connect Portal (K153074)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 20, 2026
Luminoah
Edgar (Rawley) Rearick
Director of Quality
1 Morton Dr., Suite 100
Charlottesville, Virginia 22903
Re: K253558
Trade/Device Name: Luminoah FLOW™ Enteral Nutrition System
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: LZH
Dated: March 20, 2026
Received: March 20, 2026
Dear Edgar (Rawley) Rearick:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253558 - Edgar (Rawley) Rearick
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253558 - Edgar (Rawley) Rearick
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jake K. Lindstrom -S
Jake Lindstrom, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253558 | ? |
| Please provide the device trade name(s). | | ? |
| Luminoah FLOW™ Enteral Nutrition System | | |
| Please provide your Indications for Use below. | | ? |
| The Luminoah Flow™ Enteral Nutrition System is intended for the delivery of supplemental nutrition and
fluids to the gastrointestinal system including nutritional formula, fluids, and/or water. The system is
intended for use by healthcare professionals (HCP) to laypersons.
The Luminoah Flow™ Enteral Nutrition System is indicated for use in individuals who are unable to obtain
sufficient nutrition through eating and or swallowing. The system is indicated for use with the Luminoah
Flow™ Administration Set and Luminoah branded accessories.
The Luminoah Flow™ System can be used by an adult layperson, caretaker, or healthcare professional, for
both adult and pediatric patients, ages 2 years and older. This system is not intended for use with neonates.
The Luminoah Flow™ Enteral Nutrition System is intended for hospital and acute care settings, as well as
for home health use. It is intended to be used in both stationary and ambulatory conditions. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
LUMINOAH
# 510(k) Summary – K253558
Applicant / Manufacturer Name: Luminoah
1 Morton Drive, Suite 100
Charlottesville, VA 22930
Corresponding Official: Edgar (Rawley) Rearick
Director of Quality
Luminoah, Inc.
Telephone Number: (540) 480-2074
Email: rrearick@luminoah.com
Date prepared: 04/14/2026
## DEVICE IDENTIFICATION
Trade Names: Luminoah FLOW™ Enteral Nutrition System
Common Name: Enteral Feeding Pump, Infusion Pump
Classification Name: Infusion pump
Regulation Number: 21CFR 880.5725
Classification: Class II
Product Code: LZH
## PREDICATE DEVICE
Trade Names: Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™
Connect Feeding Sets and Kangaroo™ Connect Portal
510(K) Number: K153074
{5}
LUMINOAH
# DEVICE DESCRIPTION
The Luminoah FLOW™ Enteral Nutrition System consists of an enteral nutrition pump and disposable enteral nutrition sets, and Luminoah FLOW™ accessories to deliver formula via rotary peristaltic pumping to provide nutrition for those who do not have the ability to orally ingest food.
The Luminoah Flow™ Administration Set is intended for products with ENFit™ Male Connectors. Do not use with non-ENFit connectors. Do not use the device with small bore connectors from other healthcare applications. The Luminoah Flow has been validated for use with nasally inserted enteral feeding tubes (NG, ND, NJ) and extension sets of size 6 French or greater with up to 2 ports and with internal volumes up to 4.3 mL. It has not been validated with giving sets that include any additional in-line resistance.
# INDICATIONS FOR USE
The Luminoah Flow™ Enteral Nutrition System is intended for the delivery of supplemental nutrition and fluids to the gastrointestinal system including nutritional formula, fluids, and/or water. The system is intended for use by healthcare professionals (HCP) to laypersons.
The Luminoah Flow™ Enteral Nutrition System is indicated for use in individuals who are unable to obtain sufficient nutrition through eating and or swallowing. The system is indicated for use with the Luminoah Flow™ Administration Set and Luminoah branded accessories.
The Luminoah Flow™ System can be used by an adult layperson, caretaker, or healthcare professional, for both adult and pediatric patients, ages 2 years and older. This system is not intended for use with neonates.
The Luminoah Flow™ Enteral Nutrition System is intended for hospital and acute care settings, as well as for home health use. It is intended to be used in both stationary and ambulatory conditions.
# SUBSTANTIAL EQUIVALENCE COMPARISON
The Luminoah FLOW™ Enteral Nutrition System, comprised of the FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets, and FLOW™ Accessories is substantially equivalent for its intended use, use conditions and use environment compared to the currently marketed predicate Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets and Kangaroo™ Connect Portal. The table below is a comparison of the subject and predicate device regarding substantial equivalence.
{6}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Indications For Use | The Kangaroo™ Connect. Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets is intended to deliver nutritional formula to the gastrointestinal system of a patient age infant and older who is physically unable to eat and swallow. Not for use with neonates. The feeding pump and feeding sets are intended to be used in clinical or home care settings by users ranging from laypersons to physicians. | The Luminoah Flow™ Enteral Nutrition System is intended for the delivery of supplemental nutrition and fluids to the gastrointestinal system including nutritional formula, fluids, and/or water. The system is intended for use by healthcare professionals (HCP) to laypersons.
The Luminoah Flow™ Enteral Nutrition System is indicated for use in individuals who are unable to obtain sufficient nutrition through eating and/or swallowing. The system is indicated for use with the Luminoah Flow™ Administration Set and Luminoah branded accessories.
The Luminoah Flow™ System can be used by an adult layperson, caretaker, or healthcare professional for both adult and pediatric patients, ages 2 years and older. This system is not intended for use with neonates.
The Luminoah Flow™ Enteral Nutrition System is intended for hospital and acute care settings as well as for home health use. It is intended to be used in both stationary and ambulatory conditions. | General purpose of the subject device, its function, indications for use, and use conditions are the same as predicate device. The FLOW Enteral Nutrition System is designed specifically for patients ages 2 years and older. |
| Sterility | Non-Sterile Feeding sets | Non-Sterile feeding sets | Same |
{7}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Technical Characteristics | The feeding sets are based on peristaltic pumping using a rotating wheel which presses against the tubing and moves the fluid at a controlled rate. The connection to the patient enteral access device is an ENFit Connector compliant to ISO 80369-3. | The feeding sets are based on peristaltic pumping using a rotating wheel which presses against the tubing and moves the fluid at a controlled rate. The connection to the patient enteral access device is an ENFit Connector compliant to ISO 80369-3. | Same |
| Design (pump) | The pump incorporates a menu controlled operating system which contains on board custom software designed to allow the user to set feed rates and volumes as well as other feeding options. The device incorporates ultrasonic sensors to detect the air and blockages in the feeding set. | The pump incorporates a menu controlled operating system which contains on board custom software designed to allow the user to set feed rates and volumes as well as other feeding options. The device incorporates an (1) optical sensor to detect the air in the tubing, and blockages in the feeding set are detected through motor monitoring of current and pulse width modulation. | Difference in design that does not introduce or raise concerns regarding the safe and effective use of the subject device. |
{8}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Design (feeding set) | The pump set incorporates 5 basic segments:
• Fluid reservoir(s), which may be an attached bag (500ml or 1000ml) or a spike for connection to a formula container
• Tubing from fluid reservoir to pump (14 inch)
• Cassette containing pump interface module (peristaltic tubing)
• Tubing from pump to patient connector (66 inches)
• Patient connector (EnFit connector compliant to ISO 80369-3) | The pump set incorporates 2 basic segments:
• Fluid reservoir(s) (250ml or 500ml)
• Pump Head (inclusive of):
- Spike Connector
- Tubing from spike connector to patient connector (15 inches)
- Pump head housing that attaches to FLOW pump
- Patient connector (EnFit connector compliant to ISO 80369-3) | Different in design only – does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Graphic Display | Color TFT (320X240 pixels) | Color TFT 2.8” display | Same |
| Ingress Rating | IP26 | IP24 | Different in design only – does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Free Flow Mechanism | Anti-Free Flow Valve | Anti-Free Flow Mechanism (Vacuum) | Different in design only – does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Feed Capability | Yes | Yes | Same |
| Feed and Flush Capability | No | Yes | Different in design – FLOW includes added functionality to allow users to flush (or ‘rinse’) the administration sets between uses |
| Thick Formula Capability | No | No | Same |
{9}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Occlusion Detection (time), 1-50 mL/hr | Upstream: <=350 minutes | Upstream: <=180 minutes | Different in design – FLOW has a shorter detection threshold (time). Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| | Downstream: <= 95 minutes | Downstream: <= 95 minutes | Same |
| Occlusion Detection (time), >50 mL/hr | Upstream: <=9 minutes | Upstream: <=15 minutes | Different in design – FLOW has increased detection threshold (time). Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| | Downstream: <= 3 minutes | Downstream: <=15 minutes | Different in design – FLOW has increased detection threshold (time). Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Accuracy | ± 5% or 0.5 mL/hr, whichever is larger | ± 5% or 0.5 mL/hr, whichever is larger | Same. The following factors may affect flow rate accuracy: ambient temperature, fluid temperature, pressure (e.g., head-height, backpressure, atmospheric pressure), fluid viscosity. |
| Delivery Rate Range | 1 – 600mL/hr in 1 mL increments | 1 – 600mL/hr in 1 mL increments | Same |
{10}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Dose Range | 1 – 3000mL in 1 mL increments | 1 – 500mL in 1 mL increments | Different in design only - The subject device has a narrower dose range, but the range remains within the predicate device range. Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Power requirements | 120V, 60Hz, 1Amp | 120V, 60Hz, .35Amp | Different in design only – does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Auto prime | Yes | Yes | Same |
| Maximum Occlusion Pressure | 20psi | 20psi | Same |
| Size | 3.9” x 6.1” x1.6” | 2.6” x 5.2” x 1.75” | Different in design only – does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Weight | 0.73lbs (0.33kg) | 0.62lbs (0.28kg) | Different in design only – does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Pole Clamp Mountable | Yes | Yes | Same |
| Battery Type | Rechargeable Lithium Ion | Rechargeable Lithium Ion | Same |
| Battery Life | 24hrs | ≥24hrs | Same |
| Battery re-charge time | 7hrs | ~4.5hrs | Different by design – FLOW has a faster recharge time. Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
{11}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Battery Operation Indicator | Yes | Yes | Same |
| Operating Temperature Range | 5° - 40°C (41° - 104°F) @ 93% R.H. (non-condensing) | 5° - 40°C (41° - 104°F) @ 90% R.H. (non-condensing) | Different - The subject device has a narrower humidity range, but the range remains within the predicate device range. does not introduce or raise concerns regarding the safe and effective use of the subject device |
| Storage Temperature Range | 0° - 50°C (32° - 122°F) @ 93% R.H. (non-condensing) | 0° - 50°C (32° - 122°F) @ 90% R.H. (non-condensing) | Different - The subject device has a narrower humidity range, but the range remains within the predicate device range. does not introduce or raise concerns regarding the safe and effective use of the subject device |
| Medical and Electrical Safety Standards | IEC 60601-1
IEC 60601-1-1
IEC 60601-1-2
IEC 60601-1-6
IEC 60601-1-8
IEC 60601-1-11
IEC 60601-2-24
IEC 62304
IEC 62366 | IEC 60601-1
IEC 60601-1-1
IEC 60601-1-2
IEC 60601-1-6
IEC 60601-1-8
IEC 60601-1-11
IEC 62304
IEC 62366 | Same |
| Degree of Protection against electrical shock | Class 2 Type BF per IEC Standards | Class 2 Type BF per IEC Standards | Same |
| Power Adapter | Yes | Yes | Same |
{12}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Pump Control Alarms and Notifications | **Alarms:**
System Error
Feed Bag Empty
Supply Tube
Blocked
Patient Tube
Blocked
Cassette Dislodged
Cassette Error
Rotor Stuck
Dead Battery
**Notifications:**
Feeding Complete
Feeding
Incomplete
Low Battery
Pump Inactive | **Alarms:**
Motor Stuck
Incomplete Feed
Low Battery
Upstream Occlusion
Downstream Occlusion
Administration Set Error
Dead Battery
Air-in-Line
Stuck Button
Temperature
Feed Interrupted
System Error
**Notifications:**
Feeding Complete
Hydration Complete
End of Priming Sequence
Insufficient Charge
Feeding Paused
Snoozed Alarm | Difference in design that does not introduce or raise concerns regarding the safe and effective use of the subject device |
| Timed Pause Feature | Displayed as KTO VOL delivered at maximum time interval | Not Included | Different – convenience feature not offered by Luminoah FLOW. Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
{13}
LUMINOAH
| Characteristic | Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets (Predicate Device) | Luminoah FLOW™ Enteral Nutrition System: FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ accessories (Subject Device) | Comparison to Predicate |
| --- | --- | --- | --- |
| Materials/Chemical composition | Polyvinyl chloride (PVC)
• Feeding bags and caps
• Tubing
Silicone
• Peristaltic tubing
• Diaphragm valve
Polycarbonate
• Cassette body
Copolyester
• Patient connector
ABS
• Spike
• Tube holder and Cap
Strontium Ferrite / nylon
• Set ID magnets | Nylon
• Pump Housing
Impact Copolymer PP
• Administration Set Housing
Silicone
• Administration Set Tubing
PET/PE blend
• Nutrition Pouch
ABS
• Spike
• Tube holder and Cap
Neodymium Magnets
• Back attachment magnets | Different – Subject device uses different materials that are supported by biocompatibility testing. Does not introduce or raise concerns regarding the safe and effective use of the subject device. |
| Service Life | 5 years | 1 year | Different by design – Difference in service life does not introduce or raise concerns regarding the safe and effective use of the subject device. |
# TECHNOLOGICAL CHARACTERISTICS COMPARISON
The primary difference between the predicate Kangaroo™ Connect Enteral Feeding Pump cleared under K153074 and the proposed Luminoah FLOW™ Enteral Nutrition System are the sensors used to monitor pump function. The sensing methodology in the FLOW system achieves equivalent monitoring of single faults related to blockage and air in line. Motor monitoring allows for an alternative method of occlusion detection while monitoring from the source of mechanical output / fluid delivery rather than an external sensor.
# PERFORMANCE TESTING
A Safety Assurance Case (SAC) as recommended by the FDA guidance document, Infusion Pumps Total Product Life Cycle is provided for the Luminoah FLOW™ Enteral Nutrition System, including the FLOW™
{14}
LUMINOAH
Durable Pump, FLOW™ Administration Aet and Luminoah FLOW™ Accessories. The stated top-level claim of the SAC is:
The Luminoah FLOW™ Enteral Nutrition System, comprised of the FLOW™ Enteral Feeding Pump, FLOW™ Administration Sets and FLOW™ Accessories are acceptably safe for its intended use, within its environment of use, when being used by intended users, over the lifecycle of the product.
The following specific evidence was included within the SAC to demonstrate that the subject device is verified and validated for its intended use, and to demonstrate substantial equivalence to the predicate device.
| Software | Software verification and validation per FDA guidance for the Content of Premarket Submissions for Device Software Functions (June, 14, 2023) for Enhance Documentation level and FDA guidance document Infusion pump total product life cycle (December 2, 2014) |
| --- | --- |
| Electrical Safety | The electrical safety evaluation of the medical electrical equipment was performed per standards IEC60601-1 Medical electrical equipment part1: General requirements for basic safety and essential performance |
| EMC | The electromagnetic compatibility was evaluated to IEC 60601-1-2: Medical electrical equipment- Part 2: General requirement for basic safety and essential performance - Collateral standard: electromagnetic compatibility - Requirements and tests and per FDA guidance document Electromagnetic Compatibility (EMC) of Medical Devices. |
| Device Performance | The performance requirements of the device (including feeding sets) were verified through performance testing in accordance with the intended use of the device and in accordance with the FDA Guidance “Infusion Pumps Total Product Life Cycle” including:
• Performance testing
• Reliability testing
• Flow rate accuracy testing across all operating conditions (in accordance with AAMI TIR101:2021)
• Alarm detection |
| Human Factors | Following FDA Guidance Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016), the human factors studies were conducted with the intended user population, use environment, and use scenarios to simulate clinical conditions. Results of the human factors testing demonstrate validation of the device per the intended use. |
| Biocompatibility | The biocompatibility test reports provided were conducted per ISO10993 series standard following Good Laboratory Practices and the representative product tested passed all acceptance criteria. |
| Battery Testing | The battery pack has been tested in accordance with IEC 62133. |
Clinical Testing
Not Applicable
{15}
LUMINOAH
# CONCLUSION
The subject device, Luminoah FLOW™ Enteral Nutrition System, FLOW™ Durable Pump, FLOW™ Administration Sets and FLOW™ Accessories, is substantially equivalent to the legally marketed predicate device, Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets and Kangaroo™ Connect Portal (K153074) with respect to the intended use/indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.